Abstract
A compact ammonia sensor based on a 10-µm single-frequency, thermoelectrically cooled, pulsed quantum-cascade laser with an embedded distributed feedback structure has been developed. To measure NH3 concentrations, we scanned the laser over two absorption lines of its fundamental ν 2 band. A sensitivity of better than 0.3 parts per million was achieved with just a 1-m optical path length. The sensor is computer controlled and automated to monitor NH3 concentrations continuously for extended periods of time and to store data in the computer memory.
© 2002 Optical Society of America
Full Article | PDF ArticleMore Like This
Anatoliy A. Kosterev, Frank K. Tittel, Rüdeger Köhler, Claire Gmachl, Federico Capasso, Deborah L. Sivco, Alfred Y. Cho, Shawn Wehe, and Mark G. Allen
Appl. Opt. 41(6) 1169-1173 (2002)
Anatoliy A. Kosterev, Frank K. Tittel, Claire Gmachl, Federico Capasso, Deborah L. Sivco, James N. Baillargeon, Albert L. Hutchinson, and Alfred Y. Cho
Appl. Opt. 39(36) 6866-6872 (2000)
Michael E. Webber, Ricardo Claps, Florian V. Englich, Frank K. Tittel, Jay B. Jeffries, and Ronald K. Hanson
Appl. Opt. 40(24) 4395-4403 (2001)