The absolute frequency of the clock transition at 237 nm was measured with an accuracy of 1.8 parts in . Using a phase-coherent frequency chain, we compared the transition with a methane-stabilized He–Ne laser at , which was calibrated against an atomic cesium fountain clock. A frequency gap of 37 THz at the fourth harmonic of the He–Ne standard was bridged by a frequency comb generated by a mode-locked femtosecond laser. The frequency of the clock transition was found to be 1 267 402 452 899.92 (0.23) kHz, the accuracy being limited by the uncertainty of the He–Ne laser reference. This result represents an improvement in accuracy of more than 2 orders of magnitude over previous measurements of the line and now stands as what is to our knowledge the most accurate measurement of an optical transition in a single ion.s
© 2000 Optical Society of America
Full Article | PDF ArticleMore Like This
Jörn Stenger, Christian Tamm, Nils Haverkamp, Stefan Weyers, and Harald R. Telle
Opt. Lett. 26(20) 1589-1591 (2001)
K. L. Corwin, I. Thomann, T. Dennis, R. W. Fox, W. Swann, E. A. Curtis, C. W. Oates, G. Wilpers, A. Bartels, S. L. Gilbert, L. Hollberg, N. R. Newbury, S. A. Diddams, J. W. Nicholson, and M. F. Yan
Opt. Lett. 29(4) 397-399 (2004)
Th. Udem, J. Reichert, R. Holzwarth, and T. W. Hänsch
Opt. Lett. 24(13) 881-883 (1999)