Determination of reduced scattering coefficient of biological tissue from a needle-like probe

Open Access Open Access

Abstract

Detection of interactions between light and tissue can be used to characterize the optical properties of the tissue. The purpose of this paper is to develop an algorithm that determines the reduced scattering coefficient (µs) of tissues from a single optical reflectance spectrum measured with a small source-detector separation. A qualitative relationship between µs and optical reflectance was developed using both Monte Carlo simulations and empirical tissue calibrations for each of two fiber optic probes with 400-µm and 100-µm fibers. Optical reflectance measurements, using a standard frequency-domain oximeter, were performed to validate the calculated µs values. The algorithm was useful for determining µs values of in vivo human fingers and rat brain tissues.

©2005 Optical Society of America

1. Introduction

In recent years, optical reflectance spectroscopy has been used to extract physiological information from human tissues [13] and organs [4]. The fiber optic methods provide a low risk, low cost, and minimally invasive way to investigate characteristics of human tissues and structures. Numerous studies have shown that the reduced scattering coefficient (µs) can be obtained using an integrating sphere in combination with a curve-fitting procedure between experimental data and Monte Carlo simulations [57]. While this technique is effective, it requires ex vivo tissues for the integrating sphere and increases computational time due to complex curve fitting. In ex vivo measurements, it is uncertain whether the optical properties of tissues are significantly affected by such factors as 1) lack of blood perfusion after excision, 2) length of time between excision and measurement, 3) the use of saline for re-hydration purposes, and 4) placement of the sample between two pieces of glass. Thus, direct in vivo measurements of light scattering from living tissues are desirable but impossible with the integrating sphere approach. It is highly desirable to develop an algorithm with the ability to quantify the optical scattering property based on optical reflectance measured from in vivo tissues in real time. In recent years, researchers have made significant efforts in developing methods to measure the optical properties of tissue in vivo [3, 5, 810].

In recent studies, we have developed a needle-like, near infrared (NIR) probe with a small source-detector separation (100–400 µm) to identify gray and white matter structures in the living human and animal brain [2, 11, 12]. The NIR probes had an outer diameter of 1.3 mm used in the human study [11] and of 0.45 mm for the animal study [12]. Both kinds of probes contained two 100-µm fibers to deliver light to and collect light from the brain tissue. We have shown that the optical reflectance spectra exhibited significant differences between gray and white matter structures [2, 11, 12]. The goal of the present paper is to develop an algorithm that allows us to extract the light scattering properties (µs’) of brain tissues in vivo from a single reflectance measurement with a needle-like probe. This is accomplished by developing a qualitative relationship between µs’ and optical reflectance using both Monte Carlo simulations and empirical calibrations. Then, the developed algorithm is further validated using a standard frequency-domain oximeter and applied for determining µs’ values of in vivo human fingers and rat brain tissues.

2. Monte Carlo simulations

With the definition of optical reflectance being the number of photons per unit area that are reflected back from a measured sample, a single, optical reflectance spectrum contains a large amount of information about the reflecting medium. Currently, few publications have presented an analytical expression that directly associates µs’ with the reflectance measurements [13, 14]. In our study, to develop such a relationship, we utilized a well tested [15, 16] and publicly available [17] Monte Carlo simulation software to simulate actual reflectance measurements in tissues at an appropriate source-detector separation with a set of input parameters. Such parameters include g=0.9 and n=1.38 [18, 19], where g is scattering anisotropy, and n is the index of refraction. In common with our previous work [2], we used an infinitely large slab with only one bulk layer, and the tissue layer thickness (d) was chosen to be 10 cm. The simulations were performed with a spatial step size of 0.05 mm (or 50 µm) and a maximum detection radius of 1.5 mm (or 1500 µm). During the simulations, µs’ was varied from 1–60 cm-1 for each of the four absorption coefficients: µa=0.01, 0.1, 0.25 and 0.5 cm-1. These absorption and scattering values were selected based on the values that may be commonly found in tissues [20, 21]. The Monte Carlo output results were used to develop a relationship between the simulated optical reflectance and µs’ for each of the four absorption coefficients. Since the output of Monte Carlo simulations for reflectance depends on the product of µs and (1-g), rather than their individual values [15, 16, 22], a fixed value of g=0.9 in the simulation still permits us to simulate a broad range of tissue scattering properties by combining different g and µs values.

Figure 1 shown below plots four sets of simulated reflectance (Rsim) versus µs’ at µa=0.01, 0.1, 0.25, 0.5 cm-1, with a source-detector separation of 400 µm. This separation matches the actual fiber separation in one of our needle probes used in the in vivo and in vitro experiments, and this probe is referred to as the 400-µm probe throughout the paper. On close inspection of this figure, we see that the reflectance values seem a little lower for higher values of absorption, as expected.

In general, a student t-test is a fundamental statistical process to compare the two means of two groups of data. Normally, if p is less than 0.05 (p <0.05), the two means of the two groups are significantly different; if p>0.05, the two means have no significant differences. After such statistical analysis, we learned that the differences among the reflectance with µa values from 0.01 cm-1 to 0.5 cm-1 are insignificant (p>0.8). This permits us to pool all the data together and to obtain a linear relationship between the simulated reflectance, Rsim, and µs’ as

Rsim(λ0)=0.404μs(λ0)+0.582.

Specifically, the four sets of data were averaged at each µs’ value, and the mean values of reflectance were used to generate Eq. (1), with a correlation coefficient of R=0.99. It is known that the µs’ values of biological tissues are weakly wavelength dependent, and this dependence is sufficiently weak over the near infrared (NIR) range. Thus, we may use Eq. (1) for wavelength values (λ 0) within 650 nm to 900 nm. Notice that a simple linear fitting is a desirable starting approach, while the data given in Fig. 1 may be fitted well with a 2nd order polynomial equation. Error analysis by the end of the algorithm development will justify if the linear approach is reasonable, or a higher order approach is needed.

 figure: Fig. 1.

Fig. 1. Simulated relationship between reflectance and the reduced scattering coefficient at varying µa values of 0.01 cm-1 (solid diamond), 0.1 cm-1 (open square), 0.25 cm-1 (solid triangle), and 0.5 cm-1 (cross), with the source-detector separation fixed at 400 microns. The unit for reflectance is the number of photons/cm2.

Download Full Size | PDF

To associate the simulated reflectance, Rsim(λ 0), and the measured reflectance, Rm(λ 0), we assumed a simple proportion between them as

Rm400(λ0)=a0400Rsim400(λ0)=a0400[0.404μs(λ0)+0.582],

where a 0–400 is an overall intensity factor for the 400-µm probe. The a 0 factor is used to relate µs’ values and the measured Rm values, as shown in Eq. (2). The experimental set-up, fiber geometry, and the diameter of the internal fibers each contribute to the a 0 term. While Monte Carlo simulations do provide a reasonable estimate of light behavior through a turbid medium, they cannot account for all influences from the various system components that contribute to the actual measured signal. Therefore, the a 0 term is determined using an empirical calibration method that is to be described in Section 3.

An optical needle probe with a source-detector separation of 100 µm was also used for both in vivo and in vitro measurements. A similar Monte Carlo simulation was performed to obtain the relationship between Rsim(λ 0) and µs’(λ 0) for the 100 µm probe, which will be referred to as the 100-µm probe throughout the paper. Using the similar methodology to that given above, the simulated data with four different µa values were averaged at each µs’ value for the 100-µm probe (figure not shown), and a linear relationship between R m-100(λ 0) and µs’(λ 0) is obtained as follows:

Rm100(λ0)=a0100Rsim100(λ0)=a0100[1.670μs(λ0)1.544].

With a chosen probe (either the 400-µm or the 100-µm), quantification of a0 becomes the key issue for final determination of µs’ from the measured Rm, as seen in Eqs. (2) and (3). An experimental approach was taken to empirically quantify a 0 for each of the two probes.

3. Algorithms to determine a0 and µs

Measurements of µs’ were obtained from a diluted solution of Intralipid (20% Intralipid, Pharmacia and Up-John, Clayton, NC) using a dual-channel, frequency-domain oximeter (ISS, Champaign, IL) [23], which will be described more in Section 4.1. Simultaneously, the 400-micron fiber optic probe was used with a portable CCD spectrometer (SD2000, Ocean Optics, FL) to obtain the reflectance at two selected wavelengths, 750 nm and 830 nm. The Intralipid solution was serially diluted to include Intralipid concentrations between 0.5% and 8% to cover a broad range of light scattering properties. Additional details for the experimental setup are given in Section 4.

3.1 Algorithms to determine a0 and µs’ for the 400-µm probe

Using Eq. (2) to solve for a 0–400, we obtain

a0400=Rm400(λ0)0.404*μs(λ0)+0.5819.

Based on the measured values of R m-400 from the portable spectrometer and µs’ from the ISS oximeter, individual a 0–400 values for each experiment can be obtained using equation (4). Figure 2(a) shows the relationship between the calculated a 0–400 values and corresponding measured R m-400 obtained from 5 individual experiments at 750 nm and 830 nm. The 5 sets of data were pooled here for the Intralipid concentrations between 0.5 and 4%.

 figure: Fig. 2. (a).

Fig. 2. (a). Overall intensity factor, a0–400, versus measured reflectance, R m-400, at 750 nm (open blue circles) and at 830 nm (filled red circles) for the 400-µm probe. An average a 0–400 value is shown as the solid green line while the solid black curve is a quadratic fit for the data.

Download Full Size | PDF

Our statistical analysis for Fig. 2(a) shows that the mean a 0–400 value at 750 nm (0.066±0.01) is not statistically different from the mean a 0–400 value at 830 nm (0.065±0.01; p>0.5). Thus, pooling the data from 750 nm and 830 nm allows the fitted equation to represent the data within the spectral range between 750 nm to 830 nm rather than fixed at a specific wavelength. Ideally, a 0–400 is expected to be constant and independent of R m-400; however, Figure 2(a) shows that a 0–400 varies slightly with Rm-400. A quadratic fitting between a 0–400 and R m-400 can be obtained in Eq. (5) with a correlation coefficient of R=0.96:

a0400(λ0)=0.0458Rm400(λ0)2+0.0808Rm400(λ0)+0.0407,

where λ 0=750 nm-830 nm.

Figure 2(a) also shows a mean a 0–400 of 0.065±0.01. This was calculated by taking a mean of all data points (at both 750 nm and 830 nm) given in Fig. 2(a). Detailed results will be given in Section 5 to compare the calculated µs’ values using both the polynomial a0–400 and the constant a 0–400 value.

Furthermore, substituting the constant a 0–400 or Eq. (5) into Eq. (2) and solving for µs’ lead to empirically derived relationships between the µs’ and the measured reflectance, R m-400, for the 400-µm probe, as given below:

μs(λ0)=Rm400(λ0)0.5819a0400(λ0)0.404a0400(λ0),
μs(λ0)=Rm400(λ0)0.5819[0.065]0.404[0.065],
μs(λ0)=Rm400(λ0)0.5819[0.0458Rm400(λ0)2+0.0808Rm400(λ0)+0.0407]0.404[0.0458Rm400(λ0)2+0.0808Rm400(λ0)+0.0407],

where Eqs. (6b) and (6c) are for the constant a 0–400 and quadratic a 0–400, respectively.

3.2 Algorithms to determine a0 and µs’ for the 100-µm probe

 figure: Fig. 2. (b).

Fig. 2. (b). Overall intensity factor, a 0–100, versus measured reflectance, R m-100, at 750 nm (open blue circles) and 830 nm (filled red triangles) for the 100-µm probe. An average a 0–400 value is shown as the solid green line, while the solid black curve is a quadratic fit for the data.

Download Full Size | PDF

Similarly, an expression for a 0–100 was developed for the 100-µm probe. Figure 2(b) shows the relationship between the a 0–100 values and corresponding measured reflectance, R m-100, obtained from 8 individual experiments at 750 nm and 830 nm. The 8 sets of data were pooled, and the a 0–100 and reflectance values were averaged at each Intralipid concentration, ranging between 1–8% in this set of experiments. Statistical analysis shows that the mean a 0–100 value at 750 nm (0.0034±0.0005) is not statistically different from the mean a 0–100 value at 830 nm (0.0034±0.0006; p>0.1).

A quadratic fitting between a 0–100 and R m-100 can be also obtained in Eq. (7) for the 100-µm probe. Figure 2(b)shows a constant value of a 0–100, i.e., 0.0034±0.0005, which was obtained by finding an average a0–100 of all the data given in Figure 2(b). In Section 5, detailed comparison will demonstrate that the constant a 0–100 gives better µs’ determination for the 100-µm probe.

a0100(λ0)=0.0126Rm100(λ0)2+0.0074Rm100(λ0)+0.0027.

Using the approach similar to that in obtaining Eq. (6), we substitute the constant a 0–100 and the quadratic a 0–100 (i.e., Eq. (7)) into Eq. (2), respectively, and solve for µs’. Then, we arrive at empirically derived relationships between µs’ and R m-100 for the 100-µm probe:

μs(λ0)=Rm100(λ0)+1.5437a0100(λ0)1.6696a0100(λ0),
μs(λ0)=Rm100(λ0)+1.5437[0.0034]1.6696[0.0034],
μs(λ0)=Rm100(λ0)+1.54370.0126Rm100(λ0)2+0.0074Rm100(λ0)+0.00271.6696[0.0126Rm100(λ0)2+0.0074Rm100(λ0)+0.0027],

where Eqs. (8b) and (8c) are for the constant a 0–100 and quadratic a 0–100, respectively.

4. Methodology

4.1 Instrumentation

The experimental set-up for Rm measurements consisted of a tungsten-halogen light source (LS-1, Ocean Optics, Inc., Dunedin, FL), bifurcated fiber optic probe, data acquisition card (DAQ-700), spectrometer (SD2000, Ocean Optics, Inc., Dunedin, FL) with wavelength grating from 350–950 nm and a laptop computer [2, 11]. The fiber optic probe contained two 400-µm (or 100-µm) diameter fibers for light delivery and light collection, respectively, with an outer diameter of 1.3 mm (or 0.4 mm). The center-to-center distance between the source and detector fibers was approximately 400 µm (or 100 µm) for the 400-µm (or 100-µm) probe. A cross-section drawing for the 400-µm probe is shown in Fig. 3(a). The reflected signal passed through the optical fiber, dispersed and detected by the CCD spectrometer, and the electrical signal for the spectrometer output was sent to the computer [22, 24]. LabView interface software (National Instruments, Austin, TX) was used between the spectrometer and laptop to display the optical reflectance in real-time.

All effects from the fiber, light source and spectrometer have been removed by dividing each optical reflectance signal by a calibration curve taken from a standard reflectance sample [25, 26]. The standard sample surface provides approximately 100% reflectance in the wavelength range of 250–2000 nm (Diffuse Reflectance Standard Specifications, Ocean Optics, FL). The calibration reflectance was obtained by placing the probes on top of the standard sample and recording the reflected signals. Moreover, it is practically important to note that in spite of turning off all room lights and covering the sample container, a small level of background due to thermal noise of the CCD array was still present. By recording this background signal prior to the actual measurement and subtracting it from the raw data, we could obtain improved a 0 values.

 figure: Fig. 3.

Fig. 3. (a). The schematic cross section of the 400-µm fiber probe. 3(b). Experimental setup for the in vivo reflectance measurements of the human middle finger. The particular probe shown above is just for the demonstration purpose and equivalent to the 400-µm probe, which was used for the determination of µs’ values of human fingers. Also, the broadband light source and CCD spectrometer are shown, as labeled.

Download Full Size | PDF

The dual-channel ISS oximeter (ISS, Champaign, IL) was the “gold standard” instrument and used to measure the reduced scattering coefficient from each sample solution. More details on the ISS oximeter can be found through ref. [23]. Basically, the device works by emitting NIR light into tissue at known distances from a collector. Light of two different wavelengths is used, and the light is modulated at an RF (radio frequency) of 110 MHz. The collected light is measured and processed, and the absorption and reduced scattering coefficients of the medium are determined. This technology allows the absolute measurement of light absorption and scattering in a highly scattering medium, such as human tissue. In our study, the two channels at 750 nm and 830 nm were used.

To compare the calculated µs’ values determined from the reflectance probes with the ISS ‘gold standard” readings, we used relative error calculations to express the deviation between them as follows:

Relativeerrorinμs=μs(probe)μs(ISS)μs(ISS)100%

4.2 In vitro measurements using Intralipid

In vitro measurements were performed using a diluted solution of Intralipid to simultaneously measure the reduced scattering coefficient and optical reflectance. An Intralipid solution was chosen since it closely simulated light scattering properties of tissue with an estimated g value >0.9 and a mean particle size of 0.5 µm. A rectangular, plastic container with the volume of 8 cm×10 cm×14 cm (height x width x length) was used to hold the testing Intralipid solution. The probes were inserted from the top into the solution during the measurement, so there was no intralipid-container or intralipid-air interfaces for the probe. The probe tip was facing vertically down towards the bottom of the container, a few centimeters below the solution surface, and a few centimeters away from the walls of the container.

All measurements were performed with an integration time of 40 msec for the 400-µm probe, and 100 msec for the 100-µm probe. The Intralipid stock solution was 20% by volume (Pharmacia and Up-John, Clayton, NC), and the diluted Intralipid solutions ranged in concentration from 0.5%–8%, depending on the optical probe being used. Five spectral readings per location were taken during the Intralipid measurements for the 400-µm probe. In general, the reduced scattering coefficients, µs’, found in tissues vary between 5 and 25 cm-1; however, larger values of µs’ have been reported in brain tissues [21, 22]. After the reflectance spectra were taken at different Intralipid concentrations, Eqs. (6b) and (6c) were used to calculate µs’ values at 750 nm and 830 nm, and they were compared with the expected µs’ values obtained from the ISS system. Similar procedures were used for the 100-µm probe, except that multiple locations (3 times) per Intralipid solution were taken for the measurement to decrease noise due to small source-detector separation (see more details in Section 5).

4.3 In vivo measurements from human fingers

To obtain the µs’ values of human fingers in vivo, the 400-µm probe was placed against the surface of the skin of a human (index or middle) finger with the tip of the probe lightly touching the skin, as shown in Fig. 3(b), without any compression. Uncompressed touch is to keep local hemoglobin oxygen levels unperturbed. The hand was rested on the table comfortably while the bifurcated fiber probe was facing down in contact with the finger to collect the in vivo data. The optical reflectance readings took place less than 100 ms.

4.4 In vivo measurements from living rat brain tissues

To obtain the µs’ values of living animal brain tissues in vivo, experiments were performed using male Sprague-Dawley albino rats. The rats were anesthetized using Nembutal and were mounted in a stereotaxic device. Lidocaine was used to locally anesthetize pressure points near the ears and snout that were in contact with the stereotaxic frame. Since the ventilation, perfusion, and blood oxygenation of the animals were normally not alternated much during the optical measurements, those parameters were not monitored. The experiments were performed in accordance with the guidelines for the ethical use of animals established by the Society for Neuroscience. The University of Texas Southwestern Medical School’s Institutional Animal Care and Use Committee approved the experiments. Following an incision in the scalp, a drill was used to bore a small hole in the skull. Through this hole passed the 100-µm probe. The NIR probe was advanced from the dorsal surface of the cerebral cortex, down to a depth of 4 to 8 mm in steps of 200 µm. It was expected that the probe would traverse the cortex (CTx), enter the corpus callosum, extend down into the caudate putamen (CPu), and end above the anterior commissure (ac) [11, 22]. The location of gray and white matter was estimated by referring to rat stereotaxic atlas [27].

5. Results

5.1 In vitro — Intralipid data

With different concentrations of Intralipid solutions, we obtained readings of µs’ and R m-400 using both the ISS oximeter and optical reflectance, respectively (Fig. 4). Both µs’ and R m-400 are linearly proportional to Intralipid concentration (%) with a correlation coefficient of 0.971 and 0.995, respectively. Each data point was obtained by averaging the data (750 and 830 nm) from five different measurements at each concentration. A similar graph was also obtained (data not shown) for the 100-µm probe.

Figure 4 shows the µs’ values, as labeled through the vertical right axis, that were obtained directly from the ISS oximeter readings. The vertical left axis shows the values of R m-400, which can be converted to µs’ (cal) using Eqs. (6b) and (6c). We employed Eq. (9) to quantify the errors of µs’ that were calculated using the reflectance readings. Figure 5 presents the error comparison between the expected µs’ (ISS) and calculated µs’ (cal) values, using the constant a 0–400 (=0.065±0.01) and the polynomial a 0–400, i.e., Eq. (5). This figure demonstrates that the relative errors of calculated µs’ range up to 35% with a mean ± standard deviation of 16±11%, when the constant a 0–400 is used. The relative errors are reduced to less than 10% when using the polynomial a 0–400 (mean ± standard deviation: 3.5±3.2%). The range of µs’ values used for this error analysis is from 5.7 cm-1 to 26.7 cm-1, which are the common µs’ values well known for biological tissues. Thus, this error comparison serves as validation and justification for the polynomial a 0–400, or Eq. (6c), to be used in the future for µs’ quantification from a reflectance measurement with the 400-µm probe.

 figure: Fig. 4.

Fig. 4. Linear relationships between 1) the Intralipid concentration and the reflectance (red circles) and 2) the Intralipid concentration and the µs’ values (blue solid squares) obtained from the ISS oximeter. The data are fitted with linear relationships for the reflectance (red line) and the µs’ (blue line), respectively, for the 400-µm probe. Specifically, the linear relationships are R m-400=0.310×[Intralipid concentration]+0.009 and µ s’=10.094× [Intralipid concentration] + 0.433 in cm-1.

Download Full Size | PDF

 figure: Fig. 5.

Fig. 5. Error comparison between expected µs’ (ISS) and calculated µs’ (cal) values, using the constant a 0–400 (=0.065±0.01) (filled red squares) and the polynomial a 0–400, i.e., Eq. (5) (open blue circles). All of the data points were based on five readings per location per Intralipid concentration. The data at 750 nm were used for this comparison.

Download Full Size | PDF

To investigate the accuracy of ignoring absorption effects, as mentioned in Section 2, we conducted the experiments without and with an ink solution mixed in the Intralipid solution so as to induce light absorption, using the 400-µm probe. The results are shown in Fig. 6, where the µa values are 0.04 cm-1 and 0.4 cm-1 for the two respective cases. Similar to the Monte Carlo results seen in Fig. 1, the reflectance values with the ink solution (i.e., µa=0.4 cm-1) are little lower than those without ink. However, the error calculation shows that the relative deviations between the two cases are about or less than 10%. Such an error range supports our approach that when a needle-like probe is utilized, light absorption effect is insignificant and can be reasonably ignored without losing too much accuracy.

 figure: Fig. 6.

Fig. 6. Linear relationships between the reduced light scattering coefficient (µs’) and the reflectance measured without ink (red diamonds) and with ink (blue circles). The absorption coefficients of Intralipid solutions without ink and with ink are 0.04 and 0.4 cm-1, respectively. The experiment was taken with the 400-µm probe, and the different values of µs’ were obtained by varying the Intralipid concentration. The data at 750 nm were used for this comparison.

Download Full Size | PDF

Additional experiments were performed using the 100-µm probe in order to validate Eqs. (8b) or (8c). In this case, the Intralipid concentrations for the experiment ranged from 1% to 8%. For each Intralipid concentration, three spectra were averaged at each probe location to collect one spectrum, and four measurements were repeated at different random locations within the solution volume for each of the concentrations. Thus, there were 12 spectra collected and averaged for the calculated µs’. Then, we performed the comparison between the expected and calculated µs’ values with the constant a 0–100 and polynomial a 0–100, i.e., using Eqs. (8b) and (8c). The range of µs’ values used in this case is from 9.7 cm-1 to 76.9 cm-1, covering a broad range of µs’ values known for biological tissues.

 figure: Fig, 7(a).

Fig, 7(a). Error comparison between expected µs’ (ISS) and calculated µs’ (cal) values, using the constant a 0–100 (=0.0034±0.0005) (shown as filled red squares) and the polynomial a 0–100, i.e., equation (7) (shown as open blue circles). The horizontal dashed lines are the mean values of the open-circle and filled-square data points, respectively. To decrease noise due to small a source-detector separation, 4 locations per Intralipid solution were taken for the measurement, and 3 readings per location were used. The data at 750 nm were used for this comparison.

Download Full Size | PDF

The data given in Fig. 7(a) illustrate that the relative errors of calculated µs’ vary below 10% (mean ± standard deviation: 2.7±2.6 %) when the constant a 0–100 is applied, whereas the relative errors increase above 10% (mean ± standard deviation: 6.6±3.0 %) when using the polynomial a 0–100. It is clearly shown that the deviation with the polynomial a 0–100 was more than that with the constant a 0–100. This analysis suggests that the constant a 0–100, or Eq. (8b), is the suitable approach to use for µs’ quantification with the 100-µm probe in the future.

To further demonstrate the necessity of multiple readings for the 100-µm probe [28], additional comparison was plotted to examine the developed algorithms, Eqs. (8b) and (8c). The experimental results shown in Fig. 7(a) are replotted in Fig. 7(b) together along the data taken with a single reading at a single location. Figure 7(b) clearly reveals that when R m-100 were calculated using data taken at multiple locations with an average of three spectra per location, the relative errors have been greatly reduced in comparison to that obtained with a single measurement at a single location. This conclusion holds for either the constant a 0–100 or polynomial a 0–100.

 figure: Fig, 7. (b).

Fig, 7. (b). Error comparison in µs’ when using a single location measurement (constant a 0–100: filled red squares with the mean value plotted by solid red line; polynomial a 0–100: filled blue circles with the mean value by solid blue line) versus the average of three measurements per location (constant a 0–100: open red squares with the mean value by dashed red line; polynomial a 0–100: open blue circles with the mean value by dashed blue line). The data at 750 nm were used for this comparison.

Download Full Size | PDF

In Section 5.1, we have experimentally tested and validated the newly developed algorithms for both the 400-µm and 100-µm probes, which can be now utilized to derive µs’ values of tissues in vivo. In Sections 5.2 and 5.3 next, we will provide examples to employ the new algorithms for quantification of tissue µs’ values.

5.2 In vivo — human data

The calculated µs’ values of the human finger tissues under the uncompressed state, using Eq. (6c), were averaged over 5 measurements at different but adjacent locations from each of 3 human subjects with the 400-µm probe. The µs’ values at 750 nm are 8.6±2.0 cm-1 (mean ± standard deviation), being consistent well with those of in vivo human breast tissue at 750 nm reported by Cerussi et al. (8.8 cm-1) and Durduran et al. (8.3±2.0 cm-1) [29, 30].

5.3 In vivo — rat brain data

Next, we applied Eq. (6c) to the reflectance data measured from the surface of a rat brain to ~6.5 mm deep in the brain, using the 400-µm probe. The NIR probe was advanced from the dorsal surface of the cerebral cortex. Figure 8 shows that the µs’ values vary between 10 and 35 cm-1 at 750 nm, depending upon the particular type of brain tissue being interrogated. According to the corresponding histology, the tissues at the first 1 to 2 mm near the brain surface are gray matter (made of nerve cell bodies and nerve terminals), and the tissue near the peak of µs’ at ~3.6 mm is located in white matter (made of myelinated and unmyelinated axons) [11]. Indeed, the histology has proven that the probe path traverses the cortex (CTx), enters the corpus callosum, extends down into the caudate putamen (CPu), and ends above the anterior commissure (ac) [11, 28].

 figure: Fig. 8.

Fig. 8. Calculated reduced scattering coefficients, µs’, from living rat brain tissues at 750 nm; the data were obtained using the 400-µm probe.

Download Full Size | PDF

Another measurement and calculation were repeated from a second rat using the 100-µm probe, and the data is shown in Fig. 9. This figure shows clearly that the µs’ values of the rat brain gradually increase from 15 to 61 cm-1 at 750 nm as the probe passes through from gray matter to white matter. Notice that there exists a dip in µs’ at around 5.4 mm and 4.5 mm in Figs. 8 and 9, respectively. This decrease in µs’ has proven to result from a piece of intracranial gray matter underlying the white matter [11]. Both of the figures taken from animal experiments in vivo strongly illustrate that the µs’ values of white matter in living animal brains can be quite large, ranging from 35 cm-1 to 60 cm-1.

 figure: Fig. 9.

Fig. 9. Calculated µs’ values from living rat brain tissues at 750 nm; the data were obtained using the 100-µm probe.

Download Full Size | PDF

6. Discussions and conclusions

By using Monte Carlo simulations and experimental methods, we were able to empirically develop analytical expressions to extract µs’ values from a single optical reflectance measurement. The key factor in deriving the expressions is to quantify the overall intensity factor, a 0, which depends strongly on the fiber probe geometry. By calibrating a 0 for both the 400-µm and 100-µm probes using the ISS oximeter, we arrived at Eqs. (6) and (8) as the appropriate expressions for µs’ quantification. Such an algorithm allows µs’ values of various tissues to be measured in vivo in real-time, which is useful during surgeries or “optical biopsies” where instant information is important. However, one should be cautious to employ them when the reduced light scattering coefficient becomes significantly low. Errors may occur when µs’ is near or smaller than 2 cm-1, which is not often seen within biological tissues. It could be possible to re-derive other empirical equations, using either linear (with a zero intercept) or higher order forms.

Several experiments with different concentrations of Intralipid were performed to gather additional data in order to validate the algorithms. The data at 750 nm were used to determine if there was a difference in µs’ calculations using a constant a 0 versus a polynomial a 0 for both of the probes. Statistical analysis showed that the polynomial a 0–400 produced improved µs’ values for the 400-µm probe, while the constant a 0–100 produced improved µs’ for the 100-µm probe.

In the calibration study, we learned that the µs’ data obtained at 750 nm and 830 nm can be pooled together to generate Eqs. (6) and (8). Given the fact that light scattering of biological tissues in the NIR range does not strongly depend on wavelength [3, 4], we suggest that Eqs. (6) and (8) can be approximately valid in the wavelength range of 700 nm to 850 nm without causing significant errors.

Why can relatively simple algorithms, such as Eqs. (6) and (8), be developed to determine µs’ using reflectance data? Our answer is that the absorption coefficient, µa, of the tissue has much less contribution to light reflectance when the light source and detector are nearby, in a few hundred microns. The Monte Carlo simulations (Fig. 1) and our experimental data (Fig. 6) clearly demonstrate that the reflectance readings did not depend on µa strongly when µa values ranged from 0.01 cm-1 to 0.4 cm-1(0.5 cm-1), which are the typical µa values seen in biological tissues. In the NIR region, µa is normally much smaller than µs’, and thus the absorption length (=1/µa) is much longer than the reduced scattering length (1/µs’). In the case of needle-like probes with a source-detector separation of a few hundred microns, the absorption length is much larger than the separation so that the effect of µa can be reasonably ignored without causing a significant error. Thus, the algorithms developed in this study have the theoretical foundation and can be applicable only in the NIR range.

It is noted that the data taken with the 100-µm probe are noisier than those with the 400-µm probe [28]. We hypothesize that the noisy feature is possibly attributed to the dimension of source-detector separation of the 100-µm probe. In many cases of Intralipid measurements, the separation was comparable to the single scattering length of the Intralipid solution, so the statistical noise became severe. Namely, we suggest that Brownian motion of the lipid particles during the time course of a measurement may contribute to the higher noise in the data obtained with the 100-µm probe. Thus, it is more important to average several measurements within the sample solution when the 100-µm probe is used. We expect that this kind of noise should decrease in solid tissues because of limited random motion of scatterers. Indeed, we observed much less deviation in measurements of biological tissues during the course of our human and animal studies [11, 12, 22]. Further experimental confirmation for this hypothesis can be performed by lowering the temperature of the solutions or adding an agent that would increase its viscosity.

The calculated µs’ values from the living rat brain tissues ranged from 10–61 cm-1 in our measurement. As mentioned in Section 5, the rat brain data reveal that the µs’ values of white matter are relatively high, ranging 35–60 cm-1 in the two animal brains reported in this paper. We have consistently observed similar µs’ values of white matter in other animal experiments (more than 8 rat brains) [22]. Indeed, to the authors’ knowledge, this is the first report to provide the µs’ values of living white matter in vivo, and the first study to show good consistency between the µs’ values taken in vivo in our study and those measured in vitro, as reported by other groups [21, 31, 32].

The results obtained from living rat brain tissue (using both probes) are consistent with those published by Cheong et al. [20], where the reduced scattering coefficients were reported to vary from 6.6 cm-1 in a calf brain at 633 nm to 57 cm-1 in a pig brain at 633 nm. It is highly possible that the low µs’ value resulted from a piece of gray matter, and the high µs’ value came from a piece of white matter. Furthermore, it is important to note that the published results were obtained all from post-mortem tissues, whereas our measured data are all from in vivo living animal brains.

Shangguan et al. conducted experiments to investigate changes in optical properties of elastin lamina, from porcine aorta, as a function of pressure [33]. Their work showed that the µs’ value increased from 53.7 to 83.1 cm-1 (by 57%) when pressure (1 kg/cm2) was applied to the tissue. Even though this work was conducted on ex vivo tissue, it indicates that as tissue is compressed, the reduced scattering coefficient increases. Since the fiber probe applies a force to the brain tissue as it advances through the rat brain, the pressure might cause an increase in µs’ within a certain degree. However, since the pressure applied at the probe tip should be much less than 1 kg/cm2, we do not expect a large increase in µs’ caused by pressure.

The potential utility of µs’ measurement of tissues is to study the origin of light scattering, leading to the determination of cellular or sub-cellular structures at the local measurement site [3,5]. This is the direction of our future study. Such morphological information, made available in the operating room, may assist the identification of important targets for brain surgery.

In summary, this paper shows that an algorithm can be empirically developed to directly calculate the reduced scattering coefficient from a single optical reflectance spectrum. However, it is important to note that the a 0 term is fiber dependent. Individual calibration procedures and experiments are needed to obtain the a 0 term for individual fiber optic probes, which contain fibers with different sizes and/or different configurations. This study demonstrates a feasible methodology to use a needle-like probe to quantify µs’ values of various living tissues in vivo in real time, which may be practically useful for a variety of biomedical applications.

Acknowledgments

This work was supported in part by the National Institutes of Health 1-R01-NS40874-01 and 1-R21-CA101098-01. The authors are also grateful to Mr. Ali N. Bahadur for his assistance of collecting certain Intralipid data.

References and Links

1. H.R. Eggert and V. Blazek, “Optical Properties of Human Brain Tissue, Meninges, and Brain Tumors in the Spectral Range of 200 to 900 nm,” Neurosurg. 21, 459–464 (1987). [CrossRef]  

2. M. Johns, C.A. Giller, and H. Liu, “Computational and In Vivo Investigation of Optical Reflectance from Human Brain to Assist Neurosurgery,” J. Biomed. Opt. 3, 437–445 (1998). [CrossRef]  

3. G. Zonios, L.T. Perelman, V. Backman, R. Manoharan, M. Fitzmaurice, J. Van Dam, and M.S. Feld, “Diffuse Reflectance Spectroscopy of Human Adenomatous Colon Polyps In Vivo,” Appl. Opt. 38, 6628–6637 (1999). [CrossRef]  

4. J.R. Mourant, I.J. Bigio, J. Boyer, T.M. Johnson, J. Lacey, A.G. Bohorhoush, and M. Mellow, “Elastic Scattering Spectroscopy as a Diagnostic Tool for Differentiating Pathologies in the Gastrointestinal Tract: Preliminary Testing,” J. Biomed. Opt. 1, 192–199 (1996). [CrossRef]  

5. A.M.K. Nilsson, C. Sturesson, D.L. Liu, and S. Andersson-Engels, “Changes in spectral shape of tissue optical properties in conjunction with laser-induced thermotherapy,” Appl. Opt. 37, 1256–1267 (1998). [CrossRef]  

6. J.S. Dam, T. Dalgaard, P.E. Fabricius, and S. Andersson-Engels, “Multiple polynomial regression method for determination of biomedical optical properties from integrating sphere measurements,” Appl. Opt. 39, 1202–1209 (2000). [CrossRef]  

7. A.M.K. Nilsson, R. Berg, and S. Andersson-Engels, “Measurements of the optical properties of tissue in conjunction with photodynamic therapy,” Appl. Opt. 34, 4609–4619 (1995). [CrossRef]   [PubMed]  

8. S.-P. Lin, L. Wang, S.L. Jacques, and F.K. Tittel, “Measurement of tissue optical properties by the use of oblique-incidence optical fiber reflectometry,” Appl. Opt. 36, 136–143 (1997). [CrossRef]   [PubMed]  

9. F. Bevilacqua, D. Piguet, P. marguet, J.D. Gross, B.J. Tromberg, and C. Depeursinge, “In vivo local determination of tissue optical properties : applications to human brain,” Appl. Opt. 38, 4939–4950 (1999). [CrossRef]  

10. A. Amelink, A.P. van den Heuvel, W.J. de Wolf, D.J. Robinson, and H.J. Sterenborg, “Monitoring PDT by means of superficial reflectance spectroscopy,”J. Photochem. Photobiol. B 79, 243–251 (2005). [CrossRef]   [PubMed]  

11. C.A. Giller, H. Liu, P. P. Gurnani, S. Victor, U. Yazdani, and D. C. German, “Validation of a Near-Infrared Probe for Detection of Thin Intracranial White Matter Structures,” J. Neurosurg. 98, 1299–1306 (2003). [CrossRef]   [PubMed]  

12. C.A. Giller, M. Johns, and H. Liu, “Use of an intracranial near-infrared probe for localization during stereotactic surgery for movement disorders,” J. Neurosurg. 93, 498–505 (2000). [CrossRef]   [PubMed]  

13. F. Bevilacqua and C. Depeursinge, “Monte Carlo study of diffuse reflectance at source-detector separations close to one transport mean free path,” J. Opt. Soc. Am. 16, 2935–2945 (1999). [CrossRef]  

14. J. S. Dam, C. B. Pedersen, T Dalgaard, P. E. Fabricius, P. Aruna, and S. Andersson-Engels, “Fiber-optic probe for noninvasive real-time determination of tissue optical properties at multiple wavelengths,” Appl. Opt. 40, 1155–1164 (2001). [CrossRef]  

15. L.H. Wang, S.L. Jacques, and L-Q Zheng, “MCML-Monte Carlo modeling of photon transport in multi-layered tissues,” Comp. Meth. Prog. Biomed. 47, 131–146 (1995). [CrossRef]  

16. L.H. Wang, S.L. Jacques, and L-Q Zheng, “CONV-Convolution for responses to a finite diameter photon beam incident on multi-layered tissues,” Comp. Meth. Prog. Biomed. 54, 141–150 (1997). [CrossRef]  

17. http://oilab.tamu.edu/mc.html

18. A. Kienle, L. Lilge, M.S. Patterson, R. HIbst, R. Steiner, and B.C. Wilson, “Spatially resolved absolute diffuse reflectance measurements for noninvasive determination of the optical scattering and absorption coefficients of biological tissue,” Appl. Opt. 35, 2304–2314 (1996). [CrossRef]   [PubMed]  

19. F.A. Duck, Physical Properties of Tissue: A Comprehensive Reference Book (Academic Press, San Diego, 1990), p.62.

20. W-F Cheong, S.A. Prahl, and A.J. Welch, “A Review of the Optical Properties of Biological Tissues,” IEEE J. of Quan. Elec. 26, 2166–2185 (1990). [CrossRef]  

21. P. van der Zee, M. Essenpreis, and D.T. Delpy, “Optical properties of brain tissue,” Proc. SPIE 1888, 454–465, (1993). [CrossRef]  

22. P. Gurnani, “Near Infrared Spectroscopic Measurement of Human and Animal Brain Structures,” Master Thesis, The University of Texas at Arlington, Arlington, TX, May, 2003.

23. http://www.iss.com/Products/oxiplex.html

24. Z. Qian, S. Victor, Y. Gu, C.A. Giller, and H. Liu, “‘Look-Ahead Distance’ of a fiber probe used to assist neurosurgery: phantom and Monte Carlo study,” Opt. Express 111844–1855, (2003). [CrossRef]   [PubMed]  

25. M. Johns, C.A. Giller, and H. Liu, “Determination of hemoglobin saturation in blood-perfused tissues using reflectance spectroscopy with small source-detector separations,” Appl. Spectrosc. 55, 1686–1694 (2001). [CrossRef]  

26. M. Solonenko, R. Cheung, T.M. Busch, A. Kachur, G.M. Griffin, T. Vulcan, T.C. Zhu, H.W. Wang, S.M. Hahn, and A.G. Yodh, “In vivo reflectance measurement of optical properties, blood oxygenation and motexafin lutetium uptake in canine large bowels, kidneys and prostates,” Phys. Med. Biol. 47, 857–73 (2002). [PubMed]  

27. G. Paxinos and C. Watson, “The rat brain in stereotaxic coordinates,” Academic Press Inc., 4th edition, London, (1998).

28. M. Johns, “Optical properties of living tissues determined in vivo using a thin fiber optic probe,” Ph.D. Dissertation, The University of Texas at Arlington, Arlington, TX, December, (2003).

29. A.E. Cerussi, A.J. Berger, F. Bevilacqua, N. Shah, D. Jakubowski, J. Bulter, R.F. Holcombe, and B.J. Tromberg, “Sources of Absorption and Scattering Contrast for Near-Infrared Optical Mammography,” Acad Radiol. 8, 211–218 (2001). [CrossRef]   [PubMed]  

30. T. Durduran, R. Choe, J.P. Culver, L. Zubkov, M.J. Holboke, J. Giammarco, B. Chance, and A.G. Yodh, “Bulk optical properties of healthy female breast tissue,” Phys. Med. Biol. 47, 2847–2861 (2002). [CrossRef]   [PubMed]  

31. H. Eggert and V. Blazek, “Optical properties of human brain tissue, meninges, and brain tumors in the spectral range of 200 to 900 nm,” Neurosurgery 21, 459–464, (1987). [CrossRef]   [PubMed]  

32. H. Schwarzmaier, A. Yaroslavsky, I. Yaroslavsky, G. Thomas, K. Thomas, U. Frank, P. Schulze, and R. Schober, “Optical properties of native and coagulated human brain structures,” Proc. SPIE 2970, 492–499 (1997). [CrossRef]  

33. H. Shangguan, S.A. Prahl, S.L. Jacques, and L.W. Casperson, “Pressure effects on soft tissues monitored by changes in tissue optical properties,” in Laser-Tissue Interaction IX , S.L. Jacques Ed., Proc. SPIE 3254, 366–371 (1998).

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (11)

Fig. 1.
Fig. 1. Simulated relationship between reflectance and the reduced scattering coefficient at varying µa values of 0.01 cm-1 (solid diamond), 0.1 cm-1 (open square), 0.25 cm-1 (solid triangle), and 0.5 cm-1 (cross), with the source-detector separation fixed at 400 microns. The unit for reflectance is the number of photons/cm2.
Fig. 2. (a).
Fig. 2. (a). Overall intensity factor, a0–400 , versus measured reflectance, R m-400, at 750 nm (open blue circles) and at 830 nm (filled red circles) for the 400-µm probe. An average a 0–400 value is shown as the solid green line while the solid black curve is a quadratic fit for the data.
Fig. 2. (b).
Fig. 2. (b). Overall intensity factor, a 0–100, versus measured reflectance, R m-100, at 750 nm (open blue circles) and 830 nm (filled red triangles) for the 100-µm probe. An average a 0–400 value is shown as the solid green line, while the solid black curve is a quadratic fit for the data.
Fig. 3.
Fig. 3. (a). The schematic cross section of the 400-µm fiber probe. 3(b). Experimental setup for the in vivo reflectance measurements of the human middle finger. The particular probe shown above is just for the demonstration purpose and equivalent to the 400-µm probe, which was used for the determination of µs ’ values of human fingers. Also, the broadband light source and CCD spectrometer are shown, as labeled.
Fig. 4.
Fig. 4. Linear relationships between 1) the Intralipid concentration and the reflectance (red circles) and 2) the Intralipid concentration and the µs ’ values (blue solid squares) obtained from the ISS oximeter. The data are fitted with linear relationships for the reflectance (red line) and the µs ’ (blue line), respectively, for the 400-µm probe. Specifically, the linear relationships are R m-400=0.310×[Intralipid concentration]+0.009 and µ s ’=10.094× [Intralipid concentration] + 0.433 in cm-1.
Fig. 5.
Fig. 5. Error comparison between expected µs ’ (ISS) and calculated µs’ (cal) values, using the constant a 0–400 (=0.065±0.01) (filled red squares) and the polynomial a 0–400, i.e., Eq. (5) (open blue circles). All of the data points were based on five readings per location per Intralipid concentration. The data at 750 nm were used for this comparison.
Fig. 6.
Fig. 6. Linear relationships between the reduced light scattering coefficient (µs’) and the reflectance measured without ink (red diamonds) and with ink (blue circles). The absorption coefficients of Intralipid solutions without ink and with ink are 0.04 and 0.4 cm-1, respectively. The experiment was taken with the 400-µm probe, and the different values of µs’ were obtained by varying the Intralipid concentration. The data at 750 nm were used for this comparison.
Fig, 7(a).
Fig, 7(a). Error comparison between expected µs ’ (ISS) and calculated µs ’ (cal) values, using the constant a 0–100 (=0.0034±0.0005) (shown as filled red squares) and the polynomial a 0–100, i.e., equation (7) (shown as open blue circles). The horizontal dashed lines are the mean values of the open-circle and filled-square data points, respectively. To decrease noise due to small a source-detector separation, 4 locations per Intralipid solution were taken for the measurement, and 3 readings per location were used. The data at 750 nm were used for this comparison.
Fig, 7. (b).
Fig, 7. (b). Error comparison in µs ’ when using a single location measurement (constant a 0–100: filled red squares with the mean value plotted by solid red line; polynomial a 0–100: filled blue circles with the mean value by solid blue line) versus the average of three measurements per location (constant a 0–100: open red squares with the mean value by dashed red line; polynomial a 0–100: open blue circles with the mean value by dashed blue line). The data at 750 nm were used for this comparison.
Fig. 8.
Fig. 8. Calculated reduced scattering coefficients, µs ’, from living rat brain tissues at 750 nm; the data were obtained using the 400-µm probe.
Fig. 9.
Fig. 9. Calculated µs ’ values from living rat brain tissues at 750 nm; the data were obtained using the 100-µm probe.

Equations (13)

Equations on this page are rendered with MathJax. Learn more.

R sim ( λ 0 ) = 0.404 μ s ( λ 0 ) + 0.582 .
R m 400 ( λ 0 ) = a 0 400 R sim 400 ( λ 0 ) = a 0 400 [ 0.404 μ s ( λ 0 ) + 0.582 ] ,
R m 100 ( λ 0 ) = a 0 100 R sim 100 ( λ 0 ) = a 0 100 [ 1.670 μ s ( λ 0 ) 1.544 ] .
a 0 400 = R m 400 ( λ 0 ) 0.404 * μ s ( λ 0 ) + 0.5819 .
a 0 400 ( λ 0 ) = 0.0458 R m 400 ( λ 0 ) 2 + 0.0808 R m 400 ( λ 0 ) + 0.0407 ,
μ s ( λ 0 ) = R m 400 ( λ 0 ) 0.5819 a 0 400 ( λ 0 ) 0.404 a 0 400 ( λ 0 ) ,
μ s ( λ 0 ) = R m 400 ( λ 0 ) 0.5819 [ 0.065 ] 0.404 [ 0.065 ] ,
μ s ( λ 0 ) = R m 400 ( λ 0 ) 0.5819 [ 0.0458 R m 400 ( λ 0 ) 2 + 0.0808 R m 400 ( λ 0 ) + 0.0407 ] 0.404 [ 0.0458 R m 400 ( λ 0 ) 2 + 0.0808 R m 400 ( λ 0 ) + 0.0407 ] ,
a 0 100 ( λ 0 ) = 0.0126 R m 100 ( λ 0 ) 2 + 0.0074 R m 100 ( λ 0 ) + 0.0027 .
μ s ( λ 0 ) = R m 100 ( λ 0 ) + 1.5437 a 0 100 ( λ 0 ) 1.6696 a 0 100 ( λ 0 ) ,
μ s ( λ 0 ) = R m 100 ( λ 0 ) + 1.5437 [ 0.0034 ] 1.6696 [ 0.0034 ] ,
μ s ( λ 0 ) = R m 100 ( λ 0 ) + 1.5437 0.0126 R m 100 ( λ 0 ) 2 + 0.0074 R m 100 ( λ 0 ) + 0.0027 1.6696 [ 0.0126 R m 100 ( λ 0 ) 2 + 0.0074 R m 100 ( λ 0 ) + 0.0027 ] ,
Relative error in μ s = μ s ( probe ) μ s ( ISS ) μ s ( ISS ) 100 %