Abstract
We present a highly accurate Raman spectrometer capable of measuring changes in atmospheric temperature as small as 0.75 K with high spatial and temporal resolution. The spectrometer is based on a laser diode tuned to the resonant absorption line of the $^{85}{\rm{Rb}}$ isotope near 780.0 nm. A heated glass cell containing Rb atoms was used as an ultranarrowband atomic absorption notch filter with $0.3 \;{\rm{c}}{{\rm{m}}^{- 1}}$ bandwidth and optical density as high as four. This filter was placed in front of the spectrograph and blocked up to 99.99% of the elastically scattered laser light, which made it possible to resolve the pure-rotational Raman spectra of molecular atmospheric gases. The relative intensities of pure-rotational Raman transitions were then used to infer atmospheric temperature changes.
© 2021 Optical Society of America
Full Article | PDF ArticleMore Like This
Yu. F. Arshinov, S. M. Bobrovnikov, V. E. Zuev, and V. M. Mitev
Appl. Opt. 22(19) 2984-2990 (1983)
Henry Misoi, Jonas I. Hölzer, and Thomas Seeger
Appl. Opt. 60(15) 4410-4417 (2021)
Andreas Behrendt and Jens Reichardt
Appl. Opt. 39(9) 1372-1378 (2000)