Evaluation of the Appropriate Sample Position in a PAS/FT-IR Experiment

Not Accessible

Your library or personal account may give you access

Abstract

The experimental selection of the distance between the sample and the window in a photoacoustic FT-IR experiment is shown to be an important parameter when one is optimizing the interferogram intensity. We have experimentally demonstrated the advantages of selecting a small distance rather than the full boundary layer thickness required to ensure complete conversion of thermal to pressure signal at the sample surface. The minimum volume of the sample chamber is a controlling factor in the optimization of the experiment. In this report we demonstrate the value of minimizing sample chamber volume through controlling sample thickness while pointing out the need for care when doing so.

PDF Article
More Like This
Self-supporting carbon glass films: a PAS standard sample?

H. Coufal
Appl. Opt. 21(1) 104-106 (1982)

Quantitative Fourier transform IR photoacoustic spectroscopy of condensed phases

Y. C. Teng and B. S. H. Royce
Appl. Opt. 21(1) 77-80 (1982)

Determination of absorption coefficients of thick semiconductor samples using photoacoustic spectroscopy

J. Fesquet, B. Girault, and M. D. M. Razafindrandriatsimaniry
Appl. Opt. 23(16) 2784-2787 (1984)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved