Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Pulse interferometry with a free-space Fabry-Pérot for shot-noise-limited detection of ultrasound

Not Accessible

Your library or personal account may give you access

Abstract

Coherence-restored pulse interferometry (CRPI) is a recently developed method for optical detection of ultrasound that achieves shot-noise-limited sensitivity and high dynamic range. In principle, the wideband source employed in CRPI may enable the interrogation of multiple detectors by using wavelength multiplexing. However, the noise-reduction scheme in CRPI has not been shown to be compatible with wideband operation. In this work, we introduce a new scheme for CRPI that relies on a free-space Fabry-Pérot filter for noise reduction and a pulse stretcher for reducing nonlinear effects. Using our scheme, we demonstrate that shot-noise-limited detection may be achieved for a spectral band of 80 nm and powers of up to 5 mW.

© 2019 SPIE/OSA

PDF Article
More Like This
Shot-noise-limited optical sensing of ultrasound using pulse interferometry with a free-space Fabry-Pérot

Oleg Volodarsky, Yoav Hazan, and Amir Rosenthal
SW3L.1 CLEO: Science and Innovations (CLEO:S&I) 2019

Optical phase shifted pulse interferometry for parallel multichannel ultrasound detection

Yoav Hazan and Amir Rosenthal
11077_17 European Conference on Biomedical Optics (ECBO) 2019

Phase modulated pulse interferometry for simultaneous multi-channel ultrasound detection

Yoav Hazan and Amir Rosenthal
SW3L.2 CLEO: Science and Innovations (CLEO:S&I) 2019

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.