Abstract
We demonstrate the feasibility of resetting and reusing dosimeters exploiting the measurement of the infrared radiation-induced attenuation (IR-RIA) in phosphosilicate optical fibers (OFs) to provide point or distributed dose measurements in radiation environments. To bleach the room temperature stable IR-RIA, we used the photobleaching (PB) phenomenon. The PB efficiency was evaluated for different wavelengths in the [400–1100] nm range. The best identified PB resetting condition consists in using a continuous-wave Argon-ion laser at 514 nm. This treatment successfully bleached ${\sim}{{97}}\%$ of the IR-RIA at 1550 nm of a 30 m-long P-doped single mode optical fiber X-ray irradiated at a dose of 100 Gy. Successive re-irradiations of the same OF sample, regenerated after each run, confirm that the dosimeter keeps the same calibration curve during the whole process.
© 2020 Optical Society of America
Full Article | PDF ArticleMore Like This
Alexander L. Tomashuk, Mikhail V. Grekov, Sergei A. Vasiliev, and Vyacheslav V. Svetukhin
Opt. Express 22(14) 16778-16783 (2014)
Qiancheng Zhao, Qun Hao, Yanhua Luo, Xiaochong Li, Shiyu Cui, Fengze Tan, Changyuan Yu, and Gang-Ding Peng
Opt. Lett. 45(19) 5389-5392 (2020)
Qiancheng Zhao, Qun Hao, Yanhua Luo, and Gang-Ding Peng
Opt. Lett. 45(5) 1152-1155 (2020)