Abstract
A nondestructive technique for the determination of thorium in optical glass has been devised using tube-excited x-ray fluorescence with an energy-dispersive detector. Capability for full calibration and sample analysis has been incorporated into a computer-coupled system. Bulk thorium concentrations as low as 87 ppm have been detected with a coefficient of variation of 27.2% at 95% confidence using NBS-certified standards. At 466 ppm the coefficient of variation was 5.6% at the 95% confidence level. Qualitative determination of thorium in optical coatings has been demonstrated also.
PDF Article
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription