Abstract
A 150-groove per millimeter grating was installed in a Baird Model B infrared spectrophotometer to provide improved resolution in the 4000 to 1000 cm−1 region. Grating orders are separated by means of a small prism monochromator installed after the exit slit of the converted grating monochromator; and modifications were made in the slit, wavelength drive, and recording systems to provide a maximum of flexibility in the operation of the instrument over narrow wavelength intervals
© 1964 Optical Society of America
Full Article | PDF ArticleMore Like This
Hiroshi Yoshinaga, Shigeo Minami, Isao Makino, Isao Iwahashi, Masao Inaba, and Koichi Matsumoto
Appl. Opt. 3(12) 1425-1430 (1964)
C. S. C. Tarbet and E. F. Daly
J. Opt. Soc. Am. 49(6) 603-608 (1959)
T. M. Hard and R. C. Lord
Appl. Opt. 7(4) 589-598 (1968)