Abstract
We explored a method to evaluate imaging performance for the optimal assembly of an endoscopic miniature lens and a sensor constituting an imaging module at the distal end of gastrointestinal endoscopy. For the assembly of the imaging module, the image sensor was precisely located at the focal plane when collimated light passed through the endoscopic lens. As another method, the distance between the lens and sensor was adjusted to obtain the highest focus index from images measured the star chart of the International Organization for Standardization (ISO) standard at various positions. We analyzed the slanted-edge modulation transfer function (MTF), corresponding depth of field, and number of line pairs for MTF 50% and 20% at each working distance within the range of 5–100 mm for imaging modules assembled in different ways. Assembly conditions of the imaging module with better MTF performance were defined for each working distance range of 5–30 mm and 30–100 mm, respectively. In addition to the MTF performance, the focus index of each assembled module was also compared. In summary, we examined the performance of imaging modules assembled with different methods within the suggested working distance and tried to establish the optimal assembly protocol.