Abstract
This study develops a handheld optical coherence tomography angiography (OCTA) system that uses a high-speed (200 kHz) swept laser with a dual-reference common-path configuration for stable and fast imaging. The common-path design automatically avoids polarization and dispersion mismatches by using one circulator as the primary system element, ensuring a cost-effective and compact design for handheld probe use. With its stable envelope (i.e., sub-µm shifts) and phase variation (corresponding to nm changes in axial displacement), the minimum detectable flow velocity is ∼ 0.08 mm/s in our experiment, which gives the common-path setup a high potential for application in a handheld OCTA system for clinical skin screening. In vivo skin structures and microvasculature networks on the dorsum of the hand and cheek of a healthy human are imaged successfully.
© 2023 Optica Publishing Group
Full Article | PDF ArticleMore Like This
Tianyu Zhang, Simon Shepherd, Zhihong Huang, Michaelina Macluskey, and Chunhui Li
Opt. Lett. 48(18) 4857-4860 (2023)
Naixing Huang, Tristan T. Hormel, Guangru B. Liang, Xiang Wei, Yukun Guo, Siyu Chen, and Yali Jia
Opt. Lett. 49(5) 1201-1204 (2024)
Kaicheng Liang, Osman O. Ahsen, Zhao Wang, Hsiang-Chieh Lee, Wenxuan Liang, Benjamin M. Potsaid, Tsung-Han Tsai, Michael G. Giacomelli, Vijaysekhar Jayaraman, Hiroshi Mashimo, Xingde Li, and James G. Fujimoto
Opt. Lett. 42(16) 3193-3196 (2017)