Comb-resolved spectroscopy with immersion grating in long-wave infrared

Open Access Open Access

Abstract

We have developed a dispersive spectrometer by using a compact immersion grating for direct frequency comb spectroscopy in the long-wave infrared region of 8-10 μm for the first time. A frequency resolution of 460 MHz is achieved, which is the highest reported in this wavelength region with a dispersive spectrometer. We also demonstrate individual comb mode-resolved imaging by cavity filtering and apply this to obtain spectra of both simple and complex molecular spectra. These results indicate that the immersion grating spectrometer offers the next advancement for sensitive, high-resolution spectroscopy of transient and large/complex molecules when combined with cavity enhancement and cooling techniques.

© 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

1.1 Mid-infrared direct frequency comb spectroscopy

Direct frequency comb spectroscopy (DFCS) in the mid-infrared spectral region find important applications in both fundamental laboratory spectroscopy and remote sensing. In particular, mid-IR DFCS is an attractive approach for molecular spectroscopy because of the afforded high sensitivity, broad spectral coverage, and rapid acquisition capabilities. The fast acquisition advantage allows for both real-time remote sensing [1] (seconds) and measurement of fast chemical kinetics [2–5] (microseconds). These combined advantages have been demonstrated in techniques of cavity-enhanced spectroscopy with a dispersive spectrometer [2–4,6]. As alternatives to a dispersive spectrometer, mid-IR DFCS has also been demonstrated with dual-comb spectroscopy [5,7–10] and Fourier transform spectroscopy [11–13]. However it is challenging to simultaneously satisfy both criteria of microsecond acquisition and broad spectral bandwidth with these spectrometers.

In a dispersive spectrometer, the frequency comb spectrum is spatially dispersed in two spatial dimensions (2D) by a combination of a virtually imaged phase array (VIPA) etalon and a reflective grating, and then imaged onto a camera [14]. In this case, the VIPA spectrometer can record spectra within 10 μs, limited by the integration time of the camera, which is required to observe transient chemical species [2–5]. Unfortunately, the VIPA etalon has the inherent disadvantage of a limited-bandwidth due to optical coatings, as well as a relatively low throughput of ~20%. The former is problematic for very broadband spectroscopy, and each etalon design must be tailor-made to the specific wavelength application.

With VIPA spectrometers, DFCS has been demonstrated in the mid-infrared region up to 5 μm. Relative to shorter mid-IR wavelengths (< 5 μm), important molecular targets in atmospheric sciences like the Criegee intermediate [15], NO3 radical [16,17], isoprene [18], and fundamental spectroscopy like buckyball (C60) [19], display significantly larger absorption intensities near 10 μm. A comparably important consideration for these larger, complex molecules is that spectroscopic probing at longer infrared wavelengths alleviates spectral congestion due to IVR (intramolecular vibrational redistribution) processes, thus enabling quantum-state resolution [6,20]. These advantages motivate the recent construction of an 8-10 μm mid-infrared frequency comb (an optical parametric oscillator (OPO) based on AgGaSe2) for DFCS [21]. Here, we report a dispersive spectrometer for 8-10 μm DFCS comprised of an immersion grating, a reflective grating, and a strained layer superlattice (SLS) detector element camera for imaging. The enabling technology for this spectrometer is the immersion grating, which has been fabricated for wavelengths > 5 μm.

1.2 Immersion grating

An immersion grating is essentially an echelle grating with a large blaze angle. However, unlike traditional echelle gratings, the diffraction surface is immersed in a high refractive index (n) material, so that the angular dispersion of the immersion grating is enhanced by n. Therefore, the physical size of immersion grating can be reduced by a factor of n to achieve the same resolving power as a reflective echelle grating. According to the Rayleigh criterion, the resolving power (RP) of an immersion grating is described as

RP=λ/Δλ=(2nwtanθ)/λ,
where w is the diameter of the input beam, θ is the blaze angle, and λ is the wavelength. w × tanθ indicates the effective beam diameter illuminating grooves and the resolving power is proportional to the input beam diameter. Although the immersion grating concept has been around for some time [22,23] for silicon [24–26], ZnSe [27], GaP [27], and bismuth germinate [27], recent advances in machining techniques of brittle crystals like CdZnTe [28], InP [29], and germanium have enabled their realization [29–31] and potential application for spectroscopy in the infrared. In the field of infrared astronomy, immersion gratings are highly desired for their high resolving power, which for a normal echelle grating would have been of much larger sizes. Immersion gratings have been already installed on international telescopes [32–34].

The immersion grating used in this work is fabricated from a single germanium crystal. The Ge immersion grating can achieve relatively high diffraction efficiency comparable to a traditional reflective grating, and its resolving power can be arbitrarily scaled by increasing the input beam size. Therefore, this immersion grating is a versatile candidate for use in a high resolution, long mid-infrared spectrometer due to its transparency from 2 μm to 16 μm and large reflective index (n = 4). A preliminary demonstration of DFCS with a Ge immersion grating was reported at 4 μm [35]. Here, we apply this grating for DFCS in the 8-10 μm spectral region. In this study, we report in-depth characterization of this dispersive spectrometer, with focused discussions on diffraction efficiency, resolving power, noise performance, its use for high resolution molecular spectroscopy, and the potential for time-resolved spectroscopy.

2. Apparatus and characterization

2.1 Immersion grating spectrometer

Figure 1(a) shows an overview of the immersion grating spectrometer. Our Ge immersion grating has dimensions of 50 × 40 × 155 mm as shown in the top panel in Fig. 1(a). The blaze angle is 75 ° and the lattice constant (groove spacing) is 476 μm. The diffraction surface is Au-coated to maximize the diffraction efficiency. The input facet of the immersion grating is AR-coated to cover the whole spectrum of the light source, which is an OPO with wavelength tunable 8-10 μm [21] (about 200 nm). The incident light enters the immersion grating at an angle about normal to the surface, which corresponds to the blaze angle (75 degrees). Due to the large blaze angle and groove spacing characteristic of echelle-type gratings as well as high refractive index of Ge, the diffraction order, m, is high (~432) and thus higher orders of the diffracted light are highly spatially overlapped. The resulting grating free spectral range (FSR), given by λ/m ~ 20 nm, is much narrower than the spectral bandwidth of the OPO. A conventional reflective grating is inserted as an orthogonal cross-disperser to map the full frequency comb spectrum onto a 2D image. Since the cross-disperser grating only needs to separate wavelengths covering one FSR (20 nm) set by the immersion grating, its required resolving power is relatively low, at about 500. Our cross-disperser grating has a groove spacing of 13 μm and an overall dimension of 68 × 68 × 9 mm. Finally, the diffraction efficiency is 50% for the immersion grating and 70% for the cross-disperser grating in a quasi-Littrow configuration.

 figure: Fig. 1

Fig. 1 DFCS spectrometer. (a) A schematic of the Ge immersion grating and an overview of the immersion grating spectrometer. d is the lattice constant of the immersion grating. The OPO is an optical parametric oscillator operating at 8-10 μm. The diffracted light from the Ge immersion grating is cross-dispersed with a reflective grating and mapped onto a camera as a 2D image. (b) Camera image of the dispersed comb light. The comb modes are dispersed vertically by the immersion grating and horizontally by the cross-disperser grating.

Download Full Size | PDF

Figure 1(b) shows a typical 2D image taken by the LN2-cooled SLS camera with 640 × 512 pixels, 20-μm pixel pitch, and frame rate of 100 Hz. Although the SLS sensor exhibits non-linearity, this effect is mostly normalized away by our measurement technique of rapidly (10 ms) subtracting the reference image without molecular absorption from the signal image with absorption. The frequency comb wavelength is vertically dispersed by the immersion grating and horizontally dispersed by the cross-disperser grating into a series of vertical fringes. The image for each fringe is not comb-tooth resolved. According to Eq. (1), the calculated frequency resolution for a beam diameter (1/e2 diameter) of 18 mm is about 570 MHz, which is larger than typical repetition rate of our fiber-based frequency comb (frep = 110 MHz), such that mode filtering is required to observe a comb mode-resolved image. The width of one fringe is about 90 μm (4.5 pixels) which is consistent with the expected beam diameter from Gaussian beam propagation through the imaging optics after the gratings. The separation between fringes, which is determined by the angular dispersion of the cross-disperser grating and the focal length of the imaging lenses, is about 280 μm (14 pixels), which is more than sufficient to avoid inter-fringe crosstalk. The entire OPO spectrum is mapped on 28 fringes that cover approximately two-thirds of the camera detector size.

2.2 Grating free spectral range (FSR)

We characterize the FSR of the immersion grating from measurement of molecular absorption features. An absorption cell is inserted before the beam expansion lenses shown in Fig. 1(a). Figure 2(a) shows recorded images of rotationally-resolved D2O and N2O absorption of the immersion grating spectrometer. In Fig. 2(a), a negative image is shown to accentuate molecular absorption indicated by bright dots. The FSR is determined by locating the repeating absorption patterns, which is most apparent in the D2O image. The repeating pattern is indicated in Fig. 2(a) with circular markers of the same color for the same absorption features. The red boundary lines in Fig. 2(a) indicate one FSR. The N2O spectrum also corroborates this observation. Since the wavelength spacing between rotational transitions of N2O is fairly constant at 6 nm (~0.8 cm−1), and one FSR is calculated to be ~20 nm (section 2.1), we would expect approximately 4 absorption dots to span one FSR. This is consistent with the data in the absorption image of N2O by counting the number of absorption dots on a single fringe. To obtain a traditional frequency-domain spectrum, the fringes are rastered from top to bottom and left to right by a fringe-finding algorithm. To avoid redundancy in the overlapping diffraction orders, the image is first cropped to discover fringes only within one FSR region.

 figure: Fig. 2

Fig. 2 Molecular spectroscopy. (a) Observed camera images (negative) of D2O and N2O absorption shown in bright dots. The marked absorption dots in the D2O image show the repeating pattern, indicating they are the same spectral feature. The vertical gap separated by the two horizontal red lines indicates the FSR of the immersion grating. (b) Observed spectrum of N2O, D2O, and dimethyl ether with the cell length of 10 cm. The frequency axis is calibrated with N2O and D2O spectra reported in the HITRAN database. The simulation is shown as inverted red traces. Absorption spectra of dimethyl ether were taken with three different sample pressures (red: 0.3 Torr; blue: 0.7 Torr; and black: 1.3 Torr).

Download Full Size | PDF

2.3 Molecular spectroscopy

Once the grating FSR has been determined, molecular absorption can be quantitatively determined according to the Lambert-Beer’s law by recording camera images with and without absorption. Figure 2(b) shows the observed single-shot, absorption spectra of N2O, D2O, and dimethyl ether retrieved from the fringe-finding algorithm. The frequency axes are calibrated with N2O and D2O rovibrational spectra reported in the HITRAN database [36]. Under-sampling causes the discrepancy in line intensity between the observed and simulated spectra.

In contrast to the 3-5 μm where predominately vibrational stretching fundamentals are observed, the 8-10 μm region provides access to lower frequency bending modes, such as those observed in the methyl rocking motion of dimethyl ether centered at ~1175 cm−1 (8510 nm). Dimethyl ether is an important molecule in astrophysics and the interstellar medium, and is also one of the simplest molecules displaying large-amplitude motions, corresponding to the internal rotations of the two methyl groups [37]. A combination of the tunneling splitting of the rotational levels caused by the methyl torsional motion and highly perturbed nature of the vibrational excited state results in a complex and intractable spectrum, which could be resolved if the molecules were rotationally and vibrationally cooled, for example by buffer gas cooling or molecular beam expansion. Here several room temperature dimethyl ether spectra are recorded at different sample pressures. Although rovibrational assignments are not possible at the current room temperature conditions, this is the first report of a rotationally-resolved dimethyl ether spectrum of this band, which demonstrates that the immersion grating spectrometer has potential for extending the applications of high-resolution spectroscopy towards larger molecules with complex structure.

2.4 Resolution

The frequency resolution of the spectrometer is set by the resolving power of the immersion grating and the imaging system. We experimentally determined the spectrometer resolution using two different approaches: 1) pressure dependent linewidth measurements of N2O absorption lines and 2) comb mode resolved imaging with a comb-mode-filtering cavity. In Fig. 3(a), the linewidth of pure N2O samples, determined from fitting absorption lines with the Voigt line profile (Fig. 3(b)), is plotted as a function of pressure. At high pressures, the linewidth of the absorption is determined by pressure broadening and its line profile is dominated by the Lorentzian component. At low pressures, the linewidth is dominated by the instrument linewidth and Doppler broadening, which is described by a Gaussian component. In the fit, the Lorentzian width is fixed to the calculated value for each pressure according to the pressure-broadening coefficient based on the HITRAN database [36]. Here, we measured the pressure-dependent linewidth of N2O at two different beam diameters incident on the immersion grating. The slope of the figure shows the pressure broadening coefficient and both slopes are about 10 MHz/Torr for N2O self-broadening, which agrees with the HITRAN database. The linewidth extrapolated to zero pressure (y-intercept) is described by the Gaussian component, which is dominated by the instrument linewidth since the Doppler width at room temperature is only about 65 MHz (FWHM). Using this method, the instrument-limited frequency resolution (FWHM) is measured to be 623(5) and 460(16) MHz with a beam diameter of 18(1) mm and 34(1) mm, respectively. A resolution of ~460 MHz is the highest reported for DFCS with a dispersive spectrometer in this wavelength region.

 figure: Fig. 3

Fig. 3 Spectrometer resolution. (a)Pressure-dependent N2O linewidth measurements with the input beam diameters of 18 mm (red) and 34 mm (blue). (b) Voigt profile (green, solid) and Lorentzian profile (purple, dashed) fit to a hot band transition near 1145.518cm−1 for a 10 torr sample of pure N2O. The root-mean-squares (RMS) of the residual for the Voigt and Lorentzian line profile fits are 0.0027 and 0.0049, respectively. (c)Comb mode-resolved 2D image. One out of nineteen comb modes is filtered with a Fabry-Perot cavity (effective repetition rate is 2.09 GHz). Each comb mode is focused onto 4.5 pixels (90 μm) and displays a linewidth of about 740 MHz when the input beam diameter is 18 mm.

Download Full Size | PDF

The expected spectrometer resolution is a convolution of the immersion grating and the imaging system instrument functions. The former was calculated using Eq. (1) in which the resolving power of the immersion grating is determined by the number of illuminated grooves, and therefore, input beam size w. The latter was calculated from the image mapping of the linear dispersion of the gratings and beam diameter on the camera detector plane. For input beam diameters of 18(1) and 34(1) mm, we obtained expected spectrometer resolutions of 650(60) MHz and 460(30) MHz, respectively, which are in good agreement with the measured values. In this current system, the beam size, thus resolving power, is limited by the input facet dimensions of the immersion grating, which is only limited by the fabrication process for producing larger Ge crystals.

As additional validation of the spectrometer resolution, we measured the comb mode-resolved camera image. To obtain the comb mode-resolved image, we use the cavity-filtering (Vernier) technique in which one comb mode is filtered out every 19 modes using a Fabry-Perot cavity, so that the resulting effective repetition rate is 2.09 GHz. The cavity comprised two highly reflective ZnSe mirrors with a finesse of 6000 at 8-9 μm. 7 cavity FSRs correspond to 2.09 GHz. This value is much larger than the resolution of the immersion grating, resulting in resolvable comb teeth shown in Fig. 3(c). To lock the comb laser frep to the Vernier filter cavity FSR, the swept-locked technique is used [38]. Here, we sweep frep by modulating the comb oscillator cavity length, and the feedback error signal is used to control the length of the ring PZT attached to one of the cavity mirrors in the Vernier cavity. The comb teeth spacing in Fig. 3(c) is the expected 2.09 GHz, and the spectrometer resolution is determined by fitting the FWHM (740(80) MHz) of each comb tooth to a Gaussian line profile. The discrepancy in the frequency resolution between the two methods originates from the slight difference in the input beam size to the immersion grating. The large uncertainty of the latter method is due to smaller number of data than the former method.

2.5 Noise characterization and sensitivity

Finally, we characterized the spectrometer’s sensitivity for direct absorption experiments. The intensity of transmitted light is given by Lambert-Beer’s Law:

I=I0exp(αL)I0(1αL)forαL1.
Here, α is the absorption coefficient, and L is the absorption length. Solving for αL in Eq. (2) yields
αL(I0I)/I0ΔI/I0.
We determined noise processes on absorption from measurement of ΔI/I0. To characterize the noise of this measurement, 100 sets of three images were recorded every 200 ms with comb light incident on the camera sensor in the absence of absorption. The three images within each set were obtained at 250 μs integration time, each separated by 10 ms, corresponding to a frame rate of 100 Hz. The first image is a background image (B) where the laser light is blocked by a chopper and the two others (S0 and S1) measure the intensity of the incident light such that S0B corresponds to I0 and S1B corresponds to I. Therefore, ΔI/I0=(S0S1)/(S0B). For a given pixel, the standard deviation of ΔI/I0 over the 100 measurements gives an estimate for the fractional intensity noise, δI. First, we determined δI as a function of laser input power. We estimate the power incident on a single pixel by first determining the gain, g, which has units of electron/count:
g=ηCtotPtottintEphoton,
where η is the quantum efficiency of the camera, Ctot is the integrated counts on illuminated pixels, Ptot is the total incident power, Ephoton is photon energy, and tint is the camera integration time. From Eq. (4), the direct conversion of the measured count to optical power for a single pixel is given by the conversion factor a, which has units of power/count:

a=Ephotontintgη.

The average δI for as a function of power per pixel is shown in Fig. 4(a). If no noise sources were present in the spectrometer, the minimum ΔI/I0 on a single pixel would be limited by shot noise and given by [39]

(ΔI/I0)minshot=2eB/Rp,
where e is the electron charge, B is the measurement bandwidth, R is the responsivity of the detector and p is the power incident on the pixel obtained from measure counts of single pixel multiplied by a in Eq. (5). Figure 4(a) shows that (ΔI/I0)minshot is roughly two orders of magnitude below δI, indicating that we are well above the shot noise limit at the current experimental optical power.

 figure: Fig. 4

Fig. 4 Spectrometer sensitivity. (a) The fractional noise as a function of power per pixel. The noise figure is much higher than the shot noise limit shown in the blue dotted line. When the incident power per pixel is lower than 5 nW, the camera noise is dominant. While above that power, intensity noise is dominant. (b) Fractional noise as a function of average time with the total input power of about 3 nW. δI3×104 per detection channel is achieved with 1000 averages.

Download Full Size | PDF

To determine the sources of noise, we analyzed the behavior of δI. By assuming that the major noise sources, (laser intensity noise, shot noise, and camera noise) are uncorrelated, we obtain

δI(AΔI)2+B2ΔI+C2/I0.
Laser intensity noise is proportional to AΔI, shot noise to B(ΔI)1/2, and camera noise (a total measurement of dark noise, readout noise, etc.) to C, for constants A, B, and C If AΔIB(ΔI)1/2,C, laser intensity noise dominates; if B(ΔI)1/2AΔI,C, shot noise dominates; if CAΔI,B(ΔI)1/2, camera noise dominates. Until ~5 nW of incident power, the slope of the log-log plot is approximately −1, indicating that detector noise dominates. At higher powers, the δI becomes independent of power, indicating that laser intensity noise dominates. This observation was independently verified by an additional laser relative intensity noise (RIN) measurement.

Noise can be further averaged down. A set of 3000 measurements of ΔI/Io were recorded with ~3 nW of total incident laser power on a pixel. N measurements of ΔI/Io were averaged together, where N=1,2,3,,1500. For a single pixel, δI was calculated as a function of N. The result for an average of 50 randomly chosen pixels is shown in Fig. 4(b). Since the slope of this line is proportional to N1/2, we conclude that the noise at this power is essentially random, and that we can achieve an absorption sensitivity of δI3×104 per detection channel for 1000 measurements, or 200 s of acquisition.

3. Conclusion

In summary, we have developed a frequency comb spectrometer in the long-wave infrared region (8-10 μm) using an immersion grating. The highest frequency resolution of 460 MHz for dispersive spectrometer is achieved. In our current implementation, the frequency resolution of the spectrometer is limited by the size of the immersion grating. However, the resolution can be significantly increased by performing comb mode-resolved spectroscopy demonstrated in this work and stepping the repetition rate of the comb, which will push the limit of resolution to the linewidth of individual comb teeth [40].

The next application of this spectrometer is to perform cavity-enhanced time-resolved spectroscopy, which will exploit high sensitivity for the detection of transient molecules like NO3, Criegee intermediates, and carbonic acid for applications in biology, geology, and atmospheric science, with time resolution of about 1 μs limited by the integration time of a camera. Moreover, we plan to incorporate the buffer gas cooling and molecular beam techniques with this spectrometer to perform quantum state-resolved spectroscopy of cold, large molecules like C60 and dimethyl ether.

Funding

Japan Society for the Promotion of Science (JSPS) Overseas Research Fellowships; Air Force Office of Scientific Research (AFOSR) (FA9550-15-1-0111); Defense Advanced Research Projects Agency (DARPA) SCOUT Program; National Science Foundation (NSF) (PHYS-1734006).

Acknowledgment

We thank P. Bryan Changala for helpful discussions.

References

1. L. Nugent-Glandorf, F. R. Giorgetta, and S. A. Diddams, “Open-air, broad-bandwidth trace gas sensing with a mid-infrared optical frequency comb,” Appl. Phys. B 119(2), 327–338 (2015). [CrossRef]  

2. A. J. Fleisher, B. J. Bjork, T. Q. Bui, K. C. Cossel, M. Okumura, and J. Ye, “Mid-infrared time-resolved frequency comb spectroscopy of transient free radicals,” J. Phys. Chem. Lett. 5(13), 2241–2246 (2014). [CrossRef]   [PubMed]  

3. B. J. Bjork, T. Q. Bui, O. H. Heckl, P. B. Changala, B. Spaun, P. Heu, D. Follman, C. Deutsch, G. D. Cole, M. Aspelmeyer, M. Okumura, and J. Ye, “Direct frequency comb measurement of OD + CO → DOCO kinetics,” Science 354(6311), 444–448 (2016). [CrossRef]   [PubMed]  

4. T. Q. Bui, B. J. Bjork, P. B. Changala, T. L. Nguyen, J. F. Stanton, M. Okumura, and J. Ye, “Direct measurements of DOCO isomers in the kinetics of OD + CO,” Sci. Adv. 4(1), 4777 (2018).

5. J. L. Klocke, M. Mangold, P. Allmendinger, A. Hugi, M. Geiser, P. Jouy, J. Faist, and T. Kottke, “Single-shot sub-microsecond mid-infrared spectroscopy on protein reactions with quantum cascade laser frequency combs,” Anal. Chem. 90(17), 10494–10500 (2018). [CrossRef]   [PubMed]  

6. P. B. Changala, B. Spaun, D. Patterson, J. M. Doyle, and J. Ye, “Sensitivity and resolution in frequency comb spectroscopy of buffer gas cooled polyatomic molecules,” Appl. Phys. B 122(12), 292 (2016). [CrossRef]  

7. G. Ycas, F. R. Giorgetta, E. Baumann, I. Coddington, D. Herman, S. A. Diddams, and N. R. Newbury, “High-coherence mid-infrared dual-comb spectroscopy spanning 2.6 to 5.2 μm,” Nat. Photonics 12(4), 202–208 (2018). [CrossRef]  

8. G. Villares, A. Hugi, S. Blaser, and J. Faist, “Dual-comb spectroscopy based on quantum-cascade-laser frequency combs,” Nat. Commun. 5(1), 5192 (2014). [CrossRef]   [PubMed]  

9. M. Yan, P.-L. Luo, K. Iwakuni, G. Millot, T. W. Hänsch, and N. Picqué, “Mid-infrared dual-comb spectroscopy with electro-optic modulators,” Light Sci. Appl. 6(10), e17076 (2017). [CrossRef]   [PubMed]  

10. M. Yu, Y. Okawachi, A. G. Griffith, N. Picqué, M. Lipson, and A. L. Gaeta, “Silicon-chip-based mid-infrared dual-comb spectroscopy,” Nat. Commun. 9(1), 1869 (2018). [CrossRef]   [PubMed]  

11. F. Adler, P. Masłowski, A. Foltynowicz, K. C. Cossel, T. C. Briles, I. Hartl, and J. Ye, “Mid-infrared Fourier transform spectroscopy with a broadband frequency comb,” Opt. Express 18(21), 21861–21872 (2010). [CrossRef]   [PubMed]  

12. A. Foltynowicz, P. Masłowski, A. J. Fleisher, B. J. Bjork, and J. Ye, “Cavity-enhanced optical frequency comb spectroscopy in the mid-infrared application to trace detection of hydrogen peroxide,” Appl. Phys. B 110(2), 163–175 (2013). [CrossRef]  

13. A. Khodabakhsh, V. Ramaiah-Badarla, L. Rutkowski, A. C. Johansson, K. F. Lee, J. Jiang, C. Mohr, M. E. Fermann, and A. Foltynowicz, “Fourier transform and Vernier spectroscopy using an optical frequency comb at 3–54 μm,” Opt. Lett. 41(11), 2541 (2016). [CrossRef]   [PubMed]  

14. L. Nugent-Glandorf, T. Neely, F. Adler, A. J. Fleisher, K. C. Cossel, B. Bjork, T. Dinneen, J. Ye, and S. A. Diddams, “Mid-infrared virtually imaged phased array spectrometer for rapid and broadband trace gas detection,” Opt. Lett. 37(15), 3285–3287 (2012). [CrossRef]   [PubMed]  

15. Y.-T. Su, Y.-H. Huang, H. A. Witek, and Y.-P. Lee, “Infrared absorption spectrum of the simplest Criegee intermediate CH2OO,” Science 340(6129), 174–176 (2013). [CrossRef]   [PubMed]  

16. J. F. Stanton, “On the vibronic level structure in the NO3 radical: II. Adiabatic calculation of the infrared spectrum,” Mol. Phys. 107(8–12), 1059–1075 (2009). [CrossRef]  

17. E. Hirota, “Vibrational assignment and vibronic interaction for NO3 in the ground electronic state,” J. Mol. Spectrosc. 310, 99–104 (2015). [CrossRef]  

18. M. V. Pinto Pereira Jr., T. P. Hoadley, M. C. Iranpour, M. N. Tran, and J. T. Stewart, “High-resolution infrared spectroscopy of the ν26 band of isoprene,” J. Mol. Spectrosc. 350, 37–42 (2018). [CrossRef]  

19. P. B. Changala, M. L. Weichman, K. F. Lee, M. E. Fermann, and J. Ye, “Quantum state resolution of the C60 fullerene,” in ArXiv e-prints (2018).

20. B. Spaun, P. B. Changala, D. Patterson, B. J. Bjork, O. H. Heckl, J. M. Doyle, and J. Ye, “Continuous probing of cold complex molecules with infrared frequency comb spectroscopy,” Nature 533(7604), 517–520 (2016). [CrossRef]   [PubMed]  

21. K. Iwakuni, G. Porat, T. Q. Bui, B. J. Bjork, S. B. Schoun, O. H. Heckl, M. E. Fermann, and J. Ye, “Phase-stabilized 100 mW frequency comb near 10 μm,” Appl. Phys. B 124(7), 128 (2018). [CrossRef]  

22. E. Hulthén and H. Neuhaus, “Diffraction gratings in immersion,” Nature 173(4401), 442–443 (1954). [CrossRef]  

23. T. Larsson and H. Neuhaus, “The immersion grating: practical considerations,” Opt. Acta (Lond.) 17(12), 915–923 (1970). [CrossRef]  

24. G. Wiedemann and D. E. Jennings, “Immersion grating for infrared astronomy,” Appl. Opt. 32(7), 1176–1178 (1993). [CrossRef]   [PubMed]  

25. U. U. Graf, D. T. Jaffe, E. J. Kim, J. H. Lacy, H. Ling, J. T. Moore, and G. Rebeiz, “Fabrication and evaluation of an etched infrared diffraction grating,” Appl. Opt. 33(1), 96–102 (1994). [CrossRef]   [PubMed]  

26. J. P. Marsh, D. J. Mar, and D. T. Jaffe, “Production and evaluation of silicon immersion gratings for infrared astronomy,” Appl. Opt. 46(17), 3400–3416 (2007). [CrossRef]   [PubMed]  

27. P. J. Kuzmenko, S. L. Little, Y. Ikeda, and N. Kobayashi, “Fabrication and testing of diamond-machined gratings in ZnSe, GaP, and bismuth germanate for the near-infrared and visible,” Proc. SPIE 7018, 70184Q (2008). [CrossRef]  

28. Y. Ikeda, N. Kobayashi, Y. Sarugaku, T. Sukegawa, S. Sugiyama, S. Kaji, K. Nakanishi, S. Kondo, C. Yasui, H. Kataza, T. Nakagawa, and H. Kawakita, “Machined immersion grating with theoretically predicted diffraction efficiency,” Appl. Opt. 54(16), 5193–5202 (2015). [CrossRef]   [PubMed]  

29. T. Sukegawa and Y. Okura, “Three types of immersion grating for next-generation infrared spectrometer,” Proc. SPIE 10100, 101000M (2017). [CrossRef]  

30. Q. Tang and L.-X. Guo, “Dispersion nonlinearity analysis of immersed gratings,” Appl. Opt. 57(7), 1548–1553 (2018). [CrossRef]   [PubMed]  

31. P. J. Kuzmenko, P. J. Davis, S. L. Little, L. M. Little, and J. V. Bixler, “High efficiency germanium immersion gratings,” Proc. SPIE 6273, 62733T (2006). [CrossRef]  

32. J. Rayner, T. Bond, M. Bonnet, D. Jaffe, G. Muller, and A. Tokunaga, “iSHELL: a 1-5 micron cross-dispersed R=70,000 immersion grating spectrograph for IRTF,” Proc. SPIE 8446, 84462C (2012). [CrossRef]  

33. G. Mace, H. Kim, D. T. Jaffe, C. Park, J.-J. Lee, K. Kaplan, Y. S. Yu, I.-S. Yuk, M.-Y. Chun, S. Pak, K.-M. Kim, J.-E. Lee, C. A. Sneden, M. Afsar, M. D. Pavel, H. Lee, H. Oh, U. Jeong, S. Park, B. Kidder, H.-I. Lee, H. A. N. Le, J. McLane, M. Gully-Santiago, J. S. Oh, S. Lee, N. Hwang, and B.-G. Park, “300 nights of science with IGRINS at McDonald Observatory,” Proc. SPIE 9908, 99080C (2016).

34. D. T. Jaffe, S. Barnes, C. Brooks, M. Gully-Santiago, S. Pak, C. Park, and I. Yuk, “GMTNIRS (Giant Magellan Telescope Near-Infrared Spectrograph): optimizing the design for maximum science productivity and minimum risk,” Proc. SPIE 9147, 914722 (2014). [CrossRef]  

35. S. Tani, T. Sukegawa, T. Nakamura, and Y. Kobayashi, “Extremely-high-order Ge immersion grating-based spectrometer for offset-free precision spectroscopy in the mid-infrared region,” in 2017 Conference on Lasers and Electro-Optics Europe & European Quantum Electronics Conference (CLEO/Europe-EQEC),(IEEE, 2017) [CrossRef]  

36. L. S. Rothman, I. E. Gordon, Y. Babikov, A. Barbe, D. C. Benner, P. F. Bernath, M. Birk, L. Bizzocchi, V. Boudon, L. R. Brown, A. Campargue, K. Chance, E. A. Cohen, L. H. Coudert, V. M. Devi, B. J. Drouin, A. Fayt, J. M. Flaud, R. R. Gamache, J. J. Harrison, J. M. Hartmann, C. Hill, J. T. Hodges, D. Jacquemart, A. Jolly, J. Lamouroux, R. J. Le Roy, G. Li, D. A. Long, O. M. Lyulin, C. J. Mackie, S. T. Massie, S. Mikhailenko, H. S. P. Muller, O. V. Naumenko, A. V. Nikitin, J. Orphal, V. Perevalov, A. Perrin, E. R. Polovtseva, C. Richard, M. A. H. Smith, E. Starikova, K. Sung, S. Tashkun, J. Tennyson, G. C. Toon, V. G. Tyuterev, and G. Wagner, “The HITRAN2012 molecular spectroscopic database,” J. Quant. Spectrosc. Radiat. Transf. 130, 4–50 (2013). [CrossRef]  

37. L. H. Coudert, P. Çarçabal, M. Chevalier, M. Broquier, M. Hepp, and M. Herman, “High-resolution analysis of the ν6, ν17, and ν21 bands of dimethyl ether,” J. Mol. Spectrosc. 212(2), 203–207 (2002). [CrossRef]  

38. F. Adler, M. J. Thorpe, K. C. Cossel, and J. Ye, “Cavity-enhanced direct frequency comb spectroscopy: technology and applications,” Annu. Rev. Anal. Chem. (Palo Alto, Calif.) 3(1), 175–205 (2010). [CrossRef]   [PubMed]  

39. J. Ye, L.-S. Ma, and J. L. Hall, “Ultrasensitive detections in atomic and molecular physics: demonstration in molecular overtone spectroscopy,” J. Opt. Soc. Am. B 15(1), 6–15 (1998). [CrossRef]  

40. G. Kowzan, D. Charczun, A. Cygan, R. S. Trawiński, D. Lisak, and P. Masłowski, “Optical Cavity Mode Measurements at Hz-Level Precision With a Comb-Based VIPA Spectrometer, ” in Conference on Lasers and Electro-Optics, OSA Technical Digest (Optical Society of America, 2018), paper STu3N.8. [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (4)

Fig. 1
Fig. 1 DFCS spectrometer. (a) A schematic of the Ge immersion grating and an overview of the immersion grating spectrometer. d is the lattice constant of the immersion grating. The OPO is an optical parametric oscillator operating at 8-10 μm. The diffracted light from the Ge immersion grating is cross-dispersed with a reflective grating and mapped onto a camera as a 2D image. (b) Camera image of the dispersed comb light. The comb modes are dispersed vertically by the immersion grating and horizontally by the cross-disperser grating.
Fig. 2
Fig. 2 Molecular spectroscopy. (a) Observed camera images (negative) of D2O and N2O absorption shown in bright dots. The marked absorption dots in the D2O image show the repeating pattern, indicating they are the same spectral feature. The vertical gap separated by the two horizontal red lines indicates the FSR of the immersion grating. (b) Observed spectrum of N2O, D2O, and dimethyl ether with the cell length of 10 cm. The frequency axis is calibrated with N2O and D2O spectra reported in the HITRAN database. The simulation is shown as inverted red traces. Absorption spectra of dimethyl ether were taken with three different sample pressures (red: 0.3 Torr; blue: 0.7 Torr; and black: 1.3 Torr).
Fig. 3
Fig. 3 Spectrometer resolution. (a)Pressure-dependent N2O linewidth measurements with the input beam diameters of 18 mm (red) and 34 mm (blue). (b) Voigt profile (green, solid) and Lorentzian profile (purple, dashed) fit to a hot band transition near 1145.518cm−1 for a 10 torr sample of pure N2O. The root-mean-squares (RMS) of the residual for the Voigt and Lorentzian line profile fits are 0.0027 and 0.0049, respectively. (c)Comb mode-resolved 2D image. One out of nineteen comb modes is filtered with a Fabry-Perot cavity (effective repetition rate is 2.09 GHz). Each comb mode is focused onto 4.5 pixels (90 μm) and displays a linewidth of about 740 MHz when the input beam diameter is 18 mm.
Fig. 4
Fig. 4 Spectrometer sensitivity. (a) The fractional noise as a function of power per pixel. The noise figure is much higher than the shot noise limit shown in the blue dotted line. When the incident power per pixel is lower than 5 nW, the camera noise is dominant. While above that power, intensity noise is dominant. (b) Fractional noise as a function of average time with the total input power of about 3 nW. δ I 3× 10 4 per detection channel is achieved with 1000 averages.

Equations (7)

Equations on this page are rendered with MathJax. Learn more.

RP=λ/ Δλ = (2nwtanθ)/λ ,
I= I 0 exp( αL ) I 0 ( 1αL )for αL1.
αL ( I 0 I )/ I 0 ΔI/ I 0 .
g= η C tot P tot t int E photon ,
a= E photon t int g η .
( ΔI/ I 0 ) min shot = 2eB/ Rp ,
δ I ( AΔI ) 2 + B 2 ΔI+ C 2 / I 0 .
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.