Abstract
An error in a previously reported procedure for correcting Raman spectra for instrument response was corrected, and then the approach was tested against reported values of Raman cross sections. The method uses luminescent standards with known emission curves of intensity vs. Raman shift, expressed in polynomial form. After acquisition of an observed emission curve from the luminescent standard, the correction of observed Raman spectra is automatic. Agreement of corrected peak areas with those calculated from literature cross sections is 5-20%. This correspondence is approximately as good as the agreement between different reported values, and much better than that observed for uncorrected spectra. Due to variations in the luminescence of the materials, it is likely that these or similar Raman intensity standards will be calibrated by the manufacturer or an independent agency.
PDF Article
More Like This
Application of ultrafast gold luminescence to measuring the instrument response function for multispectral multiphoton fluorescence lifetime imaging
Clifford B. Talbot, Rakesh Patalay, Ian Munro, Sean Warren, Fulvio Ratto, Paolo Matteini, Roberto Pini, H. Georg Breunig, Karsten König, Antony C. Chu, Gordon W. Stamp, Mark A. A. Neil, Paul M. W. French, and Chris Dunsby
Opt. Express 19(15) 13848-13861 (2011)
Optical calibration of the SuperCam instrument body unit spectrometers
Carey Legett, Raymond T. Newell, Adriana L. Reyes-Newell, Anthony E. Nelson, Pernelle Bernardi, Steven C. Bender, Olivier Forni, D. M. Venhaus, Samuel M. Clegg, A. M. Ollila, Paolo Pilleri, V. Sridhar, S. Maurice, and Roger C. Wiens
Appl. Opt. 61(11) 2967-2974 (2022)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription