Transport-of-intensity model for single-mask x-ray differential phase contrast imaging

Open Access Open Access

Abstract

X-ray phase contrast imaging holds great promise for improving the visibility of light-element materials such as soft tissues and tumors. The single-mask differential phase contrast imaging method stands out as a simple and effective approach to yield differential phase contrast. In this work, we introduce a model for a single-mask phase imaging system based on the transport-of-intensity equation. Our model provides an accessible understanding of signal and contrast formation in single-mask x-ray phase imaging, offering a clear perspective on the image formation process, for example, the origin of alternate bright and dark fringes in phase contrast intensity images. Aided by our model, we present an efficient retrieval method that yields differential phase contrast imagery in a single acquisition step. Our model gives insight into the contrast generation and its dependence on the system geometry and imaging parameters in both the initial intensity image as well as retrieved images. The model validity as well as the proposed retrieval method are demonstrated via both experimental results on a system developed in house as well as Monte Carlo simulations. In conclusion, our work not only provides a model for an intuitive visualization of image formation but also offers a method to optimize differential phase imaging setups, holding tremendous promise for advancing medical diagnostics and other applications.

© 2024 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement

1. INTRODUCTION

Conventional x-ray imaging relies on the variations of x-ray attenuation properties among different tissue types. However, it has limited contrast for low atomic number materials such as organs, tumors, and other soft tissue [14]. In recent years, x-ray phase contrast imaging (PCI) has gained much attention for its potential to enhance this soft tissue contrast by utilizing relative phase changes with x-ray propagation through the object. Among the various techniques available, the single-mask differential phase contrast imaging method stands out as a simple and effective approach yielding higher contrast than optic-free methods such as propagation-based phase imaging. Other single-optical element methods, such as the speckle tracking technique, also yield favorable outcomes. However, they necessitate the use of high-resolution detectors and may potentially demand a higher x-ray dose [5,6].

Simple propagation-based (PB) phase contrast imaging does not require any additional optics in the beam path, but only an increase in the object-to-detector distance and a partially coherent source [7,8]. At a longer propagation distance, the wavefront distortions caused by the object are recorded as intensity variations on the detector plane. These variations can be modeled by the approximated form of the transport-of-intensity equation (TIE) [9]:

$$I(z,\vec r) = I(0,\vec r) - \frac{z}{k}({\nabla _ \bot}I(0,\vec r) \cdot {\nabla _ \bot}\phi (\vec r) + I(0,\vec r)\nabla _ \bot ^2\phi (\vec r)).$$
Here $I(z,\vec r)$ and $I(0,\vec r)$ are the x-ray intensity at the object plane and detector plane, respectively, $\phi (\vec r)$ is the beam’s phase shift caused by the object, $z$ is the object-to-detector distance, $k$ is the wave number, and $\vec r$ is the coordinate in the $x$-$y$ plane. In most applications of interest with predominantly soft materials in the beam path, we can assume the intensity variation is slow in the $x$ and $y$ directions, so the second term ${\nabla _ \bot}I(0,\vec r) \cdot {\nabla _ \bot}\phi (\vec r)$ can be neglected [9,10]. Hence the equation becomes
$$I(z,\vec r) = I(0,\vec r) - \frac{z}{k}I(0,\vec r)\nabla _ \bot ^2\phi (\vec r).$$

Thus in addition to the attenuation signal ($I(0,\vec r)$), the intensity at each detector pixel is predominantly influenced by the Laplacian of x-ray phase shift caused by the object. This Laplacian phase signal manifests as bright and dark borders along the edges, leading to edge enhancement.

A single-mask phase imaging technique [11] is similar to PB phase imaging but with an added periodic x-ray absorption mask positioned between the source and the object, in close proximity to the object [Fig. 1(a)]. The mask creates x-ray beamlets by periodically blocking ${\rm x}$ rays with thin strips of heavy-element materials such as gold. The mask is aligned with respect to the detector such that the center of each thin and long strip of beamlet is aligned to every other pixel boundary [11,12]. Hence, with a proper alignment, in the absence of an object in the beam path, the signal intensity on each detector pixel column is uniform, showing no discernible patterns [Fig. 1(b)]. When the object is introduced, the heterogeneities within the object induce refraction effects that alter the original directions of the beamlets. Thus, intensity differences appear between neighboring pixels, resulting in the appearance of bright and dark fringes on the detector. Figure 1(b) shows the schematic and Fig. 2(b) shows experimental results to be described in detail later. These relative intensity variations can allow disentangling differential phase information from x-ray intensity variations on the detector plane when the appropriate light-transport model is known.

 figure: Fig. 1.

Fig. 1. Schematic of the single-mask phase imaging method. (a) Top view of the setup. The x-ray beam propagates in the $z$ direction, and the detector pixels are placed in the $x - y$ plane. (b) Diagram of mask alignment with detector pixels. The mask strips are along the $y$ direction.

Download Full Size | PDF

 figure: Fig. 2.

Fig. 2. PCI intensity images of a PMMA rod and their cross-section profile (blue curve) with (a) propagation-based method and (b) single-mask method. The $y$ axis ticks on the left indicate positions in pixels, while the ticks on the right correspond to the cross-section profile curve.

Download Full Size | PDF

In terms of wave optics, this can also be explained as the modification of the Fresnel diffraction pattern of the periodic mask with the introduction of the object. However, given the x-ray wavelength and the mask geometry, the resolution of fine structure induced by the diffraction effect is at submicrometer scale, a dimension considerably smaller than the detector pixel size. Moreover, these fine structure features from mask diffraction also get blunted due to any focal spot width of the source. Consequently, these fine structures from diffraction have minimal contribution to the signal measured by the detector. Thus, for the relatively low-resolution detectors (tens of micrometers pixel sizes), used for the current x-ray imaging systems, with an initial mask alignment with the detector as shown in Fig. 1(b), a uniform illumination is observed. The shift in Fresnel diffraction patterns with the introduction of the object yields intensity variations shown as bright and dark fringes that will also be explained by the TIE model derived in this paper. Here again, one does not capture the fine details of the diffraction patterns, but rather the shift in intensity with and without the object, which can be easily captured by even the lower resolution detectors.

We note that the single-mask PCI method is a significantly simplified version of the double-mask edge-illumination (EI) method developed earlier [13,14] and also avoids “wasting” the large number of photons that have already transmitted through the object.

The formulation of single-mask PCI has been previously attempted using both refraction [11] and wave-optics [15] models. However, these existing models have limitations in terms of providing intuitive visualizations of signal and contrast formation in the images. In this paper, we present a new model based on TIE and show how this model can be used for efficient retrieval of absorption and differential phase. Furthermore, we show a single-shot (only one acquisition with no movement of object or optical components), low-dose phase-imaging that yields multiple image features and contrast types. While our prior work has shown efficient phase retrieval methods with spectral data (using photon counting detectors) [1620], the retrieval shown here does not require spectral data.

2. METHODS

A. Formulation

Our formulation for single-mask PCI starts with the TIE [Eq. (1)]. Unlike the propagation-based method, here we have a high-contrast periodic absorption mask, so the term ${\nabla _ \bot}I(0,\vec r) \cdot {\nabla _ \bot}\phi (\vec r)$ can no longer be neglected. Here, the transmitted intensity at the object plane is $I(0,\vec r) = T(\vec r) \cdot M(x)$, where $T(\vec r)$ and $M(x)$ represent the transmission function of the object and the mask, respectively. Therefore,

$${\nabla _ \bot}I(0,\vec r) = T{\nabla _ \bot}M + M{\nabla _ \bot}T \approx T{\partial _x}M.$$
Here we applied the approximation that ${\nabla _ \bot}I(0,\vec r)$ is mainly contributed by the mask so $M{\nabla _ \bot}T$ can be neglected. After substituting Eq. (3) into Eq. (1), the x-ray intensity measured by each detector pixel can be calculated by integrating Eq. (1) over the range of the corresponding pixel:
$$\begin{split}{I_n} &= \int_{{x_n}}^{{x_{n + 1}}} T \cdot M {\rm d}x - \frac{z}{k}\int_{{x_n}}^{{x_{n + 1}}} T \cdot {\partial _x}M \cdot {\partial _x}\phi {\rm d}x \\&\quad- \frac{z}{k}\int_{{x_n}}^{{x_{n + 1}}} T \cdot M \cdot \nabla _ \bot ^2\phi {\rm d}x,\end{split}$$
where $n$ is the pixel index in the horizontal direction when the masks with slits in the vertical direction are used, and ${x_n}$ and ${x_{n + 1}}$ are the coordinates of the left and right boundary of the corresponding pixel, respectively.

Here we assume that the attenuation, phase, and differential phase of the sample vary slowly within the range of a pixel. Then Eq. (4) becomes

$$\begin{split}{I_n} & = {{T_n}\int_{{x_n}}^{{x_{n + 1}}} M(x) {\rm d}x - \frac{z}{k}{T_n}{\partial _x}{\phi _n}\int_{{x_n}}^{{x_{n + 1}}} {\partial _x}M(x) {\rm d}x}\\&\quad - {\frac{z}{k}{T_n}\nabla _ \bot ^2{\phi _n}\int_{{x_n}}^{{x_{n + 1}}} M(x) {\rm d}x}\\ & = { {T_n}(1 - {L_n})\int_{\textit{np}}^{(n + 1)p} M(x) {\rm d}x - {T_n}{D_n}\int_{\textit{np}}^{(n + 1)p} {\partial _x}M(x) {\rm d}x.}\end{split}$$
Here ${T_n} = T({x_n})$, which represents the object attenuation function averaged within each pixel; ${L_n} = \frac{z}{k}\nabla _ \bot ^2\phi ({x_n})$, which is the Laplacian of phase shift caused by the object; and ${D_n} = \frac{z}{k}{\partial _x}\phi ({x_n})$, which is the gradient of phase shift, and is proportional to the x-ray refraction angle.

In the case of a perfect mask, as we have demonstrated in our previous paper [18,21], the mask transmission function $M(x)$ can be expressed as a square wave. Considering the imperfection of the mask, a more general form of its transmission function can be expressed as a Fourier series:

$$M(x) = \sum\limits_m {C_m}\cos \frac{{2\pi mx}}{{(2p)}},$$
where $p$ is the detector pixel size, which means the period of the mask is two times of detector pixel size [see Fig. 1(a) for reference on mask versus detector period]. Then we have
$$\int_{\textit{np}}^{(n + 1)p} M(x) {\rm d}x = \int_{\textit{np}}^{(n + 1)p} \sum\limits_m {C_m}\cos \frac{{\pi mx}}{p} {\rm d}x = {C_0}p,$$
$$\begin{split}\int_{\textit{np}}^{(n + 1)p} {\partial _x}M(x) {\rm d}x &= M((n + 1)p) - M(np) \\&= - 2\sum\limits_m {C_{2m + 1}} \cdot {(- 1)^n}.\end{split}$$

The results of Eq. (7) are related to the mask transmission function and are not related to the object property. Thus, Eq. (5) becomes

$${I_n} = {w_e}{T_n}(1 - {L_n}) - \alpha {(- 1)^n}{T_n}{D_n}.$$
One can see from this equation that the signal is a combination of two distinct effects. The first term in Eq. (8), which we refer to as the propagation-based (PB) part, shares the same form as the propagation-based PCI [Eq. (2)].

The second term, referred to as the differential phase contrast (DPC) term, gives rise to the characteristic bright and dark fringes within the image, as demonstrated in the example depicted in Fig. 2(b). This is because it contains the factor ${(- 1)^n}$, where $n$ denotes the pixel column index. The magnitude of these fringes is directly proportional to the DPC signal ${D_n}$.

Also, the two parts of the signal are multiplied by two mask-related coefficients ${w_e}$ and $\alpha$, respectively. The coefficients’ values are determined by combining Eqs. (5) and (7):

$${w_e} = {C_0}p$$
$$\alpha = 2\sum_m {C_{2m + 1}}.$$

According to Eq. (7a), the coefficient ${w_e}$ represents the integration of the mask-transmission function within a pixel, which corresponds to the average transmission of the mask. Thus, it can be interpreted as the effective transparent width or aperture size. It is also similar to the ${w_e}$ in our previous model for the double-mask method [21]. On the other hand, $\alpha$ is a unit-less coefficient that depends on the odd Fourier coefficients of the mask’s transmission function. It can be understood as the attenuation contrast between the blocked and transmitted region of the mask.

The two coefficients can be interpreted as separate filters for the PB part and the DPC part independently. In comparison with the PB method, the intensity of the PB signal in the single-mask method is reduced by the coefficient ${w_e}$. This implies that the mask selectively reduces the x-ray intensity that contributes to the PB part, thereby allowing for a reduction in x-ray radiation dose to the sample without affecting the signal intensity of the DPC part. The second coefficient, $\alpha$, which represents the contrast of the mask, determines the efficiency of obtaining the DPC signal.

B. Retrieval Method

From the last section we could see that the attenuation, Laplacian phase, and differential phase have contributions to the measured intensity. Among them, the DPC part is shown as high-frequency fringes in Fig. 2(b). A retrieval process is needed to separate the PB part and the DPC part.

 figure: Fig. 3.

Fig. 3. Retrieved (a) PB image and (b) DPC image of a PMMA rod taken with single-mask method in experiment, together with their average cross-section profiles (light blue curve).

Download Full Size | PDF

In an experimental realization, a single image is taken with the object and the mask in the beam path. This image (represented as ${I_{n(M + S)}}$) can be compared with the image with mask only (flat field ${I_{n(M)}}$). The formula for the mask-and-sample image ${I_{n(M + S)}}$ is shown in Eq. (8); for the mask-only (or flat-field) image, ${I_{n(M)}} = {w_e}$. After doing flat-field correction, we obtain

$${\bar I_n} = \frac{{{I_{n(M + S)}}}}{{{I_{n(M)}}}} = {T_n}(1 - {L_n}) - {(- 1)^n}\frac{\alpha}{{{w_e}}}{T_n}{D_n}.$$
Thus we can write the corrected intensity for $n$th and ($n + {1}$)th pixels in the same row:
$$\begin{split}{{\bar I}_n} & = {{T_n}(1 - {L_n}) - {{(- 1)}^n}\frac{\alpha}{{{w_e}}}{T_n}{D_n}},\\{{\bar I}_{n + 1}} & ={ {T_{n + 1}}(1 - {L_{n + 1}}) + {{(- 1)}^n}\frac{\alpha}{{{w_e}}}{T_{n + 1}}{D_{n + 1}}}.\end{split}$$
We can separate the PB and DPC signals by adding and subtracting the intensity values on $n$th and ($n + {1}$)th pixels in each row:
$${\bar I_n} + {\bar I_{n + 1}} \approx 2{T_n}(1 - {L_n}),$$
$${\bar I_n} - {\bar I_{n + 1}} \approx 2(- {1)^n}\frac{\alpha}{{{w_e}}}{T_n}{D_n}.$$

From Eq. (12a), we can easily have the retrieval of the PB part [Eq. (13a)]. In order to retrieve ${D_n}$, if we consider the intensity of the Laplacian of phase to be relatively weak compared with 1, we can apply the approximation of $1 - {L_n} \approx 1$ when calculating differential phase ${D_n}$. Then we can arrive at the retrieval formula for propagation-based PCI and differential phase:

$${T_n}(1 - {L_n}) \approx \frac{{{{\bar I}_n} + {{\bar I}_{n + 1}}}}{2},$$
$${D_n} \approx {(- 1)^n}\frac{{{w_e}}}{\alpha}\frac{{{{\bar I}_n} - {{\bar I}_{n + 1}}}}{{{{\bar I}_n} + {{\bar I}_{n + 1}}}},$$
where Eq. (13a) is the retrieved PB image and Eq. (13b) is the retrieved DPC image. As one can observe, the strength of the retrieved differential phase signal is related to the effect slit width ${w_e}$ and the attenuation contrast of the mask ($\alpha$).

C. Experiment

We used a polychromatic micro-focus x-ray tube (Hama-matsu L8121-03) operating with a focal spot of 7 µm and the tube voltage of 40 kV. The source-to-object and object-to-detector distance were both around 60 cm. The sample in consideration is a PMMA rod with a diameter of 3 mm. We used a mask with gold strips, approximately 52 µm in periodicity, fabricated on a silicon substrate. The data was collected using a silicon photon-counting detector with the pixel size of 55 µm [22], which was carefully calibrated and corrected [23,24]. While spectral data is available with this detector, the methods presented here do not use this spectral information and treat it as an energy integrating detector. The raw image is shown in Fig. 2(b), and the retrieved PB and DPC images are shown in Fig. 3. The results also include images of a multimaterial sample shown in Fig. 4, as well as a dried wasp specimen shown in Fig. 5.

 figure: Fig. 4.

Fig. 4. Retrieved images of the middle part of a plastic tube with water and a PMMA rod inside. (a) Picture of the sample; (b) retrieved PB image; and (c) retrieved DPC image.

Download Full Size | PDF

 figure: Fig. 5.

Fig. 5. Retrieved images of a wasp specimen taken with single-mask method in experiment. (a) Retrieved PB image; (b) retrieved DPC image.

Download Full Size | PDF

3. RESULTS AND DISCUSSION

The raw image [Fig. 2(b)] obtained from the single-mask method reveals distinct signal components, including attenuation, the Laplacian of phase, and differential phase. The presence of attenuation results in darker regions in the middle of the cylinder. The Laplacian phase manifests as bright and dark borders along the edges. Additionally, the differential phase appears as bright and dark fringes specifically in regions with nonzero phase gradient. As the differential phase signal varies across the sample, it gives rise to variations in the intensity of the fringes. These observed signal components align well with the outcomes predicted by our newly proposed model [Eq. (8)], validating its reliability in capturing and explaining the underlying physics of the single-mask phase imaging method.

Additionally, the retrieved PB image [Fig. 3(a)] obtained from our model closely resembles the image captured using the propagation-based method [Fig. 2(a)]. The minor difference between the attenuation levels can be attributed to the shift of the spectrum induced by the mask’s silicon substrate. Furthermore, the retrieved differential phase contrast (DPC) image [Fig. 3(b)] exhibits excellent contrast and visibility. These results indicate that our proposed retrieval method, based on our formulated model, effectively separates and provides visualization of different signal components all from a single image.

The multimaterial sample we used was a plastic tube filled with water and a 3 mm diameter PMMA rod. We could see in the PB image [Fig. 4(b)] that we can identify the wall of the tube between water, but the PMMA rod inside is almost invisible. Conversely, in the DPC image [Fig. 4(c)], the PMMA rod is easily recognizable, although with a lower contrast between the tube wall and water compared to the PB image. Both images provide unique and complementary information about the materials within the sample. However, we can observe jagged artifacts along the edges of the sample. This occurs because, for objects with edge features, the differential phase signal—which represents the first-order derivative—mathematically diverges at the edges. Essentially, near-edge signal intensity changes significantly within a small pixel range. When sampled at pixelated resolution, the captured signal intensity depends on the edge’s relative position with pixels, leading to periodic intensity changes along the edges.

The retrieved PB and DPC images of the dried wasp using our model are shown in Fig. 5. Both of these images, retrieved from a single-shot of single-mask phase contrast intensity image, show fine details of the specimen. Both of these images exhibit a high sensitivity to tissue boundaries within the sample, due to the visualization of Laplacian and gradient of phase, respectively. Note that the retrieved PB image is a combination of the Laplacian of phase and the attenuation.

Also, discernible differences exist between the two images. The retrieved PB image captures edge information in every orientation in the 2D plane. Conversely, the retrieved DPC image accentuates features that align perpendicular to the mask strips.

In addition, it is interesting to note that the DPC images have higher sensitivity to features with slower variations, such as the bubbles within the adhesive used to affix the specimen shown at the lower region of the image.

For further verification of our TIE model, we compared the results obtained from the TIE model and the Monte-Carlo simulation [25]. For the TIE model [Eq. (10)], the calculation is based on known ${w_e}$ and $\alpha$ without using the mask transmission function. In contrast, the Monte-Carlo simulation employs the mask’s transmission function. Both methods model the same experimental geometry and sample. The results are shown in Fig. 6. From the figure, we observe that the results obtained via our TIE model calculation align consistently with the Monte-Carlo simulation outcomes with different mask selections. It shows that our new model demonstrates overall accuracy in comparison with Monte-Carlo simulations.

 figure: Fig. 6.

Fig. 6. Comparison of TIE model calculation and Monte-Carlo simulation with different mask transmission functions. (a) Plot of transmission functions, where Mask 1 represents a perfect square wave, while Mask 2 is defined as $M(x) = 0.5 + 0.5\cos (\frac{{\pi x}}{p})$. (b), (c) Comparison of the flat-field corrected raw image between TIE model calculation and Monte-Carlo simulation of Mask 1 and Mask 2.

Download Full Size | PDF

According to our model, the final signal depends not on the specific mask transmission function but rather on two mask parameters: effective aperture size (${w_e}$) and mask contrast ($\alpha$). The flat-field corrected raw image in Fig. 6 reveals components corresponding to Eq. (10), including attenuation, DPC, and Laplacian phase. Notably, smaller ${w_e}$ values yield higher DPC signal contrast, due to increased filtration of photons contributing to the PB term. This enhances the DPC signal proportion relative to the PB signal, thus improving x-ray dose efficiency. Furthermore, it is essential to fabricate the mask using heavy-element materials such as gold. This choice is driven by the need for sufficient x-ray attenuation while maintaining a relatively small thickness, thereby ensuring a larger contrast parameter ($\alpha$). Also, it is important to note that smaller ${w_e}$ values may present challenges in mask manufacturing and potentially require longer exposure times to maintain image quality. Careful consideration of trade-offs between dose efficiency, mask fabrication feasibility, and exposure time is crucial in practical single-mask method applications.

4. CONCLUSION

The single-mask method presents a promising technique for x-ray phase contrast imaging, specifically tailored for x-ray detectors with pixel sizes in the tens of micrometers range. This method enables the acquisition of both attenuation and high-contrast differential phase images in a single exposure from a simple system design, without the requirement for ultra-high-resolution detectors.

We have presented a novel light-transport model for a single-mask (SM) x-ray phase contrast imaging system. The measured x-ray intensity with the SM method combines attenuation, Laplacian phase, and differential phase effects. Our proposed model provides intuitive understanding of the relative contributions of these effects to the detector pixel intensities. Our model also shows how these effects depend on the design parameters of the imaging system. In particular, our newly derived model [Eq. (8)] gives a clear understanding of the unique bright and dark fringes in phase contrast intensity images.

Aided by our new model, we show an effective retrieval method yielding a PB image (combining attenuation with Laplacian phase) and a differential phase contrast image in a single acquisition, thus yielding images with two types of edge enhancement and shape-based contrast.

Our TIE model suggests that the mask transmission function can be characterized by two parameters that have a significant influence on the final signal. By considering these two parameters, one finds a flexibility and adaptability in mask design and performance optimization in practical applications. Our single-shot retrieval method combined with the simple system design yields multiple contrast. This offers a pathway for practical translatability of PCI for a broad range of applications.

Funding

National Institute of Biomedical Imaging and Bioengineering (R01 EB EB029761); Congressionally Directed Medical Research Programs (BC151607); National Science Foundation (1652892).

Disclosures

The authors declare no conflict of interest.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

REFERENCES

1. R. A. Lewis, “Medical phase contrast x-ray imaging: current status and future prospects,” Phys. Med. Biol. 49, 3573 (2004). [CrossRef]  

2. R. Fitzgerald, “Phase-sensitive x-ray imaging,” Phys. Today 53(7), 23–26 (2000). [CrossRef]  

3. S. Tao, C. He, X. Hao, et al., “Principles of different x-ray phase-contrast imaging: a review,” Appl. Sci. 11, 2971 (2021). [CrossRef]  

4. S. D. Auweter, J. Herzen, M. Willner, et al., “X-ray phase-contrast imaging of the breast—advances towards clinical implementation,” Br. J. Radiol. 87, 20130606 (2014). [CrossRef]  

5. K. S. Morgan, D. M. Paganin, and K. K. W. Siu, “X-ray phase imaging with a paper analyzer,” Appl. Phys. Lett. 100, 124102 (2012). [CrossRef]  

6. T. Zhou, I. Zanette, M.-C. Zdora, et al., “Speckle-based x-ray phase-contrast imaging with a laboratory source and the scanning technique,” Opt. Lett. 40, 2822–2825 (2015). [CrossRef]  

7. S. W. Wilkins, T. E. Gureyev, D. Gao, et al., “Phase-contrast imaging using polychromatic hard X-rays,” Nature 384, 335–338 (1996). [CrossRef]  

8. Y. Suzuki, N. Yagi, and K. Uesugi, “X-ray refraction-enhanced imaging and a method for phase retrieval for a simple object,” J. Synchrotron. Radiat. 9, 160–165 (2002). [CrossRef]  

9. C. Zuo, J. Li, J. Sun, et al., “Transport of intensity equation: a tutorial,” Opt. Lasers Eng. 135, 106187 (2020). [CrossRef]  

10. D. Paganin, S. C. Mayo, T. E. Gureyev, et al., “Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object,” J. Microsc. 206, 33–40 (2002). [CrossRef]  

11. F. Krejci, J. Jakubek, and M. Kroupa, “Hard x-ray phase contrast imaging using single absorption grating and hybrid semiconductor pixel detector,” Rev. Sci. Instrum. 81, 113702 (2010). [CrossRef]  

12. G. K. Kallon, P. C. Diemoz, F. A. Vittoria, et al., “Comparing signal intensity and refraction sensitivity of double and single mask edge illumination lab-based x-ray phase contrast imaging set-ups,” Appl. Phys. 50, 415401 (2017). [CrossRef]  

13. A. Olivo and R. Speller, “A coded-aperture technique allowing x-ray phase contrast imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007). [CrossRef]  

14. A. Olivo, K. Ignatyev, P. R. T. Munro, et al., “Noninterferometric phase-contrast images obtained with incoherent x-ray sources,” Appl. Opt. 50, 1765–1769 (2011). [CrossRef]  

15. F. A. Vittoria, P. C. Diemoz, M. Endrizzi, et al., “Strategies for efficient and fast wave optics simulation of coded-aperture and other x-ray phase-contrast imaging methods,” Appl. Opt. 52, 6940–6947 (2013). [CrossRef]  

16. D. Gürsoy and M. Das, “Single-step absorption and phase retrieval with polychromatic x rays using a spectral detector,” Opt. Lett. 38, 1461–1463 (2013). [CrossRef]  

17. M. Das, “Single step x-ray phase imaging,” U.S. patent 9237876 (19 January 2016).

18. M. Das and Z. Liang, “Spectral x-ray phase contrast imaging for single-shot retrieval of absorption, phase, and differential-phase imagery,” Opt. Lett. 39, 6343–6346 (2014). [CrossRef]  

19. M. Das, “Single step differential phase contrast x-ray imaging,” U.S. patent 9445775 (20 September 2016).

20. I. Vazquez, I. E. Harmon, J. C. R. Luna, et al., “Quantitative phase retrieval with low photon counts using an energy resolving quantum detector,” J. Opt. Soc. Am. A 38, 71–79 (2021). [CrossRef]  

21. M. Das and Z. Liang, “Approximated transport-of-intensity equation for coded-aperture x-ray phase-contrast imaging,” Opt. Lett. 39, 5395–5398 (2014). [CrossRef]  

22. R. Ballabriga, J. Alozy, G. Blaj, et al., “The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging,” J. Instrum. 8, C02016 (2013). [CrossRef]  

23. M. Das, B. Kandel, C. S. Park, et al., “Energy calibration of photon counting detectors using x-ray tube potential as a reference for material decomposition applications,” Proc. SPIE 9412, 263–268 (2015). [CrossRef]  

24. S. Vespucci, C. S. Park, R. Torrico, et al., “Robust energy calibration technique for photon counting spectral detectors,” IEEE Trans. Med. Imaging 38, 968–978 (2019). [CrossRef]  

25. J. Giersch and J. Durst, “Monte Carlo simulations in X-ray imaging,” Nucl. Instrum. Methods Phys. Res. A 591, 300–305 (2008). [CrossRef]  

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1.
Fig. 1. Schematic of the single-mask phase imaging method. (a) Top view of the setup. The x-ray beam propagates in the $z$ direction, and the detector pixels are placed in the $x - y$ plane. (b) Diagram of mask alignment with detector pixels. The mask strips are along the $y$ direction.
Fig. 2.
Fig. 2. PCI intensity images of a PMMA rod and their cross-section profile (blue curve) with (a) propagation-based method and (b) single-mask method. The $y$ axis ticks on the left indicate positions in pixels, while the ticks on the right correspond to the cross-section profile curve.
Fig. 3.
Fig. 3. Retrieved (a) PB image and (b) DPC image of a PMMA rod taken with single-mask method in experiment, together with their average cross-section profiles (light blue curve).
Fig. 4.
Fig. 4. Retrieved images of the middle part of a plastic tube with water and a PMMA rod inside. (a) Picture of the sample; (b) retrieved PB image; and (c) retrieved DPC image.
Fig. 5.
Fig. 5. Retrieved images of a wasp specimen taken with single-mask method in experiment. (a) Retrieved PB image; (b) retrieved DPC image.
Fig. 6.
Fig. 6. Comparison of TIE model calculation and Monte-Carlo simulation with different mask transmission functions. (a) Plot of transmission functions, where Mask 1 represents a perfect square wave, while Mask 2 is defined as $M(x) = 0.5 + 0.5\cos (\frac{{\pi x}}{p})$. (b), (c) Comparison of the flat-field corrected raw image between TIE model calculation and Monte-Carlo simulation of Mask 1 and Mask 2.

Equations (17)

Equations on this page are rendered with MathJax. Learn more.

I ( z , r ) = I ( 0 , r ) z k ( I ( 0 , r ) ϕ ( r ) + I ( 0 , r ) 2 ϕ ( r ) ) .
I ( z , r ) = I ( 0 , r ) z k I ( 0 , r ) 2 ϕ ( r ) .
I ( 0 , r ) = T M + M T T x M .
I n = x n x n + 1 T M d x z k x n x n + 1 T x M x ϕ d x z k x n x n + 1 T M 2 ϕ d x ,
I n = T n x n x n + 1 M ( x ) d x z k T n x ϕ n x n x n + 1 x M ( x ) d x z k T n 2 ϕ n x n x n + 1 M ( x ) d x = T n ( 1 L n ) np ( n + 1 ) p M ( x ) d x T n D n np ( n + 1 ) p x M ( x ) d x .
M ( x ) = m C m cos 2 π m x ( 2 p ) ,
np ( n + 1 ) p M ( x ) d x = np ( n + 1 ) p m C m cos π m x p d x = C 0 p ,
np ( n + 1 ) p x M ( x ) d x = M ( ( n + 1 ) p ) M ( n p ) = 2 m C 2 m + 1 ( 1 ) n .
I n = w e T n ( 1 L n ) α ( 1 ) n T n D n .
w e = C 0 p
α = 2 m C 2 m + 1 .
I ¯ n = I n ( M + S ) I n ( M ) = T n ( 1 L n ) ( 1 ) n α w e T n D n .
I ¯ n = T n ( 1 L n ) ( 1 ) n α w e T n D n , I ¯ n + 1 = T n + 1 ( 1 L n + 1 ) + ( 1 ) n α w e T n + 1 D n + 1 .
I ¯ n + I ¯ n + 1 2 T n ( 1 L n ) ,
I ¯ n I ¯ n + 1 2 ( 1 ) n α w e T n D n .
T n ( 1 L n ) I ¯ n + I ¯ n + 1 2 ,
D n ( 1 ) n w e α I ¯ n I ¯ n + 1 I ¯ n + I ¯ n + 1 ,
Select as filters