Intravascular optical coherence tomography imaging at 3200 frames per second

Not Accessible

Your library or personal account may give you access

Abstract

We demonstrate intravascular optical coherence tomography (OCT) imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system, with an imaging depth of 3.7 mm (in air), is based on a Fourier domain mode locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter is 1.1 mm at the tip. Ex vivo images of human coronary artery (78.4 mm length) were acquired at a pullback speed of 100mm/s. True 3D volumetric imaging of the entire artery, with dense and isotropic sampling in all dimensions, was performed in <1 second acquisition time.

© 2013 Optical Society of America

Full Article  |  PDF Article
More Like This
Heartbeat OCT: in vivo intravascular megahertz-optical coherence tomography

Tianshi Wang, Tom Pfeiffer, Evelyn Regar, Wolfgang Wieser, Heleen van Beusekom, Charles T. Lancee, Geert Springeling, Ilona Krabbendam, Antonius F.W. van der Steen, Robert Huber, and Gijs van Soest
Biomed. Opt. Express 6(12) 5021-5032 (2015)

High frame-rate intravascular optical frequency-domain imaging in vivo

Han Saem Cho, Sun-Joo Jang, Kyunghun Kim, Alexey V. Dan-Chin-Yu, Milen Shishkov, Brett E. Bouma, and Wang-Yuhl Oh
Biomed. Opt. Express 5(1) 223-232 (2014)

Real-time volumetric lipid imaging in vivo by intravascular photoacoustics at 20 frames per second

Min Wu, Geert Springeling, Matija Lovrak, Frits Mastik, Sophinese Iskander-Rizk, Tianshi Wang, Heleen M. M. van Beusekom, A. F. W. van der Steen, and Gijs Van Soest
Biomed. Opt. Express 8(2) 943-953 (2017)

Supplementary Material (1)

Media 1: AVI (7841 KB)     

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Figures (3)

You do not have subscription access to this journal. Figure files are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved