Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Optimizing ultrafast illumination for multiphoton-excited fluorescence imaging

Open Access Open Access

Abstract

We study the optimal conditions for high throughput two-photon excited fluorescence (2PEF) and three-photon excited fluorescence (3PEF) imaging using femtosecond lasers. We derive relations that allow maximization of the rate of imaging depending on the average power, pulse repetition rate, and noise characteristics of the laser, as well as on the size and structure of the sample. We perform our analysis using ~100 MHz, ~1 MHz and 1 kHz pulse rates and using both a tightly-focused illumination beam with diffraction-limited image resolution, as well loosely focused illumination with a relatively low image resolution, where the latter utilizes separate illumination and fluorescence detection beam paths. Our theoretical estimates agree with the experiments, which makes our approach especially useful for optimizing high throughput imaging of large samples with a field-of-view up to 10x10 cm2.

© 2016 Optical Society of America

1. Introduction

Imaging utilizing two-photon excitation of fluorescence (2PEF) and, since recently, three-photon excited fluorescence (3PEF), has significantly enhanced biological microscopy by improving image quality and acquisition speed, reducing photo-damage to the sample, as well as by allowing deeper sample depth penetration [1–11]. While a broad variety of new imaging modalities continue to be demonstrated that take advantage of the high peak pulse intensity of different femtosecond laser sources, optimizing the illumination and detection conditions still poses an issue. One reason for this is because the average rate of multiphoton excitation depends not only on the average incident power, as in the case of linear (one-photon) excitation, but also on the instantaneous flux of photons, i.e. peak intensity of the pulses. This imposes constraints on the photon budget, which, along with such factors as the sample damage threshold, maximum laser power available, and various sources of experimental noise, makes finding the optimal illumination- and detection conditions a challenging task.

In this paper we provide quantitative analysis of the maximum imaging rate that can be achieved under realistic conditions in femtosecond multiphoton excited fluorescence systems. Faster imaging is imperative for capturing real-time processes such as the activity of individual nerve cells, which occur on a sub-millisecond time scale, and is also critical for reducing long scan times that currently limit multi-photon imaging of large-area samples. We achieve our goal by taking into account different illumination and focusing conditions, sample properties, and the signal-to-noise ratio (SNR) of the image data, where the latter may be affected by photon shot noise, laser pulse energy fluctuations and fluorescence detection noise. As the first step, we consider a generic multiphoton microscope arrangement, with coaxial illumination- and fluorescence detection beam paths, using the tightly-focused beam of a 100 MHz pulse rate femtosecond oscillator. We evaluate the maximum achievable shot noise limited rate of imaging depending on the two-photon absorption (2PA) and three-photon absorption (3PA) cross section values of the fluorophores and other key parameters such as average laser power, sample damage threshold etc. Some applications such as early cancer detection [12] and high throughput screening [13] require much larger samples, that cannot be addressed by conventional coaxial illumination setups. For this purpose, we model a wide field-of-view (FOV) imaging system, where the illumination- and the fluorescence detection beam paths are separated, and where the illumination uses a ~kHz pulse rate amplified femtosecond laser. In order to verify our model, we compare the theoretical results with published data about different multiphoton microscope systems. We also perform experiments, where we determine the maximum imaging rate in a CCD camera based, ~10x10 cm2 FOV, 2PEF imaging system using different 1 kHz amplified laser sources, and show that our theoretical analysis has a good correlation with the experiments. To our best knowledge, this paper presents the first quantitative analysis of femtosecond illumination for wide FOV 2PEF imaging as well as the first theoretical treatment of imaging speed optimization for multi-photon fluorescence microscopes.

2. Results and discussion

2.1 2PEF and 3PEF imaging with coaxial illumination

Figure 1(a) shows the femtosecond laser beam that emanates from the output aperture of a high-NA microscope objective and converges onto a diffraction-limited 3D spot inside a sample of thickness, d. If we would neglect potential side-effects associated with the femtosecond illumination, such as photo-bleaching and saturation of the 2PA transition, then the number of the fluorophores excited by a single laser pulse may be expressed as Eq. (1) [3,9,14],

N2PA=  σ2PA2VC(x,y,z)[IL2(x,y,z,t)dt]dxdydz,
where σ2PA is the 2PA cross section (in units of GM = 10−50 cm4 s photon−1) of the fluorophores at the illumination wavelength, λL, IL is the photon flux (number of photons per unit area and per unit time), C is the concentration of 2PEF-active fluorophores, and the integration is carried out over the duration of the pulse and over the entire volume of the sample. If we assume, for simplicity, that the fluorophore concentration is constant throughout the sample, then most of the 2PEF signal will be produced by the volume confined inside the focal region (voxel), where the photon flux is the highest. In the case of a Gaussian spatial beam profile, the voxel may be defined by the closed surface, where the peak photon flux decreases by factor e−2 from its maximum value,
Veff  53.0ω04λL,
where ω0 is the beam waist related to the numerical aperture of the objective, ω0 = λLNA)−1, and the length is given in meters. If we approximate the temporal intensity profile with a Gaussian with FWHM value, τp, and if the Rayleigh length of the focus, ΔzR = (πω02)/λL, is much less than the sample thickness, ΔzR << d, e.g. as shown in Fig. 1(a), then, by carrying out the integral over the voxel volume, Eq. (1) simplifies to:
N2PA1.24Cσ2PAλL8ln2τp(Paveπghc)2,
where Pave is the average excitation power in W, g is the pulse repetition rate in s−1, h is Planck’s constant and c is the speed of light (both in MKS units). The above relation illustrates the well-known fact that the number of excited fluorophores is proportional to the square of the incident power and to the 2PA cross section. It also shows, provided that the voxel is entirely contained inside the sample, the independence of the excitation rate on NA, or on how tightly the beam is focused. Equation (3) may be recast in a more familiar way, that was first introduced by Denk et al. [3]:
N2PA 1.30VeffC[σ2PAτp(PaveNA2ghcλL)2],
where the quantity enclosed in the square brackets is the number of photons absorbed per fluorophore per pulse.

 figure: Fig. 1

Fig. 1 Schematic of a generic coaxial illumination- and fluorescence detection microscope. (a) The laser beam emanates from the output aperture of a high-NA objective and converges onto the focus voxel (dashed line); d is the sample thickness, ΔzR is the Rayleigh length of the focus, λL is the laser wavelength, and λF is the fluorescence wavelength. (b) Simplified raster scan pattern, where the step size is equal to the lateral size of the focus spot, Δx.

Download Full Size | PDF

For multiphoton microscopy the focus point is scanned in the x-y directions until it covers the entire FOV area, e.g. in a raster pattern shown in Fig. 1(b), and a 2D image is built pixel-by-pixel from the recorded 2PEF signals. The lateral resolution is determined either by the focus spot size or by the raster step size, whichever of the two is larger. Here, for simplicity, the 2D focus spot is shown as a square pixel of area, ΔS = Δx2, and the raster-scan step size is set equal to Δx. The average number of the fluorophores excited in the nth pixel is then proportional to the number of laser pulses delivered at that pixel:

(N2PA)n= gτdwN2PA,
where τdw is the pixel dwell time. The total number of pixels in the FOV is,
MFOV= FOVΔS.
The inverse of the minimum time required to scan the whole FOV is called the frames-per-second, or FPS, and may be expressed as:
FPS=1τdwMFOV.
In the coaxial arrangement, the objective used for illumination also collects the fluorescence photons, which are then passed on to a photo-detector (not shown in Fig. 1). The number of fluorescence photons detected from one laser pulse is:
(F2PEF)n=ηMOεeff(N2PA)n ,
where εeff is the fluorescence detection efficiency and ηMO is the fluorescence photon collection efficiency. Assuming that the fluorescence photons are not scattered by the sample, and using the definition of the numerical aperture; NA = nsin(θ), where n is the index of refraction of the sample and θ is the half angle of the cone of light collected by the objective, we can express the fluorescence collection efficiency as [15]:
ηMO=12[1(1NA2n2)1/2],
Spectral band-pass filters are used to reject scattered laser light as well as any other spurious photons, while allowing the fluorescence photons to pass through to the detector. The overall fluorescence detection efficiency may be expressed as:
εeff= ΦFL(λF)ϕF(λF)κD(λF)dλF,
where ΦFLF) is the differential fluorescence emission quantum efficiency of the fluorophores, ϕFF) is the spectral filter transmission function and κDF) is the quantum efficiency of the photo-detector. The integration in Eq. (10) is performed over all fluorescence wavelengths.

Ultimately, the utility of any image depends not only on the highest attainable resolution and fastest FPS, but also on whether the sought-after information may be extracted in the presence of the inevitable noise. In order to quantify the latter, we introduce the per-pixel SNR defined as:

SNR=(F2PEF)nδshot2+δlaser2+δel2,
where δshot stands for the photon shot noise, δlaser is the noise due to laser pulse-to-pulse energy fluctuations and δel represents detection noise, e.g. due to electronic amplifiers. The pulse-to-pulse energy fluctuations of mode-locked femtosecond lasers operating at high pulse rates, g > 1 MHz, are usually small, while δel can be minimized by using photon counting detection schemes. This leaves δshot as the main contributing factor to the noise, in which case, Eq. (11) may be expressed as:
SNR=(F2PEF)n.
By solving Eq. (5) for τdw, and using the relations Eqs. (8), (12), and (3), we can express Eq. (7) for the maximum attainable frame rate as:

FPSmax1.24ηMOεeffCσ2PAλL8ln(2)gτpMFOVSNRmin2(Paveπhc)2.

Equation (13) illustrates the well-known fact that the highest attainable imaging rate in a 2PEF microscope system is limited by the maximum incident power, i.e. excitation photon budget [7]. Figure 2(a) presents the value of FPSmax, plotted as a function of Pave, for g = 100 MHz, SNRmin = 10, λL = 790 nm, and FOV = 100x100 μm2. Other parameter values used in this calculation are listed in the figure caption. Increasing the average power allows for a decreased dwell time, which in turn increases the frame rate. However, damage to the samples usually limits the average laser power to about, Pave < 50 mW. On the other hand, if the fluorophores possess large σ2PA, then the 2PA transition may approach saturation. To obtain a quantitative estimate of the saturation effect, we numerically solve the kinetic equations for the excited state population in the two-level approximation [16,17]:

dρf(t)dt= 12σ2PAIL2(12ρf(t)) γf0ρf(t),
where ρf is the probability of finding the chromophore in the excited state and γf0 is the spontaneous excited state population relaxation time back to the ground state, 0. Because in our case the excitation pulse duration is much less than γfo −1, Eq. (14) predicts that a maximum of half of the fluorophores can be excited during one laser pulse, which prompts us to set the saturation limit at half of the maximum value. In Fig. 2(a) the two-photon saturation limit is represented by the horizontal dashed line. With all these factors taken into account, we obtain, FPSmax ~103. At this point it is useful to introduce the rate of imaging, right vertical axis of Fig. 2(a), defined as,
ROIm=FOV×FPS.
The estimated maximum attainable rate is then, ~10−1 cm2 s−1.

 figure: Fig. 2

Fig. 2 Solid lines - Dependence of FPSmax on the average power, Pave. (a) 2PEF FPSmax calculated using Eq. (13) with g = 100 MHz and λL = 790 nm. (b) 3PEF FPSmax calculated using Eq. (18) with g = 1 MHz and λL = 1700 nm. Different σ2PA and σ3PA values are indicated above the corresponding lines. The common parameters used in both these calculations are: τp = 150 fs, NA = 0.9, n = 1, C = 2x1021 m−3, and ηMO*εeff = 0.1. The red dotted horizontal line represents the saturation limit, when 25% of the fluorophores are excited during one pulse. Imaging rates estimated from literature data are represented by: (a): (○) – Denk et al. [3], (●) – Fan et al. [18], (■) – Lee et al. [19], (□) – Clancy et al. [20], (◊) - Voigt et al. [21]; (▲) – Tal et al. [22]; (∆) – Murayama et al. [8]. (b): Horton et al. [23] (♦) – at sample surface, (◊) –0.8 mm in sample, and (▼)-0.8-1.4 mm in sample.

Download Full Size | PDF

For comparison, in Fig. 2(a) we have collected some ROIm values, and FPSmax values assuming FOV = 100x100 μm, deduced from published descriptions of different 2PEF microscope systems, including the pioneering demonstration by Denk et al. (hollow circle) [3], where the size of the image and the frame rate were both relatively small, FOV ~40x30 μm2, FPSmax ~0.01 s−1. Later, Fan et al. (solid circles) [18] and Lee et al. (solid square) [19] took advantage of high-speed resonance scanners to facilitate faster image acquisition over a larger FOV~200x200 μm2. Clancy et al. (hollow square) [20] used a commercial two-photon microscope (Sutter Instruments MOM) to image neural activity in a behaving animal with FOV = 160x160 μm2. Voigt et al. (hollow diamond) [21] used a custom-built system to simultaneously address two cortical areas in a live mouse brain, each with FOV = 200x300 μm2. Note that even though different implementations obviously represent quite different experimental conditions, including variation of the fluorophore concentration, cross section etc., the corresponding ROIm values agree with our theoretical predictions.

Instantaneous three-photon absorption (3PA) is a process where the transition from the ground to the final excited state occurs due to simultaneous absorption of three longer-wavelength photons. The version of multiphoton microscopy using three-photon excitation of fluorescence (3PEF) takes advantage of the ability of longer-wavelength photons to penetrate even deeper into scattering tissues [23], and can also potentially achieve a tighter excitation confinement than 2PEF. In the case of degenerate 3PA, the 3PA equivalent of Eq. (1) is:

N3PA=  σ3PA3VC(x,y,z)[IL3(x,y,z,t)dt]dxdydz,
where σ3PA is the 3PA cross section at the wavelength λL. Integrating over the voxel volume gives the number of fluorophores excited by one laser pulse:
N3PA0.60Cσ3PA8ln2(λLω0τp)2(Paveπghc)3.
If we assume, as before, that the accuracy of measuring the 3PEF signal is limited by the photon shot noise, then the corresponding maximum attainable frame rate is:
FPSmax0.60ηMOεeffCσ3PA8ln(2)MFOVSNRmin2(λLω0τpg)2(Paveπhc)3.
Peak values for the three-photon cross section of organic fluorophores are rather small, typically in the range, σ3PA ~10−82 −10−84 cm6 s2 photon−2 [11,24,25]. In practical terms, this means that the relative efficiency of 3PEF is lower compared to that of 2PEF, assuming similar illumination conditions. On the other hand, the maximum average power is still limited by the sample damage threshold, Pave < 50 mW. A suitable solution to this issue was demonstrated by Xu et al. [11], who used a femtosecond laser with about two orders of magnitude lower pulse repetition rate, which boosted the peak photon flux by the same factor. Figure 2(b) presents the dependence of FPSmax on Pave for g = 1 MHz, and for different σ3PA values. The saturation limit is represented again by the horizontal dashed line. Other parameter values are listed in the figure caption. The estimated maximum imaging rate, ~0.6x10−2 cm2 s−1, turns out to be about one order of magnitude less compared to the 2PEF rate at the same average power. Unfortunately, there is a lack of commercially available, near-IR few-MHz pulse rate femtosecond lasers providing sufficient output power. The symbols in Fig. 2(b) represent the ROIm values deduced from the work of Horton et al. [23], where they used a self-built 1 MHz laser, operating at 1675 nm, for 3PEF imaging of an intact mouse brain. 3 mW of average power was sufficient for high-resolution imaging at the surface (filled diamond), whereas increasing the power facilitated access to increasingly deeper layers, up to 0.8 mm (empty diamond) and 1.4 mm (filled triangle). Beyond 1.4 mm there were too few fluorescence photons to achieve a meaningful SNR.

2.2 2PEF and 3PEF imaging using separated illumination- and fluorescence beam paths

Our goal here is to perform a similar photon budget and maximum imaging rate analysis with a much larger FOV. For example, if the sample encompasses a 10 cm diameter Petri dish, then a straightforward extrapolation of the above estimates predicts that it would take at least 103 s to acquire a two-photon image of the whole plate. Even if we would disregard the excessively long acquisition time, aberrations caused by large beam scan angles [26,27] would render the coaxial illumination arrangement rather incompatible with large objects.

An alternative approach is to increase N2PA by increasing the peak photon flux. This may be achieved using amplified femtosecond lasers that operate at ~kHz pulse rates. Such lasers have been commercially available since the 1990-ies and are capable of delivering up to 106 times higher peak power compared to mode-locked femtosecond oscillators. However, to take full advantage of this high power, an entirely different illumination- and detection system design strategy is on order. First of all, diffraction-limited focusing becomes impractical because the resulting extremely high peak intensity may damage not only the sample, but also the objective along with other setup components. Secondly, with a loosely focused illumination beam path, it becomes advantageous to separate it from the fluorescence collection beam path because the former no longer needs high NA optics. Figure 3(a) shows the principle of a wide FOV imaging setup, where the multiphoton excitation is induced by a loosely-focused amplified femtosecond laser beam, while the fluorescence is collected by a camera objective. Due to the small effective NA of the illuminating beam, the corresponding Rayleigh length is now comparable to or even exceeds the sample thickness, ΔzR > d. As will be discussed below, the shape of the beam will be optimized according to the available laser power and properties of the sample, but its lateral size rarely exceeds ~0.1 - 2 cm. To cover the entire FOV, the beam is moved in the x-y directions, e.g. in a raster scan pattern depicted in Fig. 3(b), where we have again assumed, for simplicity, a rectangular beam area, ΔS = ΔxΔy, and that the scan proceeds in equidistant steps without overlap between the neighboring positions. The fluorescence is collected by a lens, as shown in Fig. 3(b), and then focused on the imaging sensor of a camera (not shown). The image resolution is now determined by the lens and the camera sensor, while the maximum FOV is determined either by the camera FOV or by the maximum beam scan area, whichever of the two is smaller. For simplicity, let us assume that the resolution equals the pixel size, {Δxpix, Δypix}.

 figure: Fig. 3

Fig. 3 The image of the fluorescence on a camera in a wide FOV imaging setup. The square illuminating laser spot (red squares) is raster scanned over the surface of a sample with a homogeneous distribution of fluorophores. (a) Side view of the sample. d is the thickness of the sample and ΔzR is the Rayleigh length of the focused laser. (b) Front view of the sample. Δy and Δx are the vertical and horizontal laser spot sizes respectively. For simplicity a square laser illumination spot with even illumination intensity is shown. The camera pixels have height Δypix, and width Δxpix.

Download Full Size | PDF

By following the procedure that led to Eq. (5), we express the number of fluorophores that are excited within the effective pixel volume as:

(N2PA)n=  VeffCσ2PAτdw2gτp(PaveλLhcΔS)2,
where, Veff ~dΔxpixΔypix, and τdw is the illumination beam dwell time on the area ΔS. The number of fluorescence photons detected per one image pixel is:
(F2PEF)n=ηCOεeff(N2PA)n,
where εeff is the camera detection efficiency and ηCO is the fluorescence collection efficiency. When considering the image noise we need to account for the fact that the output of femtosecond amplifiers is noisier than the output of mode-locked oscillators. For example, a typical regenerative Ti:Sapphire amplifier displays short-term root mean square pulse-to-pulse energy fluctuations of about 5 - 20% (RMSlaser = 0.05 – 0.2). Let us assume, for simplicity, that the photon shot noise and the electronic camera detection noise are small and may be neglected compared to the random laser pulse energy fluctuations. In this case, the image pixel SNR increases in proportion to the square root of the number of laser pulses that are averaged at each pixel, which leads us directly to the following simplified expression for the maximum imaging rate:

ROImmaxgΔS4RMSlaser2(SNR)min2.

Inserting into Eq. (21), the typical parameter values, g = 103 Hz, ΔS = 0.1 cm2, RMSlaser = 0.2 and (SNR)min = 10, yields ROImmax ≈6 cm2 s−1. Even though this value exceeds the maximum coaxial 2PEF imaging rate by about two orders of magnitude, one should remember that this is accomplished at the expense of a much lower spatial resolution. Tal et al. [22] demonstrated a CCD -based detection system combined with single-axis line-scan illumination, and achieved FPS = 10 with FOV ~120x80 μm2, at Pave ~10 mW. Maruyama et al. [8] used light-sheet illumination at a 100 kHz laser pulse rate and camera-based detection and demonstrated FOV~1x1 mm2 at Pave ~30 mW. The estimated ROIm values are, correspondingly, ~10−3 cm2 s−1 and ~0.05 cm2 s−1, and are shown in Fig. 2(a). Because these practical ROIm values fall substantially short of the estimation, we take this as a clue that a rectangular beam shape may not be accurate enough. The excitation is better described by Gaussian functions, which in our case may be expressed as:

IL= 2[ln(16)]1/2λLPaveπ3/2ΔxΔyτpghcexp[4ln(2)(t2τp2)2(y2Δy2+x2Δx2)],
where Δx and Δy are the beam spot sizes. It is also important to consider that the illumination beam position is moving continuously over the FOV, rather than in discrete, non-overlapping steps. In practical terms, this means that each pixel is being illuminated not just by one pulse but by multiple pulses, with a different effective photon flux every time, as the spatial distribution, given by Eq. (22), moves over the corresponding area. Furthermore, because the number of photons available per illumination pulse is limited by the maximum power of the laser, it is imperative to optimize the beam size (ΔS = ΔxΔy) as well as the scan speed (dwell time), such that, for each image pixel, the desired minimum SNR is achieved. In other words, a realistic simulation requires an illumination scheme with a moving, low pulse repetition rate Gaussian beam, where each pulse is added up and averaged individually. These rather involved, inter-dependent conditions may be summarized in the following relation:
βs=PavegτpΔSconst.,
where the empirical parameter βs represents the optimal average power density that is to be determined for each particular experiment and/or type of sample. All the above aspects are taken into account by our numerical model that is described in detail in the Appendix. The results of the numeric simulation are presented in Fig. 4(a). The calculation assumes the following parameter values: σ2PA = 10 GM, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, and SNRmin = 10. The power parameter is, βs ~2.0x1014 W m−2, which corresponds to the optimal illumination intensity for imaging of E. coli bacteria colonies expressing fluorescent protein mutants [13]. As before, the lower horizontal axis of the figure corresponds to the average power, which is in this case limited to 4W. The upper horizontal axis shows the corresponding optimal beam spot area determined from Eq. (23). As expected, the maximum imaging rate depends strongly on the laser noise and on the available maximum power, and in the case of RMSlaser = 0.2, with σ2PA ~10 GM peak cross section fluorophores, reaches, ~0.01 – 0.1 cm2 s−1. According to this calculation, the minimum number of pulses needed to achieve SNR = 10, is 50, 200, or 800 pulses for RMSlaser ~5%, 10%, and 20%, respectively. We may conclude that given a typical amplified femtosecond laser, a 2PEF image may be collected from an entire 10 cm diameter Petri dish within less than an hour.

 figure: Fig. 4

Fig. 4 The rate of imaging with amplified laser pulses as a function of average power (lower horizontal axis). Top axis shows the corresponding optimized illumination spot area. (a) Solid red lines - 2PEF in a spatially homogeneous sample with different laser noise levels. SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 790 nm; Solid blue line - 2PEF in a spatially inhomogeneous sample comprised of fluorescent features with an ideal noise-less laser. δel = 2,000, SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 790 nm, σ2PA = 10 GM, Δrk = 1 mm, Vk = 1.1 mm3. Dashed black line – the same sample parameters as the blue line but in the case of a femtosecond oscillator with g = 100 MHz. Comparison to experimental implementations: (●) - Clark MXR CPA-1000 Ti:Sapphire regenerative amplifier; (□) - Coherent Legend-HE Ti:Sapphire regenerative amplifier ;(○) - Light Conversion TOPAS-C optical parametric amplifier. (b) 3PEF in a spatially inhomogeneous sample comprised of fluorescent features with an ideal noise-less laser in the case of different three-photon cross sections. δel = 2,000, SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 1700 nm, Δrk = 1 mm, Vk = 1.1 mm3.

Download Full Size | PDF

Typical scientific-grade cameras have a read-out noise of about ~10 −100 counts per pixel. If we would assume that the laser noise is absent, then, to guarantee a minimum per-pixel SNR = 10, the fluorescence signal should be at least ~102 – 103 counts per pixel. Note that at such a relatively high signal level the shot noise may be disregarded. However, a notable complication arises if the distribution of the fluorophores in the sample is not uniform. Figure 5(a) shows a representative case, where the fluorophores are concentrated primarily in distinct features, such as bacterial colonies expressing fluorescent proteins [13]. The goal of the multiphoton imaging experiments is, in this case, to determine the total fluorescence emitted by each feature, which are randomly distributed over the entire FOV. Therefore, we need to augment our model by including the illumination of each feature. Figure 5(b) provides the x-y view of the model sample, along with the outline of the asymmetric Gaussian beam and the corresponding raster scan pattern. The fluorescent features are depicted as circles, where the radius is larger than the image pixel size, but less than the illumination spot size.

 figure: Fig. 5

Fig. 5 The object plane of a 2PEF imaging setup. The illuminating laser spot is raster scanned over the surface of the sample. (a) Side view. (b) Front view.

Download Full Size | PDF

If the sample is homogeneous, or if the pulse repetition rate is high, then the ROIm would not need to account for the details of the illumination pattern. In our case, however, we need to calculate explicitly how much fluorescence is emitted by each feature excited with each pulse. To facilitate this, we divide the 10x10 cm2 FOV into an illumination matrix comprised of imax elements in the x direction and jmax elements in the y direction, where each element {i, j} represents one excitation beam position. The number of fluorophores excited in the kth feature may then be then expressed as:

(N2PA)k=  σ2PA2i=0imax j=0jmax Ck(x,y)IL2(xxi,yyj,t)dxdydt,
where the photon flux IL(x, y) is given by Eq. (22) and Ck(x,y) is the effective fluorophore concentration in the kth feature. Details of the numerical simulation are given in the Appendix. Briefly, we start by creating a random pattern of fluorescent features to be detected, and determine the optimal beam area, ΔS, corresponding to the average power value, Pave, using Eq. (23). We set the Gaussian beam parameters such that ΔxΔy = ΔS, and the degree of asymmetry matches the experimental conditions. Because the FOV is constant, the choice of imax and jmax determines the raster step size in the x- and y-direction, respectively. We now calculate the total number of fluorescence photons detected from each feature using Eq. (24) and (20) for a range of different scan speeds, or equivalently, different imaging matrix sizes, which allows us to determine when the minimum SNR per feature is achieved. The results are shown as the blue line in Fig. 4(a), and the corresponding parameter values are listed in the figure caption.

At this point we may conclude that, given an ideal, noise-free 1 kHz amplified femtosecond laser, one should achieve rather fast 2PEF imaging of large area samples at a rate of at least ROIm ~1 cm2 s−1, even at moderate average powers < 0.1 - 1 W.

It is also of interest to estimate how a 100 MHz source would perform under similar conditions. This simulation is shown as the dashed line in Fig. 4(a). Even though relatively high average power > 1W will be needed to achieve ROIm ~0.1 cm2 s−1, this type of illumination source may also be potentially considered, provided that issues associated with the requirement of rapid scanning of a tightly-focused beam over a large area without loss of the beam position accuracy can be resolved.

Figure 4(b) shows the maximum imaging rate obtained by applying the same calculation to three-photon excitation of fluorescence with different σ3PA values. Since a further increase of the peak photon flux per pulse is not practical, useful imaging rates may be achieved only for fluorophores with exceptionally large 3PA cross sections, on the order of 10−80 cm6 s2 or more.

2.3 Experimental implementation of 2PEF imaging with wide FOV illumination

Figure 6 shows the principle of the 2PEF imaging experimental setup, which was described, in part, in [13,28]. The output beam of a 1 kHz pulse rate femtosecond laser is shaped by two lenses (L1, L2) and is scanned over the area of a standard 10-cm diameter Petri dish using a bi-axial motorized mirror mount (Zaber T-OMG). The fluorescence emitted from the sample is collected with an objective lens and is detected with a TE-cooled CCD camera (Hamamatsu C4742-98) with an estimated image resolution of 200 μm. A stack of band-pass filters (F1) placed in front of the camera objective cut off scattered laser light. A reference photodiode (PD) was used to monitor and correct for long-term (~minutes) changes of the laser power. As the model sample, we use E. coli colonies expressing EGFP-type fluorescent proteins [13,28]. The beam steering and data collection were controlled by a LabView routine, and the data was analyzed using custom MatLab code as described in [13].

 figure: Fig. 6

Fig. 6 Schematic of the wide FOV femtosecond 2PEF imaging setup. L1, cylindrical lens; L2, spherical lens; PD, photodiode; SM, scanning mirror; F1, fluorescence band-pass filters.

Download Full Size | PDF

The symbols in Fig. 4(a) show the maximum imaging rate obtained with three different 1 kHz femtosecond lasers sources: (a) Clark MXR CPA-1000 Ti:Sapphire regenerative amplifier (solid circle); (b) Coherent Legend-HE Ti:Sapphire regenerative amplifier (empty rectangle) and (c) Light Conversion TOPAS-C optical parametric amplifier (empty circle). In both regenerative amplifier laser systems, the pulse-to-pulse instability was, RMSlaser ~0.1. Because the maximum average power of the first system was limited to Pave ~500 mW at λL = 790 nm, the resulting maximum imaging rate was about ~0.08 cm2 s−1, while the maximum average power of the second laser system, Pave ~1300 mW, allowed, at the same excitation wavelength λL = 790 nm, about twice as fast imaging rate, ~0.2 cm2 s−1. In the third case, the wavelength was converted by OPA to λL = 925 nm, which is closer to the peak of the 2PA spectrum of EGFP than λL = 790 nm [13]. However, because the OPA had less maximum output power, Pave ~80 mW, and also because this source had more pulse-to-pulse instability, RMSlaser > 0.15, the maximum imaging rate was lower, ~0.007 cm2 s−1. Figure 7 shows plots of the corresponding beam profiles. The beam area was determined in all three cases according to Eq. (23), but the aspect ratio, i.e. Δx vs. Δy was chosen based on convenience of steering the beam over the sample.

 figure: Fig. 7

Fig. 7 Contour plots of the laser beam at the sample. (a) Clark MXR CPA-1000 beam, λL = 790 nm, Pave = 0.5 W. (b) Coherent Legend-HE beam, λL = 790 nm, Pave = 1.3 W. (c) Light Conversion TOPAS-C beam, λL = 925 nm, Pave = 0.08 W.

Download Full Size | PDF

3. Conclusions

We have analyzed the performance of femtosecond multiphoton excited fluorescence imaging systems using different illumination- and focusing modalities, while taking into account the minimum required SNR and dominant sources of noise, such as photon shot noise, laser pulse energy fluctuations and fluorescence detection noise. In the case of a tightly-focused beam in the coaxial arrangement, using a 100 MHz illumination source to produce 2PEF, the maximum rate of imaging is limited by the fluorescence photon shot noise, by light-induced damage to the sample, and by saturation of the two-photon transition, and may reach about ~10−1 cm2 s−1, depending on the 2PA cross section and concentration of the fluorophores. In the case of coaxial 3PEF imaging with a 1 MHz illumination source, the effect of saturation limits the maximum imaging rate to about ~10−2 cm2 s−1. However, due to aberrations, in a coaxial illumination system it is difficult to access a FOV much larger than a few mm2. To achieve wide FOV 2PEF imaging, up to 10x10 cm2, one can take advantage of ~kHz pulse rate femtosecond amplifiers and use separated illumination and fluorescence imaging beam paths. We show that the highest rate of imaging, up to 1 cm2 s−1, may be achieved with a virtually noise-free amplified laser, and is limited just by the camera detection noise. If the laser pulse-to-pulse fluctuations are significant, then the maximum imaging rate decreases due to time averaging of the signal, leading to a maximum ROIm ~0.1 - 1 cm2 s−1. We perform experiments using three different 1 kHz pulse rate femtosecond lasers with different characteristics and find good agreement with our theoretical estimates.

Our results may be used to optimize multiphoton imaging systems that are designed to perform challenging tasks such as high throughput screening of two-photon fluorescent proteins, detection of early-stage cancer, or real-time monitoring of the brain cell activity in behaving animals. Similar analysis to that performed here could benefit other novel imaging modalities such as photoacoustic imaging [29].

Appendix

Numerical evaluation of wide FOV 2PEF imaging

We define imax and jmax to be;

imax=Sh;jmax=gτ2PEF,
where Sh is defined as the number of steps taken by the scanning laser in the x direction and τ2PEF is the time it takes to scan the laser in the y direction. The position of the {i, j} laser pulse is then given by;
xi=idhSh;yj=jdvgτ2PEF,
where dh and dv are the dimensions of the FOV being imaged. To guarantee that the entire sample area is more or less uniformly illuminated, we determine the number of steps according to the empirical relation, Sh ~(10*dh)/(Δx), which allows sufficient overlap between the neighboring laser spots. With this choice of illumination matrix, the maximum imaging rate is given by:
ROImmax=FOVShτ2PEF.
In order to find the imaging rate, the time it takes to scan the laser over the sample, τ2PEF, is calculated for the two cases discussed in the text.

If the SNR is limited by the noise of the illuminating laser source, the exposure time, τ2PEF, is expressed in terms of the number of laser pulses that need to be averaged, npulses, in order to achieve the desired SNR.

τ2PEF=npulsesdvgΔy
In practice, the number of pulses that needs to be averaged, will depend on the experimental conditions and the noise statistics of the particular laser used. We take an approach, where we keep adjusting τ2PEF until the desired SNR is reached. In the experimental systems described here, npulses = 200 and npulses = 400 was enough to achieve SNR = 10 for the regenerative amplifier and OPA systems respectively.

For a low noise laser source, the exposure time is found in terms of the total noise contribution of the detection system and the number of fluorescence photons emitted by each fluorescent feature (e.g. colony of fluorescent protein expressing E. coli). Assuming that the fluorophore concentration inside the feature is constant, Eq. (24) can be written as:

(N2PA)k= Cσ2PA2i=0Sh j=0gτ2PEF dk(x,y)IL2(xdhShi,ydvgτ2PEFj,t)dxdydt,
where dk(x,y) is the thickness of the kth colony, which itself may be expressed assuming a Gaussian profile as:
dk(x,y)= d0,kexp[4ln(2)(xx0)2+(yy0)2Δrk2],
where Δrk is the diameter and d0,k is the maximum thickness at the center, {x0, y0}. Note that as long as x0, and y0 lie within the scan range of the laser (dv = dh = 10 cm), the position of the colony does not affect the calculation. The volume of the kth colony is given by:
Vk= d0,kexp[4ln(2)(xx0)2+(yy0)2Δrk2]dxdy= d0,kΔrk2πln(16).
Using Eqs. (22), (30), and (31), the integral in Eq. (29) is solved and the number of the fluorophores excited in the kth colony is given by:
(N2PA)k= VkCσ2PA2ln(16)3/22π5/2τp(Paveλ2PAghcΔxΔy)2i=0Sh j=0gτ2PEF  exp{4ln(2)[(idhShx0)2Δrk2+Δx2ln(2)+(jdvgτ2PEFy0)2Δrk2+Δy2ln(2)]}[(ln(2)+Δrk2Δx2)(ln(2)+Δrk2Δy2)]1/2 
The number of 2PEF photons detected from a single colony is calculated using Eq. (20) with Eq. (32) substituted in for (N2PA)n.

To find the appropriate exposure time, τ2PEF, the number of 2PEF photons detected is set to equal some minimum necessary detected signal, (F2PEF)k,min, that will satisfy the condition, (F2PEF)k,min/δel = SNRmin > 10. This leads to the following relation:

SNRmin= σ2PAVkCηCOεeffln(16)3/2π5/2δelτp(PaveλLghcΔxΔy)2 0gτ2PEFexp{4ln(2)[(jdvgτ2PEFy0)2Δrk2+Δy2ln(2)]}[(ln(2)+Δrk2Δx2)(ln(2)+Δrk2Δy2)]1/2dj,
where the laser is scanned over the center of the colony, idk/Sh = x0, and the sum over the laser pulses is approximated as an integral. Completing the integral yields:
Foverlap=π2[ln(2)+Δrk2Δx2]1/22ln(2)Δy{Erf[2y0(Δy2+Δrk2ln(2))1/2]Erf[2(y0dv)(Δy2+Δrk2ln(2))1/2]},
We solve Eq. (33) for τ2PEF. The appropriate exposure time for a detection noise limited wide FOV imaging system is thus given, in terms of an overlap function, Foverlap, by:
τ2PEF= SNRminδeldvτpgVkCσ2PAηCOεeff(hcΔxΔyPaveλL)2Foverlap(dv,Δx,Δy,y0,Δrk).
Assuming the detection system uses 100 pixels to image one colony and there is roughly 20 counts of noise per pixel, we estimate that the total noise equals, δel = 2,000.

Equation (27) was then numerically evaluated using MATLAB software and the parameter values as described in the text. Different values of τ2PEF were found using Eq. (28) and (35), which then gave the maximum imaging rates presented in Fig. 5(a).

Acknowledgments

This work was supported by NIH grants R01GM098083 and 1-U01-NS094246. We would like to thank Thomas E. Hughes for loan of the CCD camera and preparing the biological test samples. We acknowledge Ting Xu for his assistance in performing the experiments.

References and links

1. S. Bovetti, C. Moretti, and T. Fellin, “Mapping brain circuit function in vivo using two-photon fluorescence microscopy,” Microsc. Res. Tech. 77(7), 492–501 (2014). [CrossRef]   [PubMed]  

2. D. M. Chudakov, M. V. Matz, S. Lukyanov, and K. A. Lukyanov, “Fluorescent proteins and their applications in imaging living cells and tissues,” Physiol. Rev. 90(3), 1103–1163 (2010). [CrossRef]   [PubMed]  

3. W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248(4951), 73–76 (1990). [CrossRef]   [PubMed]  

4. B. F. Grewe, D. Langer, H. Kasper, B. M. Kampa, and F. Helmchen, “High-speed in vivo calcium imaging reveals neuronal network activity with near-millisecond precision,” Nat. Methods 7(5), 399–405 (2010). [CrossRef]   [PubMed]  

5. B. F. Grewe, F. F. Voigt, M. van ’t Hoff, and F. Helmchen, “Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens,” Biomed. Opt. Express 2(7), 2035–2046 (2011). [CrossRef]   [PubMed]  

6. E. E. Hoover and J. A. Squier, “Advances in multiphoton microscopy technology,” Nat. Photonics 7(2), 93–101 (2013). [CrossRef]   [PubMed]  

7. A. H. Marblestone, B. M. Zamft, Y. G. Maguire, M. G. Shapiro, T. R. Cybulski, J. I. Glaser, D. Amodei, P. B. Stranges, R. Kalhor, D. A. Dalrymple, D. Seo, E. Alon, M. M. Maharbiz, J. M. Carmena, J. M. Rabaey, E. S. Boyden, G. M. Church, and K. P. Kording, “Physical principles for scalable neural recording,” Front. Comput. Neurosci. 7, 137 (2013). [CrossRef]   [PubMed]  

8. A. Maruyama, Y. Oshima, H. Kajiura-Kobayashi, S. Nonaka, T. Imamura, and K. Naruse, “Wide field intravital imaging by two-photon-excitation digital-scanned light-sheet microscopy (2p-DSLM) with a high-pulse energy laser,” Biomed. Opt. Express 5(10), 3311–3325 (2014). [CrossRef]   [PubMed]  

9. P. T. So, C. Y. Dong, B. R. Masters, and K. M. Berland, “Two-photon excitation fluorescence microscopy,” Annu. Rev. Biomed. Eng. 2(1), 399–429 (2000). [CrossRef]   [PubMed]  

10. K. Svoboda and R. Yasuda, “Principles of two-photon excitation microscopy and its applications to neuroscience,” Neuron 50(6), 823–839 (2006). [CrossRef]   [PubMed]  

11. C. Xu, W. Zipfel, J. B. Shear, R. M. Williams, and W. W. Webb, “Multiphoton fluorescence excitation: new spectral windows for biological nonlinear microscopy,” Proc. Natl. Acad. Sci. U.S.A. 93(20), 10763–10768 (1996). [CrossRef]   [PubMed]  

12. J. R. Starkey, N. S. Makarov, M. Drobizhev, and A. Rebane, “Highly sensitive detection of cancer cells using femtosecond dual-wavelength near-IR two-photon imaging,” Biomed. Opt. Express 3(7), 1534–1547 (2012). [CrossRef]   [PubMed]  

13. C. R. Stoltzfus, L. M. Barnett, M. Drobizhev, G. Wicks, A. Mikhaylov, T. E. Hughes, and A. Rebane, “Two-photon directed evolution of green fluorescent proteins,” Sci. Rep. 5, 11968 (2015). [CrossRef]   [PubMed]  

14. N. S. Makarov, M. Drobizhev, and A. Rebane, “Two-photon absorption standards in the 550-1600 nm excitation wavelength range,” Opt. Express 16(6), 4029–4047 (2008). [CrossRef]   [PubMed]  

15. J. P. Zinter and M. J. Levene, “Maximizing fluorescence collection efficiency in multiphoton microscopy,” Opt. Express 19(16), 15348–15362 (2011). [CrossRef]   [PubMed]  

16. R. W. Boyd, Nonlinear Optics (Academic press, 2003).

17. R. L. Sutherland, Handbook of Nonlinear Optics (CRC press, 2003).

18. G. Y. Fan, H. Fujisaki, A. Miyawaki, R.-K. Tsay, R. Y. Tsien, and M. H. Ellisman, “Video-rate scanning two-photon excitation fluorescence microscopy and ratio imaging with cameleons,” Biophys. J. 76(5), 2412–2420 (1999). [CrossRef]   [PubMed]  

19. A. M. D. Lee, H. Wang, Y. Yu, S. Tang, J. Zhao, H. Lui, D. I. McLean, and H. Zeng, “In vivo video rate multiphoton microscopy imaging of human skin,” Opt. Lett. 36(15), 2865–2867 (2011). [CrossRef]   [PubMed]  

20. K. B. Clancy, A. C. Koralek, R. M. Costa, D. E. Feldman, and J. M. Carmena, “Volitional modulation of optically recorded calcium signals during neuroprosthetic learning,” Nat. Neurosci. 17(6), 807–809 (2014). [CrossRef]   [PubMed]  

21. F. F. Voigt, J. L. Chen, R. Krueppel, and F. Helmchen, “A modular two-photon microscope for simultaneous imaging of distant cortical areas in vivo,” Proc. SPIE 9329, 93292C (2015). [CrossRef]  

22. E. Tal, D. Oron, and Y. Silberberg, “Improved depth resolution in video-rate line-scanning multiphoton microscopy using temporal focusing,” Opt. Lett. 30(13), 1686–1688 (2005). [CrossRef]   [PubMed]  

23. N. G. Horton, K. Wang, D. Kobat, C. G. Clark, F. W. Wise, C. B. Schaffer, and C. Xu, “In vivo three-photon microscopy of subcortical structures within an intact mouse brain,” Nat. Photonics 7(3), 205–209 (2013). [CrossRef]   [PubMed]  

24. L.-C. Cheng, N. G. Horton, K. Wang, S.-J. Chen, and C. Xu, “Measurements of multiphoton action cross sections for multiphoton microscopy,” Biomed. Opt. Express 5(10), 3427–3433 (2014). [CrossRef]   [PubMed]  

25. M. Drobizhev, A. Rebane, Z. Suo, and B. Spangler, “One-, two- and three-photon spectroscopy of π-conjugated dendrimers: cooperative enhancement and coherent domains,” J. Lumin. 111(4), 291–305 (2005). [CrossRef]  

26. J. N. Stirman, I. T. Smith, M. W. Kudenov, and S. L. Smith, “Wide field-of-view, twin-region two-photon imaging across extended cortical networks,” bioRxiv011320 (2014).

27. P. S. Tsai, C. Mateo, J. J. Field, C. B. Schaffer, M. E. Anderson, and D. Kleinfeld, “Ultra-large field-of-view two-photon microscopy,” Opt. Express 23(11), 13833–13847 (2015). [CrossRef]   [PubMed]  

28. C. Stoltzfus, L. Barnett, A. Rebane, T. Hughes, M. Drobizhev, G. Wicks, and A. Mikhailov, “A multidimensional screening method for the selection of two-photon enhanced fluorescent proteins,” Proc. SPIE 8956, 895611 (2014). [CrossRef]  

29. S.-Y. Lee, Y.-H. Lai, K.-C. Huang, Y.-H. Cheng, T.-F. Tseng, and C.-K. Sun, “In vivo sub-femtoliter resolution photoacoustic microscopy with higher frame rates,” Sci. Rep. 5, 15421 (2015). [CrossRef]   [PubMed]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (7)

Fig. 1
Fig. 1 Schematic of a generic coaxial illumination- and fluorescence detection microscope. (a) The laser beam emanates from the output aperture of a high-NA objective and converges onto the focus voxel (dashed line); d is the sample thickness, ΔzR is the Rayleigh length of the focus, λL is the laser wavelength, and λF is the fluorescence wavelength. (b) Simplified raster scan pattern, where the step size is equal to the lateral size of the focus spot, Δx.
Fig. 2
Fig. 2 Solid lines - Dependence of FPSmax on the average power, Pave. (a) 2PEF FPSmax calculated using Eq. (13) with g = 100 MHz and λL = 790 nm. (b) 3PEF FPSmax calculated using Eq. (18) with g = 1 MHz and λL = 1700 nm. Different σ2PA and σ3PA values are indicated above the corresponding lines. The common parameters used in both these calculations are: τp = 150 fs, NA = 0.9, n = 1, C = 2x1021 m−3, and ηMO*εeff = 0.1. The red dotted horizontal line represents the saturation limit, when 25% of the fluorophores are excited during one pulse. Imaging rates estimated from literature data are represented by: (a): (○) – Denk et al. [3], (●) – Fan et al. [18], (■) – Lee et al. [19], (□) – Clancy et al. [20], (◊) - Voigt et al. [21]; (▲) – Tal et al. [22]; (∆) – Murayama et al. [8]. (b): Horton et al. [23] (♦) – at sample surface, (◊) –0.8 mm in sample, and (▼)-0.8-1.4 mm in sample.
Fig. 3
Fig. 3 The image of the fluorescence on a camera in a wide FOV imaging setup. The square illuminating laser spot (red squares) is raster scanned over the surface of a sample with a homogeneous distribution of fluorophores. (a) Side view of the sample. d is the thickness of the sample and ΔzR is the Rayleigh length of the focused laser. (b) Front view of the sample. Δy and Δx are the vertical and horizontal laser spot sizes respectively. For simplicity a square laser illumination spot with even illumination intensity is shown. The camera pixels have height Δypix, and width Δxpix.
Fig. 4
Fig. 4 The rate of imaging with amplified laser pulses as a function of average power (lower horizontal axis). Top axis shows the corresponding optimized illumination spot area. (a) Solid red lines - 2PEF in a spatially homogeneous sample with different laser noise levels. SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 790 nm; Solid blue line - 2PEF in a spatially inhomogeneous sample comprised of fluorescent features with an ideal noise-less laser. δel = 2,000, SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 790 nm, σ2PA = 10 GM, Δrk = 1 mm, Vk = 1.1 mm3. Dashed black line – the same sample parameters as the blue line but in the case of a femtosecond oscillator with g = 100 MHz. Comparison to experimental implementations: (●) - Clark MXR CPA-1000 Ti:Sapphire regenerative amplifier; (□) - Coherent Legend-HE Ti:Sapphire regenerative amplifier ;(○) - Light Conversion TOPAS-C optical parametric amplifier. (b) 3PEF in a spatially inhomogeneous sample comprised of fluorescent features with an ideal noise-less laser in the case of different three-photon cross sections. δel = 2,000, SNRmin = 10, g = 1 kHz, τp = 150 fs, C = 2x1021 m−3, ηCO*εeff = 1x10−4, λL = 1700 nm, Δrk = 1 mm, Vk = 1.1 mm3.
Fig. 5
Fig. 5 The object plane of a 2PEF imaging setup. The illuminating laser spot is raster scanned over the surface of the sample. (a) Side view. (b) Front view.
Fig. 6
Fig. 6 Schematic of the wide FOV femtosecond 2PEF imaging setup. L1, cylindrical lens; L2, spherical lens; PD, photodiode; SM, scanning mirror; F1, fluorescence band-pass filters.
Fig. 7
Fig. 7 Contour plots of the laser beam at the sample. (a) Clark MXR CPA-1000 beam, λL = 790 nm, Pave = 0.5 W. (b) Coherent Legend-HE beam, λL = 790 nm, Pave = 1.3 W. (c) Light Conversion TOPAS-C beam, λL = 925 nm, Pave = 0.08 W.

Equations (35)

Equations on this page are rendered with MathJax. Learn more.

N 2PA =   σ 2PA 2 V C(x,y,z)[ I L 2 (x,y,z,t)dt ]dxdydz ,
V eff   53.0 ω 0 4 λ L ,
N 2PA 1.24 C σ 2PA λ L 8ln2 τ p ( P ave πghc ) 2 ,
N 2PA  1.30 V eff C[ σ 2PA τ p ( P ave N A 2 ghc λ L ) 2 ],
( N 2PA ) n = g τ dw N 2PA ,
M FOV =  FOV ΔS .
FPS= 1 τ dw M FOV .
( F 2PEF ) n = η MO ε eff ( N 2PA ) n  ,
η MO = 1 2 [ 1 ( 1 N A 2 n 2 ) 1/2 ],
ε eff =  Φ FL ( λ F ) ϕ F ( λ F ) κ D ( λ F )d λ F ,
SNR= ( F 2PEF ) n δ shot 2 + δ laser 2 + δ el 2 ,
SNR= ( F 2PEF ) n .
FP S max 1.24 η MO ε eff C σ 2PA λ L 8ln(2) g τ p M FOV SN R min 2 ( P ave πhc ) 2 .
d ρ f (t) dt =  1 2 σ 2PA I L 2 ( 12 ρ f (t) )  γ f0 ρ f (t),
ROIm=FOV×FPS.
N 3PA =   σ 3PA 3 V C(x,y,z)[ I L 3 (x,y,z,t)dt ]dxdydz ,
N 3PA 0.60C σ 3PA 8ln2 ( λ L ω 0 τ p ) 2 ( P ave πghc ) 3 .
FP S max 0.60 η MO ε eff C σ 3PA 8ln(2) M FOV SN R min 2 ( λ L ω 0 τ p g ) 2 ( P ave πhc ) 3 .
( N 2PA ) n =   V eff C σ 2PA τ dw 2g τ p ( P ave λ L hcΔS ) 2 ,
( F 2PEF ) n = η CO ε eff ( N 2PA ) n ,
ROIm max gΔS 4RM S laser 2 ( SNR ) min 2 .
I L =  2 [ ln(16) ] 1/2 λ L P ave π 3/2 ΔxΔy τ p ghc exp[ 4ln(2)( t 2 τ p 2 )2( y 2 Δ y 2 + x 2 Δ x 2 ) ],
β s = P ave g τ p ΔS const.,
( N 2PA ) k =   σ 2PA 2 i=0 i max   j=0 j max   C k (x,y) I L 2 (x x i ,y y j ,t)dxdydt ,
i max = S h ; j max =g τ 2PEF ,
x i =i d h S h ; y j =j d v g τ 2PEF ,
RO Im max = FOV S h τ 2PEF .
τ 2PEF = n pulses d v gΔy
( N 2PA ) k =  C σ 2PA 2 i=0 S h   j=0 g τ 2PEF   d k (x,y) I L 2 (x d h S h i,y d v g τ 2PEF j,t)dxdydt,
d k ( x,y )=  d 0,k exp[ 4ln(2) (x x 0 ) 2 + (y y 0 ) 2 Δ r k 2 ],
V k =  d 0,k exp[ 4ln(2) (x x 0 ) 2 + (y y 0 ) 2 Δ r k 2 ]dxdy=  d 0,k Δ r k 2 π ln(16) .
( N 2PA ) k =  V k C σ 2PA 2ln (16) 3/2 2 π 5/2 τ p ( P ave λ 2PA ghcΔxΔy ) 2 i=0 S h   j=0 g τ 2PEF    exp{ 4ln(2)[ ( i d h S h x 0 ) 2 Δ r k 2 +Δ x 2 ln(2) + ( j d v g τ 2PEF y 0 ) 2 Δ r k 2 +Δ y 2 ln(2) ] } [ (ln(2)+ Δ r k 2 Δ x 2 )(ln(2)+ Δ r k 2 Δ y 2 ) ] 1/2  
SN R min =  σ 2PA V k C η CO ε eff ln (16) 3/2 π 5/2 δ el τ p ( P ave λ L ghcΔxΔy ) 2   0 g τ 2PEF exp{ 4ln(2)[ ( j d v g τ 2PEF y 0 ) 2 Δ r k 2 +Δ y 2 ln(2) ] } [ (ln(2)+ Δ r k 2 Δ x 2 )(ln(2)+ Δ r k 2 Δ y 2 ) ] 1/2 dj,
F overlap = π 2 [ ln(2)+ Δ r k 2 Δ x 2 ] 1/2 2ln(2)Δy{ Erf[ 2 y 0 ( Δ y 2 + Δ r k 2 ln(2) ) 1/2 ]Erf[ 2( y 0 d v ) ( Δ y 2 + Δ r k 2 ln(2) ) 1/2 ] } ,
τ 2PEF =  SN R min δ el d v τ p g V k C σ 2PA η CO ε eff ( hcΔxΔy P ave λ L ) 2 F overlap ( d v ,Δx,Δy, y 0 ,Δ r k ).
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.