Abstract
In this paper, a doubled-period grating (DPG) method is proposed to measure the scan angle error in scanning beam interference lithography (SBIL), together with a high-resolution two-dimensional stage and phase-stepping algorithms. A reference grating, which was the doubled period of the interference field, is adapted to diffract the incident left and right beams to form an interferogram captured by a CCD camera. The phase-stepping algorithm was applied to calculate the phase of the interferogram. First, by translating the stage, the reference grating lines were adjusted parallel to the scan direction by comparing the phase of the interferogram between the starting point and the ending point. Next, by rotating the stage step by step, the phase of the interferogram was obtained at each step as well as the phase slope. The scan angle error was considered to be least when the slope was minimum. Finally, a grating mask with a size of was fabricated to verify the feasibility of the method. The scan angle error was measured with a precision of less than 12.65 μrad, which indicates the effectiveness of the proposed method to fabricate high-quality gratings in the future.
© 2019 Optical Society of America
Full Article | PDF ArticleMore Like This
Wei Wang, Ying Song, Shan Jiang, Mingzhong Pan, and Bayanheshig
Appl. Opt. 56(14) 4138-4145 (2017)
Yubo Li, Shan Jiang, Xingshuo Chen, Zhaowu Liu, Wei Wang, Ying Song, and Bayanheshig
Opt. Express 31(17) 28145-28160 (2023)
Shan Jiang, Bo Lü, Ying Song, Zhaowu Liu, Wei Wang, Li Shuo, and Bayanheshig
Appl. Opt. 59(19) 5830-5836 (2020)