Abstract
A simple method is described for the approximate measurement of the ion energy distribution in an inductively coupled plasma-mass spectrometer (ICP-MS) with a continuum flow sampling interface. The average ion kinetic energy, kinetic energy spread, and maximum kinetic energy are evaluated from a plot of ion signal as a function of retarding voltage applied to the quadrupole mass analyzer. The effects of plasma operating parameters on ion signals and energies are described. In particular, kinetic energy is a sensitive function of aerosol gas flow rate. This behavior is attributed to a nonthermal, possibly electrical, interaction between the plasma and the sampling interface, which is induced by the presence of the axial channel in the ICP.
PDF Article
More Like This
Multielemental analysis of prehistoric animal teeth by laser-induced breakdown spectroscopy and laser ablation inductively coupled plasma mass spectrometry
Michaela Galiová, Jozef Kaiser, Francisco J. Fortes, Karel Novotný, Radomír Malina, Lubomír Prokeš, Aleš Hrdlička, Tomáš Vaculovič, Miriam Nývltová Fišáková, Jiří Svoboda, Viktor Kanický, and Javier J. Laserna
Appl. Opt. 49(13) C191-C199 (2010)
Propagation characteristics of terahertz wave in inductively coupled plasma
Jinhai Sun, Yan Zheng, Jielin Shi, Yarui Zhao, Yu Li, Ding Wu, He Cai, Xutao Zhang, Xianli Zhu, Yongqiang Liu, Xinxue Sun, Zengming Chao, Hongcheng Yin, Wenqi Lu, and Hongbin Ding
Opt. Express 29(22) 35837-35847 (2021)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription