Sensitivity optimization of the one beam Z-scan technique and a Z-scan technique immune to nonlinear absorption

Open Access Open Access

Abstract

It is presented a criteria for selecting the optimum aperture radius for the one beam Z-scan technique (OBZT), based on the analysis of the transmittance of the aperture. It is also presented a modification to the OBZT by directly measuring the beam radius in the far field with a rotating disk, which allows to determine simultaneously the non-linear absorptive coefficient and non-linear refractive index, much less sensitive to wave front distortions caused by inhomogeneities of the sample with a negligible loss of signal to noise ratio. It is demonstrated its equivalence to the OBZT.

© 2013 Optical Society of America

1. Introduction

Since the publication of the one beam Z-scan technique (OBZT) by Bahae-Sheik et al. in 1989 [1], it has become very popular for determining nonlinear absorptive coefficient and refractive index of a thin sample, because its simplicity and sensitivity. The nonlinear refractive index is determined from the transmittance of the sample measured through an aperture in the far field as the sample is displaced around the focal position of a Gaussian beam propagating along z axis.

However, the aperture introduces important disadvantages, namely, it must be aligned with the axis of the beam (z axis) to measure on axis transmitted power (P), second, P is only a small fraction of the total laser power (P0), compromising the signal to noise ratio (SNR), and perhaps the most important of all, when the sample exhibits nonlinear absorption, there is no way to know whether changes in P due to changes in the refractive index or absorption when the sample is displaced, forcing to repeat the experiment without the aperture for measuring the nonlinear absorption and deduct it, as will be seen later.

Since then, several modifications have been proposed to improve the sensitivity of the OBZT, namely, modifying the beam profile to enhance the aperture transmittance [2], substituting the aperture [3] or eliminating the aperture [4] where the beam width is directly measured using a laser beam profiler based on CCD camera.

Although the effect of the aperture has been studied for Gaussian beams [5, 6] little attention has been paid to the selection of the aperture radius for optimizing the signal to noise ratio of the OBZT.

The present paper presents a criteria for selecting the optimum aperture radius, based on the analysis of the physical variable of the OBZT, it is presented also a Z-scan technique based on a direct measurement of the beam width in the far field, fast and cheap, it is demonstrated its equivalence, the signal to noise ratio of the OBZT with the optimum aperture radius is compared with the technique presented, such technique allows to measure simultaneously the nonlinear absorption and refractive index.

2. Description of the chopper-width technique

Many techniques have been developed for measuring the beam width of a Gaussian beam, such as the slit scan technique [7, 8], the pinhole technique [9], but among all, the most extended is the knife-edge technique [10, 11], because is an inexpensive, and accurate beam profiler. In this technique, a knife located at (ζ) eclipses the beam transversely to its propagation axis (z axis), the beam width (W(ζ)) is determined from the derivative plot of the transmitted power versus the knife-edge position. This method is simple but a bit tedious, slow and noise sensitive. These issues can be overcome with the technique described below.

The chopper-width measurement technique basically consists in to extract W(ζ), from the rise or fall time, of a electric signal [12] generated by switch- on and switch-off the beam light, with a slotted rotating disk (named chopper) at the position (ζ) [13]. This technique is alike to the knife-edge technique but faster, allowing real-time measurements. The width W(ζ) measured in this technique is larger than the obtained with the knife-edge technique, due the curved trajectory to cut the beam, rather than perpendicular. This drawback can be overcome calibrating the chopper-width technique, this is presented in the appendix at the end of this work. Let us consider a Gaussian beam propagating along the z axis impinging on a photoreceptor, the chopper is located at the ζ position where the W(ζ) is to be measured, like the chopper has l slots and rotates with constant angular frequency Ω0, as shown in Fig. 1. To each blade will take τ=θΩ0 seconds to sweep the angle θ subtended by the beam. If RW(ζ), τ is given by

τ2W(ζ)Ω0R
where R is the distance from the axis of rotation to the beam axis, therefore, Eq. (1) provides a way to measure W(ζ) from the time τ.

 figure: Fig. 1

Fig. 1 Front view of a slotted disk which rotates at a frequency Ω0 and eclipses a Gaussian beam shown in red. R is the distance from the rotation axis to the center of the beam, W is the beam width, and θ is the angle subtended by the spot beam.

Download Full Size | PDF

Then, it is proposed to use the width-measurement technique, based on Eq. (1), to obtain Z-scan curves (we shall name chopper Z-scan technique) like Tsigaridas [4], but in a simpler and inexpensive way. It is important to mention that until now, choppers have been used in the OBZT as modulation instruments [14, 15], in order to improve the detection of P, however, here the chopper is used as meter width. Before experimentally demonstrate the feasibility and advantages of this approach, let us present its justification and the comparison of the sensitivities of chopper and OBZT.

3. Equivalence and sensitivities

In the OBZT (on axis measurement) for CW radiation, the normalized transmittance (T(z)) of the aperture located at ζ fixed, is defined as [1]

T(z)=P(W(z),ρ0)P(W(),ρ0)
where z is the position of the nonlinear sample referred to the minimum waist position (z = 0) and the power transmitted by the aperture is given by
P(W,(z),ρ0)=P0(1exp(2ρ02W(z)2))
ρ0 is the aperture radius, W(z) and W(∞) = WL are the beam radius at the position of the aperture when the sample is located at z and far away (∞) from the focus (linear regime) respectively, P0 is the optical power of the beam incident to the aperture. Henceforth, we shall assume the implicit dependence of W on z in order to simplify the notation.

Taking into account that the reference is the unity, i.e. T(∞) = 1, then, the effect of the presence of sample is measured from this value. Also, it is easy to realize that ΔT(z)T() represents the relative change of the transmitted power in the linear regime, i.e. ΔPP(WL,ρ0), also this fact allows to establish the following relation

ΔT(z)T()=ΔPP0S1
where the relative optical change and linear transmittance are respectively defined as
ΔPP0=P(W,ρ0)P(WL,ρ0)P0
and
S=1exp(2ρ02WL2)

The power transmitted (Eq. (3)) is what physically is measured (physical variable), hence, the sensitivity of Z-scan technique is determined by the minimum value of this quantity can be discerned (or discriminated) due to the presence of noise in the measurement.

3.1. The signal of the OBZT

In a Z-scan experiment due to the presence of the nonlinear sample, the beam width changes from WL to W, therefore, the beam width change (ΔW) is

ΔW=WWL
this causes that the transmitted power (P), changes by ΔP, let us calculate the optimum value of ρ0 to detect ( ΔPΔW), the equation to be solved is
2ρ0WP(W,ρ0)=0
which gives the following result
ρ0=WL2
i.e., when the aperture radius is 70.7% of WL (using the criteria of 1e2 for measuring the beam width), it is achieved the highest sensitivity for detecting changes in P. To reinforce this assertion, in Fig. 2, it is shown an experimental plot of ΔP as function of the ratio ρ0WL, as can be seen in such plot the maximum sensibility effectively occurs when ρ0 ≅ 0.71WL, corroborating Eq. (9), the experimental details are given in section 4.

 figure: Fig. 2

Fig. 2 Sensitivity of an aperture of radius (ρ0) to detect changes in its transmitted power (ΔP = P(W, ρ0) − P(WL, ρ0)), when the incident beam width changes from (WLW).

Download Full Size | PDF

For this optimum aperture, Eq. (5) gives the following result

ΔPP0=exp(1)exp(WL2W2)
for small changes (ΔWWL), Eq. (10) can be approximated to
ΔPP02exp(1)(ΔWWL)
Eq. (11) is important because allows to relate beam width changes (ΔW) with changes in the transmitted power (Eq. (4)). Substituting Eq. (6), Eq. (9) and Eq. (11) into Eq. (4) gives
ΔT(z)T()1.16(ΔWWL)
Eq. (12) provides the equivalence of the chopper technique. The minus sign in Eq. (12), means that the chopper Z-scan curve is inverted with respect to the OBZT Z-scan, because when there is a positive increment of beam width (ΔW > 0) the beam intensity and therefore the aperture transmittance diminishes (ΔT < 0). In the following we shall work with ΔWWL instead of ΔT(z)T().

To conclude this subsection, using Eq. (12), substituting Eq. (9) into Eq. (6) and using the following result [1] (assuming the same approximations, i.e. an incident circular Gaussian beam, small incident power, small phase changes, and thin sample.)

ΔTpv0.406(1S)0.25|ΔΦ0|
the changes on-axis phase shift at the focus (ΔΦ0) can be related to the beam width changes, as follows:
ΔWpvWL0.273|ΔΦ0|
where ΔWpv is the difference peak-valley of the chopper Z-scan curve.

3.2. The signal for the chopper Z-scan

For the chopper Z-scan technique, the elapsed time τ for eclipsing the beam width (W) is given by Eq. (1), when W changes due to the presence of the sample, τ also changes by Δτ, given by:

Δτ=2Ω0RΔW
therefore, the minimum Δτ that noise allows to discriminate, determines, the minimum ΔW that can be resolved with this technique.

3.3. Calculus of noise power for both techniques

As mentioned before, the optical power is the physical variable to be measured for the OBZT through a photoreceptor system, it usually consists of a photodiode and a trans-impedance amplifier (see Fig. 3(a)), for this system the voltage Vph (which only depends on the sample position z) is:

Vph=RLP
RL is the amplifier zero frequency gain in Ohms, ℜ is the responsivity of the photodiode in AmpereWatt and P is the optical power incident to the photodiode in Watt.

 figure: Fig. 3

Fig. 3 A) One beam Z-scan set up, B) Chopper Z-scan set up.

Download Full Size | PDF

In the case of small P, the photoreceptor is limited by thermal noise [16], therefore, the signal to noise ratio ( SN) in this regime is

SN=RL(P)4kTRLΔf
where k is the Boltzmann constant, T is the temperature in Kelvin and Δf is the detection bandwidth of the photoreceptor in Hz given by [17]
Δf=1.572πRLCL
the expression in parentheses in Eq. (17) is the photo-current generated by the photodiode.

The minimum signal detectable occurs when ( SN=1), also, assuming stationary noise sources and non-memory measurements, then, the minimum change (ΔPmin) detectable is

ΔPmin=24kTRLΔfRL
to obtain an expression for the magnitude of the minimum change of width detectable ( ΔWminOB the superscript comes from One Beam), Eq. (19) is substituted into Eq. (11)
|ΔWminOBWL|=exp(1)4kTRLΔfRLP0
the expression under the square root symbol, represents the thermal noise power (Pn), of course, if other additive noise sources were present, Pn would represent the total noise power, hence,
|ΔWminOBWL|=2.72PnRLP0
the sensitivity of the OBZT is given by Eq. (21).

For the chopper Z-scan, the total laser power (P0) impinges on the photodiode, therefore, a lower gain value of RL is needed, lowering the noise power due to this cause, however, the bandwidth requirement (Δf) is greater than the OBZT, somewhat compensating the previous advantage, as shown in the following subsection.

3.3.1. Bandwidth need of the chopper z-scan

For the chopper technique, the incident optical power impinging on the photodiode Pch(t) is a periodic function of the time t, due to periodic eclipsing; Pch(t) is given by:

Pch(t)=P02(1erf(2Rsin(Ω0t)W))
let us define the following ratio
β=RW
for Ω0t ≪ 1 the raising part of Eq. (22) can be approximated to
Pch(t)P02(1erf(2βΩ0t))
the Fourier series of Pch(t) using the approximation (Eq. (24)) is
Pch(t)=122πm=012m+1exp(18(2m+1β)2)sin((2m+1)Ω0t)
if the series is cut up to m ≥ 2β (m integer) harmonic, the error is less than 0.6%, therefore the chopper Z-scan bandwidth need (Δfc) in Hz is:
Δfc=4βf0
where it has been used the fact that Ω0 = 2πf0 and Ω0t ≪ 1, hence, the photoreceptor must satisfy the following requirement
ΔfcΔf
where Δf is given in Eq. (18).

The Eq. (26) can be approximately obtained under the following reasoning: the eclipsing time τ was calculated in Eq. (1), therefore, the photoreceptor must at least be as fast as the eclipsing time, consequently, its bandwidth Δfc must be

Δfc1τ=Ω0β2
the following is also meet Ω0 = 2πf0 where f0 is the rotation frequency in Hz, thus, Eq. (28) can be expressed as
Δfπβf0

3.3.2. Minimum change of width detectable with the chopper Z-scan technique

The eclipsing time τ is the time elapsed for Vph(t) to vary from Vth1 to Vth2 as shown in Fig. 4, therefore, Vph(t) is compared with the threshold voltages Vth1 and Vth2, hence, Vph(t) fluctuations due to noise (n(t)) result in uncertainty in the comparison process (jitter), in this case the following relation is satisfied between the variance in the metric j of the time of the threshold crossing [18]

var(j(tc))var(n(tc))(dVph(tc)dt)2
where tc is the expected time of the threshold crossing and Vph(t) is the voltage delivered by the photoreceptor system. In order to compare both techniques let us assume that the same amplifier than the OBZT is used and Eq. (27) is satisfied, then,
Vph(t)RLPch(t)
Pch (t) is given by Eq. (22). For noise with zero mean, var (n(t)) = Pn[19], Pn is the noise power of n(t), therefore, equation (30) can be expressed as
var(j(tc))Pn(dvph(tc)dt)2
from Eq. (1) and Eq. (32) we can establish the following relation
var(W)Ω0R2Pn|dVph(tc)dt|
Note that dVph(t)dt is proportional to the Gaussian profile, referring to Eq. (33) and Fig. 4 the uncertainty in measuring ΔW is minimized if Vth1 and Vth2 are set as close as possible to the maximum value of the profile, hence, if they are established where the profile drops to ±WL3 (i.e. Vth1 ≅ 0.25Vph(t) and Vth2 ≅ 0.75Vph(t)), then from Eq. (1)
tc=±2WL3Ω0R
using Eq. (31), Eq. (24) and Eq. (34) the derivative of Vph is
|ddtVph(±tc)|=2πe89βΩ0(RLP0)
substituting Eq. (35) into Eq. (33) and taking into account that two comparisons are done (at Vth1 and Vth2), the minimum relative change of the beam width ( WminCh the superscript comes from Chopper) that this technique can resolve is
ΔWminChWL=3PnRLP0
comparing Eq. (36) with Eq. (21) under the same power of noise both techniques are almost equally sensitive, however, the chopper technique does not require an aperture, eliminating the need to align it with the beam axis and due to its larger bandwidth, a measurement is carried out in fraction of second, allowing to perform an average over a much larger sample under the same observation time, and perhaps the most important feature it is absorption immune.

 figure: Fig. 4

Fig. 4 Establishment of the threshold voltages Vth1 and Vth2 using e−2 criteria for measuring the beam width W from Vph(t) (black line) and the beam profile (red line).

Download Full Size | PDF

Before turning to the experimental results it should be noted, that for practical purposes, due to the physical variable for the chopper technique is time, then, according to Eq. (1), the relative changes in the width are equal to the relative changes in the rising time (i.e. ΔττL=ΔWWL), therefore it can be obtained a Z-scan curve plotting τ versus z, and to use Eq. (37) derived from Eq. (1) and Eq. (13) to estimate the phase change

ΔτpvτL0.273|ΔΦ0|
where Δτpv is the peak-valley difference of the Z-scan curve, and τL is the rising time of Vph(t) in the linear regime.

4. Experimental results

Equation (9) was verified using a single mode laser JDS 1145P 30mW HeNe, a positive lens with 3.5cm focal length, an adjustable aperture, a power meter Thorlabs PM100 disposed according to Fig. 5 under the following reasoning: Without the sample, the power of the light (P) passing through the aperture is function of its position (ζ) along beam axis and its radius (ρ0), i.e. P(W(ζ), ρ0), thus, the aperture was located at a distance (ζ) where the beam width was (W(ζ) = 4.5mm); then P(W(ζ), ρ0) was measured for ten increasing radius ρ0, after, the aperture was displaced at (ζ1) where the width was (W(ζ1) = 4mm), P (W(ζ1), ρ0) was measured for the same values of ρ0 as before. In Fig. 2 it is shown a plot of ΔP = P(W(ζ1), ρ0)−P(W(ζ), ρ0) versus ρ0WL where it can be seen that the optimum ρ0 for detecting changes in the transmitted power is W(ζ)(2).

 figure: Fig. 5

Fig. 5 Set up for determining the optimum aperture radius for measuring transmitted power changes.

Download Full Size | PDF

To carry out the OBZT, it was used a bacteriorhodopsin film 8μm thick, the JDS 1145P laser, a 5 cm lens, a Thorlabs PIN photodiode SM1PD1A, a home-made trans-resistance amplifier with RL = 82kΩ, CL = 10nF and operational amplifier OP27, an oscilloscope model TDS1012C-EDU for monitoring Vph, disposed according to the setup shown in Fig. 3(A), the distance from the output laser to the lens was 50cm, the distance of the lens to the aperture was 50cm; to perform the chopper technique the same elements were used arranged according to Fig. 3(B), replacing the aperture with a disk with 10 slots rotating at Ω0 = 18.85 rads (chopper Thorlabs model MC1000A).

4.1. Discussion of results

For both techniques in order to minimize disruption to the sample, the laser power was attenuated up to 40μW.

4.1.1. OBZT

For nonlinear absorptive samples, the OBZT curve is a mixture of nonlinear refractive and absorptive contributions [20], therefore, in order to obtain a free absorption Z-scan curve, it is proceeded as follows: A Z-scan experiment is performed without the aperture, the normalized curve obtained is shown in the inset of Fig. 6, subsequently, a second Z-scan experiment is done with the aperture, at the same positions as the inset of Fig. 6, the values measured are divided by their corresponding values of the curve in the inset; in Fig. 6 it is shown the free nonlinear absorption Z-scan curves for different aperture radii, where ΔV = VphVL, VL is the voltage Vph measured when the sample is far from the focus (linear regime). Again, the optimum radius turned out to be determined by the Eq. (9).

 figure: Fig. 6

Fig. 6 Z-scan curves free of nonlinear absorption for a bacteriorhodopsin sample, using different aperture radii (ρ0), WL = 3mm, the laser power was P0 = 40μW. In the inset is shown the normalized Vph measured without aperture, useful to obtain the free nonlinear absorption curves.

Download Full Size | PDF

4.1.2. Chopper Z-scan

In Fig. 7 it is shown the raising part of Vph(t) recorded with the oscilloscope, for illustrative purposes different curves were displaced to the left as the position of the sample was increased, from such figure it is clear how the chopper technique works: from the amplitude of Vph(t) the nonlinear absorption is estimated and from the raising time (τ) the refractive index is estimated because τ depends only on the beam width (see Fig. 4).

 figure: Fig. 7

Fig. 7 Rising part of Vph(t), notice the change of amplitude and slope as function of the sample position, the arrow indicates the increasing sample position z.

Download Full Size | PDF

In Fig. 8 it is shown τ vs z (the position of the sample), also in the inset is shown the amplitude of the pulses (such curve represents the transmitance).

 figure: Fig. 8

Fig. 8 Rising time (τ) of Vph(t) as function of the sample position (z), in the inset it is shown the amplitude of Vph(t) versus z (which represents the nonlinear transmitance of the sample).

Download Full Size | PDF

From Fig. 8, it is demonstrated the capacity of chopper technique to estimate both nonlinear absorption and refractive index in one experiment.

In order to highlight the advantages of the chopper technique, in Fig. 9 are presented the Z-scan curves of the relative changes of the physical variables obtained with the chopper and OBZT techniques, for the OBZT the ratio ρ0WL was adjusted as small as our iris allowed ( ρ0WL=0.1 in the small aperture limit [5] in order to minimize the nonlinear absorption effects), the laser power was attenuated up to 10μW; from Fig. 9 it is evident that the nonlinear absorption distorts the OBZT Z-scan curve [20] even with this aperture radius, unlike the chopper technique.

 figure: Fig. 9

Fig. 9 Z-scan curves of the physical variables of the two techniques when P0 = 10μW, for the OBZT ρ0 = 0.1WL and a lock-in amplifier was needed, PL and τL are the values measured for OBZT and chopper techniques respectively when the sample was located far from the focus.

Download Full Size | PDF

In Fig. 9, the OBZT curve is twice than the chopper curve, because of the following reason: the Eq. (11) for the case of ρ0WL2 is

ΔPP0(4ρ02WL2exp(2ρ02WL2))ΔWWL
using ΔWWL=ΔττL, ρ0WL=0.1 and Eq. (3) for calculating P(W(∞), ρ0) = PL, the Eq. (38) gives the following result:
ΔPPL=1.98(Δτ)τL
one might think that the OBZT is more sensitive than the chopper technique, however, it should not be confused for the result of Eq. (39), the optical power is fully utilized in the chopper technique, achieving a better signal to noise ratio (SNR) than the SNR of the OBZT, for example, for the opening radius ρ0 = 0.1WL, the optical power detected with the OBZT (Eq. (3)) technique is 50 times less than the maximum power (P0) detected with the chopper technique, this is the reason of the use of a device that improves the SNR (lock-in). The OBZT increases its SNR as ρ0 is increased up to WL(2) (value at which both techniques have the same SNR), however, for this value, the contribution of the nonlinear absorption is very dominant, that it is necessary a second experiment to estimate it and deduct it, as was done for obtaining Fig. 6.

Finally, in Fig. 10 it is shown Vph(t) measured when a sample distorts the beam profile, such distortion induces fast changes in Vph(t), which are filtered by the finite bandwidth of the amplifier (Eq. (18)), for this reason, it is advisable to keep the bandwidth as low as possible, i.e.

ΔfΔfc

 figure: Fig. 10

Fig. 10 Picture of a Gaussian beam profile distorted by an inhomogeneous sample, the graph shows the detected signal Vph(t), since the chopper technique works with the integral of the Gaussian profile and thanks to the limited bandwidth of the trans-impedance amplifier, the speckle affects little to the measurement.

Download Full Size | PDF

5. Conclusion

We have demonstrated that a direct measurement of the beam width through a chopper, the Z-scan technique is simplified without compromising its sensitivity, also the speed of data acquisition increases remarkably, allowing to perform averages with larger data, resulting in neater Z-scan curves.

We have established that the aperture radius plays an important role in the sensitivity and the optimum value is WL(2) where WL is the beam width (using the criteria of 1e2) in the linear regime.

For OBZT, the Z-scan curves are distorted by the presence of nonlinear absorption accentuated by increasing the radius of the aperture, forcing to correct them with the nonlinear absorption curve as can be seen in Fig. 6, whereas the proposed technique is immune to this effect.

A reliable method of measurement Gaussian beam by means of a chopper is given and confirmed experimentally as can be seen in Fig. 12.

 figure: Fig. 12

Fig. 12 Equivalence in W measurements given by the calibrated chopper and knife edge techniques.

Download Full Size | PDF

We have found that the chopper technique is much less sensitive to wavefront distortions caused by inhomogeneities of the sample unlike OBZT which strongly depends on it.

6. Appendix

As we mentioned, the chopper describes a curved trajectory rather than a straight one as with the knife edge technique, this causes that the beam width measured with the chopper (Wch) is larger than that of the knife (Wkn), this difference is negligible if condition (RW) is met, if not, the difference can be deducted if an expression for it is found, by measuring W(z) of a Gaussian beam, propagating along z-axis with both techniques at different positions (ζ) under the same experimental conditions. In Fig. 11, Wkn is plotted versus Wch, the data fitting Eq. (41) allows to find the relation between both techniques.

Wkn=0.8Wch+74.8μm
In Fig. 12 it is show the width measured of a the laser used in this paper, where it can be seen an excellent agreement between knife and calibrated chopper techniques.

 figure: Fig. 11

Fig. 11 Calibration of the chopper width measurement.

Download Full Size | PDF

Acknowledgments

This work has been sponsored by CONACyT, grants 51757 and 84008.

References and links

1. M. Sheik-Bahae, A. A. Said, T.-H. Wei, D. J. Hagan, and E. W. Van Stryland, “Sensitive measurement of optical nonlinearities using a single beam,” IEEE J. Quant. Electron. 26(4), 760–769 (1990) [CrossRef]  .

2. J. Wang, B. Gu, Y. M. Xu, and H. T. Wang, “Enhanced sensitivity of Z-scan technique by use of flat-topped beam,” Appl. Phys. B 95(4), 773778 (2009) [CrossRef]  .

3. T. Xia, D. J. Hagan, M. Sheik Bahae, and E. Van Stryland, “Eclipsing Z-scan measurement of λ/104 wave-front distortion,” Opt. Lett. 19(5), 317–319 (1994) [CrossRef]   [PubMed]  .

4. G. Tsigaridas, M. Fakis, I. Polyzos, P. Persephonis, and V. Giannetas, “Z-scan technique through beam radius measurements,” Appl. Phys. B 76(1), 83–86 (2003) [CrossRef]  .

5. P. B. Chapple, J. Staromlynska, J. A. Hermann, T. J. Mckay, and R. G. Mcduff, “Single-beam Z-scan: Measurement techniques and analysis,” J. Nonlinear Opt. Phys. Mater 6(3), 251–293 (1997) [CrossRef]  .

6. I. A. Ryasnyansky and B. Palpant, “Theoretical Investigation of the off-axis z-scan technique for nonlinear optical refraction measurement,” Appl. Opt. 45(12), 2773–2776 (2006) [CrossRef]   [PubMed]  .

7. R. L. McCally, “Measurement of Gaussian beam parameters,” Appl. Opt. 23(14), 2227–2227 (1984) [CrossRef]   [PubMed]  .

8. P. B. Chapple, “Beam waist and M2 measurement using a finite slit,” Opt. Eng. 33(7), 2461–2466 (1994) [CrossRef]  .

9. P. J. Shayler, “Laser beam distribution in the focal region,” Appl. Opt. 17(17), 2673–2674 (1978) [CrossRef]   [PubMed]  .

10. S. Nemoto, “Determination of waist parameters of a Gaussian beam,” Appl. Opt. 21(21), 3859–3863 (1986) [CrossRef]  .

11. J. M. Khosrofian and B. A. Garetz, “Measurement of a Gaussian laser beam diameter through the direct inversion of knife-edge data,” Appl. Opt. 22(21), 3406–3410 (1983) [CrossRef]   [PubMed]  .

12. J. A. Arnaud, W. M. Hubbard, G. D. Mandeville, B. delaClaviére, E. A. Franke, and J. M. Franke, “Technique for fast measurement of Gaussian laser beam parameters,” Appl. Opt. 10(12), 2775–2776 (1971) [CrossRef]   [PubMed]  .

13. Y. Suzaki and A. Tachibana, “Measurement of the μm sized radius of Gaussian laser beam using the scanning knife-edge,” Appl. Opt. 14(12), 2809–2810 (1975) [CrossRef]   [PubMed]  .

14. A. Nag, A. Kumar De, and D. Goswami, “Two-photon cross-section measurements using an optical chopper: z-scan and two-photon fluorescence schemes,” J. Phys. B: At. Mol. Opt. Phys. 42(6), 065103, (2009) [CrossRef]  .

15. I. Bhattacharyya, S. Priyadarshi, and D. Goswami, “Molecular structure-property correlations from optical nonlinearity and thermal-relaxation dynamics,” Chem. Phys. Lett. 469, 104–109, (2009) [CrossRef]   [PubMed]  .

16. A. Rogalski and Z. Bielecki, “Detection of optical radiation,” Bull. Pol. Ac.: Tech. 52(1), 43–66 (2004).

17. C. D. Motchenbacher and J. A. Connelly, Low-noise Electronic System Design (John Wiley & Sons Inc, 1993).

18. J. Phillips and K. Kundert, “Noise in mixers, oscillators, samplers, and logic an introduction to cyclostationary noise,” Proceedings of the IEEE custom integrated circuits conference , 431–439, (2000).

19. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics (John Wiley & Sons Inc, 1991) [CrossRef]  .

20. X. Liu, S. Guo, H. Wang, N. Ming, and L. Hou, “Investigation of the influence of finite aperture size on the Z-scan transmittance curve,” J. Nonlinear Opt. Phys. Mater. 10(4), 431–439 (2001) [CrossRef]  .

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (12)

Fig. 1
Fig. 1 Front view of a slotted disk which rotates at a frequency Ω0 and eclipses a Gaussian beam shown in red. R is the distance from the rotation axis to the center of the beam, W is the beam width, and θ is the angle subtended by the spot beam.
Fig. 2
Fig. 2 Sensitivity of an aperture of radius (ρ0) to detect changes in its transmitted power (ΔP = P(W, ρ0) − P(WL, ρ0)), when the incident beam width changes from (WLW).
Fig. 3
Fig. 3 A) One beam Z-scan set up, B) Chopper Z-scan set up.
Fig. 4
Fig. 4 Establishment of the threshold voltages Vth1 and Vth2 using e−2 criteria for measuring the beam width W from Vph(t) (black line) and the beam profile (red line).
Fig. 5
Fig. 5 Set up for determining the optimum aperture radius for measuring transmitted power changes.
Fig. 6
Fig. 6 Z-scan curves free of nonlinear absorption for a bacteriorhodopsin sample, using different aperture radii (ρ0), WL = 3mm, the laser power was P0 = 40μW. In the inset is shown the normalized Vph measured without aperture, useful to obtain the free nonlinear absorption curves.
Fig. 7
Fig. 7 Rising part of Vph(t), notice the change of amplitude and slope as function of the sample position, the arrow indicates the increasing sample position z.
Fig. 8
Fig. 8 Rising time (τ) of Vph(t) as function of the sample position (z), in the inset it is shown the amplitude of Vph(t) versus z (which represents the nonlinear transmitance of the sample).
Fig. 9
Fig. 9 Z-scan curves of the physical variables of the two techniques when P0 = 10μW, for the OBZT ρ0 = 0.1WL and a lock-in amplifier was needed, PL and τL are the values measured for OBZT and chopper techniques respectively when the sample was located far from the focus.
Fig. 10
Fig. 10 Picture of a Gaussian beam profile distorted by an inhomogeneous sample, the graph shows the detected signal Vph(t), since the chopper technique works with the integral of the Gaussian profile and thanks to the limited bandwidth of the trans-impedance amplifier, the speckle affects little to the measurement.
Fig. 12
Fig. 12 Equivalence in W measurements given by the calibrated chopper and knife edge techniques.
Fig. 11
Fig. 11 Calibration of the chopper width measurement.

Equations (41)

Equations on this page are rendered with MathJax. Learn more.

τ 2 W ( ζ ) Ω 0 R
T ( z ) = P ( W ( z ) , ρ 0 ) P ( W ( ) , ρ 0 )
P ( W , ( z ) , ρ 0 ) = P 0 ( 1 exp ( 2 ρ 0 2 W ( z ) 2 ) )
Δ T ( z ) T ( ) = Δ P P 0 S 1
Δ P P 0 = P ( W , ρ 0 ) P ( W L , ρ 0 ) P 0
S = 1 exp ( 2 ρ 0 2 W L 2 )
Δ W = W W L
2 ρ 0 W P ( W , ρ 0 ) = 0
ρ 0 = W L 2
Δ P P 0 = exp ( 1 ) exp ( W L 2 W 2 )
Δ P P 0 2 exp ( 1 ) ( Δ W W L )
Δ T ( z ) T ( ) 1.16 ( Δ W W L )
Δ T p v 0.406 ( 1 S ) 0.25 | Δ Φ 0 |
Δ W p v W L 0.273 | Δ Φ 0 |
Δ τ = 2 Ω 0 R Δ W
V p h = R L P
S N = R L ( P ) 4 k T R L Δ f
Δ f = 1.57 2 π R L C L
Δ P min = 2 4 k T R L Δ f R L
| Δ W min OB W L | = exp ( 1 ) 4 k T R L Δ f R L P 0
| Δ W min O B W L | = 2.72 P n R L P 0
P ch ( t ) = P 0 2 ( 1 erf ( 2 R sin ( Ω 0 t ) W ) )
β = R W
P ch ( t ) P 0 2 ( 1 erf ( 2 β Ω 0 t ) )
P ch ( t ) = 1 2 2 π m = 0 1 2 m + 1 exp ( 1 8 ( 2 m + 1 β ) 2 ) sin ( ( 2 m + 1 ) Ω 0 t )
Δ f c = 4 β f 0
Δ f c Δ f
Δ f c 1 τ = Ω 0 β 2
Δ f π β f 0
var ( j ( t c ) ) var ( n ( t c ) ) ( d V p h ( t c ) d t ) 2
V ph ( t ) R L P ch ( t )
var ( j ( t c ) ) P n ( d v ph ( t c ) dt ) 2
var ( W ) Ω 0 R 2 P n | d V ph ( t c ) d t |
t c = ± 2 W L 3 Ω 0 R
| d d t V ph ( ± t c ) | = 2 π e 8 9 β Ω 0 ( R L P 0 )
Δ W min Ch W L = 3 P n R L P 0
Δ τ p v τ L 0.273 | Δ Φ 0 |
Δ P P 0 ( 4 ρ 0 2 W L 2 exp ( 2 ρ 0 2 W L 2 ) ) Δ W W L
Δ P P L = 1.98 ( Δ τ ) τ L
Δ f Δ f c
W k n = 0.8 W c h + 74.8 μ m
Select as filters