Optical coherence tomography (OCT) sensors traditionally use scanning optical delay lines with moving parts and a single detector. OCT systems with a linear detector array (linear OCT or L-OCT) are simple and robust, but a detector with approximately is needed for an imaging depth of , which is necessary for most biomedical applications. We present a new optical setup for L-OCT with an increased measurement range. An additional grating performs a reduction of the spatial frequencies of the fringe pattern on the detector without loss in the signal-to-noise ratio, so the signal can be sampled with a minimal number of pixels. The theory for this approach is addressed and the first measurements are presented.