Abstract
An electrospray ionization (ESI) source was used to transfer oxo-anions (XO<sub>3</sub>-) of nitrogen, chlorine, bromine, and iodine to the gas phase for analysis by a combination of high-field asymmetric waveform ion mobility spectrometry (FAIMS) and mass spectrometry. The combination of ESI with FAIMS and mass spectrometry (ESI-FAIMS-MS) yields mass spectra that are virtually free of common spectral interferences. The unique separation and focusing properties of the FAIMS analyzer improved signal-to-background ratios (S/B) for all these analytes by 3-4 orders of magnitude compared with conventional ESI-MS and produced parts-per-trillion (pptr) detection limits for ClO<sub>3</sub>-, BrO<sub>3</sub>-, and IO<sub>3</sub>- in methanolic solution.