Abstract
The instrument described is a newly designed reflectometer. Measurements are made with a polarized light beam as a function of wavelength and angle of incidence. Calibration tests have outlined the possibility of using the instrument for reflectance measurements on mirrors with good reproducibility (0.2%) and accuracy (better than 1%). This has been obtained by the alignment system of the sample and the absolute method used for the determination of its reflectance.
© 1990 Optical Society of America
Full Article | PDF ArticleMore Like This
Victor R. Weidner and Jack J. Hsia
Appl. Opt. 19(8) 1268-1273 (1980)
Dan Sheffer, Uri P. Oppenheim, and Adam D. Devir
Appl. Opt. 29(1) 129-132 (1990)
A. Bittar and J. D. Hamlin
Appl. Opt. 23(22) 4054-4057 (1984)