Abstract
We have developed an instrument for directly measuring the emissivity of reflective surfaces at near-millimeter wavelengths. The thermal emission of a test sample is compared with that of a reference surface, allowing the emissivity of the sample to be determined without heating. The emissivity of the reference surface is determined by one’s heating the reference surface and measuring the increase in emission. The instrument has an absolute accuracy of Δ∊ = 5 × 10−4 and can reproducibly measure a difference in emissivity as small as Δ∊ = 10−4 between flat reflective samples. We have used the instrument to measure the emissivity of metal films evaporated on glass and carbon fiber-reinforced plastic composite surfaces. We measure an emissivity of (2.15 ± 0.4) × 10−3 for gold evaporated on glass and (2.65 ± 0.5) × 10−3 for aluminum evaporated on carbon fiber-reinforced plastic composite.
© 1995 Optical Society of America
Full Article | PDF ArticleMore Like This
J. J. Bock, A. E. Lange, H. Matsuhara, T. Matsumoto, T. Onaka, and S. Sato
Appl. Opt. 34(13) 2268-2277 (1995)
S. Padin
Appl. Opt. 53(20) 4431-4439 (2014)
Sønnik Clausen, Axel Morgenstjerne, and Ole Rathmann
Appl. Opt. 35(28) 5683-5691 (1996)