Abstract
A confocal reflectance theta line scanning microscope demonstrates imaging of nuclear and cellular detail in human epidermis in vivo. Experimentally measured line-spread functions determine the instrumental optical section thickness to be and the lateral resolution to be . Within human dermis (through full-thickness epidermis), the measured section thickness is and the lateral resolution is . An illumination line is scanned directly in the pupil of the objective lens, and the backscattered descanned light is detected with a linear array, such that the theta line scanner consists of only seven optical components.
© 2006 Optical Society of America