Realization of mutually unbiased bases for a qubit with only one wave plate: theory and experiment

Open Access Open Access

Abstract

We consider the problem of implementing mutually unbiased bases (MUB) for a polarization qubit with only one wave plate, the minimum number of wave plates. We show that one wave plate is sufficient to realize two MUB as long as its phase shift (modulo 360°) ranges between 45° and 315°. It can realize three MUB (a complete set of MUB for a qubit) if the phase shift of the wave plate is within [111.5°, 141.7°] or its symmetric range with respect to 180°. The systematic error of the realized MUB using a third-wave plate (TWP) with 120° phase is calculated to be a half of that using the combination of a quarter-wave plate (QWP) and a half-wave plate (HWP). As experimental applications, TWPs are used in single-qubit and two-qubit quantum state tomography experiments and the results show a systematic error reduction by 50%. This technique not only saves one wave plate but also reduces the systematic error, which can be applied to quantum state tomography and other applications involving MUB. The proposed TWP may become a useful instrument in optical experiments, replacing multiple elements like QWP and HWP.

© 2015 Optical Society of America

1. Introduction

Measurement bases with a special geometry structure in Hilbert space, such as 2-designs [1,2], weighted 2-designs [3], tight frames [4] and Platonic solids [5], have attracted wide attention in the community of quantum information science in recent years. A typical example of 2-designs is mutually unbiased bases (MUB) [6, 7]. The special geometry for MUB is the equal Hilbert-Schmidt overlap between any two projective operators corresponding to two MUB. Hence, measurement results of one operator provide no information about the measurement results of its MUB at all. Due to this geometry, MUB have found applications from quantum key distribution [810], entropy uncertainty relations [1113] to quantum state estimation [7, 1417].

A complete set of MUB in a multi-qubit system involves nonlocal measurements that are difficult to be realized in the laboratory. In many scenarios such as quantum state tomography [1821], the tensor product of a complete set of single-qubit MUB is a preferred choice as the measurement bases for multi-qubit systems [5, 17]. Therefore, it is important to implement MUB for a single qubit in experiments [2224].

In the context of polarization optics, a specific type of polarization transformation poses constraints on the minimum number of optical elements that must be used. Three wave plates are the minimum number to realize any SU(2) polarization transformations [25] and the visual tool kit can be found in [26]. As for transformations between a given pair of nonorthogonal polarization states, two QWPs are enough (see a sketched proof in [27] and a recent constructive demonstration by Zela in [28]). When the initial state is linearly polarized, a quarter-wave plate (QWP) and a half-wave plate (HWP) can realize any polarization state [27]. The combination of QWP and HWP is used in most of current optical experiments [29] where the linear polarization is generated by a polarizing beam splitter (PBS). In the setting composed of a QWP, a HWP and a PBS, the PBS acts as σz and by adjusting the optic axis angles of the QWP and HWP (a two-parameter setting), the Bloch vector of σz can be unitarily rotated to any point on the unit Bloch sphere [27]. However, when the problem is restricted to realize MUB rather than arbitrary bases, is one wave plate sufficient to transform the initial σz to a set of MUB, especially a complete set of three MUB? If the answer is positive, not only one wave plate is saved but also the systematic error [29, 30] can be expected to decrease as the parameter uncertainties of the devices (i.e., wave plates in our case) are the sources of systematic error. Intuitively, the less measurement devices, the smaller the systematic error. So far, people know how to construct two MUB with one HWP or one QWP. For example, one HWP itself can be used to realize σz and σx by setting its rotation angles as 0° and 22.5°; one QWP can be used to realize σz and σy by setting its rotation angles as 0° and 45°. One wave plate is also used to perform Fourier transform tomography [31] and the optimal phase of the wave plate is numerically calculated in [32]. As to a complete set of three MUB, this problem is barely considered. In this paper, we show that it is indeed feasible to employ only one wave plate to realize a complete set of single-qubit MUB with reduced systematic error.

Here is the organization for the rest of the paper. In section 2, the conditions to realize two MUB and three MUB using only one wave plate are considered, respectively. Section 3 calculates the systematic error in the realization of MUB with only one wave plate. In section 4, third-wave plates are used to perform MUB measurements in both single qubit and two-qubit tomography experiments and the results demonstrate an error reduction by 50%. Section 5 concludes the paper.

2. Theory

2.1. Transforming polarization states with QWP-HWP setting

In optical experiments, arbitrary projective measurements on polarization qubits are often implemented by a QWP-HWP setting; see Fig. 1(a). Define |H〉 = (1, 0)T, |V〉 = (0, 1)T and their corresponding eigenvalues are ±1, where T denotes transpose. Then the PBS acts as σz or (0, 0, 1)T in the Bloch representation. This setup transforms |H〉 to

|ψ=U(q,π2)U(h,π)|H=(cosqcos(q2h)+isinqsin(q2h)sinqcos(q2h)icosqsin(q2h)),
where q, h are, respectively, the optic axis angles of QWP and HWP deviated from the horizontal direction, and U is a unitary transformation operator on polarization by a phase plate with phase δ and rotation angle θ,
U(θ,δ)=(cos2θ+eiδsin2θ12(1eiδ)sin2θ12(1eiδ)sin2θsin2θ+eiδcos2θ).
Or equivalently this setup transforms the original operator σz to the Pauli operator r(q, h) · σ with
r(q,h)=(sin2qcos(4h2q),sin(4h2q),cos2qcos(4h2q))T
and σ = (σx, σy, σz), where σx, σy and σz are three Pauli operators.

 figure: Fig. 1

Fig. 1 MUB measurement with two different settings: (a) MUB measurement with QWP-HWP setting consisting of a QWP, a HWP and a PBS. (b) MUB measurement with a wave plate (WP) and a PBS.

Download Full Size | PDF

Two variables of rotation angles q and h correspond to a plane and can rotate the initial Bloch vector (0, 0, 1)T to any direction or position on the unit Bloch sphere. Thus, this QWP-HWP setting is used to realize any one qubit projective measurement basis. For example, the most popular set of MUB is composed of Bloch vectors (1, 0, 0)T, (0, 1, 0)T and (0, 0, 1)T (corresponding to three Pauli operators) which are realized by choosing q = 45°, 0°, 0° and h = 22.5°, 22.5°, 0° correspondingly.

In order to realize MUB for a qubit, only two or three specific Bloch vectors need to be implemented and it is not necessary to cover the Bloch sphere with two rotation angles of two wave plates. We will show below that one wave plate with appropriate phases is sufficient to realize single-qubit MUB.

2.2. Realizing two mutually unbiased bases with one wave plate

Two sets of orthogonal bases {|ψi1}i=1d and {|ψj2}j=1d in a d-dimensional Hilbert space are mutually unbiased if |ψi1|ψj2|2=1d for any i and j. There are at most d + 1 bases with every two of them mutually unbiased, which are called a complete set of MUB. So far, how to construct a complete set of MUB is known only in systems with dimensions which are powers of primes [6, 7]. For general cases, the existence of such a set is still an open problem even in the simple case of d = 6 [6, 33]. For a qubit with dimension d = 2, the bases {|ψij}i=12 are the eigenvectors of a unit Pauli operator rj · σ with ‖rj‖ = 1. In the Bloch representation, the jth basis is directly related to its Bloch vector rj as |ψijψij|=I±rjσ2 and ± corresponds to i = 1, 2. Thus, the requirement for two bases to be mutually unbiased is that their Bloch vectors are normal to each other

r1r2=0.

In Fig. 1(b), a wave plate (WP) with a phase difference δ between ordinary (o) and extraordinary (e) components is placed in front of a PBS. Its unitary transformation operator on polarization is expressed in Eq. (2) where θ is the deviation angle of the optic axis from the horizontal direction. This wave plate with rotation angle θ transforms the initial Bloch vector r0 = (0, 0, 1)T into

r(θ)=(sin2δ2sin4θsinδsin2θ,12sin2δ2sin22θ)T.
From Eq. (4), this setting with one wave plate can realize two MUB if there exist θ1 and θ2 such that their Bloch vectors are normal to each other, i.e., r(θ1) · r(θ2) = 0. The existence of θ1 and θ2 obviously depends on δ. One trivial example is δ = 0° where the wave plate does not affect the initial polarization and r(θ1) · r(θ2) = 1 for any θ1, θ2. Another example is δ = 180°. The choice of θ1 = 0° and θ2 = 22.5° makes r(θ1) = (0, 0, 1)T, r(θ2) = (1, 0, 0)T, and thus r(θ1) · r(θ2) = 0. The maximum of r(θ1) · r(θ2) is 1, which can be achieved at θ1 = θ2. Because r(θ1) · r(θ2) is a continuous function and its maximum is positive, there exist θ1 and θ2 to make r(θ1) · r(θ2) = 0 if the minimum of r(θ1) · r(θ2) is negative or zero. This problem is then converted into finding the minimum of r(θ1) · r(θ2). From Eq. (5), the first two elements of r are odd functions of θ and the third term is an even function of θ. That is, r1(−θ) = −r1(θ), r2(−θ) = −r2(θ) and r3(−θ) = r3(θ). From Eq. (5), the minimum and maximum of r3(θ) is cos δ and 1. If cos δ is non-positive, then there exists θ1=12arcsin(22cscδ2) that makes r3(θ1) = 0 and we choose θ2 = −θ1 to make r(θ1) = −r(θ2). We obtain the minimum r(θ1) · r(θ2) = −1. Thus, in this case, two MUB can be realized such as θ1=12arcsin(22cscδ2) and θ2 = 0°. If cosδ is positive, the minimum of r(θ1) · r(θ2) is cos2δ obtained at θ1 = −θ2 = 45°. Thus, the requirements that the minimum of r(θ1) · r(θ2) is non-positive in these two cases together give 45° ≤ δ ≤ 315°. As long as the phase δ (modula 360°) of the wave plate is within [45°, 315°], it can be used to realize at least two MUB. As there are two parameters and one equation, the solutions of Eq. (4) form a line. One solution is θ1=12arcsin(222cscδ2) and θ2 = −θ1, which make r3=22. The Bloch vectors of the two MUB are axially symmetric about the initial state (0, 0, 1)T, and the angles between MUB and (0, 0, 1)T are 45°.

2.3. Realizing a complete set of MUB with one wave plate

Three rotation angles θi, i = 1, 2, 3, are chosen to realize three MUB, which consist of a complete set of MUB for a qubit,

r(θi)r(θj)=0,
where i, j = 1, 2, 3 and ij. For a complete set of MUB, it is much more complicated and difficult to obtain theoretical solutions. Hence, the problem is investigated numerically in this paper. Borrowed from frame theory [2, 34], the frame potential of a wave plate is defined as
Φ(δ)=minθ1,θ2,θ3[r(θ1)r(θ2)]2+[r(θ2)r(θ3)]2+[r(θ1)r(θ3)]2.
The solution of Eq. (6) exists only if the phase δ of the wave plate makes the frame potential Eq. (7) vanishing. The frame potential is numerically computed by a MATLAB solver lsqnonlin, which is intended to solve nonlinear least-squares problems. One hundred different initial points of θ1, θ2 and θ3 are taken for us to avoid local minimum and find all the possible solutions. The numerical result of the frame potential for different phases is shown in Fig. 2. From Eq. (5), Φ(δ) has a period of 360°. Thus, we only consider 0° ≤ δ < 360°. Since r(θi, δ) · r(θj, δ) = r(θi, 360° − δ) · r(θj, 360° − δ), we have Φ(δ) = Φ(360° − δ), meaning that the frame potential Φ(δ) is symmetric about δ = 180°.

 figure: Fig. 2

Fig. 2 Frame potential Φ(δ) with respect to δ. The frame potential is symmetric about 180°. The inset is the amplified interval of δ where the frame potential vanishes to zero. Only if δ is in the range [111.5°, 141.7°] or its symmetric range about 180° can a complete set of MUB be realized by one wave plate.

Download Full Size | PDF

With δ between 111.5° and 141.7°, the frame potential is zero and the solutions of Eq. (6) are shown in Fig. 3. From Eq. (5), r(θ) = r(θ + 180°), we only consider rotation angles within [0°, 180°). Eq. (5) also shows r(θi) · r(θj) = r(90° ± θi) · r(90° ± θj) = r(180° − θi) · r(180° − θj). That is to say, if θ1, θ2, θ3 (red in Fig. 3) is the solution, the modules of 90° ± θi and 180° − θi (i = 1, 2, 3) by 180° (represented as blue, green and black in Fig. 3 correspondingly) are also the solutions. Thus, all the four sets of MUB are considered as one class of solutions.

 figure: Fig. 3

Fig. 3 The numerical solutions of rotation angles in Eq. (6) with respect to the phase δ. There are two classes of solutions with δ between 111.5° and 126.3°, represented by solid and dotted lines, respectively, and only one class of solutions in (126.3°, 141.7°]. Each class of solutions contains four complete sets of MUB, denoted by four different colors.

Download Full Size | PDF

As shown in Fig. 3, solutions from different sets of MUB in the same class intersect around δ = 120°, 126.3° and 141.7°. Using the symmetries represented by the colors, the four variables (i.e. δ, θ1, θ2 and θ3) reduce to two and we can theoretically calculate the solutions at these intersections. The phases at these intersections are also rigourously found to be δ = 120°, 126.32° and 141.76° (see Appendix A.1 and A.2). Here we compare our results with those in [32]. The optimal phase was numerically calculated in [32] to be 7π/10 (i.e. 126°). The figure of merit times the total number of counts in [32] at this optimal phase equals 10.03, which is very close to 10, the bound achieved by MUB. This phase falls within our calculated range [111.5°, 141.7°]. The reason why they only found one phase rather than an available interval of phases and the optimal performance at this phase was slightly worse than the bound is their restriction of six equally spaced rotation angles.

In the special case of δt = 120°, called third-wave plate (TWP), the Bloch vector is

r(t)=(34sin4t,32sin2t,34cos4t+14)T,
where t is the rotation angle of the optic axis of TWP deviated from horizontal direction. From Appendix A.1, solutions of Eq. (6) are r(0°) = r(90°) = r(180°) = (0, 0, 1)T, r(t0)=r(180°t0)=12(1,1,0)T, r(t0+90°)=r(90°t0)=12(1,1,0)T, where t0=14arccos(13)27.37°.

3. Systematic error in the realization of MUB in one wave plate setting

Imperfect measurement devices are the main sources of the systematic error in the realization of MUB. Here we consider the systematic error due to the parameter uncertainties of wave plates. The realized bases are denoted by their Bloch vectors as r(δ, θ), where δ is the real phase of the wave plate in the one wave plate setting in Fig. 1(b). The systematic error in the realization of r is

(Δr)2=ξrξ2(Δξ)2,
where rξ=rξ, ξ = δ, θ. From Eq. (5),
rδ2=sin22θ=(r2)2sin2δ,rθ2=16sin2δ24sin2δsin22θ=16sin2δ24(r2)2.
For a complete set of MUB (rj, j = 1, 2, 3), the systematic error sums up to
ε2=j=13(Δrj)2=j=13ξrξj2(Δξ)2=ξ(εξ)2(Δξ)2,(εξ)2=j=13rξj2.
As rj is orthogonal to each other, from Eq. (10),
(εδ)2=j=13(r2j)2sin2δ=1sin2δ,(εθ)2=j=1316sin2δ24(r2j)2=48sin2δ24.
Thus, the systematic error in the one wave plate setting is
ε2=1sin2δ(Δδ)2+(48sin2δ24)(Δθ)2.
As ε2 in Eq. (12) is an increasing function of δ in the interval [111.5°, 141.7°], the minimum and maximum systematic error in the realization of three MUB with one wave plate setting is 1.16(Δδ)2 + 28.80(Δθ)2 and 2.60(Δδ)2 + 38.83(Δθ)2, achieved at δ = 111.5° and δ = 141.7°. For a third-wave plate with δt = 120°,
ε2=1.33(Δδt)2+32(Δt)2.
Under the assumption that (Δδh)2 = (Δδq)2 = (Δδt)2 = (Δδ)2 and (Δh)2 = (Δq)2 = (Δt)2 = (Δθ)2, the systematic error in the realization of MUB is 1.33(Δδ)2 + 32(Δθ)2 in the TWP setting and averaged as (2.5Δδ)2 + 68(Δθ)2 in the QWP-HWP setting in Eq. (26). Thus, the TWP setting outperforms the QWP-HWP setting by about a factor of two.

Measurements based on single-qubit MUB are preferable choices in quantum state tomography. In qubit state estimation, a complete set of single-qubit MUB is used to extract information of the qubit optimally. In multi-qubit quantum state tomography, the product measurements of single-qubit MUB on each photon are used to reduce estimation error due to statistical fluctuation. When the copies of states ρ are infinite, the estimated state ρ̂ based on the measurement data should be the same as the real state ρ. However, since single-qubit MUB are imperfectly realized, the estimated state ρ̂ no longer converges to the real state ρ and tr(ρ̂ρ)2 is defined as the systematic error in state estimation. Generally, tr(ρ̂ρ)2 depends on ρ. Averaged over unitarily equivalent states, 〈tr(ρ̂ρ)2〉 in both single-qubit and multi-qubit state estimation is proportionate to the systematic error of the realized bases [35]. As the systematic error in the realization of multi-qubit product bases is the sum over the systematic error of single MUB for each qubit, the systematic error in the realized multi-qubit product bases with one wave plate for each qubit is still a half of that with the QWP-HWP combination. This systematic error reduction effect in quantum state tomography is experimentally verified in both single-qubit and two-qubit tomography experiments in the next section.

4. Quantum state tomography experiments with third-wave plates

4.1. Qubit tomography experiments

The experimental setup, shown in Fig. 4, includes two parts: state preparation and MUB measurement. A 40 mW, V-polarized beam at 404 nm from a semiconductor laser pumps a type I phase-matched β-barium borate (BBO) crystal. After the spontaneous parametric down-conversion (SPDC) process, a pair of 808 nm H-polarized photons are created. One photon passes through a 3 nm interference filter and is detected by a single photon detector to herald the presence of its twin photon. The quantum state of the heralded photon is prepared by HWP0, HWP1, QWP1 and a 770λ quartz crystal which is much larger than the coherence length of about 270λ with λ = 808 nm. HWP0 with rotation angle h0 and the quartz crystal with optic axis aligned horizontally together prepare the quantum state ρ wih Bloch vector s = cos4h0(0, 0, 1)T; HWP1 and QWP1 can transform s to arbitrary direction. This part is capable of preparing arbitrary qubit state. In the MUB measurement part, a complete set of MUB is performed with two methods: a TWP with rotation angles set as 0°, 27.37° and 117.37° in Fig. 4(a); the conventional QWP-HWP setting with rotation angles set as (45°, 22.5°), (0°, 22.5°) and (0°, 0°) in Fig. 4(b).

 figure: Fig. 4

Fig. 4 Experimental setup for qubit tomography. The apparatus consists of two parts: state preparation (green) and MUB measurement (pink). The MUB measurement part consists of a polarizing beam splitter and a wave-plate combination which has two choices: (a) TWP and (b) QWP-HWP combination.

Download Full Size | PDF

As the systematic error in quantum state tomography experiments depends on the state to be measured, we prepared three states ρ = p|ϕ〉 〈ϕ| + (1 − p)I/2 at p = 0.92 (by setting HWP0’s rotation angle h0 = 5.8°) and |ϕ〉 = |H〉, |D〉 and |R〉 with a fidelity of more than 0.998. The theoretical systematic error tr(ρ̂ρ)2 in the estimation of these three states is derived in Appendix C as [8(Δ h)2 + 4(Δ q) + 0.25(Δδh)2]p2, [2(Δ q) + 0.25(Δδh)2 + 0.5(Δδq)2]p2 and [16(Δ h)2 + 4(Δ q)]p2 in the QWP-HWP setting and [8(Δt)2+13(Δδt)2]p2, [112(Δt)+16(Δδt)2]p2 and [52(Δt)2+16(Δδt)]p2 in the TWP setting. Under the assumption that (Δδh)2 = (Δδq)2 = (Δδt)2 = (Δδ)2 and (Δh)2 = (Δq)2 = (Δt)2 = (Δθ)2, the total systematic error of these three states is [(Δδ)2 + 34(Δθ)2]p2 in the QWP-HWP setting and [23(Δδ)2+16(Δθ)2]p2 in one wave plate setting. That is, the systematic error in realizing the corresponding MUB using the TWP setting is around a half of that using the QWP-HWP setting.

As phase errors are determined by the manufacture and wavelength, without a variable wavelength we can only experimentally verify the relationship between the systematic error of the estimated state and the angle errors of QWP, HWP and TWP for all these three states. In the experiment, MUB measurements are performed on 3 × 106 photons with two different settings in Figs. 4(a) and 4 (b). We first measure the state with the well-calibrated setting and assume the estimated state ρ as the real state. Then we intentionally mis-calibrate the optic axes of the wave plates with an angular error, and obtain an estimation ρ̂. Thus, the systematic error due to this angular error is calculated as tr(ρ̂ρ)2. The estimated states by the well-calibrated TWP setting have a fidelity of over 99.9% with those by the well-calibrated QWP-HWP setting for all the three states above, validating each other. In terms of systematic error, experimental results (dots) and the theoretical results (solid lines) are shown in Fig. 5, and they match very well. The performance of these two settings depend on ρ and neither always outperforms the other. For example, for states in Figs. 5(a) and 5(c), TWP beats the QWP-HWP combination while reversely for the state in Fig. 5(b). However, there are more states where TWP performs better. The total systematic error in the estimation of these three states in the TWP setting adds up to be about two times smaller than that in the QWP-HWP setting as shown in Fig. 5(d).

 figure: Fig. 5

Fig. 5 Systematic error in TWP and QWP-HWP setting for qubit tomography. The dependence of the systematic error in the TWP setting (Fig. 4(a)) and the QWP-HWP setting (Fig. 4(b)) is experimentally measured with respect to the angle errors of TWP (red), QWP (black) and HWP (blue) for three states in Figs. 5(a)–5(c) at p = 0.92. The total systematic error (green) in the QWP-HWP setting is the sum of that due to QWP and HWP. The experimental results denoted as dots coincide with the theoretical calculations (solid lines). Fig. 5(d) plots the total systematic error for all these three states in the two settings and shows that the TWP setting beats the QWP-HWP setting by a factor of two. Error bars are the standard deviation of 100 trials in Monte Carlo simulation with binomial distribution of counting statistics.

Download Full Size | PDF

4.2. Two-qubit tomography experiments

In Fig. 6, a 100 mW, H-polarized beam at 404 nm from a continuous laser pumps a pair of type I phase-matched β-barium borate (BBO) crystals whose optic axes are normal to each other. After the spontaneous parametric down-conversion (SPDC) process, a pair of 808 nm photons are created. When the optic axis of half-wave plate (HWP0) at 404 nm is deviated 22.5° from horizontal direction, the twin SPDC photons are maximally entangled. HWP1 and HWP2 rotate H and V to the fast and slow axes of the single mode fibers. At the output ports of the fibers, HWP3 and HWP4 rotate the polarization direction back to horizontal and vertical. QWP1 is tilted to compensate the phase of the entangled states to a singlet state. In the MUB measurement part, a complete set of MUB on either photon is performed with two methods: TWP setting in Fig. 4(a) and QWP-HWP setting in Fig. 4(b).

 figure: Fig. 6

Fig. 6 Experimental setup for two-qubit tomography. A singlet state is prepared via SPDC process with a fidelity of 98%. In quantum state tomography, single-qubit MUB measurements are implemented on both photons with TWP or the combination of QWP and HWP in Figs. 4 (a) and 4 (b). Coincidence events are recorded by a coincidence circuit.

Download Full Size | PDF

In the two-qubit case, the singlet state ρ = |Ψ〉 〈Ψ| is chosen for three reasons: firstly, systematic error in the estimation of product states is a direct sum of that of single-qubit states; secondly, entangled states reveal the peculiar features of quantum systems and are valuable quantum resources; the last reason is that the systematic error in the estimation of Werner states is proportionate to the sum of the systematic error in the realization of single-qubit MUB for either photon [35], which is similar to the systematic error averaged over unitarily equivalent states. The systematic error for Werner states [35] is

tr(ρρ^)2=14p2(ε12+ε22),
where ρ = p〉 〈Ψ| + (1 − p)I/4, |Ψ=12(|HV|VH) and εi2 is the systematic error of the realized single-qubit MUB on photon i, i = 1, 2.

Similar to the qubit tomography experiment, we only experimentally measure the dependence of the systematic error on the angle errors of wave plates in the branches of both photons. Then product measurements of MUB are performed on 9 × 105 pairs of prepared singlet states. The estimated state by the well-calibrated TWP setting has a fidelity of over 99.8% with that by the well-calibrated QWP-HWP setting, agreeing well with each other. From Eqs. (13), (14) and (25), angle errors of wave plates for either photon theoretically contribute 8(Δt)2 in the TWP setting and 12(Δh)2 + 5(Δq)2 in the QWP-HWP setting. However, numerical results in Fig. 7 (a) show that angle errors only cause 10.5(Δh)2 + 4.6(Δq)2 in the QWP-HWP setting and 6.9(Δt)2 in the TWP setting. This gap is due to the positive semi-definite conditions of density matrices, which arises when states are singular. Both the experimental results and numerical results in Fig. 7 (b) show that the systematic error in the TWP setting is only about a half of that in the QWP-HWP setting. The experimental results are slightly smaller than the numerical results because the prepared state is not exactly the expected singlet state, which only has a fidelity of 98%.

 figure: Fig. 7

Fig. 7 Systematic error in TWP and QWP-HWP setting for two-qubit tomography. In (a), systematic errors due to angle errors of QWP (black), HWP (blue), both of them (green) and TWP (red) are numerically simulated and denoted as solid lines. The experimental results due to angle errors of wave plates are plotted as dots for photon 1 and circles for photon 2. The total systematic error for the two photons is plotted in (b). From (b), The TWP setting beats the QWP-HWP setting by about a factor of two. Error bars are the standard deviation of 100 Monte Carlo simulations with multinomial distribution of the counting statistics.

Download Full Size | PDF

5. Conclusion

We have found that one wave plate is sufficient to realize two MUB as long as its phase is within [45°, 315°]. It is capable of realizing a complete set of MUB if the phase is within [111.5°, 141.7°] or the symmetric interval about 180°. The systematic error in the realization of MUB in one wave plate setting is calculated to be twice smaller than that in the conventional QWP-HWP setting. TWPs are applied to single-qubit and two-qubit quantum state tomography experiments and experimentally show an error reduction by 50% compared with the QWP-HWP combination. Other applications of TWP and arbitrary phase plates in the realization of any SU(2) and polarization state transformations need to be explored in the future.

A. Theoretical solutions at intersections

A.1. Theoretical solutions at the intersection around δ = 120°

From the numerical solutions in Fig. 3, some solutions cross at about δ = 120°. Here solutions at this intersection are analytically calculated and the phase at this intersection is indeed 120°. We denote the three red solid lines from bottom to top as θi, i = 1, 2, 3. The green, blue and black solid lines correspond to 90 − θi, 90 + θi and 180 − θi, i = 1, 2, 3. Numerical solutions in Fig. 3 show that crossing lines at the intersection should have the same coordinates, i.e.,

90°θ1=90°+θ1,θ2=θ390°.
From Eq. (15), θ1 = 0°. Substituting θ1 into Eq. (5), one obtains r(θ1) = (0, 0, 1)T. With these relations, the conditions that r(θi), i = 1, 2, 3 should be orthogonal to each other give two independent equations
cosδsin22θ2+cos22θ2=0,sin2δsin22θ2=12.
The solutions of Eq. (16) are δ = 120°, θ2=14arccos(13)27.37° and θ3 = 117.37°. Other solutions at this intersection represented by green, blue and black can be calculated from their symmetries about the red lines.

A.2. Theoretical solutions at intersections around δ = 126.3° and 141.7°

Numerical solutions in Fig. 3 also show some solutions intersect near δ = 126.3° and 141.7°. Here we analytically calculate the solutions and phases at the intersections. The three red lines are denoted as θi, i = 1, 2, 3 from bottom to top. In Fig. 3 numerical solutions at the intersection around δ = 141.7° give

θ1+θ2=90°,θ390°=180°θ3.
From Eq. (17), θ3 = 135° for the intersection around δ = 141.7°. For the intersection at about δ = 126.3°, one has
θ1+θ3=90°,90°θ2=θ2.
Eq. (18) gives θ2 = 45° for the intersection around δ = 126.3°. With Eqs. (17) and (18), conditions of a complete set of MUB at the two intersections give the same equations
cos2δsin22θ1+cosδcos22θ1sin2δsin2θ1=0,(1cosδ)2sin22θ1cos22θ1=sin2δsin22θ1+(cosδsin22θ1+cos22θ1)2.
The two equations in Eq. (19) are equivalent to
x(x1)y2+x=(1x2)y,(x1)2y4=(x1)2y212
with x = cosδ and y = sin2θ1. The square of the first equation in Eq. (20) is
x2(x1)2y4+2x2(x1)y2+x2=(x1)2(x+1)2y2.
Replacing (x − 1)2y4 in Eq. (21) with the second equation in Eq. (20), one obtains
y2=x22(1x2).
Substituting Eq. (22) into the second equation in Eq. (20), we have
3x4+4x+2=0.
This four-order equation has two real solutions x = −0.5923 and x = −0.7854, corresponding to δ = 126.32° and 141.76°, θ1 = 15.66° and 31.90°. From Eq. (18), θ3 = 74.34° for δ = 126.32° and Eq. (17) gives θ2 = 58.10° for δ = 141.76°.

B. Systematic error in the realization of MUB in the QWP-HWP setting

In the QWP-HWP setting, the realized bases are denoted as r(δq, δh, q, h), where δq, δh are the real phases of QWP and HWP in Fig. 1(a). From Eq. (3) and [36], at δq=π2 and δh = π,

rδq2=sin2(4h2q)=(r2)2,rδh2=sin22h,rq2=4+4cos2(4h2q)=84(r2)2,rh2=16.
From Eqs. (11) and (24), one obtains
(εδq)2=j=13(r2j)2=1,(εq)2=j=1384(r2j)2=20,(εh)2=48m(εδh)2=j=13sin22hj.
(εδh)2 depends on the choices of MUB. In the realization of three Pauli operators with h1 = h2 = 22.5° and h3 = 0°, (εδh)2 = 1 and
ε2=48(Δh)2+20(Δq)2+(Δδh)2+(Δδq)2.
For simplicity, hj is assumed to be uniformly distributed within [0°, 360°] in the consideration of general MUB. Then (εδh)2 is averaged as 32, and
ε2=48(Δh)2+20(Δq)2+1.5(Δδh)2+(Δδq)2.

C. Systematic error in qubit state estimation

From [35, 36], the systematic error in qubit state estimation with a complete set of MUB is

tr(ρρ^)2=12ξRTξs2(Δξ)2,
where Rξ=(r(1)ξ,r(2)ξ,r(3)ξ) and s is the Bloch vector of ρ.

In the TWP setting with rotations angles at 0°, 27.37° and 117.37°, from Eqs. (5) and (8),

Rt=(31131102222),Rδt=(066660666603333).
Thus, the systematic error tr(ρρ̂)2 in the estimation of the three states ρ = p|ϕ〉 〈ϕ| + (1 − p)I/2 with |ϕ〉 = |H〉, |D〉, |R〉 is calculated from Eqs. (27) and (28) to be [8(Δt)2+13(Δδt)2]p2, [112(Δt)+16(Δδt)2]p2 and [52(Δt)2+16(Δδt)]p2.

In the QWP-HWP setting, from [36], the systematic error tr(ρρ̂)2 in the estimation of the three states can be calculated to be [8(Δh)2 + 4(Δq) + 0.25(Δδh)2]p2, [2(Δq) + 0.25(Δ δh)2 + (Δ δq)2]p2 and [16(Δh)2 + 4(Δq)]p2.

Acknowledgments

The work was supported by National Fundamental Research Program (Grants No. 2011CBA00200 and No. 2011CB9211200), National Natural Science Foundation of China (Grants No. 61108009 and No. 61222504) and in part by the Australian Research Council’s Discovery Projects funding scheme under Project DP130101658.

References and links

1. G. Zauner, “Quantum designs: Foundations of a noncommutative design theory,” Int. J. Quantum Inf. 09, 445–507 (2011). [CrossRef]  

2. J. M. Renes, R. Blume-Kohout, A. J. Scott, and C. M. Caves, “Symmetric informationally complete quantum measurements,” J. Math. Phys. 45, 2171–2180 (2004). [CrossRef]  

3. A. Roy and A. J. Scott, “Weighted complex projective 2-designs from bases: Optimal state determination by orthogonal measurements,” J. Math. Phys. 48, 072110 (2007). [CrossRef]  

4. A. J. Scott, “Tight informationally complete quantum measurements,” J. Phys. A Math. Gen. 39, 13507–13530 (2006). [CrossRef]  

5. M. D. de Burgh, N. K. Langford, A. C. Doherty, and A. Gilchrist, “Choice of measurement sets in qubit tomography,” Phys. Rev. A 78, 052122 (2008). [CrossRef]  

6. T. Durt, B-G Englert, I. Bengtsson, and K Życzkowski, “On mutually unbiased bases,” Int. J. Quantum Inf. 8, 535–640 (2010). [CrossRef]  

7. W. K. Wootters and B. D. Fields, “Optimal state–determination by mutually unbiased measurements,” Ann. Phys. 191, 363–381 (1989). [CrossRef]  

8. C. H. Bennett and G. Brassard, “Quantum cryptography: Public key distribution and coin tossing,” in Proceedings of IEEE International Conference on Computers, Systems and Signal Processing, 175 (1984).

9. N. Gisin, G. Ribordy, W. Tittel, and H. Zbinden, “Quantum cryptography,” Rev. Mod. Phys. 74, 145–195 (2002). [CrossRef]  

10. M. A. Nielsen and I. L. Chuang, Quantum Computation and Quantum Information (Cambridge university, 2010). [CrossRef]  

11. S. Wehner and A. Winter, “Entropic uncertainty relations–a survey,” New J. Phys. 12, 025009 (2010). [CrossRef]  

12. S. J. Wu, S. X. Yu, and K. Mølmer, “Entropic uncertainty relation for mutually unbiased bases,” Phys. Rev. A 79, 022104 (2009). [CrossRef]  

13. M. A. Ballester and S. Wehner, “Entropic uncertainty relations and locking: Tight bounds for mutually unbiased bases,” Phys. Rev. A 75, 022319 (2007). [CrossRef]  

14. G. Lima, L. Neves, R. Guzmán, E. S. Gómez, W. A. T. Nogueira, A. Delgado, A. Vargas, and C. Saavedra, “Experimental quantum tomography of photonic qudits via mutually unbiased basis,” Opt. Express 19, 3542–3552 (2011). [CrossRef]   [PubMed]  

15. H. Zhu, Quantum State Estimation and Symmetric Informationally Complete POMs (National University of Singapore, 2012).

16. R. B. A. Adamson and A. M. Steinberg, “Improving quantum state estimation with mutually unbiased bases,” Phys. Rev. Lett. 105, 030406 (2010). [CrossRef]   [PubMed]  

17. B. Qi, Z. Hou, L. Li, D. Dong, G. Xiang, and G.-C. Guo, “Quantum state tomography via linear regression estimation,” Sci. Rep. 3, 3496 (2013). [CrossRef]   [PubMed]  

18. M. Paris and J. Řeháček, Quantum State Estimation (Springer, 2004). [CrossRef]  

19. J. Fan, M. D. Eisaman, and A. Migdall, “Quantum state tomography of a fiber-based source of polarization-entangled photon pairs,” Opt. Express 15, 18339–18344 (2007). [CrossRef]   [PubMed]  

20. M. Cramer, M. B. Plenio, S. T. Flammia, R. Somma, D. Gross, S. D. Bartlett, O. Landon-Cardinal, D. Poulin, and Y. K. Liu, “Efficient quantum state tomography,” Nat. Commun. 1, 149 (2010). [CrossRef]  

21. J. B. Altepeter, E. R. Jeffrey, and P. G. Kwiat, Photonic State Tomography, Advances in AMO Physics (Elsevier, 2006).

22. D. H. Mahler, L. A. Rozema, A. Darabi, C. Ferrie, R. Blume-Kohout, and A. M. Steinberg, “Adaptive quantum state tomography improves accuracy quadratically,” Phys. Rev. Lett. 111, 183601 (2013). [CrossRef]   [PubMed]  

23. K. S. Kravtsov, S. S. Straupe, I. V. Radchenko, N. M. T. Houlsby, F. Huszár, and S. P. Kulik, “Experimental adaptive Bayesian tomography,” Phys. Rev. A 87, 062122 (2013). [CrossRef]  

24. J. Z. Salvail, M. Agnew, A. S. Johnson, E. Bolduc, J. Leach, and R. W. Boyd, “Full characterization of polarization states of light via direct measurement,” Nat. Photon. 7, 316–321 (2013). [CrossRef]  

25. R. Simon and N. Mukunda, “Minimal three-component SU(2) gadget for polarization optics,” Phys. Lett. A 143, 165–169 (1990). [CrossRef]  

26. B. N. Simon, C. M. Chandrashekar, and S. Simon, “Hamilton’s turns as a visual tool kit for designing single-qubit unitary gates,” Phys. Rev. A 85, 022323 (2012). [CrossRef]  

27. J. N. Damask, Polarization Optics in Telecommunications (Springer, 2004).

28. F. De Zela, “Two-component gadget for transforming any two nonorthogonal polarization states into one another,” Phys. Lett. A 376, 1664–1668 (2012). [CrossRef]  

29. D. F. V. James, P. G. Kwiat, W. J. Munro, and A. G. White, “Measurement of qubits,” Phys. Rev. A 64, 052312 (2001). [CrossRef]  

30. D. Rosset, R. Ferretti-Schöbitz, J.-D. Bancal, N. Gisin, and Y.-C. Liang, “Imperfect measurement settings: Implications for quantum state tomography and entanglement witnesses,” Phys. Rev. A 86, 062325 (2012). [CrossRef]  

31. M. Mohammadi, A. M. Brańczyk, and D. F. V. James, “Fourier-transform quantum state tomography,” Phys. Rev. A 87, 012117 (2013). [CrossRef]  

32. M. Mohammadi and A. M. Brańczyk, “Optimization of quantum state tomography in the presence of experimental constraints,” Phys. Rev. A 89, 012113 (2014). [CrossRef]  

33. S. Brierley and S. Weigert, “Constructing mutually unbiased bases in dimension six,” Phys. Rev. A 79, 052316 (2009). [CrossRef]  

34. J. J. Benedetto and M. Fickus, “Finite normalized tight frames,” Adv. Comput. Math. 18, 357–385 (2003). [CrossRef]  

35. Z. Hou, in preparation.

36. Z. Hou, H. Zhu, G.-Y. Xiang, C.-F. Li, and G.-C. Guo, “Error-compensation measurements on polarization qubits,” (2015). http://arxiv.org/abs/1503.00263.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (7)

Fig. 1
Fig. 1 MUB measurement with two different settings: (a) MUB measurement with QWP-HWP setting consisting of a QWP, a HWP and a PBS. (b) MUB measurement with a wave plate (WP) and a PBS.
Fig. 2
Fig. 2 Frame potential Φ(δ) with respect to δ. The frame potential is symmetric about 180°. The inset is the amplified interval of δ where the frame potential vanishes to zero. Only if δ is in the range [111.5°, 141.7°] or its symmetric range about 180° can a complete set of MUB be realized by one wave plate.
Fig. 3
Fig. 3 The numerical solutions of rotation angles in Eq. (6) with respect to the phase δ. There are two classes of solutions with δ between 111.5° and 126.3°, represented by solid and dotted lines, respectively, and only one class of solutions in (126.3°, 141.7°]. Each class of solutions contains four complete sets of MUB, denoted by four different colors.
Fig. 4
Fig. 4 Experimental setup for qubit tomography. The apparatus consists of two parts: state preparation (green) and MUB measurement (pink). The MUB measurement part consists of a polarizing beam splitter and a wave-plate combination which has two choices: (a) TWP and (b) QWP-HWP combination.
Fig. 5
Fig. 5 Systematic error in TWP and QWP-HWP setting for qubit tomography. The dependence of the systematic error in the TWP setting (Fig. 4(a)) and the QWP-HWP setting (Fig. 4(b)) is experimentally measured with respect to the angle errors of TWP (red), QWP (black) and HWP (blue) for three states in Figs. 5(a)–5(c) at p = 0.92. The total systematic error (green) in the QWP-HWP setting is the sum of that due to QWP and HWP. The experimental results denoted as dots coincide with the theoretical calculations (solid lines). Fig. 5(d) plots the total systematic error for all these three states in the two settings and shows that the TWP setting beats the QWP-HWP setting by a factor of two. Error bars are the standard deviation of 100 trials in Monte Carlo simulation with binomial distribution of counting statistics.
Fig. 6
Fig. 6 Experimental setup for two-qubit tomography. A singlet state is prepared via SPDC process with a fidelity of 98%. In quantum state tomography, single-qubit MUB measurements are implemented on both photons with TWP or the combination of QWP and HWP in Figs. 4 (a) and 4 (b). Coincidence events are recorded by a coincidence circuit.
Fig. 7
Fig. 7 Systematic error in TWP and QWP-HWP setting for two-qubit tomography. In (a), systematic errors due to angle errors of QWP (black), HWP (blue), both of them (green) and TWP (red) are numerically simulated and denoted as solid lines. The experimental results due to angle errors of wave plates are plotted as dots for photon 1 and circles for photon 2. The total systematic error for the two photons is plotted in (b). From (b), The TWP setting beats the QWP-HWP setting by about a factor of two. Error bars are the standard deviation of 100 Monte Carlo simulations with multinomial distribution of the counting statistics.

Equations (30)

Equations on this page are rendered with MathJax. Learn more.

| ψ = U ( q , π 2 ) U ( h , π ) | H = ( cos q cos ( q 2 h ) + i sin q sin ( q 2 h ) sin q cos ( q 2 h ) i cos q sin ( q 2 h ) ) ,
U ( θ , δ ) = ( cos 2 θ + e i δ sin 2 θ 1 2 ( 1 e i δ ) sin 2 θ 1 2 ( 1 e i δ ) sin 2 θ sin 2 θ + e i δ cos 2 θ ) .
r ( q , h ) = ( sin 2 q cos ( 4 h 2 q ) , sin ( 4 h 2 q ) , cos 2 q cos ( 4 h 2 q ) ) T
r 1 r 2 = 0 .
r ( θ ) = ( sin 2 δ 2 sin 4 θ sin δ sin 2 θ , 1 2 sin 2 δ 2 sin 2 2 θ ) T .
r ( θ i ) r ( θ j ) = 0 ,
Φ ( δ ) = min θ 1 , θ 2 , θ 3 [ r ( θ 1 ) r ( θ 2 ) ] 2 + [ r ( θ 2 ) r ( θ 3 ) ] 2 + [ r ( θ 1 ) r ( θ 3 ) ] 2 .
r ( t ) = ( 3 4 sin 4 t , 3 2 sin 2 t , 3 4 cos 4 t + 1 4 ) T ,
( Δ r ) 2 = ξ r ξ 2 ( Δ ξ ) 2 ,
r δ 2 = sin 2 2 θ = ( r 2 ) 2 sin 2 δ , r θ 2 = 16 sin 2 δ 2 4 sin 2 δ sin 2 2 θ = 16 sin 2 δ 2 4 ( r 2 ) 2 .
ε 2 = j = 1 3 ( Δ r j ) 2 = j = 1 3 ξ r ξ j 2 ( Δ ξ ) 2 = ξ ( ε ξ ) 2 ( Δ ξ ) 2 , ( ε ξ ) 2 = j = 1 3 r ξ j 2 .
( ε δ ) 2 = j = 1 3 ( r 2 j ) 2 sin 2 δ = 1 sin 2 δ , ( ε θ ) 2 = j = 1 3 16 sin 2 δ 2 4 ( r 2 j ) 2 = 48 sin 2 δ 2 4 .
ε 2 = 1 sin 2 δ ( Δ δ ) 2 + ( 48 sin 2 δ 2 4 ) ( Δ θ ) 2 .
ε 2 = 1.33 ( Δ δ t ) 2 + 32 ( Δ t ) 2 .
tr ( ρ ρ ^ ) 2 = 1 4 p 2 ( ε 1 2 + ε 2 2 ) ,
90 ° θ 1 = 90 ° + θ 1 , θ 2 = θ 3 90 ° .
cos δ sin 2 2 θ 2 + cos 2 2 θ 2 = 0 , sin 2 δ sin 2 2 θ 2 = 1 2 .
θ 1 + θ 2 = 90 ° , θ 3 90 ° = 180 ° θ 3 .
θ 1 + θ 3 = 90 ° , 90 ° θ 2 = θ 2 .
cos 2 δ sin 2 2 θ 1 + cos δ cos 2 2 θ 1 sin 2 δ sin 2 θ 1 = 0 , ( 1 cos δ ) 2 sin 2 2 θ 1 cos 2 2 θ 1 = sin 2 δ sin 2 2 θ 1 + ( cos δ sin 2 2 θ 1 + cos 2 2 θ 1 ) 2 .
x ( x 1 ) y 2 + x = ( 1 x 2 ) y , ( x 1 ) 2 y 4 = ( x 1 ) 2 y 2 1 2
x 2 ( x 1 ) 2 y 4 + 2 x 2 ( x 1 ) y 2 + x 2 = ( x 1 ) 2 ( x + 1 ) 2 y 2 .
y 2 = x 2 2 ( 1 x 2 ) .
3 x 4 + 4 x + 2 = 0 .
r δ q 2 = sin 2 ( 4 h 2 q ) = ( r 2 ) 2 , r δ h 2 = sin 2 2 h , r q 2 = 4 + 4 cos 2 ( 4 h 2 q ) = 8 4 ( r 2 ) 2 , r h 2 = 16 .
( ε δ q ) 2 = j = 1 3 ( r 2 j ) 2 = 1 , ( ε q ) 2 = j = 1 3 8 4 ( r 2 j ) 2 = 20 , ( ε h ) 2 = 48 m ( ε δ h ) 2 = j = 1 3 sin 2 2 h j .
ε 2 = 48 ( Δ h ) 2 + 20 ( Δ q ) 2 + ( Δ δ h ) 2 + ( Δ δ q ) 2 .
ε 2 = 48 ( Δ h ) 2 + 20 ( Δ q ) 2 + 1.5 ( Δ δ h ) 2 + ( Δ δ q ) 2 .
tr ( ρ ρ ^ ) 2 = 1 2 ξ R T ξ s 2 ( Δ ξ ) 2 ,
R t = ( 3 1 1 3 1 1 0 2 2 2 2 ) , R δ t = ( 0 6 6 6 6 0 6 6 6 6 0 3 3 3 3 ) .
Select as filters