Abstract
In this paper we will report on a study of surface plasmon resonance biosensors. The biosensors consist of a 50 nm metal thin film, with one or more layers of graphene attached to the lower surface. Surface plasmons is excited by an incident transverse magnetic field on the upper surface of the metal. This incident field is considered to be in a glass material and at an angle of incidence that will excite surface plasmons, as in the Kretschmann configuration. There is high biomolecule absorption in the graphene and this is shown to lead to a significant enhancement in the sensitivity of the surface plasmon resonance biosensor. In this paper we show, using a computational model, the increase in sensitivity that can be obtained from single and multilayer graphene. In the model we also consider different metal thin films, such as gold, silver and copper. The reflection of visible light coupled into a SPR mode propagating along a thin metal-graphene layer is calculated and compared with a conventional SPR biosensor where there is not a graphene layer present. It is shown that the reflection obtained is strongly dependent upon both the number of graphene layers and the thickness of metal thin film that is used.
© 2015 Japan Society of Applied Physics, Optical Society of America
PDF ArticleMore Like This
A.T Toloue, Anthony Centeno, and M. T. Ahmadi
17p_C3_7 JSAP-OSA Joint Symposia (JSAP) 2014
Pradeep Kumar Maharana, Triranjita Srivastava, and Rajan Jha
TPo.21 International Conference on Fibre Optics and Photonics (Photonics) 2012
Satyendra Kumar Mishra, Bing Zou, Kar Pong Lor, and Kin Seng Chiang
C4H_4 Conference on Lasers and Electro-Optics/Pacific Rim (CLEO/PR) 2020