Abstract
Two geometries of excitation/emission optics are compared for two-photon excitation of time-resolved fluorescence in capillaries. The test fluorophore is the enzymatic tag β-naphthylamine (BNA). Two 0.25 NA (numerical aperture) microscope objectives are used in the 90° geometry: This configuration has a sensitivity of 1.9 X 10<sup>10</sup> count M<sup>-1</sup> min<sup>-1</sup> and a detection limit of 2.5 nM [signal-to-noise (S/N) = 3]. The epi-geometry uses single objectives of 0.25, 0.50, or 1.25 NA. This second configuration has a sensitivity of 7.6 X 10<sup>10</sup> count M<sup>-1</sup> min<sup>-1</sup> and a detection limit of 1.7 nM (S/N = 3) for the 1.25 NA objective. The similar performance of the two configurations is in stark contrast to the factor of 46 enhancement predicted by geometric optics for epi-excitation. The discrepancy is attributed to large increases in collected background counts, which make detection of the signal difficult.
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription