Abstract
An interband cascade laser (ICL) operating at 3.7 μm has been used to perform multimode absorption spectroscopy, MUMAS, at scan rates up to 10 kHz. Line widths of individual modes in the range 10–80 MHz were derived from isolated lines in the MUMAS signatures of HCl. MUMAS data for methane covering a spectral range of 30 nm yielded a detection level of 30 μbar·m for 1 s measurement time at 100 Hz. Simultaneous detection of methane, acetylene, and formaldehyde in a gas mixture containing all three species is reported.
© 2015 Optical Society of America
Full Article | PDF ArticleMore Like This
Carl Borgentun, Clifford Frez, Ryan M. Briggs, Mathieu Fradet, and Siamak Forouhar
Opt. Express 23(3) 2446-2450 (2015)
Lukasz A. Sterczewski, Jonas Westberg, Mahmood Bagheri, Clifford Frez, Igor Vurgaftman, Chadwick L. Canedy, William W. Bewley, Charles D. Merritt, Chul Soo Kim, Mijin Kim, Jerry R. Meyer, and Gerard Wysocki
Opt. Lett. 44(8) 2113-2116 (2019)
Lukasz A. Sterczewski, Mahmood Bagheri, Clifford Frez, Chadwick L. Canedy, Igor Vurgaftman, Mijin Kim, Chul Soo Kim, Charles D. Merritt, William W. Bewley, and Jerry R. Meyer
Opt. Lett. 44(23) 5828-5831 (2019)