Waveguide-integrated chip-scale optomechanical magnetometer

Open Access Open Access

Abstract

Optomechanical magnetometers enable highly sensitive magnetic field sensing. However, all such magnetometers to date have been optically excited and read-out either via free space or a tapered optical fiber. This limits their scalability and integrability, and ultimately their range of applications. Here, we present an optomechanical magnetometer that is excited and read-out via a suspended optical waveguide fabricated on the same silicon chip as the magnetometer. Moreover, we demonstrate that thermomechanical noise limited sensitivity is possible using portable electronics and laser. The magnetometer employs a silica microdisk resonator selectively sputtered with a magnetostrictive film of galfenol (FeGa) which induces a resonant frequency shift in response to an external magnetic field. Experimental results reveal the retention of high quality-factor optical whispering gallery mode resonances whilst also demonstrating high sensitivity and dynamic range in ambient conditions. The use of off-the-shelf portable electronics without compromising sensor performance demonstrates promise for applications.

© 2023 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement

1. Introduction

In recent years, optomechanical sensors have emerged as a powerful new type of sensor for stimuli ranging from temperature [1] to pressure [2], force [3,4] and acceleration [5]. Such sensors leverage both optical and mechanical resonances to enable high sensitivity and high spatial resolution. Optomechanical magnetometers are one example [69], that are attractive due to the crucial role that high-sensitivity magnetometers play in applications ranging from fundamental research to medical diagnostics [10,11], mineral exploration and surveying [12,13], magnetic anomaly detection [14,15], and navigation [1618]. Owing to their photonic nature, they are light-weight, small-sized, low power [7,16,19], and can be exceedingly resilient to detriments such as electrical interference and radiation.

At their current state of development, optomechanical magnetometers have achieved tens-of-micron spatial resolution with sensitivities ranging from several n$\mathrm {T}/\sqrt {\mathrm {Hz}}$ down to tens of p$\mathrm {T}/\sqrt {\mathrm {Hz}}$ [7,8,16,20]. The demonstrated sensitivity is competitive with SQUID and diamond magnetometers of similar size but without the need for cryogenics or high-powered optical and RF components [16,21,22], with theoretical models suggesting that sensitivities in the low, or even sub-, femtotesla may be possible in the future [23].

To-date, optomechanical magnetometers have used free-space or tapered-fiber coupling for optical excitation and readout [9,22,24]. This prevents them from being fully integrated on a silicon chip. Furthermore, those demonstrations used the magnetostrictive material Terfenol-D to convert magnetic fields into a measurable strain [6,7]. Terfenol-D has be shown to be difficult to deposit reproducibly and vulnerable to corrosion and oxidation [25]. Works with this material have also relied on high performance laser and electronic systems [19]. Together, this introduces significant challenges for applications outside of laboratory environments. The work reported here seeks to address these challenges.

We develop an optomechanical magnetometer that is efficiently coupled to an on-chip suspended waveguide, by employing galfenol (Fe$_{82}$Ga$_{18}$) for the first time to convert magnetic fields to a measurable mechanical signal. This provides low-cost sputter-coated thin-films with improved resilience to corrosion and oxidation [25] and good magnetostriction ($\sim$400 ppm) at lower saturation fields [25,26]. We also use portable electronic and laser systems to control and read-out the magnetometer, showing that they allow performance that is limited by fundamental thermomechanical noise rather than laser or electronic noise. Together, this represents progress towards robust, portable and high performance mangetometers that could be employed in diverse research and industrial settings.

2. Design and simulation

2.1 Device design and functionality

The device design concept is depicted in Fig. 1. It is based around a 100 $\mathrm {\mu }$m-diameter silica microdisk cavity on a silicon chip. This microdisk is capable of supporting optical whispering galley modes (WGMs) throughout the visible and near-infrared spectrum as well as megahertz frequency mechanical resonances. A 1.5 $\mathrm {\mu }$m wide silica waveguide is fabricated from the same layer of silica as the microdisk. Both microdisk and waveguide are undercut so that the optical modes are confined to the silica, rather than leaking into the higher-refractive index silicon substrate. The microdisk is suspended from a central silicon pedestal. The waveguide is suspended using thin silica tethers that are patterned along its length to reduce warping of the waveguide that can be caused by the strains originating from the compressive stress intrinsically present within the SiO2 films as a result of the wafer fabrication process. Buckling of the waveguide could lead to severe bending losses, and out-of-plane buckling can lead to inconsistent coupling between the waveguide and optical cavity. The tethers are sub-wavelength (240 nm width) in order to minimise optical scattering of light propagating within the waveguide. As silica is lower refractive index than many other waveguide material (e.g., silicon) the guided wavelength is longer resulting is less scattering losses from surface roughness introduced by fabrication processes [27]. The waveguide is broadened (inverse-tapered) at input and output to efficiently mode-match light into and out-of tapered optical fibers.

 figure: Fig. 1.

Fig. 1. Design of the integrated SiO2 magnetometer. a) Schematic of the device. Here, laser light is coupled to the trumpet waveguide via a tapered fiber. The waveguide is narrow at the centre to optimize evanescent field coupling to the disk cavity. The galfenol is sputtered on top of the cavity. The tethers support the waveguide, preventing buckling. b) The cross-sectional schematic of the device (the tethers supporting the waveguide are ommited). c) Magnetostrictive strain of the galfenol changes the radius of the disk facilitating the magnetic sensing. d) The change in radius induces an intensity modulation of a sideband-detuned transmission measurement caused by the shift in optical resonance.

Download Full Size | PDF

The microdisk is coated with galfenol in a disk of diameter sufficiently smaller than the disk diameter so as not to introduce optical absorption. Galfenol is chosen because of its high magnetostriction, low fragility, and low volatility [25]. When a magnetic field is applied, the expansion of the galfenol induces strain in the microdisk, changing the optical path length and hence the optical resonance frequency. The magnetostrictive response is amplified at frequencies close to mechanical resonances of the microdisk, leading to enhanced modulation of the intracavity optical field.

The light is coupled into the disk evanescently from the suspended waveguides which are designed to support single mode propagation around 1550 nm. The coupling of light from an optical fiber into the on-chip waveguide despite the geometric mismatch (SMF-28 optical fiber has core and cladding diameters of 10 $\mathrm {\mu }$m compared to the 300 nm thickness of the waveguide) is facilitated by mode-matched taper-drawn fibers (similar to those presented in [28]) and trumpeted waveguides (i.e. width tapered waveguides 4 $\mathrm {\mu }$m down to 1.5 $\mathrm {\mu }$m over a length of 30 $\mathrm {\mu }$m). This allows adiabatic coupling of light to and from the waveguides. Further, the 4 $\mathrm {\mu }$m wide flared section acts as a semi-rigid anchor point for the fiber, and its size reduces the requirement for extremely precise positioning. This allows greater tolerance to imperfections in important integration processes, such as bonding the fiber tip in place.

2.2 Finite-element simulations

Finite Element Method (FEM) simulations performed with Ansys-Lumerical software for the optical performance of the device are presented in Fig. 2 (a). Here the coupling from fiber to waveguide is studied by monitoring the optical mode cross-section ($yz$) along the propagation direction ($x$). The cross-sections shown in (i-iii) correspond to cross-sections of only fiber (i), fiber & waveguide (ii), and only waveguide (iii). For simplicity, both the fiber taper and the waveguide were considered uniform at the coupling region. The taper was chosen to have a 1 $\mathrm {\mu }$m diameter with the optimum waveguide width of 4 $\mathrm {\mu }$m then found using recursive simulations. From the simulated cross-sections, we see that the optical mode migrates from the fiber into the waveguide. We obtain a fiber-to-waveguide coupling efficiency of 60% by taking the ratio of the optical power contained within the fiber at point (i) and within the waveguide at point (iii).

 figure: Fig. 2.

Fig. 2. FEM simulations of the integrated optomechanical magnetometer. a) The optical properties of the device. The cross-sectional optical mode profiles (i-iii) at their corresponding green rectangles demonstrate the evolution of the optical modes from fiber to SiO2 trumpet waveguide and (iv) the cross-sectional optical mode inside the microdisk. (v) Depicts the optical mode at the planar cross-section (yellow rectangle) of the device, and (vi) shows the optical transmission spectrum of the microdisk resonator. (b) Mechanical simulation revealing the resonance frequencies and their flexual mode-shapes.

Download Full Size | PDF

Within the same simulation the waveguide-to-disk coupling and disk resonances were also studied. Here the optical excitation frequency was swept and the optical intensity across the geometry was recorded at each frequency. The transmission efficiency across the resonator can then be calculated by comparing the integrated intensity over the cross-sections of the input and output of the waveguide (i.e., $T = I_{\text {out}}/I_{\text {in}}$, as measured at points (iii) & (vi)). Fig. 2(a)(vi) shows the expected periodic transmission dips when the frequency of the light matches WGMs of the microdisk (the small sharp peaks above the baseline are an artifact of the finite-time FEM). The observed simulated transmission drops of as much as 70% are indicative that efficient coupling of light into the WGMs is possible. The coupling efficiency was then further refined during the fabrication process to produced the results seen in Section 4. The correspondence of the dips with WGMs is confirmed in Fig. 2 (a)(v), which shows confinement of the light to the periphery of the disk when driven on resonance (at a wavelength of 1551 nm in this case). To further corroborate the confinement of the WGMs we performed an axisymmetric eigenmode-solver FEM simulation in COMSOL Multiphysics (Fig. 2 (a) (iv)). This confirmed that the WGM is contained within the outer 5 $\mathrm {\mu }$m of the disk, as expected.

The Free Spectral Range (FSR) of the optical resonances and corresponding coupling were calculated from the simulated transmission in Fig. 2(a)(vi). We find a simulated FSR of approximately 7 nm. This compares well to the expected FSR given the circumference of the microdisk of:

$$\Delta \lambda \approx \frac{\lambda^2}{n(\lambda)L} \approx 7.1 \;\mathrm{nm},$$
where $n(\lambda )$ is the effective refractive index of the cavity mode, which was determined from Lumerical eigenmode simulations to be $\sim$1.08 (consistent with previous report for SiO$_\text {2}$ structures of similar dimensions [29,30]), and $L$ is the length of the cavity i.e., $L =$ 100$\mathrm {\pi }\,\mathrm {\mu }\textrm {m}$.

As evidenced by the results in Fig. 2 (a) (iv & v), the optical field of the Whispering Gallery Mode (WGM) extends negligibly into the centre of the SiO2 disk. Hence, the addition of the optically metallic magnetostrictive layer to the disk’s centre should not significantly affect the quality of the optical modes contained therein [20].

Using COMSOL Multiphysics, we performed further simulations to assess the mechanical properties of the microdisk. As shown in Fig. 2 (b), we found the mechanical eigenmodes by using fully three-dimensional geometry of the released devices. This was necessitated because of the inclusion of stress-release slots (discussed in section 3) that break the axial symmetry of the mechanical modes. The physical properties of the galfenol were taken from the datasheet supplied by TdVib LLC. The lowest frequency mechanical flexural mode and two lowest frequency crown modes are shown in Fig. 2 (b), with mechanical frequencies of 3.12, 3.21, and 3.26 MHz, respectively.

3. Device fabrication

The fabrication process used to produce the devices is outlined in Fig. 3 (a) (i). SiO2 (300 nm) on Si substrate wafers (500 $\mathrm {\mu }$m, 4") were diced into square 15$\times$15 mm chips, large enough to fit more than 100 devices per chip. Electron beam lithography was used to define patterns for galfenol deposition and markers for subsequent lithography steps in the following way. Two layers of PMMA resist were spin-coated (PMMA 950k and 495k at 2500 RPM) onto of the SiO2/Si substrate, then patterned with an Electron-beam Pattern Generator (EBPG) (Raith EBPG 5150) with 100 kV accelerating voltage and a 1200 $\mathrm {\mu }$C/cm2 dosage. Post exposure, the chips were developed in methyl isobutyl ketone (MIBK) and rinsed with Isopropyl Alcohol (IPA).

 figure: Fig. 3.

Fig. 3. Representation of the fabrication process. a) PMMA e-beam resist deposition. b) EBPG exposure and development. c). Galfenol sputter deposition. d) lift-off. e) ARP e-beam resist deposition. f) Second EBPG exposure and development. g) Reactive-ion Etching (RIE) of SiO2. h) Resist removal. i) Released devices after undercutting by XeF2 etching of the Si layer. SEM images of the final devices depicting j) trumpeted wave guide and k) the optomechanical cavity with galfenol layer in its centre and coupling waveguide.

Download Full Size | PDF

To produce the markers, 5 nm of Ti and 50 nm of Au were e-beam evaporator deposited (Temescal FC-2000) followed by a lift-process via submersion into acetone and IPA. The galfenol films were then sputtered by magnetron DC sputtering in an argon atmosphere (150 W, 2 mTorr) with a (3 inches dia.) galfenol target. A seeding layer (5 nm Ta, plus a 16 nm Cu) and capping layer (5 nm Ta) were used to promote adhesion and inhibit corrosion respectively. Afterwards, the lift-off process was repeated, resulting in a 300 nm thick, 60 $\mathrm {\mu }$m diameter circular thin-film of galfenol on top of the SiO2 layer. This film-thickness is consistent with previous reports for high-piezomagnetic responses [31].

With markers produced and galfenol deposited, the waveguide and the disk cavity structures were then defined. For this, 20 nm thick Ti prime adhesion helper was spin-coated (4000 rpm) and baked (150$^{\circ }$C, 15 min) follow by a layer of ARP 6200.09 (CSAR-62, All Resist) 350 nm thick (1250 rpm spin-coat, 180$^{\circ }$C for 5 min bake). The chip was then patterned with the Raith EBPG 5150 (100 kV, 260 $\mathrm {\mu }$C/cm2). Proximity effect correction was performed using GenISys Beamer software to ensure precision and reproducibility in the EBPG process. Post exposure, the patterns were developed with All Resist AR600-546 and rinsed with o-xylene and IPA.

RIE was used to remove the unwanted SiO2 using an Oxford Instruments PlasmaPro 80. Here 25 sccm CHF3 and 45 sccm Ar at 200 W RF power for 12 min anisotropically etched all the way through the SiO2 layer exposing the silicon substrate. 50 sccm O2 at 50 W was then used to remove any residual resist. Finally, the SiO2 structures were undercut by etching of the supporting silicon with Xenon Difluoride (XeF2) gas (SPTS XACTIX e2 Series XeF2 etcher). Here, 10 pulses of XeF2 gas at a pressure of 2 torr provides an isotropic etch rate of about 1.4 $\mathrm {\mu }$m per pulse with a selectivity of >1000:1. This removed the Si beneath the silica waveguide and WGMs of the microdisk whilst leaving both the silica layer and galfenol unmarred.

SEM (Jeol 7800) imaging of the devices was performed in order to assess their structural integrity. Figure 3 (j) & (k) shows SEMs of the trumpet-shaped waveguide ends for fiber coupling and the waveguide near the resonator, supported by tethers to the main body of the wafer. It is apparent that the waveguide shows no signs of buckling or collapse after the release process. The SEM further reveals superficial stress marks on the surface of the galfenol pad, However as these features appear to be only surface level, it is inferred that they do not significantly alter the magnetometers' performance. It can also be observed that the undercut beneath the silica layer is approximately 18 $\mathrm {\mu }$m. This undercut extends under the disk, leaving behind a silicon (Si) pedestal which is obscured by the galfenol coating above. Measurements on a device with no galfenol revealed a pedestal width of 15 $\mathrm {\mu }$m (measured with a Zeta 300 3D optical profiler).

Stress release slots in the resonator were found to be necessary to prevent buckling of the disk due to the inherent compressive strain within the freestanding SiO2 structure. Further, these slots simultaniously aid in strain compensation from the galfenol film. However, as discussed in section 2, these slots are expected to have negligible effect on the optical modes because they are outside of the region containing appreciable intensity. The mechanical simulation of Fig. 2 (b) fully accounts for the effect of the slots on the mechanical eigen-frequencies.

A critical parameter for consideration during fabrication of the device is the distance between the disk and the waveguide at the coupling region ($d$). As the light is coupled evanescently the coupling efficiency ($\kappa$) follows the relation $\kappa \propto e^{-d}$ [32]. Devices with a range of waveguide-microdisk coupling distances were fabricated in order to produce resonators with optimum coupling strengths. The devices with near-critical coupling were further investigated.

4. Device performance

The experimental setup used to assess the performance of the integrated magnetometers is depicted in Fig. 4(a). Here, a continuously tuneable laser (EXFO T100S-HP) supplied light to the resonator via tapered optical fibers with a house-built test rig featuring two 3-axis translation stages for precise positioning of the fibers. The transmitted light was then guided to a Thorlabs (PDA100-CS) photodector (PD) and the photocurrent was analyzed with a spectrum analyzer (Siglent SSA 3021x). A function generator (Rigol DG-1022) was used to directly drive a home-wound coil (8 turns, 10 mm dia.) held approximately 1 mm above the chip, producing a field of $\sim$50 $\mathrm {\mu }$T$_{\text {PP}}$ (determined from the Biot-Savart relation and corroborated with a Lakeshore Cryotronics F71 Hall-probe) at the surface of the chip (drive voltage of 10 V$_{\text {PP}}$).

 figure: Fig. 4.

Fig. 4. Experimental performance of the magnetometers. a) Schematic demonstrating the experimental setup with corresponding photographs of the laser system b), and devices on test-rig c) used for the measurements. d) Optical transmission spectra of the devices with accompanying high-resolution spectra around one of the WGM resonances (inset of e) which was used for the sensing investigations. e) Power spectrum of the transmitted light with and without externally applied magnetic field. A 17.5 dB SNR is observed when the field is applied. With Eq. (3) this reveals a sensitivity of $2\;\mathrm {\mu }\textrm {T}/\sqrt {\text {Hz}}$.

Download Full Size | PDF

The emission wavelength of the laser was swept and the voltage output from the PD (and hence power via the known responsivity and transimpedance gain of the PD) was recorded to characterise the optical mode spectrum of fabricated devices. The optical transmission spectrum of a typical device is presented in Fig. 4 (d), revealing many dips in the transmission, with the most prominent series of features being attributed to the fundamental WGMs (e.g. the resonance depicted in Fig. 4(e) inset). The observed FSR for this family of modes is $\approx$ 7 nm which is in good agreement with the FSR as determined from the FEM seen in Section 2. and Fig. 2(a)(vi). Also observed were series of features which are less pronounced however share comparable FSRs, eluding to higher order WGMs being excited. On this device (with a designed waveguide-microdisk separation of 550 nm) we find that the WGM at a wavelength of 1551 nm (enclosed by the dashed box in Fig. 4 (d)) is close to critically coupled, with a transmission dip of $\sim 95{\%}$.

Because the 1551 nm WMG mode is close to critically coupled, we select it to perform magnetic field measurements. The sensors performance is governed by the slope of the Lorentzian shaped transmission dips ($d I / d \lambda$) and this, near critical coupling provides the best possible signal from the device. A high-resolution sweep across the mode is shown in the inset of Fig. 4(e). From this the optical $Q$ of the cavity is estimated to be:

$$Q_{opt}\approx \frac{\lambda_0}{\Delta\lambda_{FWHM}} \approx10^5 .$$

For many applications, it is desirable to use a low cost, low power, and compact laser source, together with compact electronic systems, rather than the high performance EXFO fiber-laser and associated electronics used in this work to date. Here, we test whether it is possible to do this without sacrificing performance. A commercially available Distributed Feedback (DFB) laser (Eblana EP1550) with a portable laser driver (Koheron CTL101) was used to couple light onto and off the chip (Fig. 4 (b)) were used for all subsequent measurements.

Tuning the DFB laser to the side of the 1551 nm WGM allows shifts in the resonance frequency to be directly observed as changes in the optical intensity. This allows optical detection of mechanical vibrations, and hence magnetic field, without the need for interferometric detection [6]. Analysing the resulting photocurrent on a spectrum analyser reveals the mechanical mode spectrum shown in Fig. 4(b). Three mechanical modes are observed at frequencies of 3.55, 3.58, and 3.64 MHz. We attribute the discrepancy between the measured and simulated mechanical frequencies to the inherent stress of the galfenol film ($\sigma$=500 MPa tensile, determined via stylus profilometry (Tencor), measurements analyzed with Stoney’s formula [33]) adding a stiffening effect to the mechanical resonances.

The noise-floor of the measurement consists of two components. At frequencies far away from the mechanical resonance frequencies it is dominated by laser noise. This is evidenced by an increase in noise when the laser is tuned to the side of the WGM compared to when it is at a frequency far away from the mode. At frequencies close to the mechanical resonance frequencies, it is dominated by the thermomechanical noise. We can therefore conclude that the compact electronic systems used introduce no degradation in performance and, close to the mechanical resonances, neither does the optical noise of the DFB laser.

To determine the magnetic field sensitivity of the device, we apply a magnetic field at the frequency of the most prevalent mechanical mode (3.55 MHz). This induces a sharp peak in the Power Spectrum Density (PSD) (Fig. 4(e), blue-trace), above the thermomechanical signal of the disk which remains unchanged by the applied field. This sharp peak evidences that magnetic fields can be detected. With this particular applied field (BAC = 50 $\mathrm {\mu }$T) we measure a Signal-to-Noise Ratio (SNR) of 17.5 dB. The magnetic sensitivity of the device at 3.55 MHz could then be calculated using:

$$S = B_{AC} \left({10^{\frac{SNR}{10}}\times \mathrm{RBW}} \right)^{{-}1/2}$$
where RBW is the resolution bandwidth of the spectrum analyser [6]. This yielded a sensitivity of 2 $\mathrm {\mu }\mathrm {T}/\sqrt {\mathrm {Hz}}$, which is comparatively less sensitive than previously demonstrated optomechanical magnetometers that present sub n$\mathrm {T}/\sqrt {\mathrm {Hz}}$ sensitivities [16].

This reduced sensitivity can be attributed to the geometric design of the device. With these devices the galfenol lies in part, above the pedestal, where the silicon greatly suppresses both mechanical motion and imbued strain. Further, the mechanical eigenmodes have very little motion where the galfenol resides, thus do not experience the maximum possible driving force from the magnetostriction. These effects provide a reduction of the force exerted onto the optical eigenmodes from magnetostrictive stress. Thus, the sensitivity could be considerably improved by employing device geometries which specifically optimize deformation of the optical path caused by the magnetostrictive strain. Such potential geometries could include double-disks with a hybridized optical mode, cantilevers with embedded photonic crystals, zipper cavities, or Fabry Perot cavities where the galfenol magnetomechanical mode directly influences or overlaps one or more optical resonance boundaries.

Despite the modest sensitivity, this work achieves thermomechanically limited sensing with suspended waveguide coupling and a galfenol thin-film atop the optomechanical resonator whilst utilising portable electronics and DFB laser.

5. Conclusion

Optomechanical magnetometers promise to enable a range of research and industrial applications. Many of these will require fully integrated magnetometers operating with compact lasers and electronics. In this work we make progress towards this goal, demonstrating an optomechanical magnetometer that is integrated on a silicon chip with a suspended optical waveguide, utilises galfenol as a magnetostrictive material to provide improved resilience to corrosion and oxidation, and achieves thermomechanical noise-limited performance using a DFB laser and compact electronic systems.

Funding

Defence Science and Technology Group; Australian Research Council; Centre of Excellence for Engineered Quantum systems (EQUS) (CE170100009); Orica Australia Pty Ltd.

Acknowledgments

The Authors acknowledge the highly valuable advice and support provided by Rodney Appleby. The authors acknowledge the facilities, and the scientific and technical assistance, of the Australian Microscopy & Microanalysis Research Facility at the Centre for Microscopy and Microanalysis, The University of Queensland. This work was performed in part at the Queensland node of the Australian National Fabrication Facility. A company established under the National Collaborative Research Infrastructure Strategy to provide nano and microfabrication facilities for Australia’s researchers. The Commonwealth of Australia (represented by the Defence Science and Technology Group) supports this research through a Defence Science Partnerships agreement. This work was financially supported by the Australian Research Council (ARC) Centre of Excellence for Engineered Quantum systems (EQUS): Grant No. CE170100009, and by Orica Australia Pty Ltd.

Disclosures

The authors declare no conflicts of interest.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

References

1. T. P. Purdy, K. E. Grutter, K. Srinivasan, and J. M. Taylor, “Quantum correlations from a room-temperature optomechanical cavity,” Science 356(6344), 1265–1268 (2017). [CrossRef]  

2. S. Basiri-Esfahani, A. Armin, S. Forstner, and W. P. Bowen, “Precision ultrasound sensing on a chip,” Nat. Commun. 10(1), 132 (2019). [CrossRef]  

3. E. Gavartin, P. Verlot, and T. J. Kippenberg, “A hybrid on-chip optomechanical transducer for ultrasensitive force measurements,” Nat. Nanotechnol. 7(8), 509–514 (2012). [CrossRef]  

4. G. I. Harris, D. L. McAuslan, T. M. Stace, A. C. Doherty, and W. P. Bowen, “Minimum requirements for feedback enhanced force sensing,” Phys. Rev. Lett. 111(10), 103603 (2013). [CrossRef]  

5. A. G. Krause, M. Winger, T. D. Blasius, L. Qiang, and P. Oskar, “A high-resolution microchip optomechanical accelerometer,” Nat. Photonics 6(11), 768–772 (2012). [CrossRef]  

6. S. Forstner, S. Prams, J. Knittel, E. D. van Ooijen, J. D. Swaim, G. I. Harris, A. Szorkovszky, W. P. Bowen, and H. Rubinsztein-Dunlop, “Cavity optomechanical magnetometer,” Phys. Rev. Lett. 108(12), 120801 (2012). [CrossRef]  

7. S. Forstner, E. Sheridan, J. Knittel, C. L. Humphreys, G. A. Brawley, H. Rubinsztein-Dunlop, and W. P. Bowen, “Ultrasensitive optomechanical magnetometry,” Adv. Mater. 26(36), 6348–6353 (2014). [CrossRef]  

8. B.-B. Li, G. Brawley, H. Greenall, S. Forstner, E. Sheridan, H. Rubinsztein-Dunlop, and W. P. Bowen, “Ultrabroadband and sensitive cavity optomechanical magnetometry,” Photonics Res. 8(7), 1064–1071 (2020). [CrossRef]  

9. C. Yu, J. Janousek, E. Sheridan, D. L. McAuslan, H. Rubinsztein-Dunlop, P. K. Lam, Y. Zhang, and W. P. Bowen, “Optomechanical magnetometry with a macroscopic resonator,” Phys. Rev. Appl. 5(4), 044007 (2016). [CrossRef]  

10. I. Savukov and T. Karaulanov, “Magnetic-resonance imaging of the human brain with an atomic magnetometer,” Appl. Phys. Lett. 103(4), 1 (2013). [CrossRef]  

11. D. R. Glenn, D. B. Bucher, J. Lee, M. D. Lukin, H. Park, and R. L. Walsworth, “High-resolution magnetic resonance spectroscopy using a solid-state spin sensor,” Nature 555(7696), 351–354 (2018). [CrossRef]  

12. H.-G. Meyer, R. Stolz, A. Chwala, and M. Schulz, “Squid technology for geophysical exploration,” phys. stat. sol. (c) 2(5), 1504–1509 (2005). [CrossRef]  

13. A. Edelstein, “Advances in magnetometry,” J. Phys. Condens. Matter 19(16), 165217 (2007). [CrossRef]  

14. C. Li, S. Huang, D. Wei, Y. Zhong, and K. Gong, “Detection range of airborne magnetometers in magnetic anomaly detection,” J. Eng. Sci. Technol. Rev. 8(4), 105–110 (2015). [CrossRef]  

15. A. Sheinker, L. Frumkis, B. Ginzburg, N. Salomonski, and B.-Z. Kaplan, “Magnetic anomaly detection using a three-axis magnetometer,” IEEE Trans. Magn. 45(1), 160–167 (2009). [CrossRef]  

16. J. S. Bennett, B. E. Vyhnalek, H. Greenall, E. M. Bridge, F. Gotardo, S. Forstner, G. I. Harris, F. A. Miranda, and W. P. Bowen, “Precision magnetometers for aerospace applications: A review,” Sensors 21(16), 5568 (2021). [CrossRef]  

17. D. Budker and M. Romalis, “Optical magnetometry,” Nat. Phys. 3(4), 227–234 (2007). [CrossRef]  

18. S. Rodriguez, S. Vinatier, D. Cordier, et al., “Science goals and mission concepts for a future orbital and in situ exploration of titan,” Responses to the ESA Voyage 2050 long-term plan (2019).

19. B.-B. Li, J. Bílek, U. B. Hoff, L. S. Madsen, S. Forstner, V. Prakash, C. Schäfermeier, T. Gehring, W. P. Bowen, and U. L. Andersen, “Quantum enhanced optomechanical magnetometry,” Optica 5(7), 850–856 (2018). [CrossRef]  

20. B.-B. Li, D. Bulla, V. Prakash, S. Forstner, A. Dehghan-Manshadi, H. Rubinsztein-Dunlop, S. Foster, and W. P. Bowen, “Invited article: Scalable high-sensitivity optomechanical magnetometers on a chip,” APL Photonics 3(12), 120806 (2018). [CrossRef]  

21. J. Zhu, G. Zhao, I. Savukov, and L. Yang, “Polymer encapsulated microcavity optomechanical magnetometer,” Sci. Rep. 7(1), 8896 (2017). [CrossRef]  

22. Y. Yu, S. Forstner, H. Rubinsztein-Dunlop, and W. P. Bowen, “Modelling of cavity optomechanical magnetometers,” Sensors 18(5), 1558 (2018). [CrossRef]  

23. W. P. Bowen and C. Yu, Cavity Optomechanical Magnetometers (Springer, 2017).

24. E. Romero-Sánchez, W. P. Bowen, M. R. Vanner, K. Xia, and J. Twamley, “Quantum magnetomechanics: Towards the ultrastrong coupling regime,” Phys. Rev. B 97(2), 024109 (2018). [CrossRef]  

25. L. R. Nivedita, P. Manivel, R. Pandian, S. Murugesan, N. A. Morley, K. Asokan, and R. T. Rajendra Kumar, “Enhancement of magnetostrictive properties of galfenol thin films,” J. Magn. Magn. Mater. 451, 300–304 (2018). [CrossRef]  

26. A. Clark, M. Wun-Fogle, J. Restorff, K. W. Dennis, T. A. Lograsso, and R. W. McCallum, “Temperature dependence of the magnetic anisotropy and magnetostriction of fe100- xgax (x= 8.6, 16.6, 28.5),” J. Appl. Phys. 97(10), 1 (2005). [CrossRef]  

27. G. Lin, R. Henriet, A. Coillet, M. Jacquot, L. Furfaro, G. Cibiel, L. Larger, and Y. K. Chembo, “Dependence of quality factor on surface roughness in crystalline whispering-gallery mode resonators,” Opt. Lett. 43(3), 495–498 (2018). [CrossRef]  

28. J. M. Ward, A. Maimaiti, V. H. Le, and S. Nic Chormaic, “Contributed Review: Optical micro- and nanofiber pulling rig,” Rev. Sci. Instrum. 85(11), 1 (2014). [CrossRef]  

29. L. Tong, J. Lou, and E. Mazur, “Single-mode guiding properties of subwavelength-diameter silica and silicon wire waveguides,” Opt. Express 12(6), 1025–1035 (2004). [CrossRef]  

30. U. Wiedemann, K. Karapetyan, C. Dan, D. Pritzkau, W. Alt, S. Irsen, and D. Meschede, “Measurement of submicrometre diameters of tapered optical fibres using harmonic generation,” Opt. Express 18(8), 7693–7704 (2010). [CrossRef]  

31. B. Adolphi, J. McCord, M. Bertram, C. Oertel, U. Merkel, U. Marschner, R. Schäfer, C. Wenzel, and W. Fischer, “Improvement of sputtered galfenol thin films for sensor applications,” Smart Mater. Struct. 19(5), 055013 (2010). [CrossRef]  

32. F. Bo, Ş. K. Özdemir, F. Monifi, J. Zhang, G. Zhang, J. Xu, and L. Yang, “Controllable oscillatory lateral coupling in a waveguide-microdisk-resonator system,” Sci. Rep. 7(1), 8045 (2017). [CrossRef]  

33. A. Kelly and C. Zweben, 3.10 - Thermomechanical Behavior of Layered MMCs and Graded Structures (Pergamon, Oxford, 2000).

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (4)

Fig. 1.
Fig. 1. Design of the integrated SiO2 magnetometer. a) Schematic of the device. Here, laser light is coupled to the trumpet waveguide via a tapered fiber. The waveguide is narrow at the centre to optimize evanescent field coupling to the disk cavity. The galfenol is sputtered on top of the cavity. The tethers support the waveguide, preventing buckling. b) The cross-sectional schematic of the device (the tethers supporting the waveguide are ommited). c) Magnetostrictive strain of the galfenol changes the radius of the disk facilitating the magnetic sensing. d) The change in radius induces an intensity modulation of a sideband-detuned transmission measurement caused by the shift in optical resonance.
Fig. 2.
Fig. 2. FEM simulations of the integrated optomechanical magnetometer. a) The optical properties of the device. The cross-sectional optical mode profiles (i-iii) at their corresponding green rectangles demonstrate the evolution of the optical modes from fiber to SiO2 trumpet waveguide and (iv) the cross-sectional optical mode inside the microdisk. (v) Depicts the optical mode at the planar cross-section (yellow rectangle) of the device, and (vi) shows the optical transmission spectrum of the microdisk resonator. (b) Mechanical simulation revealing the resonance frequencies and their flexual mode-shapes.
Fig. 3.
Fig. 3. Representation of the fabrication process. a) PMMA e-beam resist deposition. b) EBPG exposure and development. c). Galfenol sputter deposition. d) lift-off. e) ARP e-beam resist deposition. f) Second EBPG exposure and development. g) Reactive-ion Etching (RIE) of SiO2. h) Resist removal. i) Released devices after undercutting by XeF2 etching of the Si layer. SEM images of the final devices depicting j) trumpeted wave guide and k) the optomechanical cavity with galfenol layer in its centre and coupling waveguide.
Fig. 4.
Fig. 4. Experimental performance of the magnetometers. a) Schematic demonstrating the experimental setup with corresponding photographs of the laser system b), and devices on test-rig c) used for the measurements. d) Optical transmission spectra of the devices with accompanying high-resolution spectra around one of the WGM resonances (inset of e) which was used for the sensing investigations. e) Power spectrum of the transmitted light with and without externally applied magnetic field. A 17.5 dB SNR is observed when the field is applied. With Eq. (3) this reveals a sensitivity of $2\;\mathrm {\mu }\textrm {T}/\sqrt {\text {Hz}}$.

Equations (3)

Equations on this page are rendered with MathJax. Learn more.

Δ λ λ 2 n ( λ ) L 7.1 n m ,
Q o p t λ 0 Δ λ F W H M 10 5 .
S = B A C ( 10 S N R 10 × R B W ) 1 / 2