Sensitive and broadband measurement of dispersion in a cavity using a Fourier transform spectrometer with kHz resolution

Open Access Open Access

Abstract

Optical cavities provide high sensitivity to dispersion since their resonance frequencies depend on the index of refraction. We present a direct, broadband, and accurate measurement of the modes of a high finesse cavity using an optical frequency comb and a mechanical Fourier transform spectrometer with a kHz-level resolution. We characterize 16000 longitudinal cavity modes spanning 16 THz of bandwidth in terms of center frequency, linewidth, and amplitude. Using the center frequencies we retrieve the group delay dispersion of the cavity mirror coatings and pure N2 with 0.1 fs2 precision and 1 fs2 accuracy, as well as the refractivity of the 3ν1 + ν3 absorption band of CO2 with 5 × 10−12 precision. This opens up for broadband refractive index metrology and calibration-free spectroscopy of entire molecular bands.

© 2017 Optical Society of America

Corrections

Lucile Rutkowski, Alexandra C. Johansson, Gang Zhao, Thomas Hausmaninger, Amir Khodabakhsh, Ove Axner, and Aleksandra Foltynowicz, "Sensitive and broadband measurement of dispersion in a cavity using a Fourier transform spectrometer with kHz resolution: erratum," Opt. Express 28, 13290-13291 (2020)
https://opg.optica.org/oe/abstract.cfm?uri=oe-28-9-13290

1. Introduction

Fabry-Perot cavities in combination with narrow linewidth continuous wave (cw) lasers are versatile tools for ultra-sensitive measurements of displacement, absorption, and dispersion. For example, high precision measurements of minute length variation of Fabry-Perot cavities have enabled detection of gravitational waves [1]. A pressure sensor based on the measurement of gas refractivity inside a cavity can outperform a manometer [2]. Cavity-enhanced molecular absorption [3, 4] and dispersion [5, 6] spectroscopies, which rely on the measurement of intracavity absorption losses and dispersion-induced cavity modes shifts, respectively, provide complementary information about the molecular transitions and high sensitivity to absorption/dispersion. However, cw lasers allow such measurements only over narrow bandwidths, typically in the sub-THz range. Optical frequency combs, whose spectra consist of thousands of equidistant narrow lines, can probe cavity modes over a much broader bandwidth. In cavity-enhanced optical frequency comb spectroscopy, absorption spectra of entire molecular bands can be acquired with high resolution in short acquisition times [7–10]. Combs are also an ideal tool for measurements of broadband dispersion induced either by the cavity mirror coatings or intracavity samples. Previous demonstrations relied on indirect measurements of the resonant condition between the comb and the cavity and required external frequency calibration [11–13]. Therefore, they did not fully benefit from the high frequency accuracy provided by the comb and suffered from poor spectral resolution (at the THz level), which was insufficient to allow observing dispersion induced by molecular transitions.

Here we present a new method for direct measurement of broadband spectra of the modes of a high finesse cavity using a frequency comb and we determine the group delay dispersion of the cavity mirror coatings and intracavity gas samples with unprecedented combination of bandwidth and precision. We achieve sub-kHz resolution in the measurement of the cavity mode frequencies and widths using a mechanical Fourier transform spectrometer (FTS) [14, 15]. Our method allows calibration-free quantification of the dispersion induced by narrow molecular transitions and opens up for improved metrology of refractive index and of entire molecular absorption bands.

2. Theory

The cavity mode frequencies, νq, fulfill the resonance condition on the round-trip phase shift of the electric field, Φ, which can be written as

Φ(νq)=ϕ0(νq)+ϕn(νq)=2πq,
where q is an integer mode index, ϕ0q) = 2πνq2L/c + ϕm is the phase shift inside an empty cavity, where L is the cavity length, c is the speed of light, and ϕm is the phase shift induced by the cavity mirror coatings, and ϕnq) = 2πνq[n(νq) – 1]2L/c is the phase shift induced by the intracavity sample with refractive index n. The cavity modes spacing, i.e. the free spectral range, FSR = 2π[(∂Φ/∂ν)]−1, is not constant because Φ varies non-linearly with frequency. To evaluate the shift of the cavity mode frequencies with respect to an equidistant grid, we define a reference scale
νq0=qFSRref0+f0,
where FSRref0 is the empty cavity FSR evaluated around the cavity mode νref, q = floor[νq/FSRref0], and f0 = νref – qrefFSRref0 is an offset frequency, where qref = floor[νref/FSRref0]. The relation between the frequency shift of the cavity modes, ∆ν = νq -νq0, and the intracavity phase shift, Eq. (1), can be found by Taylor expanding Φ around νref to first order, which yields
Δν=νqνq0=FSRref2π[2πqΦ(νq0)],
where FSRref is the cavity FSR evaluated around the cavity mode νref . The intracavity group delay dispersion (GDD) is then calculated as

GDD(νq0)14π22Φν2=1FSRref2Δνν2.

The mode shift in an empty cavity, Δν0, originates from the dispersion of the cavity mirror coatings. When the cavity is filled with a gas, an additional shift, Δνn, occurs because of the frequency dependence of the refractive index

Δνn=νq0[n(νq0)1].
The refractive index n has two contributions n = nna + nabs, where nna is the slowly varying refractive index of a non-absorbing gas, given by e.g. the Sellmeier equation [16], while nabs is the refractive index of molecular transitions, defined as
nabs(νq0)1=c4πνq0ρiSiIm[χi(νq0)],
where ρ is the absorbing sample gas density, Si is the line intensity and χi is the complex line shape function of the ith transition.

3. Experimental setup

The experimental setup is depicted in Fig. 1(a). The cavity with a finesse of ~1700 consisted of two dielectric mirrors with 5 m radius of curvature separated by L = 45 cm, yielding an FSR of 333 MHz. The mirrors were glued to two ends of a stainless steel tube, connected to a gas system, and a ring piezo-electric transducer (PZT) was inserted between one of the mirrors and the tube to control the cavity length. The cavity modes were probed by an amplified Er:fiber frequency comb with a repetition rate (frep) of 250 MHz, whose beam was mode-matched to the TEM00 modes of the cavity. Because of the difference between the cavity FSR and the comb frep the cavity acted as a filter for the comb, as shown in Fig. 1(b), and the repetition rate in cavity transmission was equal to frepT = 4frep = 3FSR = 1 GHz. The comb offset frequency, fceo, was stabilized by locking the output of an f-2f interferometer to a frequency provided by a GPS-referenced Rubidium clock, fclk = 20 MHz, via feedback to the current of a diode laser pumping the Er:fiber oscillator. The frequency of one of the comb lines was stabilized to a narrow-linewidth cw Er:fiber laser locked to the Pe(8) CO2 transition at λcw = 1576.9396 nm using sub-Doppler noise-immune cavity-enhanced optical heterodyne molecular spectroscopy (NICE-OHMS [17], not shown in the figure). The linewidth of the cw laser was 2.8(6) kHz, estimated using the method described in ref [18]. The two laser beams were combined in a fiber and the beat note between the cw laser and the closest comb line was detected in free space after dispersing the spectrum by a diffraction grating. The beatnote was locked to a radio frequency, fDDS, generated by a tunable direct digital synthesizer (DDS) referenced to the Rb clock, via feedback to an intracavity PZT and an electro-optic modulator (EOM) controlling the frep. This optical lock transferred the linewidth of the cw laser to the closest comb line, which was verified by measuring the width of the optical beat note below the Hz level using a spectrum analyzer. The cavity length was stabilized by locking one of the cavity resonances to the cw laser using the Pound-Drever-Hall (PDH) technique. This involved phase modulation of the electric field of the cw laser at a frequency fPDH = 20 MHz using an EOM and phase sensitive detection of the cavity reflected light, which was dispersed using a diffraction grating to avoid saturation of the detector by the reflected comb intensity. The correction signal was fed to the cavity PZT with a closed-loop bandwidth of 1.1 kHz. Due to this relatively low bandwidth a frequency jitter of the cavity modes remained, yielding a mode broadening of 20 kHz, calculated from the power spectral density of the closed-loop error signal [18].

 figure: Fig. 1

Fig. 1 (a) Experimental setup: f-2f - f-2f interferometer beat note; DBM - double-balanced mixer; DDS - direct digital synthesizer; FC - fiber collimator; λ/2 - half-waveplate; PBS - polarizing beam splitter; λ/4 - quarter-waveplate; PZT - piezoelectric transducer; FTS - Fourier transform spectrometer; EOM - electro-optic modulator; G - diffraction grating; (b) Matching of the comb lines (red) and the cw laser (blue) to the cavity modes (black). The reflected comb lines are shown with the red dashed lines.

Download Full Size | PDF

Setting fDDS close to 83 MHz, as shown in Fig. 1(b), brought the comb lines close to resonance with the cavity modes. The transmitted light, composed of overlapping comb and cw laser beams, was analyzed with a fast-scanning FTS equipped with an auto-balancing detector [8, 19]. To achieve kHz resolution we used the method described in refs [14, 15], in which the nominal resolution of the FTS, given by the inverse of the optical path difference [20], is set equal to the comb line spacing in order to allow precise measurement of the comb line intensities without distortion induced by the instrumental line shape. We set the nominal resolution to 4frep/3, which yielded sampling points spaced by FSR and allowed sampling of the intensities of the transmitted comb lines as well as the cw laser. To measure the profiles of the cavity modes we scanned the comb lines across the cavity modes by tuning fDDS in steps of 20 kHz. This in turn caused tuning of frep via the feedback to the comb oscillator. The 20 kHz step was chosen to yield more than 10 points per cavity mode width and it could be made smaller if needed. Two interferograms, each lasting 2.5 s, were recorded and averaged for each frep value, and 150 steps were acquired, with a total acquisition time of 20 min. The 150 spectra obtained after fast Fourier transform were interleaved [21] to yield a spectrum containing discrete spectral pieces separated by frepT = 1 GHz, each covering 3 MHz.

4. Results

The spectrum of an empty cavity, spanning from 1500 to 1640 nm and containing 16000 resonance modes, is shown in Fig. 2(a). The peak intensities of the cavity modes follow the comb spectral envelope and a discrete peak is visible at λcw. The inset of Fig. 2(a) shows a zoom of a part of the spectrum containing 3 cavity modes separated by 3FSR. The spectrum of a cavity mode at 1600 nm is further enlarged in Fig. 2(b) (black markers). A model based on a Lorentzian function and a linear baseline is fitted to the data (red curve, with residuum in the lower panel). The fitted parameters were the amplitude Aq, the center frequency νq, and the width Γq of the Lorentzian function, as well as the offset and slope of a linear baseline. The fit returned Aq = 1.019(3), νq = 187697701022.4(4) kHz, and Γq = 214.4(9) kHz. To verify the accuracy of the mode width measurement, we measured the cavity ring-down (RD) time at 1600 nm using the comb as the light source, an acousto-optic modulator before the cavity to interrupt the beam, and a monochromator with 200 GHz resolution to disperse the transmitted spectrum. The measurement yielded a mode width of ΓRD = 191(2) kHz, which agrees with the value obtained from the fit to the mode profile when considering the mode broadening due to the remaining cavity jitter (20 kHz) and the comb linewidth (2.8 kHz).

 figure: Fig. 2

Fig. 2 (a) Spectrum of the empty cavity transmission spanning 16 THz consisting of 16000 discrete modes. Inset: enlargement of 3 modes separated by 3FSR. (b) A zoom of a single mode at 1600 nm (black markers) together with a Lorentzian fit (red line) and the residuum (lower panel).

Download Full Size | PDF

The center frequency of each of the 16000 cavity modes was determined from fits of a Lorentzian line shape function with an uncertainty ranging from 0.4 to 3 kHz depending on the signal-to-noise ratio of the considered mode. Figure 3(a) shows the shift of the cavity modes, calculated using Eqs. (2) and (3), for the empty cavity [∆ν0, black curve, left y-axis in MHz] and the cavity filled with pure N2 at 750(1) Torr and 296(3) K [∆νN2, red curve, right y- axis in GHz]. The FSRref0 in Eq. (2) was set equal to 4/3 times the frep value that maximized the transmission of the comb lines through the cavity modes around νref = c/λcw. This yielded FSRref0 = 333.5730693(1) MHz and f0 = 63.33(5) MHz. When the cavity was filled with N2 the change of the intracavity refractive index had to be compensated by adjusting the cavity length in order to keep the cavity modes in resonance with the comb lines. This in turn changed the value of FSRref0 and f0. The new value of FSRref0 = 333.5795270(2) MHz was found from the frep value that maximized the transmission of the comb lines around νref, divided by the refractive index of N2 at νref, calculated using the Sellmeier equation with coefficients from [22]. The new offset frequency was f0 = 219.40(5) MHz. Since both curves were referenced to νref, they cross 0 Hz at λcw.

 figure: Fig. 3

Fig. 3 (a) Shift of the cavity mode frequencies Δν measured when the cavity is empty (black curve, left y-axis) and when the cavity is filled with pure N2 at 750 Torr (red curve, right y-axis) plotted together with a calculated shift based on the Sellmeier equation for N2 (blue markers). Note the three orders of magnitude difference between the two y-axis scales. (b) Residuum of a polynomial fit to the shift of the empty cavity modes. (c) Residuum of a polynomial fit to the mode shift of the cavity filled with N2.

Download Full Size | PDF

To verify the accuracy of the measurement, we calculated the shift induced by N2 using Eq. (5) and the frequency dependence of the refractive index of N2 from [22], subtracted the value of νref (nref −1) to make ∆νN2 = 0 Hz at λcw, and added the experimentally determined empty cavity shift. The resulting curve is shown with blue markers in Fig. 3(a), and it is in excellent agreement with the measured curve (relative error of the slopes is 4 × 10−3).

To retrieve the GDD of the cavity mirror coatings and N2, we fitted Eq. (3) to the two Δν curves in Fig. 3 with Φ assumed as a fifth order polynomial function. The residuals of the fits are shown in Figs. 3(b) and 3(c), confirming the validity of the model and the kHz precision of the determination of the mode shift. The fitted curves were differentiated twice [see Eq. (4)] to yield the GDD of the empty cavity and the cavity filled with N2. The results are shown in Fig. 4 by the black and red solid curves, respectively. The uncertainty of the GDD is below 0.1 fs2 over the entire range, originating from the uncertainty of the determination of νq and of the fit coefficients. The GDD of the cavity filled with N2 is compared with a theoretical curve (dashed curve) calculated as the sum of the experimentally determined empty cavity GDD and the GDD of N2 calculated using the frequency-dependent refractive index of N2 from [22]. Both curves agree within 1 fs2.

 figure: Fig. 4

Fig. 4 Group delay dispersion (GDD) of the empty cavity (black solid curve), and the cavity filled with pure N2 at 750 Torr (red solid curve). The dashed curve is the sum of the GDD of N2 calculated using the Sellmeier equation and the experimentally determined empty cavity GDD (dashed black curve).

Download Full Size | PDF

The high precision of the retrieved cavity mode frequencies allows also the measurement of dispersion induced by molecular transitions. Figure 5(a) shows the refractivity of the 3ν1 + ν3 band of CO2 (black markers) obtained using Eq. (5) and the cavity mode frequencies measured when the cavity was filled with 1.00(5)% of CO2 in N2 at 750(1) Torr at room temperature [296(3) K]. The red solid curve shows a fit of the molecular refractivity, Eq. (6), calculated using the imaginary part of the complex Voigt profiles and CO2 line parameters from the HITRAN database [23], with the sample density as the fitting parameter. The slowly varying background originating from the N2 and cavity mirror dispersion was removed by fitting a fifth order polynomial function together with the model. The residual of the fit, shown in Fig. 5(b), has a standard deviation equal to 5 × 10−12, demonstrating the high precision of the measurement and a good agreement with theory. The CO2 concentration retrieved from the fit was 0.95(1)%. The relative error of 5% with respect to the specified concentration, and the structure remaining in the residuum, are presumably caused by the fact that the pressure broadening and shift parameters in HITRAN are defined for CO2 in air while our sample was CO2 in N2.

 figure: Fig. 5

Fig. 5 (a) Refractivity of the 3ν1 + ν3 absorption band of 1% of CO2 in N2 at 750 Torr (black markers) together with a fit (red curve). (b) Residual of the fit.

Download Full Size | PDF

5. Conclusions and outlook

To summarize, we measured cavity transmission spectra spanning 16 THz of bandwidth and containing 16000 modes with ~200 kHz linewidth using a comb-based Fourier transform spectrometer. These measurements provide evidence that the resolution of a comb-based mechanical FTS is given by the comb linewidth, and not by the maximum optical path difference, provided that the nominal resolution of the FTS is matched to the comb line spacing. Each cavity mode was fully characterized in terms of amplitude, width, and center frequency. From the shift of the center frequencies we retrieved the GDD of the cavity mirror coatings and of pure N2, as well as the refractivity of the entire CO2 absorption band. The precision of the GDD measurement was at the 0.1 fs2 level and the accuracy of the N2 measurement was within 1 fs2 of a model based on the Sellmeier equation, while the precision of the refractive index of the molecular transitions was 5 × 10−12.

Our method offers at least one order of magnitude improvement on the signal-to-noise ratio compared to the previous demonstration of direct measurement of cavity resonance modes using a dual-comb spectrometer [24], allowing retrieval of the cavity mode parameters with significantly improved precision. The demonstrated precision of the dispersion measurement is more than an order of magnitude better than that obtained using white light sources [25] and comparable to that achieved with other comb-based methods [11–13]. However, our approach offers at least 5 orders of magnitude improvement in frequency resolution by directly measuring the cavity transmission spectrum. This in turn enables the measurement of cavity mode shifts induced by molecular transitions, which was previously possible only with cw laser-based techniques [5, 6]. The ability to measure dispersion of entire molecular bands without any prerequisite on the knowledge of the cavity parameters will allow determination of transition line parameters with improved accuracy.

In conclusion, direct measurement of cavity resonance modes using a comb-based Fourier transform spectrometer provides means to measure simultaneously the group delay dispersion of cavity mirror coatings, the dispersion of the refractive index of gases, and the resonant refractivity of entire molecular bands. Complementary information about the molecular transitions can be obtained from the linewidth and intensity of the cavity modes, opening up for full characterization of the real and imaginary parts of molecular bands and fundamental tests of the Kramers-Kronig relations.

Funding

Swedish Research Council (2016-03593) and (621-2015-04374); Swedish Foundation for Strategic Research (ICA12-0031); and the Knut and Alice Wallenberg Foundation (KAW 2015.0159).

Acknowledgment

Gang Zhao is supported by the China Scholarship Council.

References and links

1. LIGO Scientific Collaboration and Virgo Collaboration, “Observation of gravitational waves from a binary black hole merger,” Phys. Rev. Lett. 116(6), 061102 (2016). [CrossRef]   [PubMed]  

2. P. F. Egan, J. A. Stone, J. E. Ricker, and J. H. Hendricks, “Comparison measurements of low-pressure between a laser refractometer and ultrasonic manometer,” Rev. Sci. Instrum. 87(5), 053113 (2016). [CrossRef]   [PubMed]  

3. J. Ye, L.-S. Ma, and J. L. Hall, “Ultrasensitive detections in atomic and molecular physics: demonstration in molecular overtone spectroscopy,” J. Opt. Soc. Am. B 15(1), 6–15 (1998). [CrossRef]  

4. G. W. Truong, K. O. Douglass, S. E. Maxwell, R. D. van Zee, D. F. Plusquellic, J. T. Hodges, and D. A. Long, “Frequency-agile, rapid scanning spectroscopy,” Nat. Photonics 7(7), 532–534 (2013). [CrossRef]  

5. A. Cygan, P. Wcisło, S. Wójtewicz, P. Masłowski, J. T. Hodges, R. Ciuryło, and D. Lisak, “One-dimensional frequency-based spectroscopy,” Opt. Express 23(11), 14472–14486 (2015). [CrossRef]   [PubMed]  

6. S. Tan, P. Berceau, S. Saraf, and J. A. Lipa, “Measuring finesse and gas absorption with Lorentzian recovery spectroscopy,” Opt. Express 25(7), 7645–7656 (2017). [CrossRef]   [PubMed]  

7. M. J. Thorpe, D. Balslev-Clausen, M. S. Kirchner, and J. Ye, “Cavity-enhanced optical frequency comb spectroscopy: application to human breath analysis,” Opt. Express 16(4), 2387–2397 (2008). [CrossRef]   [PubMed]  

8. A. Foltynowicz, T. Ban, P. Masłowski, F. Adler, and J. Ye, “Quantum-noise-limited optical frequency comb spectroscopy,” Phys. Rev. Lett. 107(23), 233002 (2011). [CrossRef]   [PubMed]  

9. L. Rutkowski and J. Morville, “Broadband cavity-enhanced molecular spectra from Vernier filtering of a complete frequency comb,” Opt. Lett. 39(23), 6664–6667 (2014). [CrossRef]   [PubMed]  

10. B. Spaun, P. B. Changala, D. Patterson, B. J. Bjork, O. H. Heckl, J. M. Doyle, and J. Ye, “Continuous probing of cold complex molecules with infrared frequency comb spectroscopy,” Nature 533(7604), 517–520 (2016). [CrossRef]   [PubMed]  

11. M. Thorpe, R. Jones, K. Moll, J. Ye, and R. Lalezari, “Precise measurements of optical cavity dispersion and mirror coating properties via femtosecond combs,” Opt. Express 13(3), 882–888 (2005). [CrossRef]   [PubMed]  

12. A. Schliesser, C. Gohle, T. Udem, and T. W. Hänsch, “Complete characterization of a broadband high-finesse cavity using an optical frequency comb,” Opt. Express 14(13), 5975–5983 (2006). [CrossRef]   [PubMed]  

13. T. J. Hammond, A. K. Mills, and D. J. Jones, “Simple method to determine dispersion of high-finesse optical cavities,” Opt. Express 17(11), 8998–9005 (2009). [CrossRef]   [PubMed]  

14. P. Masłowski, K. F. Lee, A. C. Johansson, A. Khodabakhsh, G. Kowzan, L. Rutkowski, A. A. Mills, C. Mohr, J. Jiang, M. E. Fermann, and A. Foltynowicz, “Surpassing the path-limited resolution of Fourier-transform spectrometry with frequency combs,” Phys. Rev. A 93(2), 021802 (2016). [CrossRef]  

15. L. Rutkowski, P. Masłowski, A. C. Johansson, A. Khodabakhsh, and A. Foltynowicz, “Optical frequency comb Fourier transform spectroscopy with sub-nominal resolution - principles and implementation,” arXiv:1612.04808v2 (2017).

16. W. Sellmeier, “Zur Erklärung der abnormen Farbenfolge im Spectrum einiger Substanzen,” Ann. Phys. (Berlin) 219(6), 272–282 (1871). [CrossRef]  

17. I. Silander, T. Hausmaninger, W. Ma, P. Ehlers, and O. Axner, “Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectrometry down to 4 × 10−13 cm−1 Hz-1/2: implementation of a 50,000 finesse cavity,” Opt. Lett. 40(9), 2004–2007 (2015). [CrossRef]   [PubMed]  

18. G. Di Domenico, S. Schilt, and P. Thomann, “Simple approach to the relation between laser frequency noise and laser line shape,” Appl. Opt. 49(25), 4801–4807 (2010). [CrossRef]   [PubMed]  

19. C. Abd Alrahman, A. Khodabakhsh, F. M. Schmidt, Z. Qu, and A. Foltynowicz, “Cavity-enhanced optical frequency comb spectroscopy of high-temperature H2O in a flame,” Opt. Express 22(11), 13889–13895 (2014). [CrossRef]   [PubMed]  

20. P. R. Griffiths and J. A. de Haseth, Fourier Transform Infrared Spectrometry (John Wiley & Sons, Inc., 2006).

21. S. Schiller, “Spectrometry with frequency combs,” Opt. Lett. 27(9), 766–768 (2002). [CrossRef]   [PubMed]  

22. E. R. Peck and B. N. Khanna, “Dispersion of Nitrogen,” J. Opt. Soc. Am. 56(8), 1059–1063 (1966). [CrossRef]  

23. “HITRAN database,” http://www.hitran.org.

24. S. Okubo, Y.-D. Hsieh, H. Inaba, A. Onae, M. Hashimoto, and T. Yasui, “Near-infrared broadband dual-frequency-comb spectroscopy with a resolution beyond the Fourier limit determined by the observation time window,” Opt. Express 23(26), 33184–33193 (2015). [CrossRef]   [PubMed]  

25. K. Osvay, G. Kurdi, J. Hebling, A. P. Kovacs, Z. Bor, and R. Szipocs, “Measurement of the group delay of laser mirrors by a Fabry-Perot interferometer,” Opt. Lett. 20(22), 2339–2341 (1995). [CrossRef]   [PubMed]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Fig. 1
Fig. 1 (a) Experimental setup: f-2f - f-2f interferometer beat note; DBM - double-balanced mixer; DDS - direct digital synthesizer; FC - fiber collimator; λ/2 - half-waveplate; PBS - polarizing beam splitter; λ/4 - quarter-waveplate; PZT - piezoelectric transducer; FTS - Fourier transform spectrometer; EOM - electro-optic modulator; G - diffraction grating; (b) Matching of the comb lines (red) and the cw laser (blue) to the cavity modes (black). The reflected comb lines are shown with the red dashed lines.
Fig. 2
Fig. 2 (a) Spectrum of the empty cavity transmission spanning 16 THz consisting of 16000 discrete modes. Inset: enlargement of 3 modes separated by 3FSR. (b) A zoom of a single mode at 1600 nm (black markers) together with a Lorentzian fit (red line) and the residuum (lower panel).
Fig. 3
Fig. 3 (a) Shift of the cavity mode frequencies Δν measured when the cavity is empty (black curve, left y-axis) and when the cavity is filled with pure N2 at 750 Torr (red curve, right y-axis) plotted together with a calculated shift based on the Sellmeier equation for N2 (blue markers). Note the three orders of magnitude difference between the two y-axis scales. (b) Residuum of a polynomial fit to the shift of the empty cavity modes. (c) Residuum of a polynomial fit to the mode shift of the cavity filled with N2.
Fig. 4
Fig. 4 Group delay dispersion (GDD) of the empty cavity (black solid curve), and the cavity filled with pure N2 at 750 Torr (red solid curve). The dashed curve is the sum of the GDD of N2 calculated using the Sellmeier equation and the experimentally determined empty cavity GDD (dashed black curve).
Fig. 5
Fig. 5 (a) Refractivity of the 3ν1 + ν3 absorption band of 1% of CO2 in N2 at 750 Torr (black markers) together with a fit (red curve). (b) Residual of the fit.

Equations (6)

Equations on this page are rendered with MathJax. Learn more.

Φ( ν q )= ϕ 0 ( ν q )+ ϕ n ( ν q )=2πq,
ν q 0 = qFSR ref 0 +f 0 ,
Δν= ν q ν q 0 = FSR ref 2π [ 2πqΦ( ν q 0 ) ],
GDD( ν q 0 ) 1 4 π 2 2 Φ ν 2 = 1 FSR ref 2 Δν ν 2 .
Δ ν n = ν q 0 [ n( ν q 0 )1 ].
n abs ( ν q 0 )1= c 4 πν q 0 ρ i S i Im[ χ i ( ν q 0 ) ] ,