Abstract
The feasibility of using fiber-optic Raman probes to identify and quantify gases in enclosures is investigated by measuring and comparing detection thresholds using several probe and enclosure designs. Unfiltered, non-imaging, fiber-optic probes are shown to achieve lower detection thresholds than a filtered, imaging, fiber-optic probe, provided that light scattering within the sample enclosure is minimized and provided that a window is not used between the probe and the analyte gas. Achievable thresholds for hydrogen, oxygen, nitrogen, carbon monoxide, and methane in gas mixtures are demonstrated to be below 1 kPa with ten seconds signal acquisition and 0.1 kPa with twenty minutes signal acquisition with the use of 0.4 W of 532-nm excitation. Ambient carbon dioxide in air (.03 kPa) is shown to be detectable in a twenty minute acquisition, and ambient water vapor is well above the detection threshold. Background signals generated within the optical fibers remain the principal factors limiting detection thresholds. Factors affecting the magnitudes of these signals reaching the detector are investigated and discussed. A flat piece of light-absorbing colored glass tilted to direct reflected light away from the fiber-optic probe performs well as a beam stop to reduce background signal in a simple, cylindrical sample enclosure.
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription