Light-field micro-endoscopy using a fiber bundle: a snapshot 3D epi-fluorescence endoscope

Open Access Open Access

Abstract

Micro-endoscopes are widely used for detecting and visualizing hard-to-reach areas of the human body and for in vivo observation of animals. A micro-endoscope that can realize 3D imaging at the camera framerate could benefit various clinical and biological applications. In this work, we report the development of a compact light-field micro-endoscope (LFME) that can obtain snapshot 3D fluorescence imaging, by jointly using a single-mode fiber bundle and a small-size light-field configuration. To demonstrate the real imaging performance of our method, we put a resolution chart in different z positions and capture the z-stack images successively for reconstruction, achieving 333-μm-diameter field of view, 24 μm optimal depth of field, and up to 3.91 μm spatial resolution near the focal plane. We also test our method on a human skin tissue section and HeLa cells. Our LFME prototype provides epi-fluorescence imaging ability with a relatively small (2-mm-diameter) imaging probe, making it suitable for in vivo detection of brain activity and gastrointestinal diseases of animals.

© 2022 Chinese Laser Press

Full Article  |  PDF Article
More Like This
Super-simplified fiber scanner for cellular-resolution endoscopic imaging

Lu He, Xijie Li, Jie Yang, Longjie Jiang, Qian Liu, and Ling Fu
Photon. Res. 11(12) 2020-2032 (2023)

High-fluence relay-based disposable photoacoustic-ultrasonic endoscopy for in vivo anatomical imaging of gastrointestinal tract

Xue Wen, Peng Lei, Shengxin Huang, Xiaoyu Chen, Yanchi Yuan, Di Ke, Rui Liu, Jiaxi Liang, Erqi Wang, Bo Wei, Kedi Xiong, and Sihua Yang
Photon. Res. 11(1) 55-64 (2023)

Endoscopic en-face optical coherence tomography and fluorescence imaging using correlation-based probe tracking

Manuel J. Marques, Michael R. Hughes, Adrián F. Uceda, Grigory Gelikonov, Adrian Bradu, and Adrian Podoleanu
Biomed. Opt. Express 13(2) 761-776 (2022)

Data Availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (11)

Fig. 1.
Fig. 1. Principle of LFME. (a) Light-field micro-endoscopy (LFME) imaging scheme. The compact LFME system mainly consists of a small-size imaging lens pair, a small-size micro-lens array (MLA), and a single-mode fiber bundle (SMFB). Zoom-in panel ① shows the relationship of a sample space point with the native image plane (NIP). Zoom-in panel ② shows the enlarged image of the excitation light passing through the fiber distal end to the MLA plane and the emission light passing back. Note that on the illumination side, we use a rotating diffuser to weaken the spatial coherence of the laser source and reduce the speckle noise. (b) Cross-section image and its enlarged view of the SMFB (top), and the real image pattern and its enlarged view of the entire LFME system (bottom). (c) Experimental performance test by imaging randomly distributed fluorescent beads. The lateral and axial maximum intensity projections (MIPs) of beads are shown, after performing the Richardson–Lucy (RL) deconvolution. The resolved FWHM is 6.06 μm in the x direction and 35.40 μm in the z direction. A zoom-in image of the green box in the zy MIP image is also shown here, which indicates two beads with 10.35μm depth distance, exhibiting the 3D imaging capability.
Fig. 2.
Fig. 2. Dictionary learning procedure of LFME.
Fig. 3.
Fig. 3. Fluorescence imaging with LFME. (a) Raw fluorescence light-field image of the human skin tissue. (b) Raw fluorescence image captured without MLA as a reference. Both images have heavy pixelation and large background fluorescence. Reconstructed fluorescence imaging in a certain slice using (c) RL deconvolution (Decov) and (d) dictionary learning after deconvolution (Decov+Dict), respectively. Benefiting from the optical design and reconstruction process of LFME, our method can resolve detailed structures of the skin tissue from (a) with a clean background. As shown in the enlarged drawings marked by the yellow box and the intensity curve in (f), dictionary learning can suppress artifacts and increase the contrast of reconstruction. (e1) is the central view image of a USAF-1951 resolution chart, while (e2) and (e3) are the reconstruction results using RL deconvolution and dictionary learning after deconvolution, respectively. Intensity curves shown in (g) correspond to the red, orange, and blue lines in (e1), (e2), and (e3), respectively.
Fig. 4.
Fig. 4. Lateral resolution analysis for LFME. (a) Simulated LFME measurements of a point source in different axial positions. (b) Deconvolved images of a point source in different axial positions. (c) Experimentally characterized lateral resolution by imaging the resolution chart. (d) The simulated modulation transfer function (MTF) varies across different depths.
Fig. 5.
Fig. 5. Fluorescence imaging by directly applying SMFB. From (a1)–(a8), the resolution chart is successively placed 0–25 μm away from the distal end of SMFB. (b1)–(b8) show corresponding enlarged drawings.
Fig. 6.
Fig. 6. 3D imaging of biological samples with LFME. (a1) Recovered lateral (xy) MIP image of HeLa cells using RL deconvolution (Decov). The xz and yz slices of the sample (marked by white dashed lines in the xy MIP image) are exhibited simultaneously. (a2) Results of HeLa cells using dictionary learning method (Decov+Dict). (b1) and (b2) show results of a human skin tissue section in area 1 by using these two methods respectively, while (c1) and (c2) show results of the skin tissue in area 2.
Fig. 7.
Fig. 7. Effectiveness of rotating diffuser in LFME. We show experimentally captured images of the fluorescent plate, HeLa cells, and skin tissue to demonstrate the effectiveness of rotating diffuser in LFME. The former three columns directly use the lens pair for acquisition without using the MLA. The first column indicates imaging without the diffuser, which has undesirable interference patterns (labeled by yellow arrows) and speckles. The second column indicates imaging by applying a diffuser, where the interference pattern is suppressed, but the speckles still exist. The third column indicates imaging by rotating the diffuser continuously to weaken the spatial coherence of the laser source, where the speckles are eliminated at this time. The last column uses both the lens pair and MLA with diffuser rotation for image acquisition, which indicates the input raw light-field measurement. (b5) and (c5) are the reconstructed images with high resolution and clean backgrounds from the raw light-field measurements.
Fig. 8.
Fig. 8. Comparison of light-field measurements before (left) and after (right) the resample step.
Fig. 9.
Fig. 9. Comparison of light-field deconvolution with and without rotating the diffuser.
Fig. 10.
Fig. 10. Comparison of system’s PSFs before and after the fiber bundle by imaging fluorescent beads.
Fig. 11.
Fig. 11. Lateral resolution testing in a larger DoF.