Abstract
In this paper, we demonstrated the design and experimental results of the near-infrared lab-on-a-chip optical biosensor platform that monolithically integrates the MRR and the on-chip spectrometer on the silicon-on-insulator (SOI) wafer, which can eliminate the external optical spectrum analyzer for scanning the wavelength spectrum. The symmetric add-drop MRR biosensor is designed to have a free spectral range (FSR) of ∼19 nm and a bulk sensitivity of ∼73 nm/RIU; then the drop-port output resonance peaks are reconstructed from the integrated spatial-heterodyne Fourier transform spectrometer (SHFTS) with the spectral resolution of ∼3.1 nm and the bandwidth of ∼50 nm, which results in the limit of detection of 0.042 RIU.
© 2023 Optica Publishing Group
Full Article | PDF ArticleMore Like This
Weiqing Cheng, Xiao Sun, Shengwei Ye, Bocheng Yuan, Yiming Sun, John H. Marsh, and Lianping Hou
Opt. Lett. 48(19) 5113-5116 (2023)
Thi Thuy Duong Dinh, Xavier Le Roux, Natnicha Koompai, Daniele Melati, Miguel Montesinos-Ballester, David González-Andrade, Pavel Cheben, Aitor V. Velasco, Eric Cassan, Delphine Marris-Morini, Laurent Vivien, and Carlos Alonso-Ramos
Opt. Lett. 47(4) 810-813 (2022)
Hugh Podmore, Alan Scott, Pavel Cheben, Aitor V. Velasco, Jens H. Schmid, Martin Vachon, and Regina Lee
Opt. Lett. 42(7) 1440-1443 (2017)