Abstract
A 6.65 m spectrometer has been fabricated indigenously using a concave grating of frequency 4800 grooves/mm for spectroscopic research of atomic and molecular gases using synchrotron radiation. Wavelength range of the spectrometer is 700Å to 2000Å with reciprocal linear dispersion of 0.3Å /mm and spectral resolution of 0.01Å.
© 2004 Optical Society of America
PDF ArticleMore Like This
Marshall L. Ginter and Charles M. Brown
THJ3 OSA Annual Meeting (FIO) 1986
John Fortna, Harry Morgan, Hailemichael Seyoum, David C. Humm, Mitchell L. Furst, and Lanny R. Hughey
WB7 High Resolution Spectroscopy (HRS) 1993
K. Yoshino, J.R. Esmond, W.H. Parkinson, A.P. Thorne, L.E. Murray, G. Cox, R.C.M. Learner, K. Ito, T. Imajo, T. Mastsui, A. S.-C. Cheung, and K.-S. Leung
FTuA.3 Fourier Transform Spectroscopy (FTS) 1997