Measuring large numerical apertures by imaging the angular distribution of radiation of fluorescing molecules

Open Access Open Access

Abstract

Exact knowledge of the numerical aperture is crucial in many applications using high-aperture objectives such as confocal microscopy, optical trapping, or advanced sub-wavelength imaging methods. We propose and apply a precise and straightforward method for measuring this fundamental parameter of microscope objectives with numerical apertures above unity. Our method exploits the peculiarities of the fluorescence emission of molecules at a glass/air interface.

©2005 Optical Society of America

Full Article  |  PDF Article
More Like This
Total internal reflection fluorescence correlation spectroscopy (TIR-FCS) with low background and high count-rate per molecule

Kai Hassler, Marcel Leutenegger, Per Rigler, Ramachandra Rao, Rudolf Rigler, Michael Gösch, and Theo Lasser
Opt. Express 13(19) 7415-7423 (2005)

Single molecule fluorescence in rectangular nano-apertures

Jérôme Wenger, Pierre-Francois Lenne, Evgueni Popov, Hervé Rigneault, José Dintinger, and Thomas W. Ebbesen
Opt. Express 13(18) 7035-7044 (2005)

Orientation imaging of single molecules by wide-field epifluorescence microscopy

Martin Böhmer and Jörg Enderlein
J. Opt. Soc. Am. B 20(3) 554-559 (2003)

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (3)

Fig. 1.
Fig. 1. (a) Angular distribution of emission into glass for isotropically oriented molecules in glass at an air/glass interface when excited by light circularly polarized within the plane of the interface. Emission maximum occurs at the angle of total internal reflection (TIR), indicated by light-red lines. (b) Intensity distribution of fluorescence within the back focal plane of an objective as a function of the distance ρ from the optical axis divided by the focal distance f of the objective. The vertical light-red lines show the position of TIR.
Fig. 2.
Fig. 2. Principal scheme of the experimental set-up used for imaging the intensity distribution of fluorescence in the back focal plane of the objective. The imaging lens images the back focal plane of the objective onto the CCD camera. In the realized experimental set-up, this imaging was achieved by using the microscope’s tube lens (not shown) plus and additional lens.
Fig. 3.
Fig. 3. (a) Measured intensity distribution in the back focal plane for the Olympus UApo/340 40× objective with nominal NA of 1.15. The barely visible black circle at the edge of the illuminated area is the determined out NA-determined cut-off of fluorescence collection, the cyan circle marks the position of the fluorescence discontinuity at the TIR-angle. (b) Same as Fig. 3(a), but for the Olympus PlanApo 60× objective with nominal NA of 1.4.

Equations (3)

Equations on this page are rendered with MathJax. Learn more.

θ TIR = asin ( n 0 n ) ,
ρ = f n sin θ
NA = ρ max ρ TIR