Accuracy of tissue oxygen saturation measurement with multidistance CW fNIRS: a phantom study

Not Accessible

Your library or personal account may give you access

Abstract

We investigate the potential of CW fNIRS with two NIR wavelengths to estimate tissue oxygen saturation by a model based on spatially resolved spectroscopy. Phantoms with optical properties similar to brain tissue are applied to study the influence of the optical parameters and the spatial probe configuration on effective attenuation, absorption, and oxygen saturation. Attenuation coefficients and saturation values can be estimated with a typical discrepancy of 5%, whereas absorption coefficients show larger deviations from the true values. The results are very sensitive against small errors in the source-detector separation.

© 2023 SPIE

PDF Article
More Like This
A broadband multi-distance approach to measure tissue oxygen saturation with continuous wave near-infrared spectroscopy

Zuzana Kovácsová, Gemma Bale, Joshua Deepak Veesa, Hamid Dehghani, and Ilias Tachtsidis
11074_29 European Conference on Biomedical Optics (ECBO) 2019

Evaluation of hyperspectral imaging measurements of changes in hemoglobin oxygenation and oxidation of cytochrome-c-oxidase using a liquid blood phantom

Charly Caredda, Frédéric Lange, Ilias Tachtsidis, and Bruno Montcel
126280Q European Conference on Biomedical Optics (ECBO) 2023

Hemoglobin Oxygen Saturation Tomography: Calibration in Phantom Studies and Patient Data Analysis

Subhadra Srinivasan, Brian W. Pogue, Shudong Jiang, Hamid Dehghani, Steven P. Poplack, and Keith D. Paulsen
TuE5 Biomedical Topical Meeting (BIOMED) 2002

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.