Absolute calibration of fiber-coupled single-photon detector

Open Access Open Access

Abstract

We show a setup for characterising the efficiency of a single-photon-detector absolutely and with a precision better than 1%. Since the setup does not rely on calibrated devices and can be implemented with standard-optic components, it can be realised in any laboratory. Our approach is based on an Erbium-Doped-Fiber-Amplifier (EDFA) radiometer as a primary measurement standard for optical power, and on an ultra-stable source of spontaneous emission. As a proof of principle, we characterise the efficiency of an InGaAs/InP single-photon detector. We verified the correctness of the characterisation with independent measurements. In particular, the measurement of the optical power made with the EDFA radiometer has been compared to that of the Federal Institute of Metrology using a transfer power meter. Our approach is suitable for frequent characterisations of high-efficient single-photon detectors.

© 2014 Optical Society of America

1. Introduction

Recently, important developments have been achieved in single-photon-counting technologies and high-efficient detectors have been developed in several laboratories [1, 2, 3]. These improvements allow us to perform challenging experiments in quantum optics, such as Device-Independent quantum-key distribution [4] and detection-loophole-free Bell tests [5, 6]. However, while the system detection efficiency is rapidly increasing, an accurate characterisation method is still not easy to access.

Conventionally, calibration of a single-photon detector is obtained by measuring the power of a classical beam with a reference power meter, then the beam is strongly attenuated and sent to the detector under test, DUT [7]. The accuracy of this method (usually between 5% and 10%) is mainly limited by the power stability of the beam, the precision of the attenuation stage and accuracy of the reference power meter. In particular, the accuracy of the absolute measurement for commercial power meters is large (∼ 5 %) so a direct calibration traceable to the primary standard is required. This process can only be performed in a metrological laboratory having the primary measurement standard, e.g. the cryogenic substitution radiometer [8], which is expensive and time-consuming.

An alternative approach has been developed on the basis of correlated photon pairs [9,10]. It allows us to estimate the detection efficiency of the DUT without relying on calibrated devices. In this scheme, the fraction of emitted photons impinging on the DUT has to be determined precisely. This is challenging due to the many spatial modes involved, particularly for fiber-coupled detectors [11]. Therefore, the conventional method is still commonly employed.

Sanguinetti et al. [12] demonstrated an absolute measurement of classical light at 1540 nm using a radiometer based on an Erbium-Doped Fiber Amplifier (EDFA). This radiometer is appealing for two reasons: it can be built with standard optical components and it is suitable for low powers in the range of nW, allowing a weaker attenuation stage.

Moreover, when characterising single-photon detectors based on semiconductor materials (SPAD), non-linearities such as afterpulsing and dead time/hold-off time have to be corrected in order to reduce the uncertainties. Time-correlated detection-efficiency measurements [13] attenuate the impact of these non-linearities at the expense of a more complicated setup.

In this paper we present a testbench for absolute characterisations of single-photon detectors with low uncertainty. Our setup is based on the conventional method, but we avoid the calibration at the metrological laboratory by using an EDFA radiometer that has been built following [12]. We calibrate the reference power meter with this device and then we verify the calibration factor at the Federal Institute of Metrology. Compared to [12], the calibration has been performed at a different wavelength (1552 nm) and includes a full error analysis. To improve stability and repeatability, we have developed a stable light source of spontaneous emission [14]. This source has been demonstrated as a precise reference for metrological applications since it is stable over days, unpolarised and incoherent. As a proof of principle, we characterise the efficiency of a SPAD analysing extensively the uncertainty on the efficiency measurement and compensating for the non-linearities.

This paper aims to provide a rigorous and reliable calibration protocol for characterising the efficiency of a SPAD. It is organized as follows: first of all we give an overview of the actual implementation devoting particular attention to the EDFA radiometer and the optical source. Then we show how to characterise the uncertainty introduced by each component. Particular attention is given to the calibration of the reference power meter traceable to the EDFA radiometer. Finally, we show how to characterise the detection efficiency of a SPAD.

2. Experimental setup

2.1. Overview

Figure 1 shows the testbench we use for the efficiency characterisation. Here we give an overview of the proposed testbench while the implementation is detailed in Sec.4.1. The beam emitted by the source is sent with an optical switch, either to the reference power meter (PMref) or to the DUT. The power in the beam is attenuated in two stages, A0 and Att, before impinging on the DUT. During the measurement PMref monitors periodically the beam power. When calibrating PMref, the light is also sent to the EDFA radiometer.

 figure: Fig. 1

Fig. 1 Conventional method: sketch.

Download Full Size | PDF

2.2. Incoherent stable source of spontaneous emission

An appropriate light source is vital for the stability and repeatability of the measurement. Although laser sources have the advantage of well-defined wavelength, the long coherence of the emitted light degrades the repeatability of the calibration because of interreflections that can occur between any couple of semi-reflecting surfaces [15]. Moreover, polarisation-dependent losses degrade the stability of the entire setup. Commercial incoherent sources are good alternatives, however their power may deviate by more than 2000 ppm after a few hours of measurement [17].

Another possible alternative is represented by an inverted atomic medium, such as Erbium-Doped Fibre (EDF) [14]. In this system, a strong 982 nm pump laser promotes all the Erbium ions in a metastable, excited state. When an ion decays to its ground level a photon is emitted. When the emission coming from the EDF is dominated by the spontaneous component, the light guided in the fibre is unpolarised and has a short coherence time as it can be deduced by the broad spectrum (centred at 1530 nm and 48 nm broad). Moreover, under strong pumping, the output optical power saturates with extremely small variations. Monteiro et al. [14] reported less than 25 ppm of power deviation after 3 days of measurement for a short-length EDF.

Based on their work, we have developed a stand-alone device (see Fig. 2) increasing the length of the EDF (ER30-4/125 by Thorlabs) to ∼18 cm (emitted power ∼ μW). The relative Allan deviation of the output power shows ≤20 ppm of deviation after one day of measurement.

 figure: Fig. 2

Fig. 2 Picture of the stand-alone device: the output fiber (on the right) is plugged into the power meter (on the left).

Download Full Size | PDF

2.3. EDFA radiometer: operating principle

As mentioned previously, the accuracy of the conventional method relies on the accuracy of the reference power meter. When measuring a light beam with an uncalibrated but linear power meter, the reading will be off by a factor k compared to the real value. Absolute optical power measurements can be obtained with the EDFA radiometer.

Like the stable source, this device is based on the EDF: the number of output photons per mode μout exiting from an EDF depends on stimulated and spontaneous emission and is described by [12]:

μout=Gμin+G1
where μin is the number of input photons per mode, and G is the gain of the medium. The term in represents the emission stimulated by the input light while the term G − 1 represents the spontaneous emission. Using the formalism of Eq. (1), the measured optical powers exiting the fibre when we inject ( Pst*) or not ( Psp*) an input light are given by
Psp*=(G1)2τchνk
Pst*=(Gμin+G1)2τchνk
where hν is the photon energy and 2/τc is the number of modes per second (1/τc is the number of temporal modes and the factor 2 corresponds to the number of polarisation modes). μin can be derived from Eq. (2,3) [18]:
μin(λ)=(11G(λ))(Pst*(λ)Psp*(λ)1)
We stress that μin is an absolute measurement since Pst*(λ)/Psp*(λ) does not depend on k. The gain G(λ) can always be deduced independently of k using:
G(λ)=Pst*(λ)Psp*(λ)Pin*(λ)
where Pin*(λ) is the reading of the incoming power. The λ dependencies introduced in Eq. (4, 5) are meant to emphasise that μin is obtained for each temporal/spectral mode.

Following [12], we have built the fibred setup shown in Fig. 3(a). The pump light is first injected into the EDF using a WDM, then it is removed with another WDM. We used a counter-propagating pumping configuration to prevent the pump light from exiting at the output of the radiometer. The input light travels through the two DWMs and the EDF where it is amplified. At the output the light is sent either to a spectrometer (Anritsu, MS9710A) or to a power meter. Two isolators are introduced at the input and output of the device to suppress any back-reflection of the telecom light that can occur outside the radiometer. Within the radiometer, back-reflections are suppressed using APC-connectors.

 figure: Fig. 3

Fig. 3 (a) Schematic of the radiometer. (b) Picture of the setup. The green line (orange arrow) enhances the Er3+ fiber (the pump laser fibre).

Download Full Size | PDF

3. Absolute calibration of the reference power meter

3.1. EDFA radiometer characterisation

To correctly determine the number of incoming photons, the transmission factors before and after the EDF have to be carefully measured. We label the transmission of the input connector as T1, T2 is the transmission up to the EDF, and T3 is the output transmission (see Fig. 3). The pump laser is off. When connecting an optical fibre to the power meter, the reading can be affected by a systematic calibration error that can be as high as 10% [15]. To reduce this uncertainty we employ an integrating sphere photodiode (Thorlabs, S144C). This photodiode, however, has a poor power sensitivity (≥1 μW) therefore we perform this measurement using a commercial amplified spontaneous emission source (Trillium Photonics) as the input source. The statistical errors are reduced by repeating several times the power measurements. To correctly calibrate T1 we connect and disconnect repeatedly the input fibre into the input connector maximising the transmission each time in order to improve the repeatability [12]. The uncertainty on T1 can be estimated from the statistics of the measured transmission factors. T2, instead, is calibrated by injecting the light backwards into the radiometer and measuring the light firstly in B, then in A. All the connectors belonging to T2 are adjusted to maximise the transmission. In this way, T2 is affected only by statistical errors which scale with the square root of the number of acquisitions. Finally, we calibrate T3 by injecting light from the input port and measuring the light in C and then at the output. Again, all the connectors belonging to T3 are adjusted in order to maximise the transmission. Also in this case the uncertainty scales with the square root of the number of acquisitions. The results are tabulated in Table 1:

Tables Icon

Table 1. Transmission factors. k indicates the coverage factor.

Once the transmission losses are characterised the radiometer is connected to the output C of the optical switch (see Fig. 1).

3.2. EDFA radiometer: gain measurement

The last quantity to be characterised is the gain of the EDF. Using the reference power meter (Thorlabs, S154C) we monitor the light at the output of the radiometer by injecting and not injecting an input beam. In the first case, the reading of the power meter corresponds to both spontaneous and stimulated emission ( Pst*), while in the second situation it is only due to the spontaneous component ( Psp*). The input light is blocked using A0 (EXFO, FVA-3150 equipped with an optical shutter). After having measured the light at the input of the radiometer always with the reference power meter ( Pin*), we can calculate G using

G¯=Pst*Psp*T3Pin*T1T2
σG=σPst*2+σPsp*2+σPin*2+σT12+σT22+σT32
and obtaining = 6.67(2). Note, in Eq. (6), all powers must be read using the same k, so we take care to use the same power-measurement-range setting during the three measurements. For simplicity, we introduce corresponding to the average gain over the spectrum of the input light. In the actual measurement we calculate the gain for each wavelength as described in [12] and in App. A.

3.3. Absolute measurement of the input power

We measure with a spectrometer (Anritsu, MS9710A) first the spontaneous emission from the EDF ( Psp*), then all the output light when the input beam is injected into the EDF ( Pst*) as a function of the wavelength, λ. With Eq. (4) we obtain μin(λi) and then we convert these numbers into optical power (Pradio) using [12]

Pradio=1T1T2λi(μin(λi)2hc2Δλλi3)
where λi is the central wavelength of each measurement point and Δλ = λi+1λi the spacing. Then we just need to sum over a wavelength range which is broader than the width of the input source. By this measurement we can calibrate PMref. We define kPM as the ratio between the power measured with the radiometer and the reading of the reference power meter, i.e. kPM=Pradio/Pin*. When calculating the effective amount of light measured with the power meter, one must take into account that a disconnected APC fibre end has an output loss of 3.3%, due to Fresnel reflection [16].

We estimate the uncertainty on kPM for a single-shot measurement. We detail the analysis in the Appendix A obtaining 0.6% of relative standard error. To test the repeatability of the measurement, we repeat the entire characterisation described in Sec.3 three times. The average between the obtained kPMs yields 95.2% with a standard deviation of 0.4%. This is consistent with the uncertainty of the single-shot measurement. As a confirmation, the calibration factor of PMref has been measured at the Federal Institute of Metrology (METAS) at 1550 nm against a transfer power meter traceable to the cryogenic radiometer. In this case, the absolute-calibration factor is measured at 100μW yielding kABS,Metas=95.3±0.7%. Considering the non-linearity coefficient between 100μW and 10 nW (measured at 1541 nm with a calibrated photodiode, kNL,Metas,10nW =99.754±0.001%), kMETAS =95.1±0.7%. We stress that the measurement with the radiometer was a blind test since METAS characterised the reference power meter only after we have measured kPM. The two values agree within less than one standard deviation, confirming the potential of the radiometer as a measurement device for the primary standard.

4. Characterisation of the measurement setup

4.1. Testbench: actual implementation

Once the reference power meter is calibrated, we proceed in characterising the efficiency measurement setup. Here we describe in details the testbench shown in Fig. 1: the light emitted by the source is filtered at λt = 1552 with 1.67 nm of FWHM (DiCon, TF500). The beam passes through A0 (EXFO, FVA-3150) and goes towards the optical switch (Lightech, LT-210) that diverts the light either to PMref (Thorlabs, S154C) or to the DUT. Att consists of two variable attenuators, A1,A2 (EXFO, FVA-60b). All the APC connectors have been polished with the same equipment to guarantee the same angle. The number of incoming photons per second on the detector, N, is derived from the measurement of the light power made with PMref using

N=PPMRDCAtthc/λt
where PPM is the power measured by PMref, hc/λt = 1.28 · 10−19 J is the photon energy, Att = A1 · A2 and RDC is the splitting ratio.

4.2. Splitting ratio

The repeatability and stability of the optical switch and the stability of Att affect the uncertainty of N. After setting Att to the minimum value, we characterise the splitting ratio, defined as RDC=PDPC where PC (PD) is the power measured at the point C (point D). We measure the light firstly in the upper path then in the bottom path over an extended period (∼10 hours). We use PMref to monitor the light at C (see Fig. 1) and an EXFO PM1100 power meter calibrated against PMref to monitor the light at D. Each power measurement is averaged for 1 second and repeated 10 times on each path. We obtain RDC=0.3441 (3) which corresponds to a relative standard error of 0.09%.

4.3. Attenuation stage

On the basis of the detector characteristics, we choose N∼20000 ph/s to guarantee enough detections without attaining saturation. Given PPM=9 nW we set A1=A2= 30 dB. The remaining attenuation is introduced adjusting A0 (see Table 2). To characterise the variable attenuator over 30 dB of attenuation we can not use the stable source since its output power is low, so we use the commercial source of amplified spontaneous emission, filtered at λt = 1552±1.67 nm again. Ai is obtained measuring the incoming light when the attenuation is set to the minimum (μW range) and when the attenuation is increased by 30 dB. Each power measurement is averaged over 1 s and repeated 10 times. The attenuation value is

Ai=P30dBm,ikNL30dBm,iP60dBm,ikNL60dBm,i
where kNLrange,i is the linearity measurement of the reference power meter performed at METAS and i = 1, 2. The standard error is
σAi=σP30dBm2+σP60dBm2+σNL30dBm2+σNL60dBm2
From that we can deduce the standard error on Att
σAtt=σA12+σA22~2σAi

Tables Icon

Table 2. Uncertainty budget of the attenuation

4.4. Overall stability

While characterising the detector, the stability of the optical components can affect the uncertainty of the measurement. We measure the Allan deviation of the optical power at D over an extended period (∼18 hours). The power stability of the source is degraded by the testbench but after one hour of measurement the power fluctuations are below 0.02% (see Table 3) of the average value. Because of that, during the efficiency characterisation we check the power every hour with 10 s of collection time.

Tables Icon

Table 3. Stability of the optical power

4.5. Uncertainty budget

Table 4 reports the uncertainty budget measured for the optical power impinging on the detector. The relative uncertainty of the testbench is 0.16%. Including the uncertainty of the reference power meter, the uncertainty of the number of photons impinging on the detector is 0.59%. This number is suitable for precise characterisations even at higher detection efficiencies.

Tables Icon

Table 4. Uncertainty budget of the testbench.

5. Measurement of the detection efficiency

The DUT is a pigtailed InGaAs/InP avalanche photodiode with a monolithic integrated negative feedback resistor. The detector is cooled with a Stirling refrigerator down to −70°C [19]. The electronic readout drives the detector in free-running conditions applying a programmable hold -off time after each detection to reduce the afterpulsing.

To quantify the stability of the overall setup, we continuously measure the detection-rate over 10 hours. In Fig. 4(a) we show the detection-rate probability distribution. Each data point corresponds to 10 seconds of collection time. The distribution, binned according to Scott’s rule [21], is Gaussian with standard deviation (184) close to the square root of the average value (191). This is consistent with a Poissonian distribution having a large average value. For such a detection rate (which corresponds to a relative uncertainty smaller than 1%), the fluctuations of the optical beam are negligible. In Fig. 4(b) we also show a quantitative analysis by means of the relative Allan deviation for the same dataset. Note that the relative Allan deviation remains below 0.2% for the entire dataset.

 figure: Fig. 4

Fig. 4 (a) Detection rate probability distribution for the DUT with collection time set to 10 s. The incoming power is not adjusted during the entire measurement (10 hours). The measured detection rate is limited only by Poissonian statistics and no significant drift of the average value is measured. This demonstrates the high stability of the setup. (b) Relative Allan deviation for the same measurement. The light blue shaded area represents the uncertainty of the measured value.

Download Full Size | PDF

After the calibration of PMref, to estimate the detection efficiency (η) of an ideal single-photon detector it would be sufficient to record, with two independent measurements, the avalanche rate sending or not photons to the detector. Then η is derived using:

η=rdetrdcN
where rdet is the avalanche rate originated by both photons and dark counts and rdc is the avalanche rate originated only by dark counts.

However, SPADs are affected by afterpulses which are countered introducing an hold-off time after each detection. This influences the measured rates: on the one hand, the detector can not detect photons during the hold-off times so, at a higher count rate, η is underestimated. On the other hand, the afterpulses increase the detection rate bringing an overestimation of η. The impact of the hold-off times on the detection rate can be corrected introducing a duty-cycle corresponding to the time when the detector is on. To correct for the afterpulsing we consider that after any detection an afterpulse is generated with probability pap. These corrections are introduced modifying rdet and rdc with

rdet(1+pap)=rdet*(1rdet*τ)
rdc(1+pap)=rdc*(1rdc*τ)
where τ is the hold-off time and rdet*(rdc*) is the measured detection rate (dark-count rate). We can now plug Eq. (14, 15) into Eq. (13) to get [13]:
η=1N(1+pap)(rdet*1rdet*τrdc*1rdc*τ)

Note, the specification sheet of a photon detector should provide η, τ and pap for different hold-off times. This would allow to calculate N in any measurement conditions simply by measuring rdet* and rdc* and using Eq. (16).

In our case, rdet*, rdc* have been measured after having integrated the avalanches over an extended period for different hold-off times (5, 10, 20 μs) and different nominal efficiencies (15% and 20%). As one may notice, it can happen that an afterpulse is originated by another afterpulse. To consider this, pap should correspond to the total afterpulsing probability and include also higher-order afterpulses pap can be measured reconstructing the temporal evolution of the avalanche rate given that an avalanche has occurred at time zero, Pc(t | 0). For very long delays there will be no correlation between two events and the probability that a pulse occurs will be determined only by the mean count rate, n. We calculate pap using

pap=075μs(Pc(t|0)n)dt
We use two independent measurement procedures (see Appendix B):
  • The first method, described in [22], records the time interval between successive avalanches. This is the easiest measurement procedure and requires only a Time To Digital Converter device to register the time-stamp.
  • The second method, described in [13], uses a pulsed laser to trigger the first avalanche. The temporal evolution is recorded with an FPGA board which controls the setup to record data only when the detector is in a well-defined condition. This method has the advantage of precise control of the detector but it requires a dedicated setup.

5.1. Results

For our detector, we measure a detection efficiency of 21.2±0.2% with a dark-count rate of 89 Hz at −70°C. However, here we are interested in the measurement uncertainty introduced by an insufficient compensation for hold-off time and the afterpulses. In our case, the uncertainty introduced by the hold-off time correction is negligible (≤ 0.04%) since the length of the hold-off time is known precisely (Δτ ≤ 100 ns) and we pay attention not to saturate the detector (the detector is inactive less than 10% the time). The afterpulse correction is more delicate as it is illustrated in Table 5. The table reports the efficiencies with different hold-off times before and after correcting for the afterpulsing (for pap obtained by the 2 methods).

Tables Icon

Table 5. Detection efficiency estimated before and after applying the afterpulsing correction. For the latter, we compared the two methods used to measure the afterpulsing probability.

Detection efficiency of a SPAD depends only on the excess bias and not on the hold off time. This is clearly not the case for the efficiencies obtained without afterpulsing correction. Looking at the values including correction, we note that the differences become much smaller. This is especially the case if Pap is measured with the second method. This indicates that the values obtained with this method are more appropriate. Note that, differences above 0.23% are significant according to Table 6. However for smaller efficiencies, Pap obtained by the simpler first method can give still satisfactory results, in particular for longer hold-off times. We expect the values for longer hold-off times being a better estimate of the real detection efficiency. The uncertainty of the detection efficiency introduced by imperfection of the afterpulses compensation depends on the settings. For 20% efficiency and 20 μs of hold-off time, we assume conservatively the introduced uncertainty to be smaller than 0.4%, which is the difference between the values obtained with 20 μs and 10 μs, respectively. Finally, we can provide the total uncertainty budget for the detection efficiency characterisation, see Table 6 for an overview.

Tables Icon

Table 6. Uncertainty budget of the efficiency characterisation. The uncertainty on the detection efficiency is obtained by adding the uncertainties of each component in quadrature.

6. Conclusions

In this paper we have shown how to perform an absolute calibration of the detection efficiency of a fibre-coupled single-photon detector using a simple testbench and an EDFA radiometer that provides a reliable primary measurement standard. We have determined the detection efficiency of a SPAD with a relative uncertainty well below 1%. If the measurement is performed carefully, correcting for afterpulses and hold-off times, the precision is determined by the absolute calibration of the reference detector. This calibration is achieved using a stable light source and the EDFA radiometer with an uncertainty as low as 0.57%. The accuracy of the radiometer is limited by the repeatability of the last connection.

The operating principle of the radiometer is general and can be applied with different rare-earth doped fiber in order to cover different ranges of wavelengths.

A. Uncertainty analysis of the calibration factor

Here we provide the analysis for the uncertainty of a single-shot absolute power measurement using the EDFA radiometer. We assume a flat gain to simplify the analysis. Equation 4 states

μin(λ)=(11G¯)(Pst*(λ)Psp*(λ)1)
σμin(λ)2=σA2+σB2=AB
where
σA=σG¯G¯(11/G¯)
σB=Δ(Pst*(λ)/Psp*(λ))Pst*(λ)/Psp*(λ)1
To estimate Δ(Pst*(λ)/Psp*(λ)) we repeatedly measure (20 times) Pi*(λ) and we calculate its standard error. The uncertainty on Pradio can be deduced from Eq. (8) using:
σPradio2=σT12+σT22+(σμin(λ)2+σΔλ2+σΔ1λ32)dλ
The calibration factor is
k=PradioPin*
σk2=σPradio2+σPin*2
where σPin* is the statistical error on PPM. Table 7 shows in details each contribution to the uncertainty of the calibration factor.

Tables Icon

Table 7. Uncertainty budget of the radiometer.

B. Afterpulsing characterisation

As mentioned before, afterpulses are usually characterised starting from the statistical distribution of the time intervals between two avalanches. The investigations about the physical nature of the phenomena are usually carried out considering the probability distribution between two subsequent avalanches [24]. On the contrary, we are interested in having the distribution between any pairs of avalanches with no assumption on what is happening between them. The difference between the two distributions is mainly due to higher order afterpulsing. We implemented two independent methods able to reconstruct the avalanche rate at time t conditioned on having an avalanche at time zero, Pc(t | 0).

In the first method [22], the timestamps of the avalanche occurrences are recorded. Then the histogram of the time delays between two pulses, h[iΔT], is built: for every pulse in the timestamp, the delays between this pulse and its successors (until a maximum delay Tmax = 75 μs) are calculated and h[iΔT] is increased by one for each of these delays. Tmax is chosen considering that the histogram becomes flat when the avalanches are uncorrelated. Pc(t | 0) is then deduced:

Pc(iΔT|0)ΔT=h[iΔT]NTot
where NTot is the total number of pulses that belong to the timestamp and ΔT is the bin-width, ∼ 300 ns. The main advantage of this method is its simplicity since it requires only a TDC to be performed. However, since the condition of the SPAD before the first avalanche is not well-defined, we can not guarantee that the distribution is independent from the history of the diode.

For this reason, a second method [13] has been developed: in this method the SPAD is prepared in a well-defined condition, i.e. that no avalanche has occurred in the previous 75μs. Then an FPGA triggers a laser pulse which is sent to the detector. When the pulse is detected, the FPGA records all the avalanches occurring in the next 75μs, building h[iΔT]. This time, NTot corresponds to the total number of avalanches originated by the laser pulses.

The total afterpulse probabilities, pap, obtained with these methods are reported in Table 8 for different settings of efficiency and hold-off times (HO). For the first method we also estimate the repeatability of the measurement doing the measurement 4 times. The two methods produce significantly different results. Since the measurement conditions can be well controlled for the second method, we believe that this method gives better results. This hypothesis is reinforced by the analysis of the results in chapter 5.1. However, more work will be necessary to understand the systematic difference between the two methods. It has to be noted that these discrepancies are small (for pap <<1) and, and finally not limiting the precision of the efficieny measurement.

Tables Icon

Table 8. Total afterpulse probabilities. The acquisition time for the first (second) method is 10 minutes (1 hours).

Acknowledgments

The authors thank A. Martin, N. Walenta, C. Barreiro, F. Do Rego Monteiro, T. Guerreiro for useful discussions. We are very grateful to Jacques Morel and Armin Gambon from the Swiss Federal Office for Metrology (METAS) for useful discussions, for hints and ideas, and for calibrating our power meter. The financial support is provided by the Swiss NCCR QSIT.

References and links

1. A. Restelli, J. C. Bienfang, and A. L. Migdall, “Single-photon detection efficiency up to 50% at 1310nm with an InGaAs/InP avalanche diode gated at 1.25 GHz,” Appl. Phys. Lett. 102(14), 141104 (2013). [CrossRef]  

2. F. Marsili, V. B. Verma, J. A. Stern, S. Harrington, A. E. Lita, T. Gerrits, I. Vayshenker, B. Baek, M. D. Shaw, R. P. Mirin, and S. W. Nam, “Detecting single infrared photons with 93% system efficiency,” Nat. Photonics 7(3), 210–214 (2013). [CrossRef]  

3. S. Miki, T. Yamashita, H. Terai, and Z. Wang, “High performance fiber-coupled NbTiN superconducting nanowire single photon detectors with Gifford-McMahon cryocooler,” Opt. Express 21(8), 10208 (2013). [CrossRef]   [PubMed]  

4. N. Gisin, S. Pironio, and N. Sangouard, “Proposal for implementing device-independent quantum key distribution based on a heralded qubit amplifier,” Phys. Rev. Lett. 105(7), 070501 (2010). [CrossRef]   [PubMed]  

5. P. M. Pearle and M. Philip, “Hidden-variable example based upon data rejection,” Phys. Rev. D 2(8), 1418–1425 (1970). [CrossRef]  

6. B. G. Christensen, K. T. McCusker, J. B. Altepeter, B. Calkins, T. Gerrits, A. E. Lita, A. Miller, L. K. Shalm, Y. Zhang, Y. S. W. Nam, N. Brunner, C. C. W. Lim, N. Gisin, and P. G. Kwiat, “Detection-loophole-free test of quantum nonlocality, and applications,” Phys. Rev. Lett. 111(13), 130406 (2013). [CrossRef]   [PubMed]  

7. J. Y. Cheung, C. J. Chunnilall, G. Porrovecchio, M. Smid, and E. Theocharous, “Low optical power reference detector implemented in the validation of two independent techniques for calibrating photon-counting detectors,” Opt. Express 19(21), 20347 (2011). [CrossRef]   [PubMed]  

8. A. C. Parr, “The candela and photometric and radiometric measurements,” J. Res. NIST 106(1), 151–186 (2000). [CrossRef]  

9. S. V. Polyakov and A. L. Migdall, “High accuracy verification of a correlated-photon- based method for determining photoncounting detection efficiency,” Opt. Express 15(4), 1390–1407 (2007). [CrossRef]   [PubMed]  

10. S. V. Polyakov and A. L. Migdall, “Quantum radiometry,” J. Mod. Opt. 569, 1045–1052 (2009). [CrossRef]  

11. M. Ware and A. L. Migdall, “Single-photon detector characterization using correlated photons: The march from feasibility to metrology,” J. Mod. Opt. 51(9), 1549–1557 (2004). [CrossRef]  

12. B. Sanguinetti, T. Guerreiro, F. Monteiro, N. Gisin, and H. Zbinden, “Measuring absolute spectral radiance using an Erbium-Doped Fiber Amplifier,” Phys. Rev. A 86(6), 062110 (2012). [CrossRef]  

13. T. Lunghi, C. Barreiro, O. Guinnard, R. Houlmann, X. Jiang, M. A. Itzler, and H. Zbinden, “Free-running single-photon detection based on a negative feedback InGaAs APD,” J. Mod. Opt. 59(17), 1481–1488 (2012). [CrossRef]  

14. F. Monteiro, T. Guerreiro, B. Sanguinetti, and H. Zbinden, “Intrinsically stable light source at telecom wavelengths,” Appl. Phys. Lett. 103(5), 051109 (2013). [CrossRef]  

15. J. Envall, P. Krh, and E. Ikonen, “Measurements of fibre optic power using photodiodes with and without an integrating sphere,” Metrologia 41(4), 353 (2004). [CrossRef]  

16. http://refractiveindex.info.

17. see e.g. EXFO Tunable laser source: IQS/FLS 2600.

18. B. Sanguinetti, E. Pomarico, P. Sekatski, H. Zbinden, and N. Gisin, “Quantum cloning for absolute radiometry,” Phys. Rev. Lett. 105, 080503 (2010). [CrossRef]   [PubMed]  

19. B. Korzh, N. Walenta, T. Lunghi, N. Gisin, and H. Zbinden, “Free-running InGaAs single photon detector with 1 cps dark count rate at 10% efficiency,” Appl. Phys. Lett. 104(8), 081108 (2014). [CrossRef]  

20. I. Vayshenker, S. Yang, X. Li, T. R. Scott, and C. L. Cromer, “Optical fiber power meter nonlinearity calibrations at NIST,” NIST special publications250–256 (2000).

21. D. W. Scott, “On optimal and data-based histograms,” Biometrika 66(3), 605–610 (1979). [CrossRef]  

22. J. W. Kindt, Geiger Mode Avalanche Photodiode Arrays: For Spatially Resolved Single Photon Counting (Delft University Press, 1999).

23. J. Zhang, R. Thew, J. D. Gautier, N. Gisin, and H. Zbinden, “Comprehensive Characterization of InGaAs-InP Avalanche Photodiodes at 1550 nm With an Active Quenching ASIC,” IEEE J. Quantum Electron. 45(7), 792–799 (2009). [CrossRef]  

24. S. Cova, A. Lacaita, and G. Ripamonti, “Trapping phenomena in avalanche photodiodes on nanosecond scale,” IEEE Electron. Dev. Lett. 12(12), 685–687 (1991). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (4)

Fig. 1
Fig. 1 Conventional method: sketch.
Fig. 2
Fig. 2 Picture of the stand-alone device: the output fiber (on the right) is plugged into the power meter (on the left).
Fig. 3
Fig. 3 (a) Schematic of the radiometer. (b) Picture of the setup. The green line (orange arrow) enhances the Er3+ fiber (the pump laser fibre).
Fig. 4
Fig. 4 (a) Detection rate probability distribution for the DUT with collection time set to 10 s. The incoming power is not adjusted during the entire measurement (10 hours). The measured detection rate is limited only by Poissonian statistics and no significant drift of the average value is measured. This demonstrates the high stability of the setup. (b) Relative Allan deviation for the same measurement. The light blue shaded area represents the uncertainty of the measured value.

Tables (8)

Tables Icon

Table 1 Transmission factors. k indicates the coverage factor.

Tables Icon

Table 2 Uncertainty budget of the attenuation

Tables Icon

Table 3 Stability of the optical power

Tables Icon

Table 4 Uncertainty budget of the testbench.

Tables Icon

Table 5 Detection efficiency estimated before and after applying the afterpulsing correction. For the latter, we compared the two methods used to measure the afterpulsing probability.

Tables Icon

Table 6 Uncertainty budget of the efficiency characterisation. The uncertainty on the detection efficiency is obtained by adding the uncertainties of each component in quadrature.

Tables Icon

Table 7 Uncertainty budget of the radiometer.

Tables Icon

Table 8 Total afterpulse probabilities. The acquisition time for the first (second) method is 10 minutes (1 hours).

Equations (25)

Equations on this page are rendered with MathJax. Learn more.

μ out = G μ in + G 1
P sp * = ( G 1 ) 2 τ c h ν k
P st * = ( G μ in + G 1 ) 2 τ c h ν k
μ in ( λ ) = ( 1 1 G ( λ ) ) ( P st * ( λ ) P sp * ( λ ) 1 )
G ( λ ) = P st * ( λ ) P sp * ( λ ) P in * ( λ )
G ¯ = P st * P sp * T 3 P in * T 1 T 2
σ G = σ P st * 2 + σ P sp * 2 + σ P in * 2 + σ T 1 2 + σ T 2 2 + σ T 3 2
P radio = 1 T 1 T 2 λ i ( μ in ( λ i ) 2 h c 2 Δ λ λ i 3 )
N = P PM R D C Att h c / λ t
A i = P 30 dBm , i k NL 30 dBm , i P 60 dBm , i k NL 60 dBm , i
σ A i = σ P 30 dBm 2 + σ P 60 dBm 2 + σ NL 30 dBm 2 + σ NL 60 dBm 2
σ Att = σ A 1 2 + σ A 2 2 ~ 2 σ A i
η = r det r dc N
r det ( 1 + p ap ) = r det * ( 1 r det * τ )
r dc ( 1 + p ap ) = r dc * ( 1 r dc * τ )
η = 1 N ( 1 + p ap ) ( r det * 1 r det * τ r dc * 1 r dc * τ )
p ap = 0 75 μ s ( P c ( t | 0 ) n ) d t
μ in ( λ ) = ( 1 1 G ¯ ) ( P st * ( λ ) P sp * ( λ ) 1 )
σ μ in ( λ ) 2 = σ A 2 + σ B 2 = A B
σ A = σ G ¯ G ¯ ( 1 1 / G ¯ )
σ B = Δ ( P st * ( λ ) / P sp * ( λ ) ) P st * ( λ ) / P sp * ( λ ) 1
σ P radio 2 = σ T 1 2 + σ T 2 2 + ( σ μ in ( λ ) 2 + σ Δ λ 2 + σ Δ 1 λ 3 2 ) d λ
k = P radio P in *
σ k 2 = σ P radio 2 + σ P in * 2
P c ( i Δ T | 0 ) Δ T = h [ i Δ T ] N Tot