Fluorescence Rejection in Resonance Raman Spectroscopy Using a Picosecond-Gated Intensified Charge-Coupled Device Camera

Not Accessible

Your library or personal account may give you access

Abstract

A Raman instrument was assembled and tested that rejects typically 98–99% of background fluorescence. Use is made of short (picosecond) laser pulses and time-gated detection in order to record the Raman signals during the pulse while blocking most of the fluorescence. Our approach uses an ultrafast-gated intensified charge-coupled device (ICCD) camera as a simple and straightforward alternative to ps Kerr gating. The fluorescence rejection efficiency depends mainly on the fluorescence lifetime and on the closing speed of the gate (which is about 80 ps in our setup). A formula to calculate this rejection factor is presented. The gated intensifier can be operated at 80 MHz, so high repetition rates and low pulse energies can be used, thus minimizing photodegradation. For excitation we use a frequency-tripled or -doubled Ti : sapphire laser with a pulse width of 3 ps; it should not be shorter in view of the required spectral resolution. Other critical aspects tested include intensifier efficiency as a function of gate width, uniformity of the gate pulse across the spectrum, and spectral resolution in comparison with ungated detection. The total instrumental resolution is 7 cm<sup>−1</sup> in the blue and 15 cm<sup>−1</sup> in the ultraviolet (UV) region. The setup allows one to use resonance Raman spectroscopy (RRS) for extra sensitivity and selectivity, even in the case of strong background fluorescence. Excitation wavelengths in the visible or UV range no longer have to be avoided. The effectiveness of this setup is demonstrated on a test system: pyrene in the presence of toluene fluorescence (λ<sub>exc</sub> = 257 nm). Furthermore, good time-gated RRS spectra are shown for a strongly fluorescent flavoprotein (λ<sub>exc</sub> = 405 nm). Advantages and disadvantages of this approach for RRS are discussed.

PDF Article
More Like This
Fluorescence suppression in Raman spectroscopy using a time-gated CMOS SPAD

Juha Kostamovaara, Jussi Tenhunen, Martin Kögler, Ilkka Nissinen, Jan Nissinen, and Pekka Keränen
Opt. Express 21(25) 31632-31645 (2013)

Fast Raman spectral mapping of highly fluorescing samples by time-gated spectral multiplexed detection

Christopher J. Corden, Dustin W. Shipp, Pavel Matousek, and Ioan Notingher
Opt. Lett. 43(23) 5733-5736 (2018)

Use of an unintensified charge-coupled device detector for low-light-level Raman spectroscopy

C. A. Murray and S. B. Dierker
J. Opt. Soc. Am. A 3(12) 2151-2159 (1986)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved