A theory of the influence of measurement conditions on temperature measurement accuracy with infrared systems has been recently presented. A comparison study of the shortwave (3–5-μm) and longwave (8–12-μm) measuring IR cameras was conducted on the basis of this theory. The results of the simulations show that the shortwave systems in typical measurement conditions generally offer better accuracy in temperature measurement than do the longwave systems. Some experiments that use a commercially available IR camera were carried out to verify the theory. The results of these experiments and a discussion about the theory limitations are presented.