Abstract
We describe a simple implementation of a slit scanning confocal microscope to obtain an axial resolution better than that of a point-scanning confocal microscope. Under slit illumination, images of a fluorescent object are captured using an array detector instead of a line detector so that out-of-focus light is recorded and then subtracted from the adjacent images. Axial resolution after background subtraction is 2.2 times better than the slit confocal resolution, and out-of-focus image suppression is calculated to attenuate with defocus faster by 1 order of magnitude than in the point confocal case.
© 2008 Optical Society of America
Full Article | PDF ArticleMore Like This
Takashi Fukano and Ichirou Yamaguchi
Opt. Lett. 25(8) 548-550 (2000)
Paul C. Lin, Pang-Chen Sun, Lijun Zhu, and Yeshaiahu Fainman
Appl. Opt. 37(28) 6764-6770 (1998)
Sunghoon Kwon and Luke P. Lee