Quantifying the x-ray dark-field signal in single-grid imaging

Open Access Open Access

Abstract

X-ray dark-field imaging reveals the sample microstructure that is unresolved when using conventional methods of x-ray imaging. In this paper, we derive a new method to extract and quantify the x-ray dark-field signal collected using a single-grid imaging set-up, and relate the signal strength to the number of sample microstructures, N. This was achieved by modelling sample-induced changes to the shadow of the upstream grid, and fitting experimental data to this model. Our results suggested that the dark-field scattering angle from our spherical microstructures deviates slightly from the theoretical model of $\sqrt {N}$, which was consistent with results from other experimental methods. We believe the approach outlined here can equip quantitative dark-field imaging of small samples, particularly in cases where only one sample exposure is possible, either due to sample movement or radiation dose limitations. Future directions include an extension into directional dark-field imaging.

© 2022 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement

1. Introduction

X-ray imaging has become a widely-adopted technique used to reveal the internal structure of a sample through a non-invasive approach [1]. It has been widely used in clinical practice, material science and security screening [2]. The image contrast in conventional x-ray imaging originates from differences in the amount of attenuation introduced by each material in the sample. For a sample made up of highly attenuating materials, such as bone or metal, conventional x-ray imaging can provide images of high quality. However, when the sample is made up of weakly attenuating materials such as soft biological tissues or plastics, the image contrast will be degraded significantly. This limitation can be addressed by phase contrast x-ray imaging (PCXI), which is a technique that converts the phase shift experienced by the x-ray wavefield as it passes through the sample into intensity modulations to reveal details that are invisible when using conventional x-ray imaging [3]. There are a number of experimental set-ups by which this phase contrast signal can be captured.

In addition to the attenuation and phase signals, the dark-field signal can also be captured using PCXI set-ups, including analyser-based [4], grating interferometers [5], edge-illumination [6,7], single-grating/speckle-based [810] and propagation-based [11]. The dark-field signal originates from small-angle or ultra-small-angle x-ray scattering (SAXS/USAXS) from unresolved sample features and provides a complementary picture of the sample properties, often showing up features that are missing in conventional and phase contrast x-ray images (e.g. [12]). The dark-field signal is useful since it reveals information about sub-micron features in a sample without having to fully resolve those features, which makes it a dose-efficient modality as a detector with large pixels (relative to the sample features) can be used. The dark-field contrast modality can also address the limitation that any measurement of the size of a sample feature will be restricted by the pixel size and the spatial resolution of the imaging system. Therefore, dark-field signal has the potential to play an important role in various fields, especially for medical diagnosis [1315], material science [16,17] and airport screening [18], with particular benefit seen from the fact it is non-invasive. While dark-field images provide significant value in a qualitative sense, it can also be useful to consider quantifying the dark-field signal to extract more information about the sample.

The dark-field signal has been successfully quantified for several dark-field imaging techniques, including measuring changes in the width of the rocking curve with analyser-based imaging (ABI) [4,1921] or equivalently in edge-illumination (EI) [22] and measuring a relative change in the visibility of a reference pattern in grating interferometry (GI) [5,2325]. The dark-field signal extracted from GI is also related to the sample microstructure size via the linear diffusion coefficient in computed tomography [24,25] and the correlation length [2628] for specific set-ups. Higher-order statistical measures of the scattering distribution, e.g. the kurtosis, can also depend on microstructure size [29]. The dark-field signal has been related to the sample microstructure properties in a few experimental tests and applications using these techniques [14,2630]. The dark-field signal described by angular spread can be related to the number of scattering interfaces via von Nardroff’s equation [31]. Note that when material microstructure is elongated and aligned, resulting in an asymmetrical dark-field signal, this directional dark-field signal has been extracted by capturing the dark-field signal at different orientations of the sample relative to the gratings [30].

In this paper, we will focus on single-grid imaging, which is an emerging dark-field imaging technique where the dark-field signal has been extracted qualitatively [8,9,32], but not yet quantitatively. The main advantage of single-grid imaging is that only a single sample exposure is required during the data acquisition process. The short data acquisition time can help in minimising any motion blur and thus makes this technique feasible for dynamic imaging and time sequence imaging. This technique is closely related to single-exposure speckle-based imaging [3335], where the reference grid is simply replaced by a piece of sandpaper to create reference speckle. In this manuscript, we demonstrate the quantification of the x-ray dark-field signal in single-grid imaging and relate the signal to the number of sample microstructures.

We first describe the mathematical modelling of the dark-field signal in single-grid imaging and derive a method for extracting this signal using an explicit cross-correlation approach. This is achieved by modelling the intensity pattern seen behind a grid with a sinusoidal function and looking at how this is reflected in the local cross-correlation between a section of the intensity pattern seen with and without the sample present. We then apply the numerical analysis to synchrotron images, converting visibility loss into an effective scattering angle and relating the angle to the number of microstructures in the sample. Finally, we discuss conditions for quantitative dark-field extraction with a single sample exposure, apply the technique to edge-scattering and to speckle-based images, and look at future directions.

2. Mathematical model for dark-field effects in single-grid imaging

This method of analysis is based on a ‘single-grid’ experimental set-up, as shown in Fig. 1, where an absorption grid is placed in the illuminating beam, some distance upstream of a detector, and a reference image captured. The sample is then introduced close to the absorption grid, such that with propagation it distorts the grid pattern according to the phase properties of the sample and blurs the grid pattern according to the dark-field properties of the sample [8,9,32,36,37].

 figure: Fig. 1.

Fig. 1. Set-up for single-grid imaging with detector being placed at multiple propagation distances, $z$. The sample can be placed immediately upstream or downstream of the grid. The linear relationship between blur width, $d$ and $z$ is described by Eq. 8. The sample introduces attenuation and significant blurring on the grid pattern formed on the detector especially at larger propagation distance. Note that the scattering angle of the x-ray has be exaggerated in this figure for visualisation purposes.

Download Full Size | PDF

2.1 Modelling the intensity behind a grid

We start by modelling the x-ray intensity seen behind the grid with and without the sample, in one dimension. Although for an ideal system, we might model the intensity behind a grid as a square wave, we modelled the intensity of the grid as a sinusoidal function here as we observed that the intensity behind the grid is better modelled by a sinusoid in the experimental data. It is worth noting that any regular intensity oscillations, for example, a smoothed square or triangular wave, captured in an experimental setting (finite point spread function and source size) are likely to be well-modelled by a sinusoid, when sampled at around 7-15 pixels per period (typical for this imaging method to balance spatial resolution and range of sensitivity). This is also an appropriate model for the periodic intensity produced by a phase grid [9,38,39] and as we see in Section 6.2, a speckle produced by a piece of sandpaper [33,34].

We define the intensity seen at the detector in the presence of the grid as a function of position $x$ across the image, $I_{g}(x)$, as a sine function,

$$I_{g}(x) = a \sin \left(\frac{2 \pi x}{p}\right)+b,$$
where $a$ represents the amplitude, $b$ represents the mean and $p$ represents the period of the grid. We will refer to $I_{g}(x)$ as the ‘grid intensity’. It is worth noting that the two parameters, a and b, allow us to model an intensity oscillation between any two values without assuming anything about the visibility of the reference pattern. When the sample is introduced, the intensity captured by the detector is shifted transversely and more importantly, has a lower mean gray value and smaller amplitude oscillations due to the presence of sample. This suggests that we can model the ‘sample-grid intensity’, $I_{sg}(x)$, which refers to the intensity seen at the detector in the presence of the grid and sample, with
$$I_{sg}(x) = A a \sin\left(\frac{2 \pi x}{p}\right) + t b,$$
where $A$ represents the change in the amplitude of the grid intensity oscillations that is introduced by the sample and $t$ represents the transmission of x-ray wavefield passing through the sample, which can have a value ranging from 0 to 1, where 0 means no x-ray wavefield passes through the sample and 1 means the x-ray wavefield is not attenuated. The phase shift of the x-ray wavefield is not taken into consideration here since the sample-grid intensity will effectively be shifted to align with the grid intensity during cross-correlation analysis, as explained later, before we perform the dark-field retrieval. A cross-correlation can be used in single-exposure quantitative single-grid or speckle-based phase retrieval [9,34], to compare the local ‘grid intensity’ with the ‘sample-grid intensity’, and we use the same approach here for dark-field retrieval, for reasons described in the discussion.

The next step is to perform a cross-correlation between $I_{sg}(x)$ and $I_{g}(x)$, and then also $I_{g}(x)$ with itself to allow normalisation of variations in the visibility and local mean intensity of the reference grid image. These two cross-correlations result in $k(i)$,

$$\begin{aligned} k(i) = (I_{sg} \star I_{g}) (i) &= \int^{p}_{0} \int^{p}_{0} \overline{I_{sg}(x-i)} I_{g}(x) \, dx \,dy\\ k(i) &= \frac{1}{2} A a^2 p^{2} \cos \left(\frac{2 \pi i}{p}\right)+b^2 p^{2} t, \end{aligned}$$
and $h(i)$,
$$\begin{aligned} h(i) = (I_{g} \star I_{g})(i) &= \int^{p}_{0} \int^{p}_{0} \overline{I_{g}(x-i)} I_{g}(x) \, dx \, dy\\ h(i) &= \frac{1}{2} a^2 p^{2} \cos \left(\frac{2 \pi i}{p}\right)+b^2 p^{2}, \end{aligned}$$
respectively. The cross-correlation is performed over one period of grid-induced intensity oscillations instead from $-\infty$ to $\infty$ since the sinusoidal function does not converge at infinity, and the cross-correlation is typically performed across one period of the grid within an experimental image.

By comparing the amplitude of $k(i)$ to $h(i)$, we can obtain

$$\frac{\text{Amplitude(k)}}{\text{Amplitude(h)}} =\frac{\frac{1}{2} A a^2 p^{2}}{\frac{1}{2} a^2 p^{2}} = A,$$
and by comparing the mean of $k(i)$ to $h(i)$, we can obtain
$$\frac{\text{Mean(k)}}{\text{Mean(h)}} =\frac{b^2 p^{2} t}{b^2 p^{2}} = t .$$

2.2 Modelling how the dark-field is reflected in the cross-correlation analysis

The dark-field signal, $DF$, extracted from a grating-based technique is defined as the relative change in visibility between the sample-grid intensity (or stepping curve), $V^{s}$, and the grid-intensity (or stepping curve), $V^{r}$ [5],

$$DF = \frac{V^{s}}{V^{r}} = \frac{\frac{I^{s}_{max} - I^{s}_{min}}{I^{s}_{max} + I^{s}_{min}}}{\frac{I^{r}_{max} - I^{r}_{min}}{I^{r}_{max} + I^{r}_{min}}} = \frac{\frac{Aa}{tb}}{\frac{a}{b}} = \frac{A}{t},$$
where we have used Michelson’s definition of visibility, $V = (I_{max} - I_{min})/(I_{max} + I_{min})$, where $I_{max}$ is the local maximum intensity and $I_{min}$ is the local minimum intensity [40].

We now have a way to measure $A$ and $t$ from the grid-only and sample-grid images (Section 2.1), and to use $A$ and $t$ to quantify changes in visibility of the reference pattern (Eq. 7). This would be sufficient to collect a visibility-reduction image to describe the dark-field, as typically used in grating interferometry. However we would like to relate the dark-field to quantities that describe the sample, so the next step is to link visibility to a less set-up-specific quantity. We do this by understanding how the set-up parameters such as the propagation distance and grid period relate to the collected visibility-reduction dark-field signal.

2.3 Modelling how the dark-field signal depends on propagation distance and blur width

The dark-field effect can be modelled as a local blurring of the collected intensity image [41,42], and we can define the blur width geometrically as shown in Fig. 1. Since we assume the effective scattering angle, $\theta$, to be very small, it can be written as,

$$\begin{aligned} \tan\left(\frac{\theta}{2}\right) \approx \frac{\theta}{2} &= \frac{d/2}{z}\\ \theta &= \frac{d}{z}, \end{aligned}$$
which can be rearranged into $d = z \theta$, where $d$ is the blur width and $z$ is the distance between the sample and detector (propagation distance).

We can model the blurring kernel applied to the grid pattern in the presence of sample as a normalised Gaussian function centred at zero, $G(x)$ [43, Fig. 5][20,44],

$$G(x) = \frac{1}{\sqrt{2 \pi } (d/2)} \exp{\left(-\frac{x^2}{2 (d/2)^2}\right)},$$
where we have set mean $\mu = 0$ and we model the blur width, $d$ as twice the root-mean-squared width of the Gaussian (equivalently the standard deviation), $\sigma$. The sample-grid intensity can then be modelled by the convolution between the grid intensity and $G(x)$,
$$I_{sg}(x) = I_{g}(x)* G(x) = a \exp\left(-\frac{2 \pi ^2 (d/2)^2}{p^2}\right) \sin \left(\frac{2 \pi i}{p}\right)+b,$$
where $i$ is the coordinate in the convolution space. The dark-field signal (defined in Eq. (7)) is given by,
$$DF = \frac{V^{s}}{V^{r}} = \frac{a \exp\left(-\frac{2 \pi ^2 (d/2)^2}{p^2}\right)}{b} \frac{b}{a} = \exp \left(-\frac{2 \pi ^2 (d/2)^2}{p^2}\right) = \exp \left(-\frac{2 \pi ^{2} z^{2} \theta^{2}}{4p^2}\right),$$
where we substituted the definition of scattering width, $d = z \theta$. Equation 11 allows us to take measurements as a function of distance to accurately extract the effective scattering angle as illustrated in Fig. 2. Note that a sin or a cos function can be used interchangeably in Eqs. 2 and 10, since the cross-correlation sinusoid is centred during the analysis process. Note also that our model can be applied to images taken with coherent or incoherent x-ray illumination since it describes changes to the intensity pattern formed at the detector rather than immediately downstream of the grid. This also means that the distance between the grid and the source does not need to appear in our model and similarly for the distance between the grid and the detector, provided the grid is placed upstream of the sample (if the opposite is true, we simply replace the sample-to-detector distance $z$ with the grid-to-detector distance in Eq. 8).

 figure: Fig. 2.

Fig. 2. A typical example of the dark-field signal measured as a loss in visibility at one pixel from 27 propagation distances (a) and (c) before and (b) and (d) after smoothing with a mean square kernel of size 13 pixels. Stronger dark-field signals are measured from pixels in the shadow of the sample ((a) and (b)) compared to pixels outside the shadow of the sample ((c) and (d)). The effective scattering angle extracted by fitting Eq. 11 to (a) and (b) were $36.94 \mu$rad and $35.25 \mu$rad, and $0.0006437 \mu$rad and $0.2716 \mu$rad on (c) and (d).

Download Full Size | PDF

2.4 Modelling how the dark-field signal relates to the sample properties

We now have a measure of the dark-field, $\theta$, which is independent of the set-up used to take that measurement. The next step is to relate that measurement to properties of the sample microstructure. It is possible to model the microstructures as many spheres, in which case the average total deviation angle of an x-ray passing through a number of spheres, $\alpha _{0}$ is described by [31]

$$\alpha_{0} = \left(4\delta^{2}N(\log{\frac{2}{\delta} + 1})\right)^{1/2},$$
where $\delta$ is the complex refractive index decrement of the sphere material and $N$ is the number of microstructures an x-ray beamlet encounters while propagating along the optical axis. By assuming the sample microstructures, which in our experiment are microspheres, to have the same diameter, we can say that the total thickness of the microspheres, $T$ is directly proportional to the number of microstructures, $N$. This implies that the effective scattering angle $\theta$ is directly proportional to $\sqrt {T}$ in the case where all the scatterers are of the same size and material.

3. Experimental details

We captured an experimental dataset to test this approach to dark-field retrieval and the quantitative nature of the extracted dark-field measure, over a range of propagation distances and hence set-up sensitivities.

3.1 Experimental setup

The setup of our experiment is the same as the typical single-grid imaging technique, with the addition that we allowed the propagation distance to change as shown in Fig. 1. The experiment was performed at the Australian Synchrotron Imaging and Medical Beamline (IMBL). An attenuating grid (a geological stainless steel sieve with holes of 90 $\mu$m and wires 61 $\mu$m thick), which attenuates 0 - 72% of the 30keV x-rays, was placed 150cm upstream of the sample (as close as the set-up permitted). The sample was allowed to move along the $x$ axis so that it could be moved out of the field of view to capture a grid-only (reference pattern) image. Both the grid and sample were placed together on a table which could be moved to adjust the distance $z$ between the sample and the detector, which was on a separate table. The energy of the x-ray beam used was 30 keV and the exposure time for each image was 650 ms (chosen to fill the dynamic range of the detector). The effective pixel size for the setup was 11.5 $\mu$m, using a 25 $\mu$m thick Gadox phosphor coupled to a pco.edge 5.5 sCMOS detector.

3.2 Sample preparation

Our sample began as a PMMA sample tube containing polystyrene microspheres suspended in 10ml of water, with the spheres making up 2.5% of the volume. Prior to imaging, the samples were spun in a centrifuge for 3 minutes or longer until most of the microspheres precipitated at the bottom of the tubes and formed a solid and an aqueous layer. We then removed the solution using a pipette and left the cap off to allow the liquid to evaporate. Two tubes were imaged, one containing microspheres of 1.0 $\mu$m diameter and the other containing microspheres of 8.0 $\mu$m diameter. Images were captured at propagation distances ranging from 0.2m to 2.7m with an increment of 0.1m and also at 0.25m to improve sampling at short distances. At each propagation distance, we took 30 exposures of the ‘grid intensity’ and 30 of the ‘sample-grid intensity’ and averaged each to improve the signal-to-noise ratio.

4. Numerical analysis/results

Before analysing the images, we performed a few pre-processing steps including adjusting for temporal changes in flux, adjusting for grid movement and calculating the grid period in pixels. We then performed cross-correlation analyses using the spatial mapping method proposed by Morgan et al. [37], which determines the distortion of reference pattern directly in real space on an area-by-area basis. We then fitted Eqs. 3 and 4 to the cross-correlation profiles in $x$- and $y$-direction to extract the amplitude and mean of each profile, while fixing $p$ as the period of the grid and setting $i=0$ as the coordinate of the maximum of the cross-correlation, with sub-pixel accuracy, to account for phase-induced shifts. Examples of this fitting process performed at pixels with and without the presence of sample, along with a histogram to show the confidence of the fitting process are shown in Figs. S2, S3 and S1 respectively in the Supplement 1.

4.1 Extracting the sample-induced visibility loss

The next step following the fitting is to extract the change in visibility induced by the sample from the cross-correlation result. It is worth mentioning that this method is designed for isotropic scattering which gives rise to dark-field signal that is not directional. This means that the dark-field signal in each direction is expected to be the same.

Since our samples are microspheres, we made a reasonable assumption that the blurring of the grid pattern imposed by the sample is symmetric. Following this assumption, we can simplify the 2D cross-correlation result into 1D by focusing on the central row or column of the phase-adjusted cross-correlation when we were trying to fit for the amplitude and mean values. In addition, performing a 2D fit would be more time-consuming and may also be more sensitive to noise. Given that most microstructure is not both elongated and oriented all in the same direction, this approach will work for most samples.

To calculate how the amplitude of the intensity oscillations changes, $A$, we calculate the RMS amplitude in each of the two directions (ie. $\textrm {Amplitude}(k)=\sqrt {\textrm {Amplitude}(k_x)^2+\textrm {Amplitude}(k_y)^2}$), and then follow Eq. 5. If there is indeed a directional dark-field signal, the RMS measure accurately extracts the typical reduction in visibility. To obtain the transmission value, $t$, we followed Eq. 6, and then simply averaged the transmission from each of the $x$- and $y$-direction fits, since no asymmetry should be present in a scalar value. The transmission $t$ was corrected in this dataset by an addition of 0.003 to account for the effect of the wide point spread function of the detector, which resulted in transmission values lower than the unity at areas where the sample is not present. This effect should ideally be corrected by performing a deconvolution between the measured point spread function and the raw images, if the measured point spread function is available. The dark-field signal is then calculated by dividing the fitted $A$ by $t$, as shown in Eq. 7.

4.2 Converting visibility measurements to effective scattering angle

The images taken at different distances are analysed numerically through the same procedure mentioned in the previous sections to give us a total of 27 dark-field images that each show the change in the visibility of the grid pattern induced by the sample. We then extracted the effective scattering angle by plotting how the dark-field visibility loss at each pixel changes with propagation distance, and fitting to Eq. 11, as shown in Fig. 2. While we retrieved the dark-field signal pixel-by-pixel, since the dark-field signal measured across the neighbouring pixels within the same period of the grid were expected to be similar, we chose to smooth the dark-field signal images at each distance with an averaging kernel of the same size as the grid period (13 pixels $\times$ 13 pixels) to reduce the noise in the retrieved dark-field images. The effect of averaging the dark-field image can be seen in Fig. 2, where the dark-field signals at different distances gave a closer fit after averaging (Fig. 2(b) and 2(d)) compared to the more noisy signals without any averaging (Fig. 2(a) and 2(c)).

The dark-field visibility signal remained constant at 1.0 when no sample is present (ie. no loss in visibility, see Fig. 2(c) and 2(d)), but decreases as the propagation distance increases with the presence of sample (Fig. 2(a) and 2(b)). This is due to the fact that the blurring width is directly proportional to the propagation distance.

4.3 Relating the effective scattering angle to sample properties

The effective scattering angle is then related to the thickness of the sample, $T$, which is directly proportional to the number of microstructures, $N$, since the microstructures in our experiment are all the same size. The thickness of the sample (Fig. 3(b)) is recovered from the transmission image obtained from images taken at a propagation distance of 2.7m (Fig. 3(a)) by using the Transport of Intensity Equation (TIE)-based phase retrieval algorithm [45], using $\beta = 9.2028 \times 10 ^{-11}$ and $\delta = 2.6286 \times 10 ^{-7}$ (polystyrene at 30 keV). This provides a way to test if our results follow the trend described by Eq. 12, since $N$, the number of microstructures in projection, is proportional to $T$.

 figure: Fig. 3.

Fig. 3. (a) Transmission image extracted from single-grid images of two tubes of polystyrene microspheres taken at a propagation distance of 2.7m, after correction for the wide point spread function. (b) Projected thickness recovered from (a) using TIE-based single-image phase retrieval [46].

Download Full Size | PDF

5. Results

We performed the curve fitting analysis on the images taken at 27 different propagation distances to extract the change in amplitude, $A$ and transmission, $t$. Examples of $A$ and $t$ extracted from images taken at two different propagation distances - $0.2$m and $1.4$m are shown in Fig. 4(a), 4(b) and Fig. 4(c), 4(d). The dark-field signal, $DF$ is then calculated by dividing $A$ by $t$ as suggested in Eq. 7 and the result is shown in Fig. 4(e) and 4(f).

 figure: Fig. 4.

Fig. 4. Dark-field imaging results obtained from single-grid images of two tubes of polystyrene microspheres, shown here for two different propagation distances - $0.2$m and $1.4$m. The change in amplitude of intensity oscillations, $A$ ((a) and (b)), is divided by the corresponding transmission of x-ray wavefield, $t$ ((c) & (d)) to obtain the dark-field signal, $DF$ ((e) and (f)). The scattering angles (from the sample tube on the left), $\theta$ shown in panel (g) and (h) are extracted from the $DF$ signal in panel (e) and (f) respectively using Eq. 11 while $\theta$ in panel (i) is extracted from the $DF$ images measured at all 27 distances shown in Fig. 2 using Eq. 11. The $\theta$ image extracted from the larger propagation distance has less noise than the smaller propagation distance, where the blurring effect is weak. The $\theta$ image retrieved using dark-field signal measured from 27 distances has less noise compared to $\theta$ image retrieved from single distance.

Download Full Size | PDF

5.1 Effective scattering angle from 27 exposures

To examine the quantitativeness of the extracted dark-field scattering angle, we look at how well this data fits the dark-field scattering model proposed by von Nardroff. Figure 5 plots the scattering angle extracted from the dark-field signal measured at 27 distances using Eq. 11 (the image shown in Fig. 4(i)) as a function of the sample thickness (see Sec. 4.3), corrected to remove the simulated sample tube thickness. We fitted the data with $\theta = m \times {T}^{1/p}$, where $m$ and $p$ were allowed to vary. From Fig. 5, we can see that the data was best fitted with $\theta = m \times ^{2.19}\sqrt {T}$ (see Fig. S5 for how the coefficient of determination ($R^{2}$) value changes with different p values) instead of $\theta = m \sqrt {T}$, where $m$ is a constant, as proposed by von Nardroff [31]. This suggested that the effective scattering angle measured here, $\theta$, is proportional to $^{2.19}\sqrt {N}$.

 figure: Fig. 5.

Fig. 5. The scattering angle, $\theta$ as a function the sample thickness, $T$, which is proportional to the number of microstructures $N$. The plot is fitted with $\theta = m ^{p}\sqrt {T}$, where the exponent of $T$ is allowed to vary, and with $\theta = m \sqrt {T}$, where $m$ is a constant, as suggested by von Nardroff [31].

Download Full Size | PDF

In Fig. 6, we aim to show how well each pixel in the image fits the model. The data points that are located within the range of 4 $\mu$rad from the line of best fit are labelled in cyan and are distributed evenly across the image.

 figure: Fig. 6.

Fig. 6. Colour-coded mapping of the pixel values within the image of the left tube, showing how well each pixel in (a) the $\theta$ image retrieved from $DF$ measured at 27 distances (i.e Fig. 4(i)) are described by our model $\theta = m ^{2.19}\sqrt {T}$. The majority of pixels in (a) is coloured in cyan, indicating that they are well described by the model and sit within the positive and negative $4\mu$rad offset from the model. Note that there are a small amount of data points coloured in pink located at the bottom right of the plot in (b). The thickness and scattering angle pixel values are shown for the entirety of the experimental image in Fig. S4.

Download Full Size | PDF

5.2 Effective scattering angle from a single sample exposure

Given the motivation to minimise imaging time and dose and the single-exposure nature of the single-grid technique, it is interesting to examine how quantitative the results are at a single distance instead of using images from multiple distances. The scattering angle, $\theta$ was extracted from images obtained at propagation distances of 0.2m, 0.7m, 1.4m, 2.1m and 2.7m. Figure 7 shows how the scattering angle extracted at each single distance relates to the sample thickness (i.e number of sample microstructures) along with the zoomed-in images of the grid intensity pattern observed at different distances with the presence of the sample. Examples of the scattering angle extracted at 0.2m and 1.4m are shown in Fig. 4(g) and 4(h) respectively.

 figure: Fig. 7.

Fig. 7. The scattering angle, $\theta$ as a function of the sample thickness, $T$ (which is proportional to $N$), extracted from dark-field signals measured at images taken at a single distance of (a) 0.2m (b) 0.7m (c) 1.4m (d) 2.1m and (e) 2.7m along with the grid intensity pattern collected at the detector behind 16 mm of the 1 $\mu$m microspheres, at corresponding distances, showing the grid is barely visible at the larger distances, hence the dark-field signal is saturating. The line of best fit is $\theta \propto ^{p}\sqrt {T}$, where $p$ corresponds to $2.465$, $2.204$, $2.244$, $2.539$ and $2.627$ respectively. The shorter distances show significant noise and the larger distances show saturation for large thicknesses, suggesting that quantitative dark-field imaging with single exposure needs to be performed with a reasonable (not too short or too long) propagation distance to obtain consistent quantitative results.

Download Full Size | PDF

A significantly wider spread in scattering angle is extracted at shorter distances, such as 0.2m and 0.7m (Fig. 7(a) and 7(b)). A saturation in the dark-field signal/scattering angle is observed from images taken at larger distances such as 2.1m and 2.7m (Fig. 7(d) and 7(e)).

The magnitude of the spread in scattering angle at a given thickness is related to the position of the data points in the plot of $DF$ against $z$ (see Fig. 2(b)). The fit is more robust to noise for data points at larger propagation distances where the blurring effect is strong and easily measured, which results in less uncertainty in the $\theta$ values extracted. However, if the propagation distance is too large and the grid pattern is so smoothed that it reaches comparable visibility to sample features, then the dark-field signal begins to saturate and hence does not correctly predict the scattering angle for those large thicknesses where the signal is strongest. This effect can be seen in the flattening of the curve in Fig. 7(e).

This suggests that the propagation distance plays an important role in quantitative single-exposure dark-field imaging. A general guide for choosing a suitable single-exposure distance is to take the images at a sufficiently large distance that when the sample is introduced, significant blurring can be observed in the intensity pattern formed at the detector (see Fig. 7(b) and 7(c)), but not so large that the intensity pattern is completely blurred out and the dark-field signal saturates (see Fig. 7(d) and 7(e)).

6. Further applications of the approach

While the key aim of this paper was to design an analytical approach to quantitatively retrieve x-ray dark-field from a single-grid set-up, we also want to show how well this approach works in a range of scenarios.

6.1 Dark-field from edges

Dark-field signal is primarily seen from microstructure, but can also be seen from sample edges. To test the ability of our technique to detect dark-field from edges, we examined some high-resolution single-grid images. Figure 8(a) shows a raw image taken using a phase grid to study a gold EM grid and PMMA spheres, also used in [47], using 25 keV x-rays and 0.722 $\mu$m effective pixel size. From the magnified image in Fig. 8(b), we can see that the grid pattern has lower visibility near the edges of the PMMA spheres due to the edge scattering effect. Figure 8(c) shows the dark-field visibility signal, $DF$, extracted from Fig. 8(a) using the algorithm presented in this paper. The blurring of the grid pattern near the edges of the PMMA spheres had been successfully detected as indicated by the stronger signals near the edges of the spheres as shown in Fig. 8(c). It is worth mentioning that dark-field from edges was not observed in the dark-field images extracted from the microsphere sample that this manuscript focuses on. We believe this is due to the fact that the dark-field signals produced by our microspheres were much stronger than the edge effect near the edges of the sample tube and thus the edge effect was not visible in our dark-field images. In addition, the pixel size used in the experiment of Fig. 8 was 0.722 $\mu$m [47], which was only 6% of the pixel size used to image the microsphere sample, and thus provided a higher sensitivity to the dark-field signal. Our algorithm is therefore able to detect the blurring of a grid pattern due to the scattering of the x-ray wavefield from the edges of an object.

 figure: Fig. 8.

Fig. 8. (a) Raw image of a phase grid illuminating a gold EM grating and PMMA spheres [47] (see main text for experimental details). (b) Magnified region from the yellow rectangular area in (a), showing reduced visibility of the grid pattern due to edge scattering. (c) The dark-field signal $DF$ extracted from (a) using the algorithm presented here.

Download Full Size | PDF

6.2 Application to speckle-based imaging

Theoretically, the approach outlined in this paper could be applied to images captured using any reference pattern. This is because the dark-field signal is extracted from the cross-correlation analyses, which works equally well with any type of reference pattern [34,37] and does not require a periodic reference in the way a Fourier technique does. Furthermore, the sine function we use to fit the local intensity is an appropriate model for any local oscillation in intensity. Therefore, we tested our algorithm on a speckle-based dataset [3335], acquired at the European Synchrotron Radiation Facility (ESRF), shown in both Berujon et al., 2016 [48] and Pavlov et al., 2020 [49], where sandpaper with grit size P800 was used in the system as a reference pattern instead of an attenuating grid. The sample used in this experiment was a red currant, imaged with 17keV x-rays, and the effective pixel size of the setup was 5.8 $\mu$m. The sample was placed 0.5m downstream of the sandpaper and 1m upstream of the detector. Images were taken with the sandpaper at 6 different transverse locations with and without the sample.

Figure 9(a) and 9(c) shows the raw speckle-only image and sample-speckle image (along with the magnified image of the yellow-rectangular area in the corresponding image) at one position of the sandpaper. In Fig. 9(d), the sample has locally reduced the visibility of the speckle pattern. We performed the analysis described in this manuscript using the speckle-only image and the sample-speckle image with the sandpaper located in the same position. Here, to reduce the noise in the images, we show the median value of each pixel over the 6 images for both the transmission (Fig. 9(e)) and dark-field (Fig. 9(f)) images. We observed dark-field signal near the edges of the sample, inside the currant where the seed/stem sits, and also from the mounting under the currant. The dark-field image also agrees qualitatively with that seen in Pavlov et al., 2020 [49], where an intrinsic approach is used to capture the dark-field, indicating that our approach is also applicable to speckle-based imaging.

 figure: Fig. 9.

Fig. 9. The raw (a) speckle-only image and (b) magnified region from the yellow rectangular area in (a) [48]. The raw (c) sample-speckle image and (d) magnified region from the yellow rectangular area in (c), where the sample is a red currant (see main text for experimental details). Images of (e) transmission, $t$ and (f) the dark-field signal, $DF$ are extracted using the approach described in this manuscript, and taking the median value over 6 images.

Download Full Size | PDF

7. Discussion

In this paper, we have designed and demonstrated a new approach to extract and quantify the dark-field signal using a single-grid imaging set-up. This has been achieved by constructing a mathematical model (Sec. 2.1) and designing a numerical algorithm (Sec. 4.) to extract quantitative dark-field images. We see qualitative agreement between the results reported in Sec. 5 and dark-field images seen at other set-ups, with signal generated by unresolved microstructure. In addition, we have shown quantitative dark-field signal retrieval from a single-grid PCXI set-up. The recovered dark-field signal is consistent with the model proposed by von Nardroff in that the scattering angle is roughly proportional to the square root of the number of microstructures through which the beam passes. We have successfully applied our algorithm on data obtained using a single-grid imaging setup and on data obtained using speckle-based imaging. Nevertheless, some unexpected behaviour was observed in the dark-field signals we extracted. We have split the remainder of the discussion into a number of sections focused on interesting observations and future directions.

7.1 Accounting for propagation-based edge effects

Some abnormal dark-field values were observed, especially at larger propagation distances, where a small number of pixels near the edges of the sample tube had visibility greater than 1, which was outside the range of the expected values for the dark-field signal ($0-1$). This may be due to the fact that the propagation-based phase contrast effect is generally not taken into consideration in the grating-based imaging techniques, including single-grid imaging. Our model/algorithm has considered the bright fringes near the edges of the sample tubes that comes from the focusing of the x-ray beam as “negative” attenuation due to the increase in mean value or amplitude. One possible way to tackle this problem is by removing the propagation-based phase contrast effect from the raw images prior to the dark-field analysis, using an approach similar to that performed by Groenendijk et al. [47]. In the approach taken by Groenendijk et al., the propagation-based phase effect was simulated by propagating a wavefield with uniform intensity numerically through the angular spectrum approach [50] using phase information retrieved from the images. The simulated propagation-based phase effect was then divided out of the attenuation images in order to perform material decomposition. Another way to solve this issue would be by incorporating the propagation-based phase contrast effects into the retrieval process using the X-ray Fokker-Planck equation [41] as suggested by Morgan & Paganin [42].

7.2 Quantitative dark-field model

A second interesting point to discuss is the extracted quantitative measures for the scattering angle as a function of the number ($N$)/thickness ($T \propto N$) of microstructures of typical thickness $a$. In Sec. 5, we have shown that the scattering angle measured by our data was best-fitted as proportional to $^{2.19}\sqrt {T} \propto ^{2.19}\sqrt {N}$ instead of $\sqrt {N}$ as suggested by von Nardroff [31]. This is not inconsistent with the result obtained by Kitchen et al. [14], who used an analyser-crystal approach to measure the scattering angle from PMMA microspheres and found the angle to be directly proportional to $^{2.2627}\sqrt {N}$. The scattering angles in both cases were proportional to a root with a factor greater than 2 of the number of microstructures in the beam path. We suspected that the deviation of our result from the model proposed by von Nardroff may be due to anomalous diffusion of the scattered x-ray wavefield. To further explore the relationship between the sample microstructures and the scattering angle, we could repeat the analysis on the images obtained with other diameter microstructures.

7.3 Optimal single-exposure imaging

In this experiment, we acquired images at 27 different propagation distances to accurately measure how the scattering angle relates to the number of microstructures in the sample. However, in practice, it is often useful to use a single sample exposure only, to minimise motion blurring due to a moving sample and radiation dose delivered to the sample. A single-exposure approach also allows time-sequence imaging to capture a dynamic process like a biological response to a treatment [51] (noting that repeated processes can use multiple images [52]). Therefore, the question then turns to whether the propagation distance between the sample and the detector can affect the accuracy of the results obtained. Generally, it is not recommended to image the sample with the detector being placed immediately downstream of the sample since this will result in a weak blurring effect and it is also not recommended to image the sample with a large propagation distance since this can result in the grid pattern being blurred out too significantly and the cross-correlation analysis may fail. In the second case, we have a saturated dark-field signal where the visibility of the grid falls below the visibility of other features in the image, and sensitivity to a greater dark-field signal is lost.

In Sections 5 and 5.2, we showed the images for the scattering angles extracted using dark-field signals measured using one distance only, for 0.2m and 1.4m (see Fig. 4(g) and (h)) and the plots showing how these dark-field effective scattering angles measured at a single distance were related to the thickness of the microspheres (see Fig. 7) respectively. At smaller propagation distances, we obtained a wider range of scattering angles for a given thickness compared to larger propagation distances, showing the low sensitivity at these short distances makes measurements less reliable. However, the scattering angle obtained with dark-field signals extracted at shorter distance still varied with sample thickness with a fit that is closely matched with the fit obtained in Fig. 5, despite having a wider range for each sample thickness. For the larger propagation distances, it seems the dark-field is saturated in the sense that we can only just see the grid in the raw images, and the visible intensity variations are largely PBI speckle from the microspheres. In Fig. 7(e), the plot is flattened out for large sample thickness when it should be increasing in dark-field, which may indicate the data is not fitting the model anymore. Figure 2(b) also shows that the visibility signal does not decrease as expected after around 2m propagation, suggesting a saturated dark-field signal. While Fig. 7(d) and 7(e) show the scattering angle measured is consistent up to 5mm thickness, the scattering angle then ‘saturates’ for larger thicknesses. Although Fig. 4(g) and 4(h) had different ranges of values, they are qualitatively similar. In summary, shorter propagation distance can result in weaker dark-field signal with more variability, while larger propagation distance typically results in a stronger dark-field signal with a smaller variability but measurements will not follow the model laid out here if the visibility signal is saturated. Essentially, we can tune the sensitivity of the system to the dark-field signal by changing the propagation distance. Custom tests would be required to determine the optimum propagation distance for each sample, but we would expect that a suitable propagation distance would show a noticable drop in grid visibility across the sample (e.g. 0.5, see Fig. 2).

7.4 Directional dark-field imaging

Our approach to quantifying the dark-field signal could be extended into two dimensions, which would be particularly useful for samples that produce a directional dark-field signal [30]. The grid-only intensity, $I_{g}(x,y)$ can be modelled as a two-dimensional sine function and the sample-grid intensity, $I_{sg}(x,y)$ can be modelled as the convolution between grid-only intensity with a zero-centred asymmetric Gaussian blurring function. A two-dimensional fitting can be performed on the cross-correlation results to extract the magnitude and direction of the blurring contributed by the sample, thus opening up a new area of study on directional dark-field imaging.

7.5 Advantages and disadvantages of this approach

The primary advantage of single-grid imaging is that only a single sample exposure is required for the data acquisition process. This makes the approach more feasible for dynamic imaging and time sequence imaging, compared to the other phase contrast x-ray imaging techniques such as analyser-based and grating-based imaging which require multiple sample exposures to extract the phase and dark-field signals. Another advantage of single-grid imaging is that it has a relatively simple setup compared to analyser-based and grating-based imaging techniques since it only requires one optical element – a grid – and no calibration or alignment is required prior to the data acquisition process. In addition, we are able to tune the sensitivity of the system to the dark-field signal, decreasing the pixel size and/or increasing the propagation distance to increase sensitivity.

We have chosen an approach that uses spatial-mapping/explicit-tracking by cross-correlation, as used for phase retrieval by Morgan et al. [37], instead of a Fourier approach [53], since the Fourier analysis method can fail when the grid frequency overlaps with the sample feature frequency [37]. The spatial mapping method can also provide better spatial resolution than the Fourier analysis (see [9, Fig. 4]) since we compare the windows pixel-by-pixel instead of only focusing on a small area in the Fourier space. The spatial mapping method can also be applied to images taken with other reference patterns, such as the sandpaper as shown in Sec. 6.2.

However, the single-grid imaging technique does have a few disadvantages. One of the major disadvantages is that this technique requires a relatively small pixel size to capture both phase and dark-field signals. Generally, the pixel size has to be smaller than the blurring width to obtain high quality dark-field images. Thus, this technique is best-suited to small samples (e.g. millimetre-to-centimetre scale) due to the requirement for a small pixel size. Although it is technically possible to image large samples, which could be achieved by using a detector with a large area but small pixel size, it may not be feasible due to the cost, availability or radiation dose.

7.6 Future directions

In addition to the directions mentioned above, there are a few other possible directions for future investigation. One subsequent investigation would be to explore how the dark-field signal and the scattering angle change with the microstructure size. This could be done by imaging additional samples over a range of microstructure sizes. We could also investigate the dependence of the dark-field signal on the microstructure material and/or the type of medium surrounding the microstructures.

Other possible future directions include exploring the relationship between the dark-field signal with x-ray energy by performing the experiment with an energy-sensitive detector [54,55]. Finally, an investigation could look at quantifying the edge scattering effects.

8. Conclusion

The x-ray dark-field signal that originates from the SAXS or USAXS of unresolved sample features can provide complementary information on sample properties that is inaccessible from the absorption- and phase-contrast images. A quantitative dark-field signal can be useful in various fields including medical diagnosis [14], material science [56] and airport security [18].

In this manuscript, we proposed a new method to extract and quantify the dark-field signal from a single-grid imaging set-up, utilising a mathematical model and a numerical analysis algorithm. The dark-field signal was quantified by curve-fitting and comparing the cross-correlation results between the grid-only image with itself and between the grid-only image and sample-grid image. The quantitative dark-field signal was then converted into the effective scattering angle and then related to the number of microstructures, $N$, that the beam passes through. The scattering angle produced by our 1.0 $\mu$m microspheres was found to be directly proportional to $^{2.19}\sqrt {N}$, which was roughly consistent with the model proposed by von Nardroff where the scattering angle is proportional to $\sqrt {N}$, and matched closely with analyser-crystal-based measurements obtained by Kitchen et al. [14]. We showed that the algorithm is able to detect edge scattering and is also compatible with speckle-based imaging [3335].

The feasibility of quantifying the dark-field signal using a single sample exposure was investigated in this manuscript. It was observed that the propagation distance has played an important role in this. We hope that this easily-implemented single-grid dark-field imaging approach can be of use in quantitatively imaging mm-cm-sized samples, particularly in the case of time-sequence imaging or dose-sensitive applications.

The analysis could be improved by incorporating the propagation-based phase contrast effects into the model through the X-ray Fokker-Planck equation. Further studies can also be done to find optimal propagation distance for single sample exposure imaging of a given sample, to extend into two-dimensions to extract a directional dark-field signal and to investigate the dark-field signal dependence on sample microstructure size, material and x-ray energy.

Funding

Australian Research Council (FT180100374).

Acknowledgements

We acknowledge useful discussions with David Paganin and recommended references from Marcus Kitchen. We thank Celebrity Groenendijk for sharing the data she had previously published [47], which was analysed in Fig. 8 here and captured at BL20XU at SPring-8 under application 2019A1151. We also thank Sebastien Berujon, Eric Ziegler and Emmanuel Brun for sharing the speckle-based x-ray dataset analysed in Fig. 9, originally published in [48] and also in [57]. The key images in this paper were captured at the Imaging and Medical Beamline at the Australian Synchrotron, part of ANSTO.

Disclosures

The authors declare no conflicts of interest.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Supplemental document

See Supplement 1 for supporting content.

References

1. W. C. Röntgen, “Uber eine neue art von strahlen,” Sitzungsber Phys Med Ges Wurtzburg 9, 132–141 (1895).

2. D. Jordan, “Handbook of x-ray imaging: Physics and technology,” Med. Phys. 47(6), 2577–2578 (2020). [CrossRef]  

3. S. Wilkins, Y. I. Nesterets, T. Gureyev, S. Mayo, A. Pogany, and A. Stevenson, “On the evolution and relative merits of hard x-ray phase-contrast imaging methods,” Philos. Trans. R. Soc., A 372(2010), 20130021 (2014). [CrossRef]  

4. L. Rigon, F. Arfelli, and R.-H. Menk, “Generalized diffraction enhanced imaging to retrieve absorption, refraction and scattering effects,” J. Phys. D: Appl. Phys. 40(10), 3077–3089 (2007). [CrossRef]  

5. F. Pfeiffer, M. Bech, O. Bunk, P. Kraft, E. F. Eikenberry, C. Brönnimann, C. Grünzweig, and C. David, “Hard-x-ray dark-field imaging using a grating interferometer,” Nat. Mater. 7(2), 134–137 (2008). [CrossRef]  

6. M. Endrizzi, P. C. Diemoz, T. P. Millard, J. Louise Jones, R. D. Speller, I. K. Robinson, and A. Olivo, “Hard x-ray dark-field imaging with incoherent sample illumination,” Appl. Phys. Lett. 104(2), 024106 (2014). [CrossRef]  

7. M. Endrizzi, F. Vittoria, L. Rigon, D. Dreossi, F. Iacoviello, P. Shearing, and A. Olivo, “X-ray phase-contrast radiography and tomography with a multi-aperture analyzer,” Phys. Rev. Lett. 118(24), 243902 (2017). [CrossRef]  

8. H. H. Wen, E. E. Bennett, R. Kopace, A. F. Stein, and V. Pai, “Single-shot x-ray differential phase-contrast and diffraction imaging using two-dimensional transmission gratings,” Opt. Lett. 35(12), 1932–1934 (2010). [CrossRef]  

9. K. S. Morgan, P. Modregger, S. C. Irvine, S. Rutishauser, V. A. Guzenko, M. Stampanoni, and C. David, “A sensitive x-ray phase contrast technique for rapid imaging using a single phase grid analyzer,” Opt. Lett. 38(22), 4605–4608 (2013). [CrossRef]  

10. I. Zanette, T. Zhou, A. Burvall, U. Lundström, D. H. Larsson, M. Zdora, P. Thibault, F. Pfeiffer, and H. M. Hertz, “Speckle-based x-ray phase-contrast and dark-field imaging with a laboratory source,” Phys. Rev. Lett. 112(25), 253903 (2014). [CrossRef]  

11. T. Gureyev, D. Paganin, B. Arhatari, S. Taba, S. Lewis, P. Brennan, and H. Quiney, “Dark-field signal extraction in propagation-based phase-contrast imaging,” Phys. Med. Biol. 65(21), 215029 (2020). [CrossRef]  

12. N. Matsunaga, K. Yano, M. Endrizzi, and A. Olivo, “Detection of individual sub-pixel features in edge-illumination x-ray phase contrast imaging by means of the dark-field channel,” J. Phys. D: Appl. Phys. 53(9), 095401 (2020). [CrossRef]  

13. S. Schleede, F. G. Meinel, M. Bech, J. Herzen, K. Achterhold, G. Potdevin, A. Malecki, S. Adam-Neumair, S. F. Thieme, F. Bamberg, K. Nikolaou, A. Bohla, A. Ö. Yildirim, R. Loewen, M. Gifford, R. Ruth, O. Eickelberg, M. Reiser, and F. Pfeiffer, “Emphysema diagnosis using x-ray dark-field imaging at a laser-driven compact synchrotron light source,” Proc. Natl. Acad. Sci. 109(44), 17880–17885 (2012). [CrossRef]  

14. M. J. Kitchen, G. A. Buckley, L. T. Kerr, K. L. Lee, K. Uesugi, N. Yagi, and S. B. Hooper, “Emphysema quantified: mapping regional airway dimensions using 2d phase contrast x-ray imaging,” Biomed. Opt. Express 11(8), 4176–4190 (2020). [CrossRef]  

15. K. Willer, A. Fingerle, W. Noichl, F. De Marco, M. Frank, T. Urban, R. Schick, A. Gustschin, B. Gleich, J. Herzen, T. Koehler, A. Yaroshenko, T. Pralow, G. Zimmermann, B. Renger, A. Sauter, D. Pfeiffer, M. Makowski, E. Rummeny, P. Grenier, and F. Pfeiffer, “X-ray dark-field chest imaging can detect and quantify emphy-sema in copd patients,” medRxiv (2021).

16. F. Yang, F. Prade, M. Griffa, I. Jerjen, C. Di Bella, J. Herzen, A. Sarapata, F. Pfeiffer, and P. Lura, “Dark-field x-ray imaging of unsaturated water transport in porous materials,” Appl. Phys. Lett. 105(15), 154105 (2014). [CrossRef]  

17. J. Glinz, J. Šleichrt, D. Kytỳř, S. Ayalur-Karunakaran, S. Zabler, J. Kastner, and S. Senck, “Phase-contrast and dark-field imaging for the inspection of resin-rich areas and fiber orientation in non-crimp vacuum infusion carbon-fiber-reinforced polymers,” J. Mater. Sci. 56(16), 9712–9727 (2021). [CrossRef]  

18. E. A. Miller, T. A. White, B. S. McDonald, and A. Seifert, “Phase contrast x-ray imaging signatures for security applications,” IEEE Trans. Nucl. Sci. 60(1), 416–422 (2013). [CrossRef]  

19. M. N. Wernick, O. Wirjadi, D. Chapman, Z. Zhong, N. P. Galatsanos, Y. Yang, J. G. Brankov, O. Oltulu, M. A. Anastasio, and C. Muehleman, “Multiple-image radiography,” Phys. Med. Biol. 48(23), 3875–3895 (2003). [CrossRef]  

20. G. Khelashvili, J. G. Brankov, D. Chapman, M. A. Anastasio, Y. Yang, Z. Zhong, and M. N. Wernick, “A physical model of multiple-image radiography,” Phys. Med. Biol. 51(2), 221–236 (2006). [CrossRef]  

21. M. J. Kitchen, D. M. Paganin, K. Uesugi, B. J. Allison, R. A. Lewis, S. B. Hooper, and K. M. Pavlov, “X-ray phase, absorption and scatter retrieval using two or more phase contrast images,” Opt. Express 18(19), 19994–20012 (2010). [CrossRef]  

22. A. Astolfo, M. Endrizzi, G. Kallon, T. P. Millard, F. A. Vittoria, and A. Olivo, “A first investigation of accuracy, precision and sensitivity of phase-based x-ray dark-field imaging,” J. Phys. D: Appl. Phys. 49(48), 485501 (2016). [CrossRef]  

23. Z.-T. Wang, K.-J. Kang, Z.-F. Huang, and Z.-Q. Chen, “Quantitative grating-based x-ray dark-field computed tomography,” Appl. Phys. Lett. 95(9), 094105 (2009). [CrossRef]  

24. M. Bech, O. Bunk, T. Donath, R. Feidenhans, C. David, and F. Pfeiffer, “Quantitative x-ray dark-field computed tomography,” Phys. Med. Biol. 55(18), 5529–5539 (2010). [CrossRef]  

25. S. K. Lynch, V. Pai, J. Auxier, A. F. Stein, E. E. Bennett, C. K. Kemble, X. Xiao, W.-K. Lee, N. Y. Morgan, and H. H. Wen, “Interpretation of dark-field contrast and particle-size selectivity in grating interferometers,” Appl. Opt. 50(22), 4310–4319 (2011). [CrossRef]  

26. F. Prade, A. Yaroshenko, J. Herzen, and F. Pfeiffer, “Short-range order in mesoscale systems probed by x-ray grating interferometry,” EPL 112(6), 68002 (2015). [CrossRef]  

27. R. P. Harti, M. Strobl, B. Betz, K. Jefimovs, M. Kagias, and C. Grünzweig, “Sub-pixel correlation length neutron imaging: Spatially resolved scattering information of microstructures on a macroscopic scale,” Sci. Rep. 7(1), 44588 (2017). [CrossRef]  

28. S. Gkoumas, P. Villanueva-Perez, Z. Wang, L. Romano, M. Abis, and M. Stampanoni, “A generalized quantitative interpretation of dark-field contrast for highly concentrated microsphere suspensions,” Sci. Rep. 6(1), 35259 (2016). [CrossRef]  

29. P. Modregger, M. Kagias, S. C. Irvine, R. Brönnimann, K. Jefimovs, M. Endrizzi, and A. Olivo, “Interpretation and utility of the moments of small-angle x-ray scattering distributions,” Phys. Rev. Lett. 118(26), 265501 (2017). [CrossRef]  

30. T. H. Jensen, M. Bech, O. Bunk, T. Donath, C. David, R. Feidenhans, and F. Pfeiffer, “Directional x-ray dark-field imaging,” Phys. Med. Biol. 55(12), 3317–3323 (2010). [CrossRef]  

31. R. von Nardroff, “Refraction of x-rays by small particles,” Phys. Rev. 28(2), 240–246 (1926). [CrossRef]  

32. E. E. Bennett, R. Kopace, A. F. Stein, and H. Wen, “A grating-based single-shot x-ray phase contrast and diffraction method for in vivo imaging,” Med. Phys. 37(11), 6047–6054 (2010). [CrossRef]  

33. S. Berujon, H. Wang, and K. Sawhney, “X-ray multimodal imaging using a random-phase object,” Phys. Rev. A 86(6), 063813 (2012). [CrossRef]  

34. K. S. Morgan, D. M. Paganin, and K. K. Siu, “X-ray phase imaging with a paper analyzer,” Appl. Phys. Lett. 100(12), 124102 (2012). [CrossRef]  

35. M.-C. Zdora, “State of the art of x-ray speckle-based phase-contrast and dark-field imaging,” J. Imaging 4(5), 60 (2018). [CrossRef]  

36. Y. Takeda, W. Yashiro, Y. Suzuki, S. Aoki, T. Hattori, and A. Momose, “X-ray phase imaging with single phase grating,” Jpn. J. Appl. Phys. 46(No. 3), L89–L91 (2007). [CrossRef]  

37. K. S. Morgan, D. M. Paganin, and K. K. Siu, “Quantitative single-exposure x-ray phase contrast imaging using a single attenuation grid,” Opt. Express 19(20), 19781–19789 (2011). [CrossRef]  

38. J. Rizzi, P. Mercere, M. Idir, P. Da Silva, G. Vincent, and J. Primot, “X-ray phase contrast imaging and noise evaluation using a single phase grating interferometer,” Opt. Express 21(14), 17340–17351 (2013). [CrossRef]  

39. A. Gustschin, M. Riedel, K. Taphorn, C. Petrich, W. Gottwald, W. Noichl, M. Busse, S. E. Francis, F. Beckmann, J. U. Hammel, J. Moosmann, P. Thibault, and J. Herzen, “High-resolution and sensitivity bi-directional x-ray phase contrast imaging using 2d talbot array illuminators,” Optica 8(12), 1588–1595 (2021). [CrossRef]  

40. A. A. Michelson, Studies in Optics (Courier Corporation, 1995).

41. D. M. Paganin and K. S. Morgan, “X-ray fokker–planck equation for paraxial imaging,” Sci. Rep. 9(1), 17537 (2019). [CrossRef]  

42. K. S. Morgan and D. M. Paganin, “Applying the fokker–planck equation to grating-based x-ray phase and dark-field imaging,” Sci. Rep. 9(1), 17465 (2019). [CrossRef]  

43. M. J. Kitchen, D. Paganin, R. A. Lewis, N. Yagi, K. Uesugi, and S. T. Mudie, “On the origin of speckle in x-ray phase contrast images of lung tissue,” Phys. Med. Biol. 49(18), 4335–4348 (2004). [CrossRef]  

44. A. Khromova, F. Arfelli, H. Besch, H. Plothow-Besch, R. Menk, and L. Rigon, “Monte carlo simulation of x-ray multiple refractive scattering from fine structure objects imaged with the DEI technique,” in IEEE Symposium Conference Record Nuclear Science 2004., vol. 6 (IEEE, 2004), pp. 4014–4018.

45. D. Paganin, Coherent X-ray optics, 6 (Oxford University Press on Demand, 2006).

46. D. Paganin, S. C. Mayo, T. E. Gureyev, P. R. Miller, and S. W. Wilkins, “Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object,” J. Microsc. 206(1), 33–40 (2002). [CrossRef]  

47. C. F. Groenendijk, F. Schaff, L. C. Croton, M. J. Kitchen, and K. S. Morgan, “Material decomposition from a single x-ray projection via single-grid phase contrast imaging,” Opt. Lett. 45(14), 4076–4079 (2020). [CrossRef]  

48. S. Berujon and E. Ziegler, “X-ray multimodal tomography using speckle-vector tracking,” Phys. Rev. Appl. 5(4), 044014 (2016). [CrossRef]  

49. K. M. Pavlov, D. M. Paganin, H. T. Li, S. Berujon, H. Rougé-Labriet, and E. Brun, “X-ray multi-modal intrinsic-speckle-tracking,” J. Opt. 22(12), 125604 (2020). [CrossRef]  

50. D. McMorrow and J. Als-Nielsen, Elements of Modern X-ray Physics (Wiley Hoboken, 2011).

51. K. S. Morgan, M. Donnelley, N. Farrow, A. Fouras, N. Yagi, Y. Suzuki, A. Takeuchi, K. Uesugi, R. C. Boucher, K. K. Siu, and D. W. Parsons, “In vivo x-ray imaging reveals improved airway surface hydration after a therapy designed for cystic fibrosis,” Am. J. Respir. Crit. Care Med. 190(4), 469–472 (2014). [CrossRef]  

52. R. Gradl, K. Morgan, M. Dierolf, C. Jud, L. Hehn, B. Günther, W. Möller, D. Kutschke, L. Yang, T. Stoeger, D. Pfeiffer, B. Gleich, K. Achterhold, O. Schmid, and F. Pfeiffer, “Dynamic in vivo chest x-ray dark-field imaging in mice,” IEEE Trans. Med. Imaging 38(2), 649–656 (2019). [CrossRef]  

53. M. Takeda, H. Ina, and S. Kobayashi, “Fourier-transform method of fringe-pattern analysis for computer-based topography and interferometry,” J. Opt. Soc. Am. 72(1), 156–160 (1982). [CrossRef]  

54. K. Taphorn, F. De Marco, J. Andrejewski, T. Sellerer, F. Pfeiffer, and J. Herzen, “Grating-based spectral x-ray dark-field imaging for correlation with structural size measures,” Sci. Rep. 10(1), 13195 (2020). [CrossRef]  

55. T. Sellerer, K. Mechlem, R. Tang, K. A. Taphorn, F. Pfeiffer, and J. Herzen, “Dual-energy x-ray dark-field material decomposition,” IEEE Trans. Med. Imaging 40(3), 974–985 (2021). [CrossRef]  

56. J. Valsecchi, M. Strobl, R. P. Harti, C. Carminati, P. Trtik, A. Kaestner, C. Grünzweig, Z. Wang, K. Jefimovs, and M. Kagias, “Characterization of oriented microstructures through anisotropic small-angle scattering by 2d neutron dark-field imaging,” Commun. Phys. 3(1), 42 (2020). [CrossRef]  

57. K. M. Pavlov, D. M. Paganin, K. S. Morgan, H. T. Li, S. Berujon, L. Quénot, and E. Brun, “Directional dark-field implicit x-ray speckle tracking using an anisotropic-diffusion Fokker-Planck equation,” Phys. Rev. A 104(5), 053505 (2021). [CrossRef]  

Supplementary Material (1)

NameDescription
Supplement 1       LaTex file that contains five figures as the supplementary material for the manuscript

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (9)

Fig. 1.
Fig. 1. Set-up for single-grid imaging with detector being placed at multiple propagation distances, $z$. The sample can be placed immediately upstream or downstream of the grid. The linear relationship between blur width, $d$ and $z$ is described by Eq. 8. The sample introduces attenuation and significant blurring on the grid pattern formed on the detector especially at larger propagation distance. Note that the scattering angle of the x-ray has be exaggerated in this figure for visualisation purposes.
Fig. 2.
Fig. 2. A typical example of the dark-field signal measured as a loss in visibility at one pixel from 27 propagation distances (a) and (c) before and (b) and (d) after smoothing with a mean square kernel of size 13 pixels. Stronger dark-field signals are measured from pixels in the shadow of the sample ((a) and (b)) compared to pixels outside the shadow of the sample ((c) and (d)). The effective scattering angle extracted by fitting Eq. 11 to (a) and (b) were $36.94 \mu$rad and $35.25 \mu$rad, and $0.0006437 \mu$rad and $0.2716 \mu$rad on (c) and (d).
Fig. 3.
Fig. 3. (a) Transmission image extracted from single-grid images of two tubes of polystyrene microspheres taken at a propagation distance of 2.7m, after correction for the wide point spread function. (b) Projected thickness recovered from (a) using TIE-based single-image phase retrieval [46].
Fig. 4.
Fig. 4. Dark-field imaging results obtained from single-grid images of two tubes of polystyrene microspheres, shown here for two different propagation distances - $0.2$m and $1.4$m. The change in amplitude of intensity oscillations, $A$ ((a) and (b)), is divided by the corresponding transmission of x-ray wavefield, $t$ ((c) & (d)) to obtain the dark-field signal, $DF$ ((e) and (f)). The scattering angles (from the sample tube on the left), $\theta$ shown in panel (g) and (h) are extracted from the $DF$ signal in panel (e) and (f) respectively using Eq. 11 while $\theta$ in panel (i) is extracted from the $DF$ images measured at all 27 distances shown in Fig. 2 using Eq. 11. The $\theta$ image extracted from the larger propagation distance has less noise than the smaller propagation distance, where the blurring effect is weak. The $\theta$ image retrieved using dark-field signal measured from 27 distances has less noise compared to $\theta$ image retrieved from single distance.
Fig. 5.
Fig. 5. The scattering angle, $\theta$ as a function the sample thickness, $T$, which is proportional to the number of microstructures $N$. The plot is fitted with $\theta = m ^{p}\sqrt {T}$, where the exponent of $T$ is allowed to vary, and with $\theta = m \sqrt {T}$, where $m$ is a constant, as suggested by von Nardroff [31].
Fig. 6.
Fig. 6. Colour-coded mapping of the pixel values within the image of the left tube, showing how well each pixel in (a) the $\theta$ image retrieved from $DF$ measured at 27 distances (i.e Fig. 4(i)) are described by our model $\theta = m ^{2.19}\sqrt {T}$. The majority of pixels in (a) is coloured in cyan, indicating that they are well described by the model and sit within the positive and negative $4\mu$rad offset from the model. Note that there are a small amount of data points coloured in pink located at the bottom right of the plot in (b). The thickness and scattering angle pixel values are shown for the entirety of the experimental image in Fig. S4.
Fig. 7.
Fig. 7. The scattering angle, $\theta$ as a function of the sample thickness, $T$ (which is proportional to $N$), extracted from dark-field signals measured at images taken at a single distance of (a) 0.2m (b) 0.7m (c) 1.4m (d) 2.1m and (e) 2.7m along with the grid intensity pattern collected at the detector behind 16 mm of the 1 $\mu$m microspheres, at corresponding distances, showing the grid is barely visible at the larger distances, hence the dark-field signal is saturating. The line of best fit is $\theta \propto ^{p}\sqrt {T}$, where $p$ corresponds to $2.465$, $2.204$, $2.244$, $2.539$ and $2.627$ respectively. The shorter distances show significant noise and the larger distances show saturation for large thicknesses, suggesting that quantitative dark-field imaging with single exposure needs to be performed with a reasonable (not too short or too long) propagation distance to obtain consistent quantitative results.
Fig. 8.
Fig. 8. (a) Raw image of a phase grid illuminating a gold EM grating and PMMA spheres [47] (see main text for experimental details). (b) Magnified region from the yellow rectangular area in (a), showing reduced visibility of the grid pattern due to edge scattering. (c) The dark-field signal $DF$ extracted from (a) using the algorithm presented here.
Fig. 9.
Fig. 9. The raw (a) speckle-only image and (b) magnified region from the yellow rectangular area in (a) [48]. The raw (c) sample-speckle image and (d) magnified region from the yellow rectangular area in (c), where the sample is a red currant (see main text for experimental details). Images of (e) transmission, $t$ and (f) the dark-field signal, $DF$ are extracted using the approach described in this manuscript, and taking the median value over 6 images.

Equations (12)

Equations on this page are rendered with MathJax. Learn more.

I g ( x ) = a sin ( 2 π x p ) + b ,
I s g ( x ) = A a sin ( 2 π x p ) + t b ,
k ( i ) = ( I s g I g ) ( i ) = 0 p 0 p I s g ( x i ) ¯ I g ( x ) d x d y k ( i ) = 1 2 A a 2 p 2 cos ( 2 π i p ) + b 2 p 2 t ,
h ( i ) = ( I g I g ) ( i ) = 0 p 0 p I g ( x i ) ¯ I g ( x ) d x d y h ( i ) = 1 2 a 2 p 2 cos ( 2 π i p ) + b 2 p 2 ,
Amplitude(k) Amplitude(h) = 1 2 A a 2 p 2 1 2 a 2 p 2 = A ,
Mean(k) Mean(h) = b 2 p 2 t b 2 p 2 = t .
D F = V s V r = I m a x s I m i n s I m a x s + I m i n s I m a x r I m i n r I m a x r + I m i n r = A a t b a b = A t ,
tan ( θ 2 ) θ 2 = d / 2 z θ = d z ,
G ( x ) = 1 2 π ( d / 2 ) exp ( x 2 2 ( d / 2 ) 2 ) ,
I s g ( x ) = I g ( x ) G ( x ) = a exp ( 2 π 2 ( d / 2 ) 2 p 2 ) sin ( 2 π i p ) + b ,
D F = V s V r = a exp ( 2 π 2 ( d / 2 ) 2 p 2 ) b b a = exp ( 2 π 2 ( d / 2 ) 2 p 2 ) = exp ( 2 π 2 z 2 θ 2 4 p 2 ) ,
α 0 = ( 4 δ 2 N ( log 2 δ + 1 ) ) 1 / 2 ,
Select as filters