Coherence-controlled holographic microscopy in diffuse media

Open Access Open Access

Abstract

Low-coherence interferometric microscopy (LCIM) enables to image through scattering media by filtration of ballistic light from diffuse light. The filtration mechanism is called coherence gating. We show that coherence-controlled holographic microscope (CCHM), which belongs to LCIM, enables to image through scattering media not only with ballistic light but also with diffuse light. The theoretical model was created which derives the point spread function of CCHM for imaging through diffuse media both with ballistic and diffuse light. The results of the theoretical model were compared to the experimental results. In the experiment the resolution chart covered by a ground glass was imaged. The experimental results are in the good agreement with the theoretical results. It was shown both by experiments and the theoretical model, that with ballistic and diffuse light we can obtain images with diffraction limited resolution.

© 2014 Optical Society of America

1. Introduction

When imaging through turbid media, such as cell structures, the image is degraded by the light scattered out of the object plane. The aim of some light microscopy techniques is to exclude this multiply scattered (diffuse) light from forming the image. In order to do that confocal microscopy [15] can be used. The other possibility, how to filter the scattered light is to use the different properties of ballistic and diffuse light.

In biology, there are many techniques which do not aim to eliminate diffuse light, but on the contrary they can use diffuse light for obtaining the information from specimen. For example, the so called continuous-wave imaging [6, 7], techniques that use diffuse photon density waves [8] etc. However, these techniques have very often much lower resolution than classic light microscopy.

Optical medical imaging techniques usually use diffuse light for imaging and they are summarized in few comprehensive publications [9,10]. The resolving power of these techniques using diffuse-light is much worse than in the case of ballistic-light techniques. Diffuse-light techniques, however, have their main advantage consisting in the fact that they can image through much thicker and denser media. The resolution and also the maximum thickness of studied objects are in the range of millimeters to centimeters depending on the particular technique used. In contrast to it the ballistic techniques have the diffraction limited resolution.

Besides confocal microscopy the interference techniques are used for imaging through diffuse media using filtering of ballistic light from diffuse light. These interference techniques, with the exception as for example phase conjugation technique [11, 12], usually use low-coherence light source. Therefore they can be called low-coherence interferometry (LCI) techniques. The mechanism responsible for filtering ballistic light from diffuse light in LCI is called coherence gating [3, 13, 14]. It means that diffuse light from the object arm interacts with light from the reference arm outside coherence volume. Therefore the two beams cannot interfere.

The ability of LCI to image through various diffuse media has been demonstrated in many publications already [3], [1321]. Typical diffuse media used in optical publications are for example the solution of milk in water [15], latex spheres in water [4]. However the most often used dry diffuse medium is ground glass [13, 16, 17]. All different types of biological material can be used as a diffuse medium, for example a collagen gel [18] or a soft tissue [14]. The historical overview of holographic techniques using low-coherence sources for imaging through diffuse media is a part of publication [14]. A very specific group of LCI is optical coherence tomography (OCT) [19] and optical coherence microscopy (OCM) [20]. Their main significance lies in the fact that they can be used in-vivo and especially OCT is nowadays a common technique used in ophthalmology.

Although the experimental results of imaging through diffuse media with LCI are currently part of many publications, the quantitative description of the overall imaging process is quite rare yet. Only few publications are the exceptions. In [14] and [21], the influence of diffuse medium on the point spread function (PSF) of LCI microscope is discussed. It is shown here schematically how PSF is altered by a diffuse medium, but there is no quantitative expression for PSF. In [3] and [15], PSF is determined experimentally by imaging a resolution test chart. In both [3] and [15], it is concluded that resolving power of LCI is reduced when imaging through diffuse media. The main reason for that is very poor level of signal to noise ratio (SNR). Therefore averaging of more images was utilized which improved resolution to the diffraction limit. Again no computation of PSF was provided. Recently, there appeared few papers which deal with computations of PSF in digital holography [2224]. Authors here are using the PSF computations for deconvolution in order to improve the resolving power. Anyway, nowhere to our knowledge, there has appeared any publication so far describing a possibility of imaging by diffuse light in an LCI microscope.

In the present publication, which links to our previous work [25] we show that in an off-axis LCI microscope which we call coherence-controlled holographic microscope (CCHM), we are able to image through diffuse media both with ballistic and diffuse light. This statement is supported both by the theoretical model of PSF and by the experimental results which prove the theory. As a result of both the theoretical and the experimental parts we show here that by imaging in CCHM we are able to image with both ballistic and diffuse light preserving the diffraction limited resolution. The largest limitation is mainly low SNR, therefore this type of imaging is possible only for certain level of the diffuser strength.

As the aim of the article is to explain the imaging principle clearly, the model of a diffuser in this paper is limited to thin (one layer) diffuser only.

2. Theory

2.1. Optical setup of CCHM

The principle of CCHM (Fig. 1) is described in detail in [26, 27]. Light coming from an extended and polychromatic source S is split by the beam splitter BS1 into two arms, the object and the reference arm. Both arms are put together again in the output plane OP where the detector D is placed. The optical path difference between the reference and the object arm, measured from beam splitter to the output plane OP is lower than the coherence length.

 figure: Fig. 1

Fig. 1 Optical setup of CCHM. Incoherent light source (S), relay lens (L), beam splitters (BS), mirrors (M), condensers (C), object plane (Sp), reference plane (R), microscope objectives (O), tube lenses (TL), diffraction grating (DG), output lens (OL), output plane (OP), detector (D). Figure adapted, with permission, from [27].

Download Full Size | PDF

In the object arm, a light beam goes through a specimen and the beam comes to the output plane in the direction of its normal line. In the reference arm, the beam goes through a reference object and the beam is diffracted on the diffraction grating DG in such a way that the reference beam enters the output plane OP at an angle β, while angle β depends on the beam wavelength. The output plane OP is optically conjugated with the plane of the diffraction grating DG and simultaneously with the object planes Sp and R of the objective lenses O1 and O2, respectively.

Due to off-axis arrangement of the microscope an image-plane hologram with a spatial carrier frequency fc is formed on the detector D. This fact enables one to reconstruct the complex amplitude of object wave just from a single hologram [17, 28]. Moreover, holograms created in CCHM are achromatic and spatially invariant [29] thanks to the fact that diffraction grating is placed in a plane conjugated with the object plane.

2.2. Principle of imaging in CCHM through diffuse medium

The principle of imaging in CCHM through diffuse media is explained with the aid of the simplified drawings in Fig. 2. To illustrate the principle more easily the output plane OP is depicted separately in each arm. It follows from the Van Cittert-Zernike theorem, that if we consider the light in the object plane of the condensers C1, C2 (see Fig. 1) to be spatially incoherent, then we can consider the sections IS1 and IS2 which are conjugated with the object planes Sp and R, respectively, to be spatially incoherent sources (see [30]).

 figure: Fig. 2

Fig. 2 Simplified drawings for explanation of principle of imaging in CCHM through diffuse medium by ballistic and diffuse light. IS1, IS2...imaginary sources in object planes of the condensers C1, C2. Sp, R...object planes of microscope objectives O1, O2. D...diffuser. AO, AR...images of point A in the output plane OP. AB...ballistic image of point A. AD...diffuse image of point A. Dotted line...scattered light, continuous line...unscattered light. a) Imaging without diffuser. b) Imaging with diffuser by ballistic light, mutual shift Δxi = 0. c) Imaging with diffuser by diffuse light, reference arm is shifted by nonzero Δxi.

Download Full Size | PDF

In the reference arm, any point A of imaginary source IS2 is imaged in the output plane as point AR. The light forming this image AR can interfere only with that light from the object arm, which is emitted from the conjugated point A of imaginary source IS1, because any point A of IS1 is mutually coherent only with the corresponding point A in IS1. This point A is imaged in the object arm as a point A0, which coincides with AR.

If the specimen is planar and if it is placed in the object plane Sp, then the coincidence of points A0 and AR is valid for all points of the sources IS1, IS2 and thus for the whole field of view (Fig. 2(a)). It means, that all the light transmitted or scattered by the specimen contributes to interference, thus forming the signal.

However the situation is changed, if a diffuser is placed behind the specimen. In the reference arm the situation remains unchanged, but in the object arm the light coming from A is scattered by the diffuser across the output plane.

Only a small portion of light goes through diffuser without being scattered. This light is called “ballistic” light and it preserves its original trajectory. The ballistic light creates the ballistic image AB which coincides with both AO and AR (Fig. 2(b)). The rest of the light passing through the diffuser is called “diffuse light”. Because of coherence only the ballistic light in LCI can interfere with the light from the reference arm and thus coherence gating takes place [13, 28].

Diffuse light therefore creates only the unwanted background. But let’s assume now, that the reference arm is laterally shifted so much that AR is shifted by Δxi in the output plane as shown in the Fig. 2(c). Then only the diffuse light which is emitted from A and scattered by the diffuser into the point AD which is coincident with AR can interfere with AR, thus creating the signal.By this way it is possible to use diffuse light for forming the image in LCI. In contrast to usual diffuse light techniques, which use both ballistic and diffuse light at once, the above described principle uses either ballistic light (when Δxi = 0) or diffuse light (when Δxi ≠0). This intuitive explanation shows the importance of the mutual shift Δxi, which will be used in following computations.

2.3. CCHM signal computation

Let us assume that extended source S is quasi-monochromatic and let uo(qti,Kt) and ur(qti,Kt)exp(i2πfcx) be complex amplitudes of object and reference waves at the point qti=(xi,yi) in the output plane due to emission of an imaginary point source which is a part of extended source S. Then we can write the intensity in qti as a result of emission of the imaginary point source, as follows

i(qti,Kt)=|uo(qti,Kt)+ur(qti,Kt)exp(i2πfcxi)|2==|uo(qti,Kt)|2+|ur(qti,Kt)|2+uo(qti,Kt)ur*(qti,Kt)exp(i2πfcxi)++uo*(qti,Kt)ur(qti,Kt)exp(i2πfcxi),
where 2πK = 2π(Kt, Kz) is a wave vector. Kt = (Kx, Ky) and K = |K| = 1/λ is a wavenumber. Term fc symbolizes a spatial carrier frequency of the hologram in the output plane [27]. Thanks to the use of Köhler illumination, the lateral component Kt = (Kx, Ky) of the vector K = (Kt, Kz) defines the position of the point source. Each of these imaginary point sources creates a plane wave in the object space with a different vector Kt. Thus the contribution of all point sources creating the whole source S can be computed by integration of (1) over all lateral components Kt. In our computations, we will not integrate the whole summation (1), but only the third term uour*, which we will call a signal. This is usual in holography, as it contains the complex object amplitude. Signal from the whole source w(qti) is computed as follows:
w(qti)=(Kx2+Ky2)1/2<NAill/λuo(qti,Kx,Ky)ur*(qti,Kx,Ky)dKxdKy,
where NAill = n sinα, n is the index of refraction in object space, α is the aperture angle of the beam entering object planes Sp and R. λ is the beam wavelength in vacuum.

2.4. The complex amplitude in the object arm

The analysis is carried out in two dimensions, x and z, as the numerical computations for three dimensional case are too demanding for a conventional computer. The lateral dimension y is dropped, however the analysis we develop is readily extended to the y dimension also.

Source, object, diffuser and output planes change into straight lines, anyway the term “planes” will be preserved for the sake of clarity. Vector K becomes two dimensional K = (Kx, Kz) and signal will be expressed then as follows:

w(xi)=NAill/λNAill/λuo(xi,Kx)ur*(xi,Kx)dKx.

In order to express the complex amplitude of the object wave in the output plane we used the Fresnel (paraxial) approximation. Analysis could be provided non-paraxially as well, which would be more precise, but too complicated for purpose of this publication. The paraxial model gives us simple but still good insight into the studied problem. The analysis was made with the aid of the simplified drawing of the object arm in Fig. 3. The objective was approximated by thin lens L with a transmission function derived by Goodman in [31]. The numerical aperture of the objective is determined by an aperture Ap with radius Rp. The aperture is placed in the back focal plane of lens L, as we assume the objective is telecentric. Aperture Ap is not depicted in Fig. 1 since it is a part of the microscope objective. The influence of aperture is expressed by transmission function rect(xf2Rp), where rect is a Rectangle function (see [31]).

 figure: Fig. 3

Fig. 3 Simplified drawing of the object arm for computation of the amplitude in the output plane when imaging a planar object placed in the object plane Sp. In this Fig. the object is a point aperture at coordinate x0 = x0N. D...diffuser, L...thin lens substituting the objective, Ap...aperture diaphragm in the back focal plane of thin lens, OP...output plane.

Download Full Size | PDF

A spherical wave emitted by a point source is transformed by condenser and collector lenses into plane wave having in object space a wave vector 2πK(Kx, Kz). This plane wave passes through a planar object placed in the object plane Sp, having the transmission function t0(x0). This wave is then scattered by a thin (one layer) diffuser D with transmission function td(xd), which is defined numerically. Then it propagates through aperture Ap into the output plane OP, where the wave is studied. The propagation between individual planes is computed with the use of the Fresnel propagators.

The output plane OP in Fig. 3 is depicted in the place of the image plane of the objective. In fact, in a CCHM there are two more optical components between objective and the output plane - output lens and tube lens. However, these two lenses can transmit all frequencies transmitted by the objective. Therefore placing the output plane into the image plane of the objective doesn’t introduce any frequency loss.

2.5. Computation of PSF in the object arm

In this part we will demonstrate the PSF of the object arm computed according to Sec. 2.4. If an opaque screen with a point aperture is placed in the object plane Sp at coordinate x0 = x0N (Fig. 3), then its transmission function is t0(x0) = δ (x0x0N). The complex amplitude of the object wave in the output plane is then denoted by ηobj(xi, Kx) and it is equal to

ηobj(xi,Kx)=Cexp(i2πKxx0N)exp(iπKx0N2ad)exp(iπKxi2z)××td(xd)Iexp[iπKad(2x0Nxd+xd2)]dxd,
where
I=F[Ee(RpB2A)Ee(RpB2A)],
A=fal2f2+12z, B=xdf+xiz, C=iK4exp[i2πK(ad+al+f+nΔ0+z)]adalfz, F=exp(iπKB22A)14KA, where n is the refraction index of the lens L and Δ0 is the lens thickness measured along the optical axis.

The function Ee(x) is a sum of Fresnel integrals [32]

Ee(x)=C(x)+iS(x)=0xexp(iπ2t2)dt

The exact meaning of ηobj(xi, Kx) is a complex amplitude in the output plane in the object arm at coordinate xi when imaging a point aperture placed in the object plane at coordinate x0N. This point aperture is illuminated by plane wave with a wave vector 2πK = 2π(Kx, Kz). In order to call ηobj PSF, both linearity and spatial invariance in ηobj must be fulfilled. The equation (4) shows, that imaging process described by ηobj(xi, Kx) is linear, but it is not spatially invariant, because ηobj(xi, Kx) doesn’t depend on the difference x0Nxi. Terms xi, x0N appear in the formula (4) independently. Anyway, as it will be shown by simulations later, within certain range of x0N the ηobj only shifts correspondingly to the relevant x0N while the shape of ηobj is preserved. For values used in experiments of this work ηobj can be regarded as PSF.

2.6. Computation of the CCHM PSF

PSF of the whole microscope hCCHM can be computed if we replace uo(xi, Kx) in Eq. (3) by ηobj(xi, Kx) (Eq. (4)). Then the PSF of the whole microscope hCCHM is expressed as

hCCHM(xi)=NAill/λNAill/ληobj(xi,Kx)ur*(xi,Kx)dKx,
where ur(xi, Kx) is the complex amplitude of the reference wave.

In contrast to the amplitude of the object wave, the reference wave is independent of the object. If we do an analogy of Eq. (4) for the reference arm, it means without diffuser (td = 1) and without any object (t0 = 1), we will obtain a simple formula.

ur(xi,Kx)=Crefexp(i2πKxxiM),
where M is an absolute value of magnification between the output plane and the object plane. Cref is a complex constant. Because of convenience from now instead of xi coordinate we will use the conjugated coordinate x0 = −xi/M. The advantage of using x0 instead of xi lies in the fact, that the results are related directly to the object plane which makes the results interpretation straightforward.

In order to simplify the formula (7) for hCCHM we will make factorization of (4)

ηobj(x0,Kx)=hobj(x0)exp(i2πKxx0N),
where hobj(x0) contains all parts of ηobj(x0, Kx) (4) which are independent of Kx. The PSF of the whole microscope can therefore be written as
hCCHM(x0)=hobj(x0)CrefNAill/λNAill/λexp[i2πKx(x0Nx0)]dKxhref(x0).
The integral in the last formula can be assigned as href because it is equal to
href(x0)=sin[NAill(x0x0N)2π/λ]NAill(x0x0N)2π/λ,
which equals the amplitude of an image of a point created by an objective with numerical aperture NAill. Here it is important to mention, that href is not PSF of the reference arm in the real sense of the word. But similar indication was used for example in publications [14,21] and we consider it as very illustrative. If we omit the constant Cref, then we can write the PSF of the whole microscope as
hCCHM(x0)=hobj(x0)href*(x0).

The use of the complex conjugation href*(x0) is pointless, while href(x0) is a real function. In the following we will therefore write only href(x0).

If no diffuser is placed in the object arm and if NAill is equal to the objective numerical aperture NAob=Rpf, then

hCCHM(x0)=|hobj(x0)|2=|href(x0)|2.

For clarity we will call the functions hobj(x0) “the object arm PSF” and hCCHM(x0) “the microscope PSF”.

2.7. Image shift

For imaging in CCHM by diffuse light we will use various lateral shifts between the reference and the object arm, as described above in 2.2 and in the 2. The mutual shift is performed by shifting the microscope objective in the reference arm laterally. If we create a mutual shift of images Δxi in the output plane, which will be denoted by conjugated Δx0, then the microscope PSF hCCHM is expressed as

hCCHM(x0,Δx0)=hobj(x0)href(x0Δx0).

2.8. Numerical computation of the object arm PSF and the microscope PSF

A thin ground glass was used in experiments as a diffuser. The transmission function of such diffuser can be described by a function td(xd) = exp[ikΦ(xd)] (see [33,34]). Φ(xd) is an optical path difference caused by the diffuser at xd, where xd is the lateral coordinate of the diffuser surface. Φ(xd) has a Gaussian distribution with a mean value μ = 0 and variance σ determining maximum range of Φ(xd). The coordinate xd is divided into small segments of length 0.1 μm which equals approximately 1/6 of the wavelength λ = 650nm used for computations. The segments of xd are therefore sufficiently small in comparison to the resolution of the microscope. By modifying σ value the strength of the diffuser is changed. The σ values used in the model are chosen within the interval 0.00 − 0.84λ. In all numerical computations as well as in the experiments the numerical aperture of illuminating beam NAill was chosen to be equal to the numerical aperture of the microscope objective NAob = 0.25. In order to compute the “microscope PSF”as well as “the object arm PSF” the above parameters were inserted into the model of PSF derived in 2.5 section.

2.9. The results of the numerical model

The parameters of the model were adjusted in such a way that they correspond to the experiment described in part 3. The results of the model for axial point x0N = 0 are depicted in Fig. 4 where in the left column of this Fig. the absolute value of the object arm PSF |hobj(x0)| together with |href(x0)| or |href(x0 − Δx0)| are shown. In the right column there is the absolute value of microscope PSF |hCCHM(x0)| = |hobj(x0)href(x0 − Δx0)|.

 figure: Fig. 4

Fig. 4 In the left column there are the absolute values of function hobj together with href mutually shifted accordingly to the relevant Δx0, in the right column there is the absolute value of their product |hCCHM| = |hobjhref| normalized to the maximum value of |hCCHM| without diffuser. The row a) shows simulation without diffuser, rows b)–d) and e) respectively represent two different realizations of diffuser. Rows b)–d) represent different mutual shifts Δx0 for the same realization of the diffuser, specifically σ = 0.42λ. In the last row e) there is the result of diffuser realization σ = 0.84λ, it means a stronger diffuser.

Download Full Size | PDF

Figures 4(a), 4(b)–4(d), and 4(e) represent three different realizations of the diffuser. One realization corresponds to one certain function Φ(xd) and therefore also to a certain function hobj(x0). Even if both σ and μ are constant, the realization can vary substantially, as σ and μ are only statistical parameters of Φ(xd).

Figure 4(a) shows the situation without diffuser, it means that σ = 0. The numerical simulations of hobj and hCCHM in Fig. 4(a) are equal to their analytical expression. In all three Figs. 4(b)–4(d) there is only one realization with σ = 0.42λ. The only difference between Figs. 4(b)–4(d) is in Δx0. Figure 4(e) belongs to a realization of a stronger diffuser with σ = 0.84λ.

Figure 4(b) shows the PSFs in ballistic light, it is with Δx0 = 0. The ballistic light PSFs look similar to the PSFs without diffuser (Fig. 4(a)), only the signal level has decreased approximately ten times when compared to the situation without diffuser.

Figure 4(c) shows the PSFs in diffuse light with mutual shift Δx0 = 16.0 μm. In this particular representation of the diffuser with this mutual shift the main maximum of the function href is in the position of a local maximum of hobj, therefore their product (right column) has the sharp maximum here as well. The microscope PSF in Fig. 4(c) is therefore very similar to microscope PSF in Figs. 4(a) and 4(b). On the other hand shifting the reference arm slightly by only 1.0 μm to the mutual shift Δx0 = 15.0 μm causes, that the microscope PSF looks very different. This is caused by the fact, that the main maximum of the href coincides now with a local minimum of hobj.

In Fig. 4(e) there is a different realization of the diffuser, with σ = 0.84λ. Here the ballistic peak becomes approximately three times smaller in comparison to the diffuser realization with smaller σ = 0.42 (Fig. 4(b)).

From the results in Fig. 4 we can conclude, that the diffuser with σ = 0.42λ provides the best imaging properties in ballistic mode, while imaging with σ = 0.84λ provides diffuse signal of comparable quality to ballistic signal. Anyway, in any case of the diffuser it is possible to change the microscope PSF substantially by choosing the right mutual shift Δx0. However the right values of Δx0 cannot be predicted by any general formula as they depend strongly on the particular realization of a diffuser. Thus also plotting or computing any dependency of image characteristics (such as RMS or resolution) on Δx0 would not give us any general information.

The computed PSFs from Fig. 4 were used for a model of imaging a non-point object (Fig. 5). First we have to prove, whether the computed functions hobj are really spatially invariant. For this purpose we have computed hobj for off-axis points x0N = −10 μm and x0N = −20 μm for one realization of diffuser from Fig. 4(b) and we have compared it to hobj for the axial point x0N = 0 μm with the same realization of diffuser. All plots in Fig. 5 illustrate only one realization which is equal to realization from Figs. 4(b)–4(d). Figures 5(a) and 5(b) show the results of simulation for mutual shift Δx = 16.0 μm, while in Figs. 5(c) and 5(d), the mutual shift is Δx = 15.0 μm.

 figure: Fig. 5

Fig. 5 Results of the PSFs for off-axis points. The first two rows a), b) represent the image with mutual shift Δx0 = 16.0 μm and following two rows c), d) Δx0 = 15.0 μm. Rows a), c) show the absolute values of functions href (red) and hobj (black) mutually shifted according to the relevant Δx0 and rows b) and d) show the absolute value of the microscope PSF |hCCHM| = |hobjhref|. Each column shows the image for different points x0N, consequently: axial point x0N = 0 μm and off-axis points x0N = −10 μm and x0N = −20 μm.

Download Full Size | PDF

The first column of Fig. 5 corresponds to an axial point x0N = 0 μm, the second column of Fig. 5 corresponds to an off-axis point x0N = −10 μm and the third column to x0N = −20 μm. Function hobj (black color) with href (red color) are depicted in Figs. 5(a)–5(c). Functions representing signal, product hobjhref, are displayed in Figs. 5 (b) and 5(d) in such a way that the main maximum is plotted at the center of the horizontal axis. By changing x0N functions href, hobj and hCCHM are only shifted in the x0 axis according to the relevant x0N, but their shape remains unchanged. Therefore we can assume that hobj, hCCHM are spatially invariant at least in the interval x0N ∈ [−20, 20] μm, thus within experiments of this article we can consider it as PSF.

In Fig. 6 there are simulated images of three identical and equidistant infinite transparent parallel slits. The transmission function of these slits is described by t(x0). t(x0) = 1 inside the slits, and t(x0) = 0 outside of slits. The images were obtained by convolution of t(x0) with the microscope PSF. The CCHM signal is then computed as

w(x0)=t(x0)hCCHM(x0x0)dx0.
The upper row of Fig. 6 shows the signal in a gray scale, the lower row is its plot profile. Depicted are only the results with the same realization of the diffuser as in Figs. 4(b)–4(d). In the first two columns of Fig. 6, the imaged slits have the spatial frequency fincoh = 2NAob/λ (resolution limit for microscopy in incoherent light and also the cut-off frequency of CCHM with NAob = NAill [27]) and fcoh = NAob/λ (resolution limit for microscopy in coherent light), respectively, and the mutual shift is Δx0 = 16.0 μm, which corresponds to PSF in Fig. 4(c). The third column of Fig. 6 displays images of slits with spatial frequency fcoh, but for mutual shift Δx0 = 15.0 μm, which corresponds to PSF in Fig. 4(d). We can conclude from the Fig. 6 that with mutual shift Δx0 = 16.0 μm we can image slits with spatial frequency fincoh. Whereas with the mutual shift Δx0 = 15.0 μm even two times lower frequency fcoh is not resolved.

 figure: Fig. 6

Fig. 6 Numerical simulation of imaging three infinite equidistant parallel slits. In the upper row there is a computed signal in the gray scale. In the bottom row, there is a plot profile of the upper row. In the left column the spatial frequency of slits is f = 2NAob/λ and the mutual shift Δx0 = 16.0 μm, in the middle column f = NAob/λ, Δx = 16.0 μm and in the right column f = NAob/λ, Δx = 15.0 μm.

Download Full Size | PDF

3. Experimental results

The experiment was carried out on CCHM. The optical setup of the microscope is shown in Fig. 1. We have used objectives Nikon 10×/0.25. The light source was a halogen lamp. The aperture angle of the illuminating beam was chosen to fill the aperture of the microscope objectives completely (NAill = NAob). In order to obtain the quasi-monochromatic light, the interference filter was used (λ0 = 650nm FWHM 10nm). This implies the coherence length (CL) approximately 42 μm. Thanks to the large coherence length together with very narrow bandwidth the effect of dispersion caused by specimen or diffuser is negligible. The fact that we use quasi-monochromatic light (the large coherence length) also explains why the longitudinal shifts in reference arm were not performed.

We used a resolution chart as a specimen and a thin ground glass of thickness 0.17mm as a diffuser. The distance between the diffuser and the specimen was ad = 4mm.

Compare now the results of simulation of the line objects image with the results of the experiment. Figure 7 shows the measured signal w in a gray scale. The original holograms, especially those made through diffuser, are not presented here with respect to their very low contrast. In the first column of Fig. 7 there is an image of the resolution chart which is displayed in detail in the second and the third column of Fig. 7. Figure 7(a) shows an image without diffuser and the Figs. 7(b)–7(e) show images made with diffuser. The only difference between images made with diffuser was in the mutual shift Δx0 implemented by lateral shift of the objective in the interferometer. The variance of td(xd) of the experimental diffuser can be estimated to be similar to a model diffuser from Figs. 4(b)–4(d). Ballistic image Δx0 = 0 is in Fig. 7(b), while diffuse light images are in Figs. 7(c)–7(e). The mutual shifts are consequently Δx0 = 44,8 μm - Fig. 7(c), Δx0 = 42.7 μm - Fig. 7(d) and Δx0 = 47.0 μm - Fig. 7(e). In each of these images the resolution is different and it is illustrated by the orange label around lines with a spatial frequency on the resolution limit. In Fig. 7(a) the resolution is flim = 0.81 fincoh, in ballistic light image, Fig. 7(b), flim = 0.73 fincoh. Theoretically it should be possible to image lines having frequency flim = fincoh, but experimentally this limit was not achieved. It has more reasons, first is that CTF has very low value for frequencies near to fincoh, the other reason rests in the fact that the image suffers from aberrations, mainly spherical. Lower resolution of ballistic image is moreover influenced also by the lower SNR level.

 figure: Fig. 7

Fig. 7 The experimental results of imaging a resolution chart. The images were performed with objectives 10×/0.25. Interference filter λ = 650nm, 10nm FWHM. a) imaging without diffuser, b) imaging with diffuser - ballistic light, c)–e) imaging with diffuser - diffuse light. The mutual shifts are c) Δx = 44.8 μm d) Δx = 42.7 μm, e) Δx = 47.0 μm. In the second and third column there are magnified details A and B respectively. Orange circles indicate the resolution limit of images in the corresponding row.

Download Full Size | PDF

The limit frequencies measured for diffuse light in CCHM are flim = 0.73 fincoh for Δx0 = 44.8 μm, flim = 0.48 fincoh for Δx0 = 42.7 μm and flim = 0.53 fincoh for Δx0 = 47.0 μm.

In order to obtain higher resolution images some attempts of averaging images from various mutual shifts Δx0 were performed. But it didn’t prove any significant results so far. At the moment we are trying to develop a new method for averaging from various Δx0.

4. Conclusions

The main aim of this publication was to show the principle and some basic properties of coherence gating of diffuse light in LCI. Therefore the simple model of PSF for imaging through thin diffuse medium by CCHM was established on the basis of the 2-D model of CCHM in Fresnel (paraxial) approximation. Although CCHM is achromatic we limited the computations as well as the experiments only for monochromatic light. By doing this the experimental results can be interpreted and compared with computations more easily.

The equation for PSF was derived analytically, nevertheless in order to obtain a quantitative solution we had to use numerical computation, since the diffuser transmission function is generated as a random function with a mesh fine enough when compared to the resolution of the microscope. The results of numerical computation were compared with the analytical expressions for CCHM without diffuser.

A significant peak was obtained as a result of simulation of PSF for imaging through weakly scattering diffuser in ballistic light. At the same time we show, that the image can be obtained in diffuse light as well. The diffuse light PSF depends on the parameter Δx0, which denotes the mutual lateral shift between the reference and the object arms in the output plane. Choosing a right value of Δx0 leads to PSF similar to the situation without diffuser. On the other hand even the small change of Δx0 from the right value can deform the PSF very significantly.

The influence of a stronger diffuser on the microscope PSF was studied as well. The ballistic peak becomes not as significant as in the case of a weaker diffuser. However, for the diffuse light imaging, the findings for PSF remain unchanged.

In the experiment we have proved the computed results for a weaker diffuser by imaging a resolution chart through a ground glass. We have also shown experimentally the strong influence of the mutual shift Δx0 on the microscope PSF. In diffuse light by choosing a right value of Δx0 we were able to image objects with spatial frequency near to the theoretical resolution limit for CCHM without diffuser 2NA/λ. However, this theoretical limit is not fully achieved mainly due to the low SNR level. When the Δx0 was changed from this right value, the PSF changed rapidly which resulted in an image of much lower quality and resolution. This is in accordance with the results of the model. However the particular values of Δx0 are not the same for the experimental and for the model results. This is caused by the fact that the model diffuser is defined by the transmission function, which is generated numerically as a matrix of random phase shifts and thus it can hardly be equal to the transmission function of a real diffuser.

Acknowledgments

The authors thank to colleagues from Experimental biophotonics group (CEITEC) for helpful discussions, especially to Věra Kollárová, Pavel Kolman, Matěj Týč, Jiří Komrska, and Zbyněk Dostál. This work was supported by Technology Agency of the Czech Republic ( TE01020229), Ministry of Industry and Trade of the Czech Republic ( FR-TI4/660), CEITEC – Central European Institute of Technology ( CZ.1.05/1.1.00/02.0068) from European Regional Development Fund and by the grant “Specific research” FSI-S-11-22 of Brno University of Technology.

References and links

1. M. Minsky, “Memoir on inventing the confocal scanning microscope,” Scanning 10, 128–138 (1988). [CrossRef]  

2. T. Wilson, Confocal Microscopy (Academic Press, 1990).

3. M. Kempe, W. Rudolph, and E. Welsch, “Comparative study of confocal and heterodyne microscopy for imaging through scattering media,” J. Opt. Soc. Am. A 13, 46–52 (1996). [CrossRef]  

4. M. Kempe, A. Genack, W. Rudolph, and P. Dorn, “Ballistic and diffuse light detection in confocal and heterodyne imaging systems,” J. Opt. Soc. Am. A 14, 216–223 (1997). [CrossRef]  

5. J. M. Schmitt, A. Knüttel, and M. Yadlowsky, “Confocal microscopy in turbid media,” J. Opt. Soc. Am. A 16(8) 2226–2235 (1994) [CrossRef]  

6. J. C. Schotland, “Continuous-wave diffusion imaging,” J. Opt. Soc. Am. A 14(1) 275–279 (1997). [CrossRef]  

7. N. L. Patel, “Relative capacities of time-gated versus continuous-wave imaging to localize tissue embedded vessels with increasing depth,” J. Biomed. Opt. 15(1) 016015 (2010). [CrossRef]  

8. B. J. Tromberg, L. O. Svaasand, T. T. Tsay, and R. C. Haskell, “Properties of photon density waves in multiple-scattering media,” Appl. Opt. 32607–616 (1993). [CrossRef]   [PubMed]  

9. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis (SPIE Press, 2007). [CrossRef]  

10. W. Rudolph and M. Kempe, “Trends in optical biomedical imaging,” J. Mod. Opt. 441617–1642 (1997). [CrossRef]  

11. E. Leith and J. Upatnieks, “Reconstructed wavefronts and communication theory,” J. Opt. Soc. Am. 521123–1128 (1962). [CrossRef]  

12. H. Kogelnik, “Holographic image projection through inhomogeneous media,” Bell Syst. Tech. J. 44, 2451–2455 (1965). [CrossRef]  

13. G. Indebetouw and P. Klysubun, “Imaging through scattering media with depth resolution by use of low-coherence gating in spatiotemporal digital holography,” Opt. Lett. 25, 212–214 (2000). [CrossRef]  

14. E. Leith, C. Chen, H. Chen, Y. Chen, D. Dilworth, J. Lopez, J. Rudd, P. Sun, J. Valdmanis, and G. Vossler, “Imaging through scattering media with holography,” J. Opt. Soc. Am. A 9, 1148–1153 (1992). [CrossRef]  

15. S. Tamano, Y. Hayasaki, and N. Nishida, “Phase-shifting digital holography with a low-coherence light source for reconstruction of a digital relief object hidden behind a light-scattering medium,” Appl. Opt. 45, 953–959 (2006). [CrossRef]   [PubMed]  

16. E. Leith and C. Kuei, “Interferometric method for imaging through inhomogeneities,” Opt. Lett. 12, 149–151 (1987). [CrossRef]   [PubMed]  

17. P. Kolman and R. Chmelík, “Coherence-controlled holographic microscope,” Opt. Express 18, 21990–22003 (2010). [CrossRef]   [PubMed]  

18. F. Dubois, C. Yourassowsky, O. Monnom, J. Legros, O Debeir, P. Van Ham, R. Kiss, and C. Decaestecker, “Digital holographic microscopy for the three-dimensional dynamic analysis of in vitro cancer cell migration,” J. Biomed. Opt. 11, 054032 (2006). [CrossRef]   [PubMed]  

19. D. Huang, E. Swanson, C. Lin, J. Schuman, W. Stinson, W. Chang, M. Hee, T. Flotte, K. Gregory, C. Puliafito, et al., “Optical coherence tomography,” Science 254, 1178–1181 (1991). [CrossRef]   [PubMed]  

20. J. Izatt, M. Hee, G. Owen, E. Swanson, and J. Fujimoto, “Optical coherence microscopy in scattering media,” Opt. Lett. 19, 590–592 (1994). [CrossRef]   [PubMed]  

21. E. Leith, “Broad-source image plane holography as a confocal imaging process,” Appl. Opt. 33, 597–602 (1994). [CrossRef]   [PubMed]  

22. Y. Cotte, M. F. Toy, N. Pavillon, and C. Depeursinge, “Microscopy image resolution improvement by deconvolution of complex fields.,” Opt. Express 18, 19462–19478 (2010). [CrossRef]   [PubMed]  

23. Y. Cotte, F. Toy, P. Jourdain, N. Pavillon, D. Boss, P. Magistretti, P. Marquet, and C. Depeursinge, “Marker-free phase nanoscopy,” Nature Photon. 7, 113–117 (2013). [CrossRef]  

24. M. Mir, S. D. Babacan, M. Bednarz, M. N. Do, I. Golding, and G. Popescu, “Visualizing Escherichia coli sub-cellular structure using sparse deconvolution spatial light interference tomography,” PLOS ONE 7, e39816 (2012). [CrossRef]   [PubMed]  

25. M. Lošt’ák, P. Kolman, Z. Dostál, and R. Chmelík, “Diffuse light imaging with a coherence controlled holographic microscope,” Proc. SPIE 7746, 77461N (2010). [CrossRef]  

26. T. Slabý, M. Antoš, Z. Dostál, P. Kolman, and R. Chmelík, “Coherence-controlled holographic microscope,” Proc. SPIE 7746, 77461R (2010). [CrossRef]  

27. T. Slabý, P. Kolman, Z. Dostál, M. Antoš, M. Lošt’ák, and R. Chmelík, “Off-axis setup taking full advantage of incoherent illumination in coherence-controlled holographic microscope,” Opt. Express 21, 14747–14762 (2013). [CrossRef]   [PubMed]  

28. R. Chmelík, “Three-dimensional scalar imaging in high-aperture low-coherence interference and holographic microscopes,” J. Mod. Opt. 53, 2673–2689 (2006). [CrossRef]  

29. E. N. Leith and B. J. Chang, “Space-invariant holography with quasi-coherent light,” Appl. Opt. 12, 1957—1963 (1973). [CrossRef]   [PubMed]  

30. M. Born and E. Wolf, Principles of Optics, 7th expanded ed. (Cambridge University, 2002).

31. J. Goodman, Introduction to Fourier Optics, 2nd ed. (McGraw-Hill, 1996)

32. W. H. Press, Numerical Recipes: The Art of Scientific Computing, 3rd ed. (Cambridge University, 2007).

33. C. N. Kurtz, “Transmittance characteristics of surface diffusers and the design of nearly band-limited binary diffusers,” J. Opt. Soc. Am. 62, 982–989 (1972). [CrossRef]  

34. R. B. Crane, “Use of a laser-produced speckle pattern to determine surface roughness,” J. Opt. Soc. Am. 60, 1658–1663 (1970). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (7)

Fig. 1
Fig. 1 Optical setup of CCHM. Incoherent light source (S), relay lens (L), beam splitters (BS), mirrors (M), condensers (C), object plane (Sp), reference plane (R), microscope objectives (O), tube lenses (TL), diffraction grating (DG), output lens (OL), output plane (OP), detector (D). Figure adapted, with permission, from [27].
Fig. 2
Fig. 2 Simplified drawings for explanation of principle of imaging in CCHM through diffuse medium by ballistic and diffuse light. IS1, IS2...imaginary sources in object planes of the condensers C1, C2. Sp, R...object planes of microscope objectives O1, O2. D...diffuser. AO, AR...images of point A in the output plane OP. AB...ballistic image of point A. AD...diffuse image of point A. Dotted line...scattered light, continuous line...unscattered light. a) Imaging without diffuser. b) Imaging with diffuser by ballistic light, mutual shift Δxi = 0. c) Imaging with diffuser by diffuse light, reference arm is shifted by nonzero Δxi.
Fig. 3
Fig. 3 Simplified drawing of the object arm for computation of the amplitude in the output plane when imaging a planar object placed in the object plane Sp. In this Fig. the object is a point aperture at coordinate x0 = x0N. D...diffuser, L...thin lens substituting the objective, Ap...aperture diaphragm in the back focal plane of thin lens, OP...output plane.
Fig. 4
Fig. 4 In the left column there are the absolute values of function hobj together with href mutually shifted accordingly to the relevant Δx0, in the right column there is the absolute value of their product |hCCHM| = |hobjhref| normalized to the maximum value of |hCCHM| without diffuser. The row a) shows simulation without diffuser, rows b)–d) and e) respectively represent two different realizations of diffuser. Rows b)–d) represent different mutual shifts Δx0 for the same realization of the diffuser, specifically σ = 0.42λ. In the last row e) there is the result of diffuser realization σ = 0.84λ, it means a stronger diffuser.
Fig. 5
Fig. 5 Results of the PSFs for off-axis points. The first two rows a), b) represent the image with mutual shift Δx0 = 16.0 μm and following two rows c), d) Δx0 = 15.0 μm. Rows a), c) show the absolute values of functions href (red) and hobj (black) mutually shifted according to the relevant Δx0 and rows b) and d) show the absolute value of the microscope PSF |hCCHM| = |hobjhref|. Each column shows the image for different points x0N, consequently: axial point x0N = 0 μm and off-axis points x0N = −10 μm and x0N = −20 μm.
Fig. 6
Fig. 6 Numerical simulation of imaging three infinite equidistant parallel slits. In the upper row there is a computed signal in the gray scale. In the bottom row, there is a plot profile of the upper row. In the left column the spatial frequency of slits is f = 2NAob/λ and the mutual shift Δx0 = 16.0 μm, in the middle column f = NAob/λ, Δx = 16.0 μm and in the right column f = NAob/λ, Δx = 15.0 μm.
Fig. 7
Fig. 7 The experimental results of imaging a resolution chart. The images were performed with objectives 10×/0.25. Interference filter λ = 650nm, 10nm FWHM. a) imaging without diffuser, b) imaging with diffuser - ballistic light, c)–e) imaging with diffuser - diffuse light. The mutual shifts are c) Δx = 44.8 μm d) Δx = 42.7 μm, e) Δx = 47.0 μm. In the second and third column there are magnified details A and B respectively. Orange circles indicate the resolution limit of images in the corresponding row.

Equations (15)

Equations on this page are rendered with MathJax. Learn more.

i ( q t i , K t ) = | u o ( q t i , K t ) + u r ( q t i , K t ) exp ( i 2 π f c x i ) | 2 = = | u o ( q t i , K t ) | 2 + | u r ( q t i , K t ) | 2 + u o ( q t i , K t ) u r * ( q t i , K t ) exp ( i 2 π f c x i ) + + u o * ( q t i , K t ) u r ( q t i , K t ) exp ( i 2 π f c x i ) ,
w ( q t i ) = ( K x 2 + K y 2 ) 1 / 2 < NA ill / λ u o ( q t i , K x , K y ) u r * ( q t i , K x , K y ) d K x d K y ,
w ( x i ) = NA ill / λ NA ill / λ u o ( x i , K x ) u r * ( x i , K x ) d K x .
η obj ( x i , K x ) = C exp ( i 2 π K x x 0 N ) exp ( i π K x 0 N 2 a d ) exp ( i π K x i 2 z ) × × t d ( x d ) I exp [ i π K a d ( 2 x 0 N x d + x d 2 ) ] d x d ,
I = F [ E e ( R p B 2 A ) E e ( R p B 2 A ) ] ,
E e ( x ) = C ( x ) + i S ( x ) = 0 x exp ( i π 2 t 2 ) d t
h CCHM ( x i ) = NA ill / λ NA ill / λ η obj ( x i , K x ) u r * ( x i , K x ) d K x ,
u r ( x i , K x ) = C ref exp ( i 2 π K x x i M ) ,
η obj ( x 0 , K x ) = h obj ( x 0 ) exp ( i 2 π K x x 0 N ) ,
h CCHM ( x 0 ) = h obj ( x 0 ) C ref NA ill / λ NA ill / λ exp [ i 2 π K x ( x 0 N x 0 ) ] d K x h ref ( x 0 ) .
h ref ( x 0 ) = sin [ NA ill ( x 0 x 0 N ) 2 π / λ ] NA ill ( x 0 x 0 N ) 2 π / λ ,
h CCHM ( x 0 ) = h obj ( x 0 ) h ref * ( x 0 ) .
h CCHM ( x 0 ) = | h obj ( x 0 ) | 2 = | h ref ( x 0 ) | 2 .
h CCHM ( x 0 , Δ x 0 ) = h obj ( x 0 ) h ref ( x 0 Δ x 0 ) .
w ( x 0 ) = t ( x 0 ) h CCHM ( x 0 x 0 ) d x 0 .