Abstract
We constructed an ultrahigh-throughput, high-resolution ultraviolet (UV) Raman spectrograph that utilizes a high-efficiency filter-stage monochromator and a high-dispersion Echelle spectrograph. The spectrograph utilizes a total of six mirrors and two gratings, with an overall efficiency at 229 nm of ?18%. The limiting resolution of our spectrometer is 0.6 cm<sup>?1</sup> full width half-maximum (FWHM), as measured for 229 nm Rayleigh scattering. Use of a 1 mm-wide entrance slit gives rise to an approximately 10 cm<sup>?1</sup> FWHM resolution at 229 nm. The ultrahigh spectrograph throughput enables ultrahigh signal-to-noise ratio, deep UV Raman spectra that allow us to monitor <1% changes in peptide bond composition. The throughput is measured to be 35-fold greater than conventional deep UV Raman spectrometers.
PDF Article
More Like This
Development of a high-resolution, broadband spatial heterodyne Raman spectrometer based on field-widened grating-echelle structure
Qihang Chu, Fuguan Li, Xiaotian Li, Ci Sun, Yuqi Sun, Jirigalantu, Nan Song, Shuo Yu, Rui Zhang, and Bayanheshig
Opt. Express 32(10) 17819-17836 (2024)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.