Quantitative technique for robust and noise-tolerant speed measurements based on speckle decorrelation in optical coherence tomography

Open Access Open Access

Abstract

Intensity-based techniques in optical coherence tomography (OCT), such as those based on speckle decorrelation, have attracted great interest for biomedical and industrial applications requiring speed or flow information. In this work we present a rigorous analysis of the effects of noise on speckle decorrelation, demonstrate that these effects frustrate accurate speed quantitation, and propose new techniques that achieve quantitative and repeatable measurements. First, we derive the effect of background noise on the speckle autocorrelation function, finding two detrimental effects of noise. We propose a new autocorrelation function that is immune to the main effect of background noise and permits quantitative measurements at high and moderate signal-to-noise ratios. At the same time, this autocorrelation function is able to provide motion contrast information that accurately identifies areas with movement, similar to speckle variance techniques. In order to extend the SNR range, we quantify and model the second effect of background noise on the autocorrelation function through a calibration. By obtaining an explicit expression for the decorrelation time as a function of speed and diffusion, we show how to use our autocorrelation function and noise calibration to measure a flowing liquid. We obtain accurate results, which are validated by Doppler OCT, and demonstrate a very high dynamic range (> 600 mm/s) compared to that of Doppler OCT (±25 mm/s). We also derive the behavior for low flows, and show that there is an inherent non-linearity in speed measurements in the presence of diffusion due to statistical fluctuations of speckle. Our technique allows quantitative and robust measurements of speeds using OCT, and this work delimits precisely the conditions in which it is accurate.

© 2014 Optical Society of America

1. Introduction

Optical coherence tomography (OCT) is a technique that provides high-resolution imaging with moderate penetration depths. It is commonly used in biomedical and clinical applications due to its combination of optical sectioning, axial resolution and high-speed imaging [1, 2]. OCT techniques have also been applied to optical inspection for non-destructive industrial applications [3, 4, 5, 6]. Signal and processing methods, such as polarization-sensitive OCT [7], spectroscopic OCT [8], OCT-based elastrography [9] and vibrography [10], have additionally been developed to provide functional and molecular contrast.

A valuable piece of information that can be acquired with OCT techniques corresponds to motion or deformation of sample. In OCT there have been two approaches traditionally applied for motion sensing: Doppler-based and speckle-based. As the tomogram provided by OCT measurements contains information about the reflected complex amplitude, Doppler OCT utilizes the phase of the reflected light wave as a function of time to derive information related to motion of the sample [11]. Doppler has been the technique of choice when quantitative measurements are important [12], but the stringent requirements on phase-stable detection makes its deployment in fast frequency-domain OCT systems difficult [13]. Additionally, robust Doppler methods are sensitive only to motion along the axis of the optical beam [11]. In comparison, speckle techniques rely on the relation between the statistics of the intensity signal and the movement of scatterers within the sample. Speckle techniques have primarily been applied to qualitative measurements, for example to characterize the morphology of vascular networks in biomedical applications [14, 15, 16]. Few studies have addressed quantitation from speckle decorrelation measurements [17, 18, 19], and we are unaware of studies validating quantitative speckle analysis. A recent report acknowledged this lack of validation and discussed the work needed to bring a particular implementation of speckle decorrelation, split-spectrum angiography, into the realm of quantitative flow measurements [20]. Despite their advances, however, the authors recognized that more insight was needed in order to realize this goal. Dynamic light scattering techniques have been used in OCT to extract speed, axial velocity and diffusion information from the complex amplitude of the OCT signal [21, 22, 23, 24]. Considering the literature on ultrasound speckle decorrelation for flow speed measurements in medical applications, it has been found that velocity estimations are strongly dependent on the noise of the data [25, 26], a point that has not been addressed in speckle OCT systematically. Interestingly, we are aware of only a recent study that approaches the issue of a thorough validation of ultrasound speckle decorrelation measurements [27].

This work aims to overcome the current limitations of speckle-decorrelation speed measurements. First we analyze the effect of background noise on the intensity autocorrelation function, and define a new autocorrelation function that is immune to the primary effect of additive noise. This autocorrelation function also provides flow contrast information, which is lacking in previous speckle-decorrelation approaches. We then extend the signal-to-noise ratio (SNR) range in which speckle decorrelation measurements work by performing a calibration of the secondary effect of background noise using a solid phantom. In order to show the validity of this technique beyond solid samples, we derive the explicit relation between decorrelation time, flow speed and diffusion, and make measurements in a flow phantom setup and compare the results with Doppler OCT. In order to have Doppler OCT measurements with high accuracy to be used as references, we implemented a posteriori correction of phase [13]. Finally, we explain the effect of statistical noise on the measurement of low flow speeds, and how the non-linear behavior of speckle decorrelation is inherent to the statistical nature of the technique.

2. Materials and methods

The experimental part of this work used an OCT system with a wavelength-swept laser source having a 134 nm sweep range, centered at 1285 nm and a repetition rate of 102 kHz (one A-line every Δt = 0.0098 ms). The signal was digitized at 180 MS/s, recording 1600 samples per sweep. The system utilized an acousto-optical frequency-shifter for depth-degeneracy removal [28] and the tomograms were reconstructed using zero-padding to have a reconstruction size of 1024 pixels in the axial direction. The system has a polarization-diverse receiver (described in [29]), and the signal from each channel was added to calculate the tomogram intensity I = |Ex|2 + |Ey|2. We define the tomogram amplitude as |E|=I.

Solid samples consisted of white rubber and liquid samples of intralipid at 0.5% volume in water. The waist radius of the lateral point spread function (PSF) of the optical probe was approximately wn = 15–18 μm in water.

We use a coordinate system defined by assigning the z axis to the light propagation direction, as shown in Fig. 1. The solid sample was placed on a motorized translation stage that moved the sample at a known speed in the x direction.

 figure: Fig. 1

Fig. 1 Measurement setup. The xyz coordinate system was defined by the light propagation direction z and its associated perpendicular plane xy (y towards the reader). The solid rubber phantom traveled in the x direction on a motorized linear stage. Inset: configuration for measuring signals from flowing intralipid. The beam was reflected almost perpendicular to the fiber after internal reflection (a rotation around y). The angle between the beam and the flow was θ = 82.5°.

Download Full Size | PDF

Liquid phantom measurements were performed using the geometry shown in the inset of Fig. 1. The flow setup consisted of a peristaltic pump, a closed 3.2 mm-diameter tubing circuit and a pulse dampener in order to have steady, laminar flow in the area of the OCT probe, which was used to take measurements from within the tube. The probe emitted light at an angle of 82.5° with respect to the flow direction, which allowed for a small Doppler signal to be acquired from the liquid. The Doppler signal was analyzed using the Kasai autocorrelation [11] in the same correlation window that was used for decorrelation measurements.

To obtain a clear Doppler signal from noisy phase measurements, the OCT system was setup in order to have two calibration reflection signals to compensate for the phase-unstable acquisition, in a similar way to the technique described in [13]. The first reflection signal, close to zero-depth, was used to compensate for the phase offset introduced by the acousto-optical frequency-shifter used for depth-degeneracy removal [28]. The second was used to compensate for the phase slope introduced by the phase jitter of the data acquisition board.

3. Speckle decorrelation of the OCT intensity

3.1. Time autocorrelation of OCT signals

The theory of dynamic light scattering applied to the OCT signal has been studied thoroughly previously [21], and it is well known that the time autocorrelation of the complex OCT signal E(t), defined as [30]

g(1)(τ)=E(t)E*(t+τ)dt,
has the following form when Mm (Ms) is the fraction of moving (static) scatterers, and all moving scatterers inside the resolution volume follow the same behavior
g(1)(τ)=Ms+Mmei2nk0vzτe4n2k02Dτevz2τ2/wz2evt2τ2/wt2.
Here the first exponential is related to the Doppler signal, the second exponential is related to the decorrelation of the amplitude signal due to the diffusion of the scatterers, and the last two exponentials are related to the decorrelation of the amplitude of the signal due to movement of the scatterers in the axial z and transverse xy directions. n is the refraction index, k0 the central wave number of the spectrum, D the diffusion constant of the scatterers, wz the axial 1/e size of the PSF of the system, and wt the transversal 1/e size of the PSF of the system. Mm can be considered as a figure of motion contrast, because areas where the OCT signal has a significant value of Mm correspond to areas with motion, such as vascular networks. We do not include the term for entering/exiting scatterers into the resolution volume in [21], because the PSF of the system is Gaussian and by definition it does not have hard edges that particles can enter or exit. In Eq. (1) and in the following equations, the integral is replaced by a summation in the discrete sampling case.

The second-order correlation function deals with the intensity I(t) = |E(t)|2 and is defined as

g(2)(τ)=I(t)I(t+τ)dt.
This function is related to the first-order autocorrelation through the Siegert relationship when the probability distribution of E is Gaussian, as in the case of speckle. If we assume Mm = 1 for simplicity, Eq. (3) becomes
g(2)(τ)=1+|g(1)(τ)|2=1+e8n2ko2Dτe2vz2τ2/wz2e2vt2τ2/wt2.
It is advantageous to consider ensemble averages when estimating g(2)(τ) experimentally. The normalized second-order correlation function makes use of the concept of observation ensemble to make a more robust estimation. It is defined as [31]
g(2)(τ)=I(t)I(t+τ)dtI(t)I(t+τ)dt,
where I is the intensity, and 〈···〉 denotes ensemble average. The ensemble average allows to take into account multiple correlation windows in the calculation, reducing the statistical fluctuations on speckle size and its effect on the calculated decorrelation time.

Although definitions Eqs.(1) and (3) are commonly used in dynamic light scattering [32, 21], in the ultrasound and OCT speckle decorrelation literature the Pearson correlation function g(P) has been traditionally used instead, either between sequential A-lines [17] or at a single depth as a function of time [33, 34, 19]. In the case of the autocorrelation of the intensity, it is defined as

g(P)(τ)=[I(t)I¯][I(t+τ)I¯]σIσIdt,
where Ī denotes the average intensity and σI the standard deviation. g(P) is defined between −1 (for anticorrelated) and 1 (for correlated signals), having a value of 0 for decorrelated signals. For this reason, comparison between the Pearson autocorrelation and g(2) requires subtraction of 1 from g(2), because decorrelated signals produce g(2) = 1. The (I(t) − Ī)/σI terms are equalizing variations in the intensity, regardless of their magnitude, increasing the effect of noise. For example, static or low speed areas with high, near constant intensity will exhibit strong decorrelation because the fluctuations of the noise determine the value of σI.

In opposition to Eq. (6), the definition in Eq. (5) carries information about the speckle contrast. In particular, g(2)(0) = 2 for fully developed, unity contrast speckle. For this reason, we define a motion contrast function C as

Cg(2)(0)g(2)(τ).
C can be used to discriminate between static and dynamic areas of the tomogram, and is a good indicator of the presence of flow, similar to the conventional speckle variance technique [14]. Experimentally, static areas with strong signal and low noise will produce a small value in C.

It is possible to use the definition in Eq. (5), replacing I(t) by A(t) = |E(t)|. This modified autocorrelation (2)(τ) behaves as

g˜(2)(τ)=1+γ|g(τ)|
in our particular case, where γ takes into account the different intercept of this function.

3.2. Background noise effects in the time autocorrelation of OCT signals

Now consider the autocorrelation of a signal that is affected by additive noise. We can express the total detected signal, up to a multiplicative constant, as

I(t)=wS(t)+(1w)N(t),
where S is the original signal normalized by its standard deviation σS, and similarly for the noise N. The relative weight of each if given by 0 ≤ w ≤ 1, and serves as a measure of the signal-to-noise ratio (SNR), which in dB equals
SNR=10*log10w1w.

We define the cross-correlation of functions f and g as Rfg. The autocorrelation of I using Eq. (3) is

RII=w2RSS+(1w)2RNN+w(1w)(RSN+RNS).

In the case of OCT, we will consider N(t) as background noise. Effects of other types of noise (such as that produced by multiple scattering) will not be considered here. It is important to note that, rigorously, noise is added to the complex OCT signal

|I(t)|2=|w[SR(t)+iSI(t)]+(1w)[NR(t)+iNI(t)]|2,
however the terms that arise are essentially the same as those in Eq. (11), with duplicity to account for the real and imaginary parts of each contribution. Because this does not change the behavior that we will describe below, we will focus on Eq. (11) for simplicity.

In the present case, the noise will have no correlation with the signal, which implies RSN = RNS = 1 as mentioned above. If we additionally assume that the frequency spectrum of the noise n(f) is white, we have

RII=w2RSS+(1w)2[1+δ(τ)]+2w(1w),
where δ denotes the impulse function. For this reason, background white noise will affect the very first point of the autocorrelation τ = 0, but τ > 0 will contain w2RSS. Because w2 is a constant, it will not modify the shape of the autocorrelation, which enables, in principle, to omit the τ = 0 point in the autocorrelation function to regain RSS. This is what we consider the primary effect of background noise. If the Pearson autocorrelation is used instead, we have
RII(P)=w2RSS(P)+(1w)2δ(τ)w2+(1w2).
The denominator arises because of the change in the standard deviation of the combined signal, and the third term vanishes because RSN(P)=RNS(P)=0.

Experimentally RII is being estimated by a limited, discrete sampling in time I(t) → It with total number of samples ns. For this reason, although N(t) → Nt will be generated by white noise, its discrete autocorrelation function, being estimated from a finite number of discrete time points, will not be a perfect impulse function. We can expect that RNN will be given by

RNN={2:τ=01+fτ(τ):otherwise
with |fτ(τ)| < 1. fτ(τ) will behave as white noise itself and oscillate around zero, but the weight of its mean square value will be directly related to the number of samples ns used to estimate the autocorrelation function. Numerical simulations suggest that the variance of fτ behaves as σfτ2=1ns. When using the Pearson definition, RNN(P)=RNN1. fτ will directly affect the estimation of the autocorrelation RII with weight (1 − w)2, and will start dominating the autocorrelation function for small values of w and ns. The autocorrelation function becomes
RII={2(1w)2+w2RSS(0)+2w(1w):τ=0(1w)2[1+fτ]+w2RSS(τ)+2w(1w):otherwise,
with a similar definition when using the Pearson autocorrelation. The primary effect of background noise is seen as an additional drop Δg in RII between the first τ = 0 and second time points τ = Δt with height
Δg=[RII(0)RII(Δt)][2RSS(Δt)]=(1w)2+(w21)[2RSS(Δt)].
If the SNR of the signal is known, Δg can be used to compensate for the primary effect of noise without losing the first point of the autocorrelation function..

Due to the random nature of fτ, the autocorrelation function RII when noise dominates will consist of oscillations with near-unity magnitude in terms of the normalized autocorrelation function. When calculating the decorrelation time defined as the crossing of the autocorrelation function below a given threshold, these oscillations will make very probable a crossing in the first few time difference samples. For this reason, once noise starts to dominate the autocorrelation function, we can expect for the correlation time to drop, reaching a point where the correlation time will be very short irrespective of the dynamics of the scatterers in the sample. This is what we refer to as the secondary effect of background noise.

As mentioned before, to compensate the primary effect of background noise we can define a new discrete autocorrelation function as

gnew(2)(τ)g(2)(τΔt)median{g(2)(τ>ns/2)}Cnew,
where Δt is the time interval between A-lines, and the new motion contrast value that replaces Eq. (7) is
Cnew=g(2)(τ=Δt)median{g(2)(τ>ns/2)}.
With these definitions, the main contribution from RNN is suppressed by omitting the value at τ = 0, and the subtraction by median{g(2)(τ > ns/2)} will allow a renormalization of the autocorrelation function between 0 and 1, by taking as 0 the median of fτ when the pure signal autocorrelation RSS is negligible. This assumes that ns/2 is larger than any time scale of the dynamics of the sample. Although this enables the calculation of the decorrelation time for medium to high SNRs, there is still a significant range of low SNRs that will be affected by the secondary noise effect. The determination of the behavior of the secondary effect, and how to account for it, will be considered in Sec. 4.2.

3.3. Relation between speckle decorrelation time and speed

Assuming an isotropic voxel (wz = wt) for simplicity, and representing the total speed as v=vz2+vt2, we can find using Eq. (4) the correlation time τc when the autocorrelation function g(2)(τ) [(2)(τ)] reaches the threshold 1+gc[1+gc] for moving and diffusing scatterers as

τc1=2g^cKD+4g^c2KD2+2g^cv2/wt2,
with the diffusion constant KD=2n2ko2D and c = −1/ln(gc), which is positive definite because gc < 1. The different definition of the threshold when using (2)(τ) from Eq. (8) makes the rest of the derivation the same for both cases. For autocorrelations defined between −1 and 1, 1 is subtracted from the threshold. Now, if the diffusion term is small compared to the speed term so that v2/wt2KD, we arrive at the familiar result
τc1=2g^cvwtv,
which is also in agreement with full-field techniques [35, 36]. Note that for most common thresholds 1 ≲ c ≲ 3.

In a practical implementation of speckle decorrelation measurements in OCT, it is important to know that OCT exhibits two speckle sizes that come from different origins. The speckle size in the axial direction is related to the axial resolution of the system. When A-lines are acquired as a function of time, the intensity at each depth will evolve in time. We can attribute a speckle size along a second axis to this temporal evolution, which will exhibit a decorrelation time given by Eq. (20), which is linked to the dynamics of the scatterers reflecting light at that depth. Figure 2 shows a clear example of this relation, where we show an OCT tomogram of a tube filled with flowing liquid (inset of Fig. 1) at two different flow rates. The tomograms are so-called M-mode measurements, which consist of the measurement of A-lines as a function of time while maintaining the OCT probe position constant. For this reason the horizontal axis in Fig. 2 corresponds to time and the vertical axis to depth. The beam enters at the top, where the reflections from the collimating lens and other surfaces can be seen. The tube material that encloses the scattering liquid can be seen in the area of weak scattering at the bottom of the tomograms. It is easily seen that the speckle-decorrelation time (and therefore the speckle size in the horizontal direction) matches the expected relationship with scatterer speed. Note how the horizontal speckle size gets larger near the interfaces due to the no-slip condition of the flow.

 figure: Fig. 2

Fig. 2 (left) M-mode tomogram of phantom fluid at 2.5 mL/min with low flow velocities and large horizontal speckle size, (right) the same sample at 80 mL/min yielding high flow speeds and small horizontal speckle size. The strong reflections at the top correspond to the lens and other surfaces of the probe, the wall of the tube is the weak reflection at the bottom. The depth scale in mm is approximate as the index of refraction of the different materials has been considered equal to the fluid’s index 1.33.

Download Full Size | PDF

For simplicity, in the rest of this work we will consider exclusively motion in the x direction, although diffusion still exists in three dimensions. In the case of movement of a liquid, Eq. (20) can be considered as the inverse decorrelation time as a function T of all the contributions to the decorrelation of the signal

τ1=KBM+KBM2+k2vx2T(vx,KBM),
where k absorbs the parameters dependent on the beam characteristics and the chosen threshold, and might be dependent on z if the beam is not well collimated in the area of interest. KBM contains the contribution from diffusion and the chosen threshold as well. It is possible have more complex scenarios in which different processes apart from scatterer translational motion affect τ and can be included inside T, especially in the case of anisotropic scatterers. For example, it is known that red blood cells (RBC) can exhibit motions such as tumbling, tank-treading, swinging and deformation under certain circumstances [37], which would add additional terms to T, and which also potentially depend on other factors such as temperature and RBC concentration.

4. Experimental results

The setup of Fig. 1 was arranged using a solid rubber phantom placed on a motorized linear stage for precise speed control. The OCT probe was placed near the surface of the sample. Because the optics of the OCT probe were designed to be used while immersed in water, we expect to have a small defocus which modifies wt in Eq. (20), and therefore k in Eq. (22). The sample was moved in front of the beam at different speeds ranging from 40 to 600 mm/s, in steps of 40 mm/s. Because the sample was a solid we have KBM = 0.

Except otherwise noted, the correlation window used was ns = 64, and the ensemble consisted of 32 correlation windows in time (2048 A-lines total) and 8 px in depth. This means that each group of 2048 samples in time and 8 px high (roughly 32 μm) produce one point in the inverse speckle decorrelation-time profile, at a rate of approximately 50 depth decorrelation profiles per second. We found that using the OCT signal amplitude |E| as described in Eq. (8) produced decorrelation profiles that were significantly more homogeneous than using the tomogram intensity. This seems to be linked to the effect that the square has on outliers of the amplitude in the statistical fluctuations of speckle intensity. We speculate that these outliers might be produced by extraneous reflective particles in the liquid. Because of γ, this modifies the range of values for C, and all autocorrelations had a maximum C around 1.15, the reason for which the threshold value chosen was gc=0.025 for the traditional autocorrelation function, and gc=0.5 for the renormalized version. A linear interpolation was used to calculate the τc time-difference crossing of the threshold at a sub-Δt accuracy.

4.1. Primary effect of background noise

Figs. 3(a) and 3(b) show the mean intensity inside the correlation window, Figs. 3(c) and 3(d) depict the traditional inverse speckle-decorrelation time τtrad1 profile using Eq. (5) (we drop the subscript c from all τ from now on, and denote this the traditional decorrelation time), and Figs. 3(e) and 3(f) show τ−1 using Eq. (18). For Figs. 3(a), 3(c) and 3(e) the sample was moving at 120 mm/s, and for Figs. 3(b), 3(d) and 3(f) at 240 mm/s. The complete profile has 128 px in the axial direction, but in Fig. 3 we show only the top 80 px in order to simplify visualization. The OCT probe presented an artifact due to multiple reflections in its interfaces, which is clearly visible in the decorrelation time profile as two lines of high τ−1 near the top of the sample. Because these reflection artifacts varied considerably depending on the particular probe, we avoided these areas in the subsequent analysis.

 figure: Fig. 3

Fig. 3 (a–b) Mean intensity of the structural image, (c–d) inverse correlation time profiles for moving solid sample using the traditional autocorrelation function and (e–f) using the newly developed normalized autocorrelation function. (g–h) Profile comparing the inverse decorrelation time using the new and the traditional autocorrelation function at a selected column. Top row corresponds to 120 mm/s, bottom row to 240 mm/s.

Download Full Size | PDF

The sample had an irregular surface and discontinuities that appear as shallow lines with low τ−1. The deep discontinuities correspond to the end of the travel of the translation stage, a point where the sample came to a stop before gaining speed in the opposite direction. The acceleration was chosen in order to go from a complete stop to nominal speed in less than 30 ms (approximately 3 time pixels in the τ−1 profile for a complete deceleration-acceleration cycle). Ignoring these discontinuities, it is easy to see that when using Eq. (5) there is a strong gradient with depth, even though the sample was a solid, moving at a well-defined speed. This gradient is more pronounced as the threshold gc increased. Note, however, that the current threshold is too low for crossing the autocorrelation function for points near the top of the sample, where the SNR is very high: although the sample starts at a depth of 20 px, measurable decorrelation is only found deeper than 24 px. When using our new autocorrelation function Eq. (18), the profile is significantly more homogeneous, which confirms our derivation on the primary effect of noise on the autocorrelation function. Also, the area with high SNR shows a similar decorrelation time. Note that the contrast function C has been used to filter out static regions in the upper part of the profiles, which are assigned a zero velocity. Figure 3, (g) and (h), show the comparison of the two first rows in the two-dimensional profiles, where the different behavior with depth is evident.

In order to appreciate more clearly the effect of noise in τ−1, the top row in Fig. 4 shows the traditional autocorrelation function at different depths and at different sample speeds, and the bottom row shows the new autocorrelation function of Eq. (18). The depth of 20 px corresponds to a point just after light enters the sample, where there is a significantly higher SNR level than in the rest of the depths shown. The dashed line represents the threshold 1+gc that was used to calculate the correlation time profile of Fig. 3. The autocorrelations show clearly the excessive drop in correlation between τ = 0 and = Δt produced by the primary effect of background noise, described by Eq. (17). This dip increases with depth as the signal decays as expected. In contrast, the new autocorrelation functions at the bottom row exhibit higher similarity at a given speed, including the autocorrelation at pixel depth 20. The secondary effect of background noise is also visible, as an oscillation for low g(2) and large depth, as well more spread in the correlation time at low speeds. This indicates that addressing the first effect of noise enables the use of speckle decorrelation in a limited range of SNRs. In the next section we will expand the SNR range by addressing the secondary effect of noise.

 figure: Fig. 4

Fig. 4 (top) Traditional autocorrelation at selected depths for the solid sample moving at (left) 120 mm/s, (center) 200 mm/s and (right) 400 mm/s. The autocorrelation for 20 px at 400 mm/s is out of the range. (bottom) The new normalized autocorrelation for the same speeds.

Download Full Size | PDF

4.2. Secondary effect of background noise

To understand the contribution from fτ and make a calibration of the proportionality constants in Eq. (22), we first had to define a way to determine the SNR at each correlation window. We estimated the noise floor given by the presence of background noise by measuring a tomogram when the sample arm of the interferometer was blocked. This depth-dependent background signal was then used as the noise level, and the average intensity at each correlation window was used as the signal intensity to calculate the SNR. The scatter plot in Fig. 5(a) shows the valid data points from the τ−1 profiles as a function of SNR, color coded with the sample speed. To asses the performance of the Pearson autocorrelation function in the presence of noise, Fig. 5(b) shows τ−1 as calculated using g(P) in Eq. (6), and modified following Eq. (18). The same correlation window was used, and ensemble averaging was performed by simple averaging among the g(P) calculated for each member of the ensemble. As described previously, the amplification of the noise fluctuations is evident at low speeds, which makes speeds indistinguishable in the dashed box area. Paradoxically, τ−1 estimations using g(P) have smaller dispersion than when using g(2) for a given speed and sufficient SNR, as seen by the more diffuse nature of points in Fig. 5(a) versus Fig. 5(b) at a given speed.

 figure: Fig. 5

Fig. 5 (a) Scatter plot of τ−1 and SNR for different sample speeds as calculated using Eq. (18). (b) Scatter plot of the same data using the Pearson autocorrelation function. (c) Parametrization F of the effective inverse correlation time as a function of inverse correlation time and SNR, Eq. (23). (d) τeff1 as a function of τ−1 in the parametrization for several representative SNRs. Note the undefined value for low SNRs and low τ−1, as well as a unity slope line 1 : 1 as a visual guide.

Download Full Size | PDF

Figure 5(a) shows that there is a small increase in τ−1 as the SNR decreases from 40 dB to 16 dB. This was mostly uniform for all sample speeds. In this region the relation between τ−1 and v is linear, with a non-linearity at higher speeds. This is expected, as speeds that decorrelate faster than the time resolution of the system should all give the same τ−1. Below 16 dB SNR the behavior changes depending on the sample speed. High speeds preserve the slow increase in τ−1, but the dispersion in τ−1 values increases. Low speeds have a sharp increase in τ−1 until the SNR reaches approximately 10 dB. At this point all the speeds are overlapped, and it is not possible to uniquely identify the speed from the inverse correlation time and the SNR. This happens because at this point fτ completely dominates the autocorrelation function. The limit on the maximum measurable τ−1 is given by the time interval between A-lines Δt and the elimination of the zero delay autocorrelation value, and is given by τmax11/(2Δt)~50ms1. This can be increased if interpolation is used.

The data in Fig. 5(a) was used to make a parametrization F of the behavior of τ−1 as a function of SNR and calibration speed vx,c for the areas in which F(τ−1, SNR) is single valued.

The function F is defined as

vx,c=kc1F(τ1,SNR)=kc1[Fτ(τ1)+FSNR(SNR)],
where we consider the effect of fτ in Eq. (16) as an additive term FSNR(SNR) when calculating the autocorrelation time. The SNR response should be independent of the PSF size, and for this reason we introduce the calibration k constant from Eq. (22) as kc=2g^c/wc, where wc is the lateral PSF during calibration. This factor implies that F(τ−1, SNR) is in inverse time units, an can be considered a noise-corrected inverse correlation time τcorr1. In our case we did not know the exact value of wc during calibration, so we used the value for the nominal lateral PSF kn=2g^c/wn
vx,c=kn1F(τ1,SNR)kn1τeff1
and replaced the corrected τ−1 by an effective inverse correlation time τeff1. The relation between the noise-corrected correlation time and the effective correlation time is
vx,c=τeff1kn1=τcorr1kc1

In the case of Fτ(τ−1), it is clear that the relation between v and τ−1 for a solid should be linear according to Eq. (21), except at low speeds where the correlation window is too small (speeds with correlation times τ > nsΔt), and at very high speeds (with respect to Δt, with correlation times ττmax). All sample speeds in Fig. 5(a) fulfilled τ < nsΔt as we are not interested in very low speeds, so the non-linearity at low speeds will not be considered here. For this reason, we proposed Fτ(τ−1) as given by polynomials of different orders. We found that the simplest polynomial that described well the experimental was given by

Fτ(τ1)=k0+k1τ1+k5τ5,
where the linear term dominates, except for correlation times near the limit of the system τmax1 where the 5-th order term starts to make significant contribution.

In the case of FSNR(SNR), we assumed there was a contribution to τ−1 from noise that increased asymptotically at low SNRs according to

FSNR(SNR)=α1(SNRα2α3τ1)α4.
The position of the asymptote had a small dependence on τ−1, as seen in Fig. 5(a).

We performed a fit of the ki and αj (i = 0...3, j = 1, 3, 4, α2 = 9 dB was fixed) parameters, minimizing the mean absolute value error. The values of the free parameters, with τ−1 given in ms−1, were

k0=24.7ms1k1=1.09k5=1.95108ms4α1=34.5ms1α2=9dBα3=1.46102dBmsα4=9.12102,
The parametrization is shown in Fig. 5(c), as well as τeff1 versus τ−1 curves for some representative SNRs in Fig. 5(d). It is clear that at τ−1 = 50 ms−1 we are approaching the resolution of the system, and the relation becomes nonlinear.

The rubber material used for this calibration presented significant multiple scattering, which is a source of noise in addition to background noise. It is feasible to expect a change in behavior at low SNRs for materials in which single scattering is predominant. For this reason, similar measurements were carried out using an intralipid solution in water concentrations ranging from 0.5% to 2.0% in volume. In this concentration range, the samples exhibited highly isotropic, single scattering. The solution was placed in a fully filled cylindrical container, and placed on the translation stage. This configuration avoided convection effects due to the acceleration. Very similar results where obtained for all concentrations, and the mean of all parameters were

k0=2.4ms1k1=1.28k5=3.6108ms4α1=5.56ms1α2=15.5dBα3=3.51102dBmsα4=5.55101,
It is clear that behavior at low SNRs changes, but the general behavior given by k1 is maintained. We attribute the small difference in k1 to a difference in the PSF due to geometrical differences: in the solid phantom, the light entered a flat surface which introduces mainly spherical aberration [38], while in the liquid phantom, light entered a curved surface which acts mainly as a lens.

Figure 6 shows the solid sample speed as determined by using the model. Figure 6 contains the data for a different set of measurements from those used in the calibration. It is clear that the determined speed is highly homogeneous, and that it is difficult to reliably determine the speed at large depths due to the decreased SNR.

 figure: Fig. 6

Fig. 6 Speed profiles for moving solid sample using τeff1. (a) 40 mm/s, (b) 100 mm/s, (c) 200 mm/s and (d) 300 mm/s. The areas in dark gray that denote where it is not possible to uniquely determine the speed due to the low SNR.

Download Full Size | PDF

4.3. Measurements in presence of diffusion

In order to test the technique in a sample with diffusion, we performed measurements using an intralipid solution in water. To account for the different divergence of the beam in this case, we will take into account the effect of the small correction factor introduced by Eq. (24) in Eq. (22). Considering that using the nominal PSF now we have

τcorr1=KBM+KBM2+kn2vx2,
using Eq. (25) to use our effective correlation time parametrized in the calibration we obtain
τeff1=F(τ1,SNR)=K˜BM+K˜BM2+k˜2vx2,
where for simplicity we defined BM = kn/kcKBM and = kn/kckn, noting that kn/kc = wc/wn. The liquid measurements used an ensemble size in the axial direction of 4 px. The measurements were carried out as shown in the inset of Fig. 1, with an angle between the optical beam and the flow direction. This angle is close enough to the perpendicular in order to expect the same behavior of speckle decorrelation time as a function of scatterer speed as in the SNR calibration, but at the same time allowing for a small Doppler signal to be acquired from the liquid.

The power of using F(τ−1, SNR) in the equations is evident in the case of measuring a liquid: all the complexity regarding the time-resolution of the system, the autocorrelation function used and the effect of noise is encapsulated in it. More complex systems can incorporate additional effects, such as those mentioned before in the case of red blood cells. The only calibration necessary at this step is to find KBM (and in our particular case, because we do not know the value of kc). Afterwards, the flow speed can be found easily by inversion of Eq. (31)

vx=1k˜(τeff1K˜BM)2K˜BM2.

In order to find the constants of Eq. (32), we performed measurements at different flow rates using the setup shown in the inset of Fig. 1. Each dataset was analyzed using both Doppler and speckle decorrelation. At this step, we included in the analysis the depths that were corrupted by the secondary reflection of the probe, manifested near pixel 30 in Fig. 3. Because this reflection artifact does not change with time, it is possible to account for it by assuming a different value of in this area. The net effect is an increase of the value of , which makes this area more sensitive to motion. The calibration process to find the constants of Eq. (32) consisted of using the Doppler signal as the absolute velocity reference in a point-per-point fashion, for different flow rates. The Doppler velocity was corrected for the line-of-sight (LOS) angle, θ, as v = vDoppler/cosθ, and phase unwrapping was required at high flow rates due to the reduced dynamic range of Doppler. In order to minimize the statistical variation of the speckle value during the calibration, each data set was time-averaged before plotting the relationship between τeff1 and v. We determined the parameter values by performing a least-squares fit to Eq. (32) as a function of depth to allow for changes in the PSF with depth as well as to account for the effect of the reflection artifact. The data and the model fit are shown in Figs. 7(a)–7(b) with the following results (averaged in depth in the area outside the artifact)

k˜=0.074μmK˜BM=1.45ms1,
which corresponds to, using the definition for kc in Eq. (23), wc = 16 μm. This agrees well with the system nominal waist radius range of 15–18 μm. Figures 7(c)–7(d) show the results of the calibration dataset as a function of time, for all flow rates. It is clear that there is an excellent agreement between the speckle decorrelation and the Doppler speed measurements, and the speckle measurements are consistent as a function of time and for different flow rates. Even the area with the reflection artifact shows good agreement after calibration, with only a slight trace of the artifact. It is important to point out that the Doppler signal had to be unwrapped, because of its reduced dynamic range. This is possible in a smooth flow profile such as the one shown here, but unwrapping is not always possible depending on the characteristics of the profile being measured.

 figure: Fig. 7

Fig. 7 (a) Scatter plot of the time-averaged effective inverse speckle-decorrelation time τeff and the measured Doppler velocity vx for different data sets in the liquid setup. The colors denote the depth, from black to green (yellow) for the experimental data (model fit). (b) Fitting parameters as a function of depth. (c) Flow speed as determined by speckle correlation of the calibration dataset. Each flow rate has 50 flow profiles in time, and all flow rates are shown in sequence. (d) Reference Doppler velocity of the calibration dataset.

Download Full Size | PDF

Figure 8 presents the results when the calibration obtained above was applied to measurements obtained with 20, 40, and 80 mL/min flow rates. The LOS-corrected speed profiles as determined by Doppler processing and after unwrapping are also presented for reference. Note that the scales for the speckle decorrelation and Doppler flow profiles for each flow rate are the same. It is clear that at these flow rates the Doppler signal needs to be unwrapped as there are speeds well beyond the Doppler resolution limit of the system (∼ ±25 mm/s, ∼ ±220 mm/s when corrected for our particular LOS). The results in Fig. 8 demonstrate that the theoretical description in this work describes very precisely the behavior of speckle-decorrelation time as a function of scatterer speed, and confirms our approach for the calibration of the secondary effect of noise and of the PSF as a function of depth in water.

 figure: Fig. 8

Fig. 8 Structural image of the first 2048 A-lines (first row), speckle decorrelation flow speed (second row), line-of-sight-corrected Doppler flow speed (third row) and time-average flow speed profiles (fourth row) for intralipid solution at 20 mL/min, 40 mL/min, and 80 mL/min flow rates in 3.2 mm-diameter tubing. The depth is measured from the surface of the sheath, which has a 0.4 mm radius. Doppler* indicates the signal after unwrapping.

Download Full Size | PDF

Figure 8 illuminates the advantages and disadvantages of speckle decorrelation motion measurements. It clear that speckle measurements are inherently more affected by noise than Doppler measurements due to the statistical nature of the measurement. More importantly, the dynamic range of speckle decorrelation is considerably larger than for Doppler: the maximum measurable speed is > 600 mm/s, or more than 24-fold greater than possible with LOS Doppler. On the other hand, measuring low speeds is challenging: the slower parts of the flow in Fig. 8 show a deviation between speckle and Doppler, either in the form of oscillations near the OCT probe, or as an increased speed near the opposite wall.

The difficulty in measuring low speeds in speckle decorrelation is an intrinsic property of the technique in the presence of Brownian motion. Consider a measurement of a liquid with no flow, where ideally τeff1=2K˜BM. There will be fluctuations in τeff1 that come from the estimation of the autocorrelation function from a finite measurement. These fluctuations can be seen as an additional term ξ(j) in the j-th correlation window, so that

τeff1(j)=2K˜BM+ξ(j).
The particular distribution of ξ is difficult to determine, because it has contributions from different sources with different statistics. The discrete nature of the autocorrelation function indicates that a Poisson distribution best describes the determination of the correlation time. We will consider, for simplicity, that we can approximate the Poisson distribution with a Gaussian distribution, and that |ξ(j)| < 2BM. Also note that the average over j, 〈ξ(j)〉j, is zero due to the absence of flow. Using Eq. (32), we obtain a distribution of velocities
vx(j)=1k˜(K˜BM+ξ(j))2K˜BM2=1k˜{ξ(j)2+2K˜BMξ(j)}=1k˜|ξ(j)|{1+2K˜BMξ(j)},
where we indicate that the real part of the square should be taken, as vx(j) can only be real. We can see that 〈vxj ∼ 〈|ξ|〉j/k̃ > 0 (except for ξ(j) = 0, ∀j), which explains why even at zero speed, on average speckle decorrelation will always measure some finite speed in a liquid. This is not necessarily the case for dynamic light scattering, because the curve fitting done in this case is able to distinguish, to some degree, between the exponential decay of diffusion and the squared exponential decay of speed. However, even in the case of phase-stable acquisition, this kind of non-linear fitting is very complex and requires ad-hoc fitting algorithms [21] for g(1). In an intensity-based dynamic light scattering approach, the fitting must be done for g(2), which is even more complex due to the crossed terms. This makes dynamic light scattering difficult to adapt for real-time measurements.

5. Conclusions

In spite of the great interest in reliable speed measurements based on decorrelation analysis of the intensity information of OCT data, previous studies have not demonstrated calibration and validation. In this work, we have shown the limitations of using a direct implementation of the traditional autocorrelation function for speed measurements in the presence of noise, and have presented an alternative autocorrelation function that is considerably more robust. Although we focused on speckle decorrelation, the same approach can be used in dynamic light scattering OCT to account for background noise contributions to decorrelation.

We derived a direct expression for the relation between speckle decorrelation time and speed and diffusion and demonstrated a calibration procedure that allowed us to perform quantitative measurements of speed using speckle intensity statistics in a wide range of SNRs. We analyzed the effects of Brownian motion when measuring speeds in liquids, and found an unavoidable non-linearity at very low flow speeds. Our technique demonstrated a high dynamic range that easily surpasses that of Doppler techniques. We validated our results in two different scenarios, using a calibrated linear stage and corrected Doppler measurements. Our experiments delimit more clearly the range of validity of the speckle decorrelation technique. Moreover, our results demonstrate that phase-unstable OCT systems, which are far easier, faster and cheaper to build and calibrate, can now be used to measure speeds quantitatively, repetitively and with a high dynamic range, thereby enabling a host of biomedical applications.

Acknowledgments

Research in this publication was supported in part by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health, award P41 EB015903, and by Terumo Corporation. MV was supported by a fellowship from the Swiss National Science Foundation.

References and links

1. B. E. Bouma and G. J. Tearney, eds., Handbook of optical coherence tomography (Marcel Dekker Inc., 2002).

2. T. Yonetsu, B. E. Bouma, K. Kato, J. G. Fujimoto, and I.-K. Jang, “Optical coherence tomography,” Circulation J. 77, 1933–1940 (2013). [CrossRef]  

3. K. Wiesauer, M. Pircher, E. Götzinger, S. Bauer, R. Engelke, G. Ahrens, G. Grützner, C. Hitzenberger, and D. Stifter, “En-face scanning optical coherence tomography with ultra-high resolution for material investigation,” Opt. Express 13, 1015–1024 (2005). [CrossRef]   [PubMed]  

4. K. Wiesauer, M. Pircher, E. Goetzinger, C. K. Hitzenberger, R. Engelke, G. Ahrens, G. Gruetzner, and D. Stifter, “Transversal ultrahigh-resolution polarizationsensitive optical coherence tomography for strain mapping in materials,” Opt. Express 14, 5945–5953 (2006). [CrossRef]   [PubMed]  

5. D. Stifter, E. Leiss-Holzinger, Z. Major, B. Baumann, M. Pircher, E. Götzinger, C. K. Hitzenberger, and B. Heise, “Dynamic optical studies in materials testing with spectral-domain polarization-sensitive optical coherence tomography,” Opt. Express 18, 25712–25725 (2010). [CrossRef]   [PubMed]  

6. G. Song and K. Harding, “OCT for industrial applications,” in Optical Metrology and Inspection for Industrial Applications II, Proc. SPIE8563, 85630N (2012). [CrossRef]  

7. J. F. de Boer, T. E. Milner, M. J. C. van Gemert, and J. S. Nelson, “Two-dimensional birefringence imaging in biological tissue by polarization-sensitive optical coherence tomography,” Opt. Lett. 22, 934–936 (1997). [CrossRef]   [PubMed]  

8. U. Morgner, W. Drexler, F. X. Kärtner, X. D. Li, C. Pitris, E. P. Ippen, and J. G. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett. 25, 111–113 (2000). [CrossRef]  

9. B. Kennedy, K. Kennedy, and D. Sampson, “A review of optical coherence elastography: Fundamentals, techniques and prospects,” IEEE J. Sel. Top. Quantum Electron. 20, 272–288 (2014). [CrossRef]  

10. E. W. Chang, J. B. Kobler, and S. H. Yun, “Subnanometer optical coherence tomographic vibrography,” Opt. Lett. 37, 3678–3680 (2012). [CrossRef]   [PubMed]  

11. E. Regar, P. W. Serruys, and T. G. V. Leeuwen, Optical Coherence Tomography in Cardiovascular Research (CRC Press, 2007). [CrossRef]  

12. X. Li, T. H. Ko, and J. G. Fujimoto, “Intraluminal fiber-optic doppler imaging catheter for structural and functional optical coherence tomography,” Opt. Lett. 26, 1906–1908 (2001). [CrossRef]  

13. B. J. Vakoc, S. H. Yun, J. F. de Boer, G. J. Tearney, and B. E. Bouma, “Phase-resolved optical frequency domain imaging,” Opt. Express 13, 5483 (2005). [CrossRef]   [PubMed]  

14. A. Mariampillai, B. A. Standish, E. H. Moriyama, M. Khurana, N. R. Munce, M. K. K. Leung, J. Jiang, A. Cable, B. C. Wilson, I. A. Vitkin, and V. X. D. Yang, “Speckle variance detection of microvasculature using swept-source optical coherence tomography,” Opt. Lett. 33, 1530–1532 (2008). [CrossRef]   [PubMed]  

15. J. Barton and S. Stromski, “Flow measurement without phase information in optical coherence tomography images,” Opt. Express 13, 5234–5239 (2005). [CrossRef]   [PubMed]  

16. Y. Jia, O. Tan, J. Tokayer, B. Potsaid, Y. Wang, J. J. Liu, M. F. Kraus, H. Subhash, J. G. Fujimoto, J. Hornegger, and D. Huang, “Split-spectrum amplitude-decorrelation angiography with optical coherence tomography,” Opt. Express 20, 4710–4725 (2012). [CrossRef]   [PubMed]  

17. X. Liu, Y. Huang, J. C. Ramella-Roman, S. A. Mathews, and J. U. Kang, “Quantitative transverse flow measurement using optical coherence tomography speckle decorrelation analysis,” Opt. Lett. 38, 805–807 (2013). [CrossRef]   [PubMed]  

18. N. Mohan and B. Vakoc, “Principal-component-analysis-based estimation of blood flow velocities using optical coherence tomography intensity signals,” Opt. Lett. 36, 2068–2070 (2011). [CrossRef]   [PubMed]  

19. Y. Wang and R. Wang, “Autocorrelation optical coherence tomography for mapping transverse particle-flow velocity,” Opt. Lett. 35, 3538–3540 (2010). [CrossRef]   [PubMed]  

20. J. Tokayer, Y. Jia, A.-H. Dhalla, and D. Huang, “Blood flow velocity quantification using split-spectrum amplitude-decorrelation angiography with optical coherence tomography,” Biomed. Opt. Express 4, 1909 (2013). [CrossRef]   [PubMed]  

21. J. Lee, W. Wu, J. Y. Jiang, B. Zhu, and D. A. Boas, “Dynamic light scattering optical coherence tomography,” Opt. Express 20, 22262–22277 (2012). [CrossRef]   [PubMed]  

22. N. Weiss, T. G. van Leeuwen, and J. Kalkman, “Localized measurement of longitudinal and transverse flow velocities in colloidal suspensions using optical coherence tomography,” Phys. Rev. E 88, 042312 (2013). [CrossRef]  

23. J. Lee, H. Radhakrishnan, W. Wu, A. Daneshmand, M. Climov, C. Ayata, and D. A. Boas, “Quantitative imaging of cerebral blood flow velocity and intracellular motility using dynamic light scattering–optical coherence tomography,” J. Cereb. Blood Flow Metab. 33, 819–825 (2013). [CrossRef]   [PubMed]  

24. B. K. Huang and M. A. Choma, “Resolving directional ambiguity in dynamic light scattering-based transverse motion velocimetry in optical coherence tomography,” Opt. Lett. 39, 521–524 (2014). [CrossRef]   [PubMed]  

25. P.-C. Li, C.-J. Cheng, and C.-K. Yeh, “On velocity estimation using speckle decorrelation [blood],” IEEE Trans. Ultrason., Ferroelectr. Freq. Control 48, 1084–1091 (2001). [CrossRef]  

26. F. A. Lupotti, C. L. De Korte, F. Mastik, and A. F. Van Der Steen, “Dynamic noise correction for IVUS quantitative volume blood flow: methods and numerical validation,” Ultrasound Med. Biol. 28, 1053–1060 (2002). [CrossRef]   [PubMed]  

27. D. W. Park, G. H. Kruger, J. M. Rubin, J. Hamilton, P. Gottschalk, R. E. Dodde, A. J. Shih, and W. F. Weitzel, “Quantification of ultrasound correlation-based flow velocity mapping and edge velocity gradient measurement,” J. Ultrasound Med. 32, 1815–1830 (2013). [CrossRef]   [PubMed]  

28. S. Yun, G. Tearney, J. de Boer, and B. Bouma, “Removing the depth-degeneracy in optical frequency domain imaging with frequency shifting,” Opt. Express 12, 4822–4828 (2004). [CrossRef]   [PubMed]  

29. S. H. Yun, G. J. Tearney, B. J. Vakoc, M. Shishkov, W. Y. Oh, A. E. Desjardins, M. J. Suter, R. C. Chan, J. A. Evans, I.-K. Jang, N. S. Nishioka, J. F. de Boer, and B. E. Bouma, “Comprehensive volumetric optical microscopy in vivo,” Nat. Med. 12, 1429–1433 (2006). [CrossRef]   [PubMed]  

30. M. Vaughan, The Fabry-Perot Interferometer: History, Theory, Practice and Applications (CRC Press, 1989).

31. Y. Bromberg, Y. Lahini, E. Small, and Y. Silberberg, “Hanbury brown and twiss interferometry with interacting photons,” Nat. Photonics 4, 721–726 (2010). [CrossRef]  

32. K. Schatzel, “Correlation techniques in dynamic light scattering,” Appl. Phys. B 42, 193–213 (1987). [CrossRef]  

33. W. Li, C. T. Lancee, E. I. Cespedes, A. F. W. v. d. Steen, and N. Bom, “Decorrelation of intravascular echo signals: Potentials for blood velocity estimation,” J. Acoust. Soc. Am. 102, 3785–3794 (1997). [CrossRef]  

34. W. Li, A. F. van der Steen, C. T. Lancée, I. Céspedes, and N. Bom, “Blood flow imaging and volume flow quantitation with intravascular ultrasound,” Ultrasound Med. Biol. 24, 203–214 (1998). [CrossRef]   [PubMed]  

35. D. D. Duncan and S. J. Kirkpatrick, “Can laser speckle flowmetry be made a quantitative tool?” J. Opt. Soc. Am. A 25, 2088–2094 (2008). [CrossRef]  

36. J. Senarathna, A. Rege, N. Li, and N. Thakor, “Laser speckle contrast imaging: Theory, instrumentation and applications,” IEEE Rev. Biomed. Eng. 6, 99–110 (2013). [CrossRef]   [PubMed]  

37. A. M. Forsyth, J. Wan, P. D. Owrutsky, M. Abkarian, and H. A. Stone, “Multiscale approach to link red blood cell dynamics, shear viscosity, and ATP release,” Proc. Natl. Acad. Sci. U. S. A. 108, 10986–10991 (2011). [CrossRef]   [PubMed]  

38. J. C. Wyant and K. Creath, “Basic Wavefront Aberration Theory for Optical Metrology,” in Applied Optics and Optical Engineering XI, R. R. Shannon and J. C. Wyant, eds. (Academic, 1992), pp. 2.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (8)

Fig. 1
Fig. 1 Measurement setup. The xyz coordinate system was defined by the light propagation direction z and its associated perpendicular plane xy (y towards the reader). The solid rubber phantom traveled in the x direction on a motorized linear stage. Inset: configuration for measuring signals from flowing intralipid. The beam was reflected almost perpendicular to the fiber after internal reflection (a rotation around y). The angle between the beam and the flow was θ = 82.5°.
Fig. 2
Fig. 2 (left) M-mode tomogram of phantom fluid at 2.5 mL/min with low flow velocities and large horizontal speckle size, (right) the same sample at 80 mL/min yielding high flow speeds and small horizontal speckle size. The strong reflections at the top correspond to the lens and other surfaces of the probe, the wall of the tube is the weak reflection at the bottom. The depth scale in mm is approximate as the index of refraction of the different materials has been considered equal to the fluid’s index 1.33.
Fig. 3
Fig. 3 (a–b) Mean intensity of the structural image, (c–d) inverse correlation time profiles for moving solid sample using the traditional autocorrelation function and (e–f) using the newly developed normalized autocorrelation function. (g–h) Profile comparing the inverse decorrelation time using the new and the traditional autocorrelation function at a selected column. Top row corresponds to 120 mm/s, bottom row to 240 mm/s.
Fig. 4
Fig. 4 (top) Traditional autocorrelation at selected depths for the solid sample moving at (left) 120 mm/s, (center) 200 mm/s and (right) 400 mm/s. The autocorrelation for 20 px at 400 mm/s is out of the range. (bottom) The new normalized autocorrelation for the same speeds.
Fig. 5
Fig. 5 (a) Scatter plot of τ−1 and SNR for different sample speeds as calculated using Eq. (18). (b) Scatter plot of the same data using the Pearson autocorrelation function. (c) Parametrization F of the effective inverse correlation time as a function of inverse correlation time and SNR, Eq. (23). (d) τ eff 1 as a function of τ−1 in the parametrization for several representative SNRs. Note the undefined value for low SNRs and low τ−1, as well as a unity slope line 1 : 1 as a visual guide.
Fig. 6
Fig. 6 Speed profiles for moving solid sample using τ eff 1. (a) 40 mm/s, (b) 100 mm/s, (c) 200 mm/s and (d) 300 mm/s. The areas in dark gray that denote where it is not possible to uniquely determine the speed due to the low SNR.
Fig. 7
Fig. 7 (a) Scatter plot of the time-averaged effective inverse speckle-decorrelation time τeff and the measured Doppler velocity vx for different data sets in the liquid setup. The colors denote the depth, from black to green (yellow) for the experimental data (model fit). (b) Fitting parameters as a function of depth. (c) Flow speed as determined by speckle correlation of the calibration dataset. Each flow rate has 50 flow profiles in time, and all flow rates are shown in sequence. (d) Reference Doppler velocity of the calibration dataset.
Fig. 8
Fig. 8 Structural image of the first 2048 A-lines (first row), speckle decorrelation flow speed (second row), line-of-sight-corrected Doppler flow speed (third row) and time-average flow speed profiles (fourth row) for intralipid solution at 20 mL/min, 40 mL/min, and 80 mL/min flow rates in 3.2 mm-diameter tubing. The depth is measured from the surface of the sheath, which has a 0.4 mm radius. Doppler* indicates the signal after unwrapping.

Equations (35)

Equations on this page are rendered with MathJax. Learn more.

g ( 1 ) ( τ ) = E ( t ) E * ( t + τ ) d t ,
g ( 1 ) ( τ ) = M s + M m e i 2 n k 0 v z τ e 4 n 2 k 0 2 D τ e v z 2 τ 2 / w z 2 e v t 2 τ 2 / w t 2 .
g ( 2 ) ( τ ) = I ( t ) I ( t + τ ) d t .
g ( 2 ) ( τ ) = 1 + | g ( 1 ) ( τ ) | 2 = 1 + e 8 n 2 k o 2 D τ e 2 v z 2 τ 2 / w z 2 e 2 v t 2 τ 2 / w t 2 .
g ( 2 ) ( τ ) = I ( t ) I ( t + τ ) d t I ( t ) I ( t + τ ) d t ,
g ( P ) ( τ ) = [ I ( t ) I ¯ ] [ I ( t + τ ) I ¯ ] σ I σ I d t ,
C g ( 2 ) ( 0 ) g ( 2 ) ( τ ) .
g ˜ ( 2 ) ( τ ) = 1 + γ | g ( τ ) |
I ( t ) = w S ( t ) + ( 1 w ) N ( t ) ,
SNR = 10 * log 10 w 1 w .
R II = w 2 R S S + ( 1 w ) 2 R N N + w ( 1 w ) ( R S N + R N S ) .
| I ( t ) | 2 = | w [ S R ( t ) + i S I ( t ) ] + ( 1 w ) [ N R ( t ) + i N I ( t ) ] | 2 ,
R II = w 2 R S S + ( 1 w ) 2 [ 1 + δ ( τ ) ] + 2 w ( 1 w ) ,
R II ( P ) = w 2 R S S ( P ) + ( 1 w ) 2 δ ( τ ) w 2 + ( 1 w 2 ) .
R N N = { 2 : τ = 0 1 + f τ ( τ ) : otherwise
R II = { 2 ( 1 w ) 2 + w 2 R S S ( 0 ) + 2 w ( 1 w ) : τ = 0 ( 1 w ) 2 [ 1 + f τ ] + w 2 R S S ( τ ) + 2 w ( 1 w ) : otherwise ,
Δ g = [ R II ( 0 ) R II ( Δ t ) ] [ 2 R S S ( Δ t ) ] = ( 1 w ) 2 + ( w 2 1 ) [ 2 R S S ( Δ t ) ] .
g new ( 2 ) ( τ ) g ( 2 ) ( τ Δ t ) median { g ( 2 ) ( τ > n s / 2 ) } C new ,
C new = g ( 2 ) ( τ = Δ t ) median { g ( 2 ) ( τ > n s / 2 ) } .
τ c 1 = 2 g ^ c K D + 4 g ^ c 2 K D 2 + 2 g ^ c v 2 / w t 2 ,
τ c 1 = 2 g ^ c v w t v ,
τ 1 = K B M + K B M 2 + k 2 v x 2 T ( v x , K B M ) ,
v x , c = k c 1 F ( τ 1 , SNR ) = k c 1 [ F τ ( τ 1 ) + F SNR ( SNR ) ] ,
v x , c = k n 1 F ( τ 1 , SNR ) k n 1 τ eff 1
v x , c = τ eff 1 k n 1 = τ corr 1 k c 1
F τ ( τ 1 ) = k 0 + k 1 τ 1 + k 5 τ 5 ,
F SNR ( SNR ) = α 1 ( SNR α 2 α 3 τ 1 ) α 4 .
k 0 = 24.7 ms 1 k 1 = 1.09 k 5 = 1.95 10 8 ms 4 α 1 = 34.5 ms 1 α 2 = 9 dB α 3 = 1.46 10 2 dB ms α 4 = 9.12 10 2 ,
k 0 = 2.4 ms 1 k 1 = 1.28 k 5 = 3.6 10 8 ms 4 α 1 = 5.56 ms 1 α 2 = 15.5 dB α 3 = 3.51 10 2 dB ms α 4 = 5.55 10 1 ,
τ corr 1 = K B M + K B M 2 + k n 2 v x 2 ,
τ eff 1 = F ( τ 1 , SNR ) = K ˜ B M + K ˜ B M 2 + k ˜ 2 v x 2 ,
v x = 1 k ˜ ( τ eff 1 K ˜ B M ) 2 K ˜ B M 2 .
k ˜ = 0.074 μ m K ˜ B M = 1.45 ms 1 ,
τ eff 1 ( j ) = 2 K ˜ B M + ξ ( j ) .
v x ( j ) = 1 k ˜ ( K ˜ B M + ξ ( j ) ) 2 K ˜ B M 2 = 1 k ˜ { ξ ( j ) 2 + 2 K ˜ B M ξ ( j ) } = 1 k ˜ | ξ ( j ) | { 1 + 2 K ˜ B M ξ ( j ) } ,
Select as filters