Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Application of a complex constraint for biological samples in coherent diffractive imaging

Open Access Open Access

Abstract

We demonstrate the application of a complex constraint in the reconstruction of images from phase-diverse Fresnel coherent diffraction data for heterogeneous biological objects. The application of this constraint is shown to improve the quality of the reconstruction of both the phase and the magnitude of the complex object transmission function.

©2013 Optical Society of America

1. Introduction

X-ray Coherent Diffractive Imaging (CDI) is rapidly developing as a routine method for biological imaging because the short wavelength and specific interaction of X-rays with matter can be exploited to image intracellular features [18] with spatial resolution beyond the limits imposed by image-forming optics [9]. Phase-diverse Fresnel CDI (FCDI) is an extension of ptychography [10] in which an object is translated in three dimensions through illumination with phase curvature introduced by a focusing optic [11]. It has been shown to provide increased contrast with reduced X-ray dose, while imaging extended areas virtually free from artefacts [1012]. Recently, phase-diverse FCDI has been shown to provide high-contrast, high-resolution images of biological samples with significantly lower X-ray dose compared to other CDI methods [7,13].

CDI involves recording a coherent diffraction pattern of an object in the far-field and reconstructing the complex exit surface wave from the object using iterative phase retrieval algorithms. These algorithms depend on a set of constraints defined by the measured intensities and known properties of the object and illumination [14,15]. The reconstructed exit surface wave provides a quantitative measure of the complex transmission function of the object. A priori knowledge of the object composition can therefore be used to define a constraint that is applied in the iterative algorithms. For example, the magnitude, T, and phase, ϕ, of the complex transmission function can be independently constrained to fall within physically reasonable limits (see, for examples [16,17]). For homogeneous areas of an object of known composition, the ratio between the logarithm of the magnitude of the transmitted wave, ln(|T|), to the phase change, ϕ, in the illuminating wave as it passes through the object can be constrained to a physically appropriate value. Known as a complex constraint, this is found to improve the convergence of the phase retrieval algorithm and the quality of the reconstructed object image [18]. However, for heterogeneous objects such as biological specimens the ratio ln(|T|)/ϕ cannot be constrained to a single value. It was previously asserted that constraining this ratio to a range that encompasses the values for all known components of a heterogeneous object will improve convergence of the phase retrieval algorithm [18]. Here we test this assertion by applying a complex constraint to whole red blood cells infected with the malaria parasite Plasmodium falciparum using phase-diverse Fresnel coherent diffractive imaging at X-ray energies of 520 eV and 2535 eV. We show that by constraining the ratio ln(|T|)/ϕ to a range encompassing the values appropriate for the cell and parasite components the quality of the reconstruction of both the phase and magnitude components of the complex transmission function are improved, achieving consistency between their values that is not otherwise possible.

2. Method

The application of the complex constraint employed here follows the description given by Clark et al. [18]. The X-ray transmission and phase retardation in the complex wavefield in each finite area (or imaging pixel) of the object are, respectively:

ln(|T|)=kβzdz,
and
φ=kδzdz
where k is the wavenumber, z is the thickness of the object, and δ and β are the decrement from unity of the real and imaginary parts of the refractive index. The average X-ray transmission and phase retardation projected through the object at each pixel of the object can therefore be given by:
ln(|TP|)=kβ¯z,
and
ϕP=kδ¯z
where TP and ϕP are the projected transmission and phase and δ¯ and β¯ are the average values for δ and β over z. Thus the transmission and phase of the complex transmission function in any given projected pixel are related by
C=ln(|TP|)ϕP=β¯δ¯.
We note that the ratio C for a given material is independent of density. For a heterogeneous object containing many complex materials, the ratio in Eq. (3) may take values within a specified range:
CminCCmax.
where Cmin and Cmax are set by the limit of the smallest and greatest ratio β¯/δ¯ based on the expected materials present.

The ratio C at each pixel can be constrained to this range at each iteration of the phase retrieval algorithm, in a way that is analogous to the application of a positivity constraint [14]. In this case, however both the phase and magnitude are constrained. Equations (3) and (4) together set upper and lower limits of the permissible range of the values of the phase and magnitude. As with other constraints, values that fall outside of this range are set to the nearest limit of the range, defined as;

ln(|T|)Cmaxϕln(|T|)Cmin
and
Cminϕln(|T|)Cmaxϕ.
where T, ϕ and T, ϕ are the magnitude and phase of the complex transmission function before and after the application of the constraint respectively. Equations (5a) and (5b) can be used together or independently, depending on the relative strength of the phase and magnitude components of the object and X-ray interactions. The constraint can be applied to a variety of CDI methods including FCDI [18,19], ptychography [10,12], and phase-diverse FCDI [11].

Phase-diverse FCDI [11] data at X-ray energies of 520 eV and 2535 eV was used to illustrate the effectiveness of the method, with the experimental and data processing details explained in [20] (520 eV) and [7] (2535 eV). In each case, the sample was a red blood cell infected with the malaria parasite P. falciparum. As the parasite develops, it consumes the hemoglobin contained in the host cell, reducing the hemoglobin volume by 75% compared to an uninfected cell [21]. The digested hemoglobin is stored as hemozoin crystals in the parasites digestive vacuole [22], and membrane protein structures such as Maurer’s clefts are observed in the host cell [23,24]. The samples were fixed and dried prior to imaging, with full details described elsewhere [7,20]. Consequently, the samples contain very little water and the level of undigested hemoglobin would likely have been small due to perforations in the host cell membrane caused during fixing and drying. As a result, we expect the samples to be heterogeneous and largely made up of proteins in cell membranes, exhibiting similar properties to many biological samples. In this case, hemozoin is likely to concentrated in moderated abundance in the digestive vacuole of the parasite, and absent from other areas. Values for the ratio C (Eq. (3)) for materials likely to be present in the sample are listed in Table 1. As any single pixel contains information from a selection of materials at relative abundances listed in Table 1, the complex constraint, Eq. (4), was chosen to limit ln(|T|)/ϕ to within a range of 0.25C0.19 for the data at 520 eV, and 0.033C0.025 for the data at 2.5keV. In addition to the complex constraint, constraints limiting the phase and magnitude π<ϕ<0 and 0<T<1respectively were applied. We also applied the modulus and support constraints, which replace the diffraction intensity at each iteration with the measured diffraction intensity, and limit the extent of the object in the sample respectively [15,18].

Tables Icon

Table 1. Relative abundance of materials likely to be found in the samples and approximate values for C (Eq. (3)) obtained from [26].

3. Results

Figure 1 shows the results of applying the complex constraint as described in Section 2 to a red blood cell infected with the malaria parasite P. falciparum at an X-ray energy of 520 eV. Figures 1(a) and 1(c) shows the effect of applying a constraint whereby the phase [Fig. 1(a)] and magnitude [Fig. 1(c)] are limited to π<ϕ<0 and 0<T<1respectively, while Figs. 1(b) and 1(d) show the effect of applying, in addition, the complex constraint, 0.25C0.19. It is evident in Figs. 1(a) and 1(c) that inconsistencies between the magnitude and the phase in the recovered complex transmission function (Fig. 1(a), arrow), are avoided in Figs. 1(b) and 1(d).

 figure: Fig. 1

Fig. 1 The phase (top) and magnitude (bottom) of the complex transmission function of a red blood cell infected with the malaria parasite P. falciparum at the trophozoite life-cycle stage. Results in A and C were obtained by independently constraining the phase and magnitude within π<ϕ<0 and 0<T<1, respectively. In B and D, the ratio, C, is applied as discussed in the text.

Download Full Size | PDF

A similar approach was applied using a higher X-ray energy of 2.5 keV applied to a red blood cell infected with the malaria parasite P. falciparum at the sexual gametocyte life-cycle stage, with results shown in Fig. 2. At this energy the ratio C is dominated by the δ component of the complex refractive index, yielding a magnitude image with little information and artifacts seen in the magnitude image that do not relate to features in the corresponding phase image (Fig. 2(c), arrow). Accordingly, in this case, we apply the complex constraint to update the magnitude information based on the phase using Eq. (5b) only, rather than updating both the phase and magnitude as in the previous case. Figures 2(a) and 2(c) used the same constraint as in Figs. 1(a) and 1(c), limiting the phase and magnitude to within π<ϕ<0 and 0<T<1respectively, while Figs. 2(b) and 2(d) were obtained by also applying the complex constraint, C as outlined in Section 2. This application resulted in the elimination of the previously mentioned artifacts from the magnitude image, and results in a magnitude image that is, as expected, more consistent with the phase image.

 figure: Fig. 2

Fig. 2 The phase (top) and magnitude (bottom) of the complex transmission function of a red blood cell infected with the malaria parasite P. falciparum at the sexual gametocyte life-cycle stage. Results in A and C were obtained by independently constraining the phase and magnitude to within π<ϕ<0 and 0<T<1 respectively. In B and D the complex constraint, C, was applied as discussed in the text.

Download Full Size | PDF

We found that in these samples the χ2 error between the measured intensity and the intensity calculated from the reconstructed transmission function was not useful discriminator between the different reconstructions, consistent with earlier findings [25]. However, as the reconstructed complex transmission function provides quantitative results of the phase and magnitude, measurements of the thickness and composition of the sample can be made using Eqs. (2a) and (2b). The relative sample thickness obtained from the phase and magnitude provides an easily interpretable comparison to judge consistency, and was calculated for a line profile (dashed line indicated in Fig. 1(c) and 2(c)). Figure 3 shows relative thickness for the images in Figs. 1(a)1(d) [Fig. 3(a)] and Figs. 2(a)2(d) [Fig. 3(b)], for the central value of C (lines) and allowing for the range in C described in Section 2 (corresponding shaded area). It is clear that the application of the complex constraint (black and grey) provides a much more self-consistent result, with the thickness determined from the phase (dashed lines) and magnitude (solid lines) in much better agreement. It is also evident from this analysis that without the application of the complex constraint (red) at lower energies, neither the phase nor magnitude independently give a reliable reconstruction. However, at higher energies, the phase reconstruction is similar with or without the application of the constraint.

 figure: Fig. 3

Fig. 3 Line profiles through the reconstructions of the two samples at the positions indicated in Fig. 1(c) [Fig. 3(a)] and Fig. 2(c) [Fig. 3(b)], plotted as the thickness using the central value of the ratio C (solid and dashed lines) and the range of C (corresponding shaded area) for the phase (dashed lines) and magnitude (solid lines). Red and grey refers to the case when the complex constraint was not applied, and the case where the complex constraint was applied respectively. The label for each line refers to the corresponding panel, (A)-(D), in Figs. 1 and 2.

Download Full Size | PDF

4. Discussion and conclusion

We have shown that although a complex constraint based on a single material cannot be applied to CDI for the biological sample used here, a complex constraint based on the expected average composition can greatly improve the self-consistency of the reconstructed complex transmission function. The measured thickness derived from measurements of the real and imaginary components of the refractive index show a much higher agreement with each other compared to the case when only the real and imaginary parts of the transmission function are independently constrained. This demonstrates an enhancement to the quantitative nature of CDI. By adjusting the range (Eq. (4)) for specific samples using limited a priori knowledge, the method can readily be applied to a variety of biological and other heterogeneous specimens. The method described here can readily be applied other two and three-dimensional CDI techniques.

Acknowledgments

The authors acknowledge support from the Australian Research Council Centre of Excellence for Coherent X-ray Science. We acknowledge travel funding provided by the International Synchrotron Access Program (ISAP) managed by the Australian Synchrotron and funded by the Australian Government.

References and links

1. A. P. Mancuso, O. M. Yefanov, and I. A. Vartanyants, “Coherent diffractive imaging of biological samples at synchrotron and free electron laser facilities,” J. Biotechnol. 149(4), 229–237 (2010). [CrossRef]   [PubMed]  

2. H. Jiang, C. Song, C.-C. Chen, R. Xu, K. S. Raines, B. P. Fahimian, C.-H. Lu, T.-K. Lee, A. Nakashima, J. Urano, T. Ishikawa, F. Tamanoi, and J. Miao, “Quantitative 3D imaging of whole, unstained cells by using X-ray diffraction microscopy,” Proc. Natl. Acad. Sci. U.S.A. 107(25), 11234–11239 (2010). [CrossRef]   [PubMed]  

3. G. J. Williams, E. Hanssen, A. G. Peele, M. A. Pfeifer, J. N. Clark, B. Abbey, G. Cadenazzi, M. D. de Jonge, S. Vogt, L. Tilley, and K. A. Nugent, “High-resolution X-ray imaging of Plasmodium falciparum-infected red blood cells,” Cytometry A 73(10), 949–957 (2008). [CrossRef]   [PubMed]  

4. D. Shapiro, P. Thibault, T. Beetz, V. Elser, M. R. Howells, C. Jacobsen, J. Kirz, E. Lima, H. Miao, A. M. Neiman, and D. Sayre, “Biological imaging by soft x-ray diffraction microscopy,” Proc. Natl. Acad. Sci. U.S.A. 102(43), 15343–15346 (2005). [CrossRef]   [PubMed]  

5. A. P. Mancuso, M. R. Groves, O. E. Polozhentsev, G. J. Williams, I. McNulty, C. Antony, R. Santarella-Mellwig, A. V. Soldatov, V. Lamzin, A. G. Peele, K. A. Nugent, and I. A. Vartanyants, “Internal structure of an intact Convallaria majalis pollen grain observed with X-ray Fresnel coherent diffractive imaging,” Opt. Express 20(24), 26778–26785 (2012). [CrossRef]   [PubMed]  

6. M. Beckers, T. Senkbeil, T. Gorniak, M. Reese, K. Giewekemeyer, S. C. Gleber, T. Salditt, and A. Rosenhahn, “Chemical contrast in soft x-ray ptychography,” Phys. Rev. Lett. 107(20), 208101 (2011). [CrossRef]   [PubMed]  

7. M. W. M. Jones, M. K. Dearnley, G. A. van Riessen, B. Abbey, C. T. Putkunz, M. D. Junker, D. J. Vine, I. McNulty, K. A. Nugent, A. G. Peele, and L. Tilley, “Rapid, low dose X-ray diffractive imaging of the malaria parasite Plasmodium falciparum,” Ultramicroscopy (2013), doi:. [CrossRef]  

8. M. W. M. Jones, G. A. van Riessen, B. Abbey, C. T. Putkunz, M. D. Junker, E. Balaur, D. J. Vine, I. McNulty, B. Chen, B. D. Arhatari, S. Frankland, K. A. Nugent, L. Tilley, and A. G. Peele, “Whole-cell phase contrast imaging at the nanoscale using Fresnel Coherent Diffractive Imaging Tomography,” Sci. Rep. 3, 2288 (2013). [CrossRef]   [PubMed]  

9. K. A. Nugent, “Coherent methods in the X-ray sciences,” Adv. Phys. 59, 1–99 (2010).

10. J. M. Rodenburg, A. C. Hurst, A. G. Cullis, B. R. Dobson, F. Pfeiffer, O. Bunk, C. David, K. Jefimovs, and I. Johnson, “Hard-x-ray lensless imaging of extended objects,” Phys. Rev. Lett. 98(3), 034801 (2007). [CrossRef]   [PubMed]  

11. C. T. Putkunz, J. N. Clark, D. J. Vine, G. J. Williams, M. A. Pfeifer, E. Balaur, I. McNulty, K. A. Nugent, and A. G. Peele, “Phase-Diverse coherent diffractive imaging: high sensitivity with low dose,” Phys. Rev. Lett. 106(1), 013903 (2011). [CrossRef]   [PubMed]  

12. D. J. Vine, G. J. Williams, B. Abbey, M. A. Pfeifer, J. N. Clark, M. D. de Jonge, I. McNulty, A. G. Peele, and K. A. Nugent, “Ptychographic Fresnel coherent diffractive imaging,” Phys. Rev. A 80(6), 063823 (2009). [CrossRef]  

13. C. T. Putkunz, J. N. Clark, D. J. Vine, G. J. Williams, E. Balaur, G. A. Cadenazzi, E. K. Curwood, C. A. Henderson, R. E. Scholten, R. J. Stewart, I. McNulty, K. A. Nugent, and A. G. Peele, “Mapping granular structure in the biological adhesive of Phragmatopoma californica using phase diverse coherent diffractive imaging,” Ultramicroscopy 111(8), 1184–1188 (2011). [CrossRef]   [PubMed]  

14. J. Miao, D. Sayre, and H. N. Chapman, “Phase retrieval from the magnitude of the Fouries transforms of nonperiodic objects,” J. Opt. Soc. Am. A 15(6), 1662–1669 (1998). [CrossRef]  

15. G. J. Williams, H. M. Quiney, A. G. Peele, and K. A. Nugent, “Fresnel coherent diffractive imaging: treatment and analysis of data,” New J. Phys. 12(3), 035020 (2010). [CrossRef]  

16. K. Giewekemeyer, M. Beckers, T. Gorniak, M. Grunze, T. Salditt, and A. Rosenhahn, “Ptychographic coherent x-ray diffractive imaging in the water window,” Opt. Express 19(2), 1037–1050 (2011). [CrossRef]   [PubMed]  

17. J. Nelson, X. Huang, J. Steinbrener, D. Shapiro, J. Kirz, S. Marchesini, A. M. Neiman, J. J. Turner, and C. Jacobsen, “High-resolution x-ray diffraction microscopy of specifically labeled yeast cells,” Proc. Natl. Acad. Sci. U.S.A. 107(16), 7235–7239 (2010). [CrossRef]   [PubMed]  

18. J. N. Clark, C. T. Putkunz, M. A. Pfeifer, A. G. Peele, G. J. Williams, B. Chen, K. A. Nugent, C. Hall, W. Fullagar, S. Kim, and I. McNulty, “Use of a complex constraint in coherent diffractive imaging,” Opt. Express 18(3), 1981–1993 (2010). [CrossRef]   [PubMed]  

19. G. J. Williams, H. M. Quiney, B. B. Dhal, C. Q. Tran, K. A. Nugent, A. G. Peele, D. Paterson, and M. D. de Jonge, “Fresnel Coherent Diffractive Imaging,” Phys. Rev. Lett. 97(2), 025506 (2006). [CrossRef]   [PubMed]  

20. M. W. M. Jones, B. Abbey, A. Gianoncelli, M. B. Luu, H. D. Couglin, E. Balaur, C. Millet, A. G. Peele, L. Tilley, and G. A. van Riessen, “Phase-diverse Fresel coherent diffractive imaging of malaria infected red blood cells in the water window,” Opt. Express (to be published).

21. E. Hanssen, C. Knoechel, M. Dearnley, M. W. A. Dixon, M. A. Le Gros, C. Larabell, and L. Tilley, “Soft X-ray microscopy analysis of cell volume and hemoglobin content in erythrocytes infected with asexual and sexual stages of Plasmodium falciparum,” J. Struct. Biol. 177(2), 224–232 (2012). [CrossRef]   [PubMed]  

22. S. Pagola, P. W. Stephens, D. S. Bohle, A. D. Kosar, and S. K. Madsen, “The structure of malaria pigment β-haematin,” Nature 404(6775), 307–310 (2000). [CrossRef]   [PubMed]  

23. E. Hanssen, C. Knoechel, N. Klonis, N. Abu-Bakar, S. Deed, M. LeGros, C. Larabell, and L. Tilley, “Cryo transmission X-ray imaging of the malaria parasite, P. falciparum,” J. Struct. Biol. 173(1), 161–168 (2011). [CrossRef]   [PubMed]  

24. C. Spycher, M. Rug, E. Pachlatko, E. Hanssen, D. Ferguson, A. F. Cowman, L. Tilley, and H.-P. Beck, “The Maurer’s cleft protein MAHRP1 is essential for trafficking of PfEMP1 to the surface of Plasmodium falciparum-infected erythrocytes,” Mol. Microbiol. 68(5), 1300–1314 (2008). [CrossRef]   [PubMed]  

25. G. Williams, M. Pfeifer, I. Vartanyants, and I. Robinson, “Effectiveness of iterative algorithms in recovering phase in the presence of noise,” Acta Crystallogr. A 63(1), 36–42 (2007). [CrossRef]   [PubMed]  

26. B. L. Henke, E. M. Gullikson, and J. C. Davis, “X-ray interactions: photoabsorption, scattering, transmission, and reflection at E=50-30000 eV, Z=1-92,” At. Data Nucl. Data Tables 54(2), 181–342 (1993). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (3)

Fig. 1
Fig. 1 The phase (top) and magnitude (bottom) of the complex transmission function of a red blood cell infected with the malaria parasite P. falciparum at the trophozoite life-cycle stage. Results in A and C were obtained by independently constraining the phase and magnitude within π < ϕ < 0 and 0 < T < 1 , respectively. In B and D, the ratio, C, is applied as discussed in the text.
Fig. 2
Fig. 2 The phase (top) and magnitude (bottom) of the complex transmission function of a red blood cell infected with the malaria parasite P. falciparum at the sexual gametocyte life-cycle stage. Results in A and C were obtained by independently constraining the phase and magnitude to within π < ϕ < 0 and 0 < T < 1 respectively. In B and D the complex constraint, C, was applied as discussed in the text.
Fig. 3
Fig. 3 Line profiles through the reconstructions of the two samples at the positions indicated in Fig. 1(c) [Fig. 3(a)] and Fig. 2(c) [Fig. 3(b)], plotted as the thickness using the central value of the ratio C (solid and dashed lines) and the range of C (corresponding shaded area) for the phase (dashed lines) and magnitude (solid lines). Red and grey refers to the case when the complex constraint was not applied, and the case where the complex constraint was applied respectively. The label for each line refers to the corresponding panel, (A)-(D), in Figs. 1 and 2.

Tables (1)

Tables Icon

Table 1 Relative abundance of materials likely to be found in the samples and approximate values for C (Eq. (3)) obtained from [26].

Equations (8)

Equations on this page are rendered with MathJax. Learn more.

ln( | T | )=k βzdz ,
φ=k δz dz
ln( | T P | )=k β ¯ z,
ϕ P =k δ ¯ z
C= ln( | T P | ) ϕ P = β ¯ δ ¯ .
C min C C max .
ln( | T | ) C max ϕ ln( | T | ) C min
C min ϕ ln( | T | ) C max ϕ .
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.