Abstract
An intravascular catheter for Raman spectroscopic detection and analysis of coronary atherosclerotic disease has been developed. The catheter, having an outer diameter of 2 mm, consisted of a side-view-type micro-Raman probe, an imaging fiber bundle, a working channel (injection drain), and a balloon. By inflating the balloon, the probe was brought close to the inner wall of a modeled blood flow system and detected a phantom target buried in the wall. Results obtained demonstrate the possibility of using the spectroscopic catheter for molecular diagnosis of coronary lesions.
© 2006 Optical Society of America
Full Article | PDF Article